Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K221218

    Validate with FDA (Live)

    Date Cleared
    2022-10-26

    (182 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Over-The-Counter Indications:
    Local management of superficial wounds, minor burns, abrasions, and lacerations.

    RX Use
    Silverlon® Wound Contact, Burn Contact Dressings are indicated for use up to 7 days for partial and full thickness wounds including traumatic wounds (donor and graft sites, incisions), first and second degree thermal burns, as well as dermal ulcers (stage I-IV pressure sores, venous stasis ulcers, arterial ulcers, diabetic ulcers), vascular access or peripheral IV sites, orthopedic external pin sites, and wound drain sites.

    Silverlon® Wound Contact, Burn Contact Dressings are indicated for use up to 7 days for decontaminated stable unroofed first and second degree mustard-induced vesicant injuries not requiring skin grafting.

    Silverlon® Wound Contact, Burn Contact Dressings are indicated for use up to 7 days for radiation dermatitis and cutaneous radiation injury through dry desquamation.

    Silverlon® Wound Contact, Burn Contact Dressings are indicated for the management of infected wounds, as the silver in the dressing provides an antimicrobial barrier that may be helpful in managing these wounds. In addition, the moist wound healing environment and control of wound bacteria within the Silverlon® Wound Contact, Burn Contact Dressings may help reduce the risk of wound infection and support the body's healing process.

    Silverlon® Wound Contact, Burn Contact Dressings may be used for the management of painful wounds. Silverlon® Wound Contact, Wound Burn Dressings are a non-adherent wound contact layer that reduces pain during dressing changes and evaporation of moisture in the dressing may soothe the wound.

    Device Description

    The Silverlon® Wound Contact, Burn Contact Dressings (WCD/BCD) are sterile, single use, single layer, non-adherent antimicrobial barrier wound dressings for up to seven day use. The dressings are comprised of Silverlon knitted nylon material plated with 99% elemental silver and 1% silver oxide. When Silverlon® dressings are moistened, the silver ions are activated, which kill wound bacteria held in the dressing which may help to reduce wound infection. The dressings are used as primary dressings that conform to the wound surface. The dressings are available in several configurations (e.g., rectangular, chest/trunk aprons, face masks, wraps, and gloves) and sizes.

    AI/ML Overview

    The provided text describes the performance data for Silverlon Wound Contact, Burn Contact Dressings but does not specify acceptance criteria or present the information in a clear "acceptance criteria vs. reported performance" table format. The study aims to demonstrate safety and feasibility, and to show non-inferiority in terms of radiation dermatitis severity compared to historical controls.

    Here's a breakdown of the requested information based on the provided text:


    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative table format for the clinical study. Instead, it focuses on demonstrating safety and feasibility, and comparing radiation dermatitis grades with a historical cohort.

    Acceptance Criteria Category (Inferred)Reported Device Performance
    SafetyNo adverse events related to the dressing were reported during or post RT (Radiation Therapy).
    FeasibilityDaily wear compliance was 99.9%.Average wear time was 22 hours/day.100% of 28 patients completing 90-day follow-up would recommend the dressing.
    Effectiveness (Radiation Dermatitis Grades)The mean RD grade for the Silverlon clinical trial group was 1.27 (SD 0.52), range 0-2 (RTOG scale).The mean RD grade for the matched cohort (standard-of-care) was 1.41 (SD 0.69), range 0-3 (RTOG or CTCAE scale).No Silverlon clinical study patients were excluded or withdrawn due to an RTOG grade of 3 or more, whereas RD grade 3 was observed in 2.8% of the matched cohort.
    Equivalence to PredicateCompared to the primary predicate, the subject device is identical for existing indications and the expanded indications have no impact on product design or technological characteristics.Compared to the secondary predicate, differences in construction/materials are not expected to raise different safety/effectiveness questions as both are occlusive dressings with the same principle of operation.

    2. Sample size used for the test set and the data provenance

    • Sample Size (Clinical Study): 30 patients were enrolled in the interventional clinical study.
    • Sample Size (Matched Cohort): 90 historical patients were matched (3 historical patients per clinical trial subject). For the comparison of RD grades, a subset of 71 matched cohort patients had documented RD grades.
    • Data Provenance: The interventional clinical study was an open label, single site, single arm, unblinded study. The text does not explicitly state the country of origin, but it is implied to be within the US due to FDA submission. The historical data for the matched cohort was from the "same study sites" within a four-year window of the clinical trial.
    • Retrospective/Prospective:
      • The Silverlon study data is prospective (interventional clinical study).
      • The matched cohort data is retrospective (historical patients).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts used to establish the ground truth or their qualifications. The "documented radiation dermatitis (RD) grades (either RTOG or CTCAE)" imply that these were assigned clinically, likely by treating physicians or radiation oncologists, but no details on their number or specific expertise are provided.


    4. Adjudication method for the test set

    The document does not describe any specific adjudication method for determining the RD grades for either the clinical trial subjects or the matched cohort. The grades were "documented," suggesting standard clinical assessment.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This study is for a wound dressing, not an AI or imaging device with human readers.


    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone (algorithm only) performance study was not done. This device is a physical wound dressing, not an algorithm.


    7. The type of ground truth used

    The ground truth for the radiation dermatitis grades was based on clinical assessment using standardized scales (RTOG or CTCAE scale) as documented in patient records.


    8. The sample size for the training set

    This section is not applicable as the document describes a physical medical device (wound dressing) and a clinical trial. There is no mention of an algorithm or a training set for an AI model.


    9. How the ground truth for the training set was established

    This section is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190343

    Validate with FDA (Live)

    Date Cleared
    2019-07-18

    (154 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Over-The Counter Indications:

    Local management of superficial wounds, minor burns, abrasions and lacerations.

    Prescription Indications:

    Silverlon® Wound Contact, Burn Contact Dressings are indicated for use up to 7 days for partial and full thickness wounds including traumatic wounds (donor and graft sites, incisions), first and second degree thermal burns, as well as dermal ulcers (stage I-IV pressure sores, venous stasis ulcers, diabetic ulcers), vascular access or peripheral IV sites, orthopedic external pin sites and wound drain sites.

    Silverlon® Wound Contact, Burn Contact Dressings are indicated for use up to 7 days for decontaminated stable unroofed first and second degree mustard-induced vesicant injuries not requiring skin grafting.

    Silverlon® Wound Contact, Burn Contact Dressings are indicated for the management of infected wounds, as the silver in the dressing provides an antimicrobial barrier that may be helpful in managing these wounds. In addition, the moist wound healing environment and control of wound bacteria within the Silverlon® Wound Contact, Burn Contact Dressings may help reduce the risk of wound infection and support the body's healing process.

    Silverlon® Wound Contact, Burn Contact Dressings may be used for the management of painful wounds. Silverlon® Wound Contact, Wound Burn Dressings are a non-adherent wound contact layer that reduces pain during dressing changes and evaporation of moisture in the dressing may soothe the wound.

    Device Description

    Silverlon Wound Contact, Burn Contact Dressings are sterile, porous, non-adherent, knitted nylon plated with 99% elemental silver oxide. Silver oxide. Silverlon Wound Contact, Burn Contact Dressings delivers antimicrobial silver ions in the dressing when activated by moisture. The silver ions in the dressing kill wound bacteria held in the dressing and provides an antimicrobial barrier to protect the wound bed.

    Silverlon Wound Contact, Burn Contact Dressings are used as the primary wound contact layer and is placed under reticulated foam, gauze, or other wound dressings. Silverlon Wound Contact, Burn Contact Dressings are non-adherent, wound care dressings designed to be used up to seven (7) days.

    AI/ML Overview

    The provided document is a 510(k) summary for the Silverlon® Wound Contact, Burn Contact Dressings. It describes the device, its indications for use, and a nonclinical (animal) study conducted to support an expanded indication.

    Here's an analysis of the provided information, focusing on acceptance criteria and the study that proves the device meets them:

    Acceptance Criteria and Device Performance

    The document describes a non-inferiority study performed on Gottingen minipigs. The acceptance criteria are based on a "non-inferiority margin" for a "histopathology composite score."

    Acceptance CriteriaReported Device Performance
    Non-Inferiority Margin: 1.5 points (or 10% of maximum possible score of 15) for the total histopathology composite score. The subject device is considered non-inferior if its score is not substantially worse than the standard of care.Moderate Partial Thickness (DD) Injuries: Mean composite histopathology score for the subject device was 0.36 points higher than standard of care.
    Superficial Partial Thickness (SD) Injuries: Mean composite histopathology score for the subject device was 0.01 points higher than standard of care.
    Conclusion:The study demonstrated that the Silverlon Wound Contact, Burn Contact Dressing is non-inferior to standard of care for both moderate and superficial partial thickness mustard-induced vesicant injuries. For DD injuries, this was demonstrated when changed every 7 days, and for SD injuries, when changed every 4 days.

    Explanation of Performance:
    Since a higher histopathology score is implicitly better (as the device score was higher than the standard of care and still met non-inferiority), the device performed as well as or slightly better than the standard of care, falling within the pre-defined non-inferiority margin.


    Study Details

    2. Sample size used for the test set and the data provenance:

    • Sample Size: 165 Gottingen minipigs.
    • Data Provenance: Nonclinical study, animal data (Gottingen minipigs). The study was a GLP (Good Laboratory Practice) nonclinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    The document does not specify the number or qualifications of experts used to establish the ground truth (histopathology scores). It only mentions that the primary endpoint was the "histopathology composite score." Typically, a pathologist would be responsible for such assessments, but no details are provided here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    The document does not describe any expert adjudication method for the histopathology scores. It only refers to a "histopathology composite score," which implies a single score derived from the histological examination.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. This was a nonclinical (animal) study comparing the wound dressing to standard of care. It was not an MRMC study and did not involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No. This was a nonclinical study of a physical wound dressing device. It does not involve an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The primary endpoint (ground truth) was a histopathology composite score of dermal wounds. This suggests a pathology-based assessment.

    8. The sample size for the training set:

    Not applicable. This was a nonclinical study of a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there was no training set.

    Ask a Question