Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161491

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-02-09

    (253 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Scopis Hybrid Navigation System EM is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.

    Example procedures include, but are not limited to the following ENT procedures:

    • · Transsphenoidal access procedures;
    • Intranasal procedures;

    · Sinus procedures, such as Maxillary antrostomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies;

    · ENT related anterior skull base procedures.

    Device Description

    The Scopis Hybrid Navigation System EM displays the position of navigated instruments on a model of the patient's anatomy based on preoperative images (CT or MRI) using electromagnetic tracking technology. The position of the instruments and the patient are localized within an electromagnetic field produced by a field generator. The navigation of instruments relative to the patient's anatomy is established via registration of the patient's anatomy to the image set via fiducial markers, anatomical landmarks, or surface matching. The position of navigated instruments is then displayed on the model from the image set.

    The Scopis Hybrid Navigation System EM consists of:

      1. Navigation Unit
      1. Electromagnetic field generator
      1. Patient tracker with integrated localizers
      1. Navigation instruments with integrated localizers
      1. Navigation software (Nova Basic)
    AI/ML Overview

    This document describes the Scopis Hybrid Navigation System EM, an image-guided surgery system for ENT procedures. The information provided focuses on its substantial equivalence to a predicate device, as demonstrated through performance data.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Predicate Device Performance)Reported Device Performance (Scopis Hybrid Navigation System EM)
    Navigation Accuracy: 0.9 ± 0.34 mmNavigation Accuracy: 0.49 ± 0.27 mm

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated. The document mentions "bench testing of the navigation accuracy was performed," suggesting a controlled laboratory environment without specifying the number of measurements or tests conducted.
    • Data Provenance: The data is a result of "bench testing," meaning it was generated in a controlled laboratory setting (likely in Germany, given the submitter's address). It is prospective data collection for the purpose of demonstrating device performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. Bench testing typically relies on metrology equipment and standardized procedures for ground truth, rather than expert human interpretation.

    4. Adjudication method for the test set:

    This information is not applicable and therefore not provided. The assessment of navigation accuracy in bench testing relies on direct measurement against a known physical standard, not on human adjudication of classifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted and is not relevant for this type of device. The Scopis Hybrid Navigation System EM is an image-guided surgery system, not a diagnostic AI system that "reads" cases or assists human readers in interpretation. Its performance is measured by its navigational accuracy.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, the performance testing described is essentially a standalone (algorithm only) assessment of the device's navigation accuracy during bench testing. No human interaction or in-vivo performance is described in the provided "Performance Data" section for establishing substantial equivalence.

    7. The type of ground truth used:

    The ground truth used for the bench testing would be a precisely measured physical standard or reference, against which the device's navigational output is compared. For example, a known physical position measured by a highly accurate metrology system would serve as the ground truth for the "0.49 ± 0.27 mm" accuracy measurement.

    8. The sample size for the training set:

    This information is not provided and is not directly applicable in the context of this submission. The Scopis Hybrid Navigation System EM is an image-guided navigation system that uses patient-specific CT or MR images as reference. While the software itself would have been developed and tested, the document focuses on the performance of the product rather than the development of deep learning models requiring a "training set" in the typical sense.

    9. How the ground truth for the training set was established:

    This information is not provided and is not directly applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1