510(k) Data Aggregation

    K Number
    K171866

    Validate with FDA (Live)

    Date Cleared
    2017-12-20

    (181 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sentrant Introducer Sheath with Hydrophilic Coating is intended to provide a conduit for the insertion of diagnostic or endovascular devices into the vasculature and minimize blood loss associated with such insertions.

    Device Description

    The Sentrant Introducer Sheath is a single-use, disposable, hydrophilic-coated catheter designed to provide a flexible and hemostatic conduit for the insertion of endovascular devices and to minimize blood loss associated with vascular procedures. The system is comprised of a dilator and an introducer sheath. The system accommodates a 0.035 in (0.89 mm) guidewire and is available in 28 and 64 cm lengths and in sizes from 12 to 26 Fr in 2 French increments.

    The dilator is radiopaque and has a flexible tapered tip that facilitates atraumatic tracking through the vasculature. There is a dilator grip at the proximal end of the dilator shaft. The dilator grip has a female Luer taper fitting on the proximal end to allow flushing of the device, and a threaded-feature at the distal end to allow the dilator to be secured to the sheath seal housing (dilator locking mechanism).

    The introducer sheath is comprised of a hydrophilic-coated, coil-reinforced catheter that is attached to a rigid seal housing containing the hemostatic valve assembly. A sideport extension assembly with a 3-way valve is attached to the seal housing. A radiopaque (RO) marker band is located at the distal tip of the sheath. The device also has a suture loop for attaching it to the patient and a strain relief to minimize kinking of the catheter where it joins to the seal housing.

    AI/ML Overview

    The provided text is a 510(k) summary for the Medtronic Sentrant Introducer Sheath with Hydrophilic Coating. This document describes the device and its equivalence to a predicate device, but it does not contain information about acceptance criteria and a study proving a specific device performance metric meets those criteria. Instead, it outlines a series of non-clinical performance tests conducted to demonstrate general safety and effectiveness for substantial equivalence to a predicate device, which had a different sterilization method.

    Therefore, many of the requested details about acceptance criteria, reported device performance (in terms of specific clinical or technical metrics for an AI/device performance study), sample sizes, ground truth establishment, expert involvement, and comparative effectiveness studies are not present in this document.

    However, I can extract the information that is available:

    1. A table of acceptance criteria and the reported device performance

    The document states that "All the predetermined acceptance criteria were met and results passed to support a determination of substantial equivalence" for the following non-clinical performance data. It does not provide specific numerical acceptance criteria or quantitative performance results for each test.

    Test PerformedReported Device Performance (General Statement)
    Design Verification TestingPredetermined acceptance criteria met; results passed to support substantial equivalence.
    Sterilization ValidationPredetermined acceptance criteria met; results passed to support substantial equivalence.
    Biocompatibility TestingPredetermined acceptance criteria met; results passed to support substantial equivalence.
    Packaging Design VerificationPredetermined acceptance criteria met; results passed to support substantial equivalence.
    Shelf Life TestingPredetermined acceptance criteria met; results passed to support substantial equivalence.

    2. Sample size used for the test set and the data provenance

    The document does not describe specific "test sets" in the context of clinical performance or algorithm evaluation with data provenance. The tests listed are non-clinical (e.g., sterilization, packaging, biocompatibility).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The document describes non-clinical tests, not expert-adjudicated ground truth for a clinical dataset.

    4. Adjudication method for the test set

    Not applicable. The document describes non-clinical tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document does not pertain to an AI-assisted device or a multi-reader multi-case study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This document does not describe an algorithm or AI performance.

    7. The type of ground truth used

    Not applicable in the context of clinical accuracy or algorithm performance. The "ground truth" for the non-clinical tests would be the established scientific and engineering standards and methods for validating those specific device characteristics (e.g., ISO standards for sterilization, biocompatibility, packaging).

    8. The sample size for the training set

    Not applicable. The document does not describe an AI or machine learning model with a training set.

    9. How the ground truth for the training set was established

    Not applicable. The document does not describe an AI or machine learning model with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123990

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-04-26

    (121 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sentrant Introducer Sheath with Hydrophilic Coating is intended to provide a conduit for the insertion of diagnostic or endovascular devices into the vasculature and minimize blood loss associated with such insertions.

    Device Description

    The Sentrant Sheath is a single-use, disposable, hydrophilic-coated catheter designed to provide a flexible and hemostatic conduit for the insertion of endovascular devices and to minimize blood loss associated with vascular procedures. The system is comprised of a dilator and an introducer sheath. The system accommodates a 0.035 in (0.89 mm) guidewire and is available in 28 and 64 cm lengths and in sizes from 12 to 26 Fr in 2 French increments. The dilator is radiopaque and has a tapered, flexible tip that facilitates atraumatic tracking through the vasculature. A female luer taper fitting is located on the distal end of the dilator grip. The proximal end of the dilator grip is threaded to allow the dilator to be secured to the sheath seal housing. The introducer sheath is comprised of a hydrophilic, coil-reinforced catheter that is attached to a rigid seal housing containing the hemostatic valve assembly. A sideport extension with a 3way valve is attached to the seal housing. A radiopaque markerband is located at the distal tip of the sheath. The device also has a suture loop for attaching it to the patient and a strain relief to minimize kinking of the catheter where it joins to the seal housing.

    AI/ML Overview

    This acceptance criteria and study information is for the Sentrant Introducer Sheath with Hydrophilic Coating (K123990).


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes the performance criteria as "predetermined acceptance criteria" which were all met. The specific quantitative values for these criteria are not detailed in the summary. However, the types of tests performed indicate the areas of performance evaluated.

    Acceptance Criteria CategorySpecific Test PerformedReported Device Performance
    Physical DimensionsOD of Sheath TubingAll predetermined acceptance criteria were met and results passed.
    OD of DilatorAll predetermined acceptance criteria were met and results passed.
    Sheath Working LengthAll predetermined acceptance criteria were met and results passed.
    Dilator Working LengthAll predetermined acceptance criteria were met and results passed.
    Hydrophilic Coating LengthAll predetermined acceptance criteria were met and results passed.
    Functional / Mechanical PerformanceGuidewire AcceptanceAll predetermined acceptance criteria were met and results passed.
    Grip to Dilator Tensile StrengthAll predetermined acceptance criteria were met and results passed.
    Seal Housing to Sheath Tubing Tensile TestAll predetermined acceptance criteria were met and results passed.
    Seal Cap to Seal Housing Tensile StrengthAll predetermined acceptance criteria were met and results passed.
    Atraumatic Tip (A-Tip) Lamination Tensile StrengthAll predetermined acceptance criteria were met and results passed.
    Sideport Torque TestAll predetermined acceptance criteria were met and results passed.
    Leak TestAll predetermined acceptance criteria were met and results passed.
    Kink TestAll predetermined acceptance criteria were met and results passed.
    Test for Liquid Leakage Under PressureAll predetermined acceptance criteria were met and results passed.
    BiocompatibilityISO MEM Elution, CytotoxicityAll predetermined acceptance criteria were met and results passed.
    ISO Maximization SensitizationAll predetermined acceptance criteria were met and results passed.
    ISO Intracutaneous ReactivityAll predetermined acceptance criteria were met and results passed.
    ISO Acute Systemic ToxicityAll predetermined acceptance criteria were met and results passed.
    USP Material Mediated PyrogenAll predetermined acceptance criteria were met and results passed.
    ASTM In-vitro HemolysisAll predetermined acceptance criteria were met and results passed.
    Complement Activation C3a and SC5b-9All predetermined acceptance criteria were met and results passed.
    ASTM Partial Thromboplastin TimeAll predetermined acceptance criteria were met and results passed.
    In-vivo ThromboresistanceAll predetermined acceptance criteria were met and results passed.
    Inductively Coupled Plasma SpectroscopyAll predetermined acceptance criteria were met and results passed.

    Study Proving Device Meets Acceptance Criteria

    The study described is a series of bench testing and biocompatibility testing.

    The type of study: Bench Testing and Biocompatibility Testing. The purpose was to evaluate its performance and verify that it meets the required performance specifications, and to assess biological compatibility per ISO10993-1.

    Outcome: "All of the predetermined acceptance criteria were met and results passed."


    Additional Information:

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the exact sample sizes (N-numbers) for each individual bench test or biocompatibility test.
      • Data Provenance: The tests were conducted by Medtronic Vascular, the device manufacturer. No information is provided regarding data country of origin, or if the data is retrospective or prospective, though bench testing is inherently prospective for device clearance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable as the study involves physical, mechanical, and biological testing of a medical device, not a diagnostic or AI-driven system requiring expert ground truth for interpretation (e.g., image analysis). The "ground truth" for these tests would be the established scientific and engineering principles, and the protocols defined by international standards (e.g., ISO, ASTM, USP).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable as the study does not involve subjective assessment or interpretation requiring adjudication by multiple readers/experts. The tests are objective measurements against predefined specifications.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. This device is an introducer sheath, not an AI-powered diagnostic tool. Therefore, no MRMC study or AI assistance evaluation was performed.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • This is not applicable. The device is a physical medical device, not an algorithm or software.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the bench testing, the "ground truth" is defined by the predetermined performance specifications and acceptance criteria that typically align with engineering standards, material science principles, and functional requirements for such devices.
      • For biocompatibility testing, the "ground truth" is established by adherence to recognized international standards such as ISO 10993-1 and other specific standards (e.g., ASTM, USP) for evaluating biological responses to medical devices.
    7. The sample size for the training set:

      • This is not applicable as the study does not involve an AI model or machine learning, and thus no "training set."
    8. How the ground truth for the training set was established:

      • This is not applicable as there is no training set for this type of device evaluation.