Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K232787

    Validate with FDA (Live)

    Date Cleared
    2023-10-06

    (25 days)

    Product Code
    Regulation Number
    868.1980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ScanNav Anatomy Peripheral Nerve Block is indicated to assist qualified healthcare professionals to identify and label the below mentioned anatomy in live ultrasound images in preparation for ultrasound guided regional anesthesia prior to needle insertion for patients 18 years of age or older.

    The highlighting of structures in the following anatomical regions is supported:

    • · Axillary level brachial plexus
    • · Erector spinae plane
    • · Interscalene level brachial plexus
    • · Popliteal level sciatic nerve
    • · Rectus sheath plane
    • · Sub-sartorial femoral triangle / Adductor canal
    • · Superior trunk of brachial plexus
    • · Supraclavicular level brachial plexus
    • · Longitudinal suprainguinal fascia iliaca plane
    • Femoral block

    ScanNav Anatomy Peripheral Nerve Block is an accessory to compatible general-purpose diagnostic ultrasound systems.

    Device Description

    ScanNav Anatomy Peripheral Nerve Block is a software (SaMD) which assists qualified healthcare professionals to identify and label relevant anatomical structures in preparation for ultrasound guided regional anesthesia prior to needle insertion for patients 18 years of age or older.

    The device receives ultrasound images in real-time from a compatible general-purpose ultrasound machine. It processes these images using deep learning artificial intelligence algorithms and highlights relevant anatomical structures. The ultrasound machine display remains unaffected, and the highlighting is only displayed on a general-purpose panel PC provided with the device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the ScanNav Anatomy Peripheral Nerve Block device, based on the provided FDA 510(k) summary:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance (Femoral Block)
    FP rate: Mis-identification rate of safety critical anatomical structures in the indicated procedures is less than 5%.1.1% (2 out of 183 scans)
    Accuracy (TP+TN) rate: Correct highlighting of safety critical anatomical structures in the indicated procedures at least 80% of the time.96.7% (177 out of 183 scans)
    FN rate: Non-identification rate of safety critical anatomical structures in the indicated procedures is less than 15%.2.2% (4 out of 183 scans)

    Note: The software tests also had acceptance criteria (e.g., successful completion of unit test, integration test, etc.) but specific quantitative performance metrics were not provided in the summary, only that "all software tests have been successfully completed without any anomalies."


    Study Details

    1. Sample Size for the Test Set and Data Provenance:

    • Sample Size: 183 scans were used for the safety and accuracy validation of the Femoral Nerve Block.
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It only states that the tests used "established protocol and acceptance criteria same as that used for the predicate device."

    2. Number of Experts Used to Establish Ground Truth and Qualifications:

    • This information is not provided in the given 510(k) summary.

    3. Adjudication Method for the Test Set:

    • This information is not provided in the given 510(k) summary.

    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • A MRMC comparative effectiveness study was not explicitly mentioned in this summary. The study described focuses on the device's standalone performance against defined criteria, rather than comparing human readers with and without AI assistance.

    5. Standalone (Algorithm Only) Performance:

    • Yes, the provided performance metrics (FP rate, Accuracy, FN rate) directly reflect the standalone performance of the algorithm for the Femoral Nerve Block. The device processes ultrasound images using AI algorithms and highlights structures, and the reported rates indicate how well the algorithm performs this task independently.

    6. Type of Ground Truth Used:

    • The summary states that the validation involved "safety and Accuracy validation" through "established test protocols." While not explicitly named (e.g., "expert consensus" or "pathology"), given the context of identifying anatomical structures in ultrasound images, it is highly probable that the ground truth was established by expert consensus (e.g., skilled sonographers or regional anesthetists manually identifying and labeling the structures) or well-defined anatomical landmarks.

    7. Sample Size for the Training Set:

    • This information is not provided in the given 510(k) summary.

    8. How the Ground Truth for the Training Set Was Established:

    • This information is not provided in the given 510(k) summary.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN220024

    Validate with FDA (Live)

    Date Cleared
    2022-10-18

    (193 days)

    Product Code
    Regulation Number
    868.1980
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ScanNav Anatomy Peripheral Nerve Block is indicated to assist qualified healthcare professionals to identify and label the below mentioned anatomy in live ultrasound images in preparation for ultrasound guided regional anesthesia prior to needle insertion for patients 18 years of age or older.

    The highlighting of structures in the following anatomical regions is supported:

    • Axillary level brachial plexus ●
    • Erector spinae plane .
    • Interscalene level brachial plexus ●
    • Popliteal level sciatic nerve .
    • Rectus sheath plane
    • Sub-sartorial femoral triangle / Adductor canal .
    • Superior trunk of brachial plexus
    • Supraclavicular level brachial plexus ●
    • . Longitudinal suprainguinal fascia iliaca plane

    ScanNav Anatomy Peripheral Nerve Block is an accessory to compatible general-purpose diagnostic ultrasound systems.

    Device Description

    ScanNav Anatomy Peripheral Nerve Block is a software medical device which assists anesthetists and other qualified healthcare professionals in the identification of anatomical structures within ultrasound images during ultrasound-guided regional anesthesia (UGRA) procedures by highlighting the relevant anatomical structures in realtime.

    The device performs the highlighting by using deep learning artificial intelligence technology based on convolutional networks (CNNs). These deep-learning models generate a colored overlay that allows the user to identify the specific anatomical structures of interest for the procedure. A separate monitor displays the highlighted images as an overlay on top of the ultrasound image, so the original view from the ultrasound machine is not affected. The deep learning models are locked, and they do not continue to learn in the field.

    The device interfaces with ultrasound machine with an external monitor output that meets the compatibility requirements. The ScanNav Anatomy Peripheral Nerve Block is run on a mobile computing platform (a commercial off the shelf panel PC) performing the processing with an integrated touchscreen monitor to display the user interface and anatomy highlighting

    The Software as a Medical Device is packaged with a tablet PC, power cable, compatible plug, and mounting bracket and instructions for mounting the tablet to the ultrasound host. This acts as a separate monitor to display the highlighted images as an overlay on top of the ultrasound image, so the original view from the ultrasound machine is not affected. The ScanNav Anatomy Peripheral Nerve Block system is composed of a software medical device and other non-medical devices such as a panel PC, power supply, an HDMI interface cable and a VESA mount.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for ScanNav Anatomy Peripheral Nerve Block

    This document outlines the acceptance criteria for the ScanNav Anatomy Peripheral Nerve Block device and details the studies conducted to demonstrate its compliance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific MetricAcceptance CriterionReported Device Performance
    Clinical Performance (Primary Endpoint)Assistance in obtaining correct ultrasound view prior to needle insertionMajority view (at least 8/15 participants) agree the device assists.63% (19/30) of participants were assisted, meeting the criterion.
    Clinical Performance (Secondary Endpoint 1)Assistance in identification of anatomical structures (up to BMI 35 kg/m2)Majority view (at least 8/15 participants) agree the device assists.70% (21/30) of participants were assisted, meeting the criterion.
    Clinical Performance (Secondary Endpoint 2)Assistance in supervision and training for anatomical structure identificationMajority view (at least 8/15 supervising experts) agree the device assists.87% (13/15) of experts were assisted, meeting the criterion.
    Clinical Performance (Secondary Endpoint 3)Improvement in operator confidenceMajority view (at least 8/15 participants) agree the device improves confidence.63% (19/30) of participants had improved confidence, meeting the criterion.
    Safety (Misidentification)Frequency of misidentification (FP Rate) of anatomical structuresNot explicitly stated as a numerical criterion for all blocks, but assessed as a primary endpoint in one study.Varies by block, ranging from 0% (ESP, Adductor) to 21.9% (SFIC). Details in section 2 below.
    Safety (Adverse Events Risk)Frequency of highlighting risking an adverse event<= 5% of total for each specified adverse event risk.Specific percentages are redacted (b)(4), but implied to be within acceptable limits as the device was granted De Novo.
    Accuracy (Correct Identification)Frequency of correct identification (TP+TN)>= 80% of total for each block.Varies by block, ranging from 76.2% (SFIC) to 98.3% (SC). Details in section 2 below.
    Human FactorsSuccessful completion of essential and critical tasks by usersAll participants complete essential and critical tasks without patterns of use failures, confusion, or difficulties.All 30 participants completed tasks. No UI design issues, use errors, or task failures were found.
    SoftwareCompliance with design and safety standardsSoftware documentation acceptable, with Major Level of Concern addressed.Documentation reviewed and accepted. Supports cybersecurity and hazard analysis.
    Electromagnetic Compatibility & Electrical SafetyCompliance with IEC 60601-1-12 and IEC 60601-1-2Test results support electrical safety and electromagnetic compatibility.Test results support compliance.

    2. Study Details for Device Performance

    The provided documentation describes two main studies relevant to device performance: a Clinical Validation Study to assess Performance and predict Adverse Events (IU2021 AG 07) and a Human Factors (HF) Study Design.

    Clinical Validation Study (IU2021 AG 07)

    • Sample Size: 40 volunteers.
    • Data Provenance: Single-center, prospective validation study conducted in the USA (Oregon Health & Science University, Portland).
    • Number of Experts for Ground Truth: Three (3) expert anesthesiologists in UGRA.
    • Qualifications of Experts: Anesthesiologists competent to perform independent UGRA.
    • Adjudication Method: Majority opinion (2/3) determined TP, TN, FP, FN, and AE rates.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No, this study was primarily a standalone validation of the device's highlighting performance against expert consensus. It did not directly compare human readers with and without AI assistance to quantify an effect size in human improvement. However, the Human Factors study (described below) did involve performance "with and without the aid of the device," providing some insight into assisted performance.
    • Standalone Performance (Algorithm-only): Yes, the study assessed the "device output" and its highlighting performance, with experts evaluating the device's interpretations in isolation. Experts answered questions on the device's highlighting performance (TP, TN, FP, FN).
    • Type of Ground Truth: Expert consensus. Three expert anesthesiologists viewed recorded ultrasound clips side-by-side with the ScanNav Anatomy PNB output and determined the correctness of the highlighting.
    • Sample Size for Training Set: Not explicitly stated in the provided text. The device uses "deep learning artificial intelligence technology based on convolutional networks (CNNs)."
    • How Ground Truth for Training Set was Established: Not explicitly stated, but typically involves expert annotation of ultrasound images for relevant anatomical structures.

    Key Performance Metrics (from IU2021 AG 07):

    • Primary Endpoint (Misidentification - FP Rate):
      • Axillary: 0.3%
      • ESP: 0.0%
      • IS: 1.3%
      • Pop: 0.6%
      • RS: 3.2%
      • Adductor: 0.0%
      • ST: 5.2%
      • SC: 0.8%
      • SFIC: 21.9%
    • Secondary Endpoint (Accuracy - Correct Identification - TP+TN Rate):
      • Axillary: 97.7%
      • ESP: 88.8%
      • IS: 94.1%
      • Pop: 98.1%
      • RS: 96.8%
      • Adductor: 90.4%
      • ST: 90.9%
      • SC: 98.3%
      • SFIC: 76.2%
    • Adverse Event Rates: Redacted sections (b)(4) indicate specific risks (PONS, LAST, Pneumothorax, Peritoneum risk) were assessed per block, aiming for <= 5% frequency.

    Human Factors (HF) Study Design

    • Sample Size: 30 anesthesiologists (15 Expert, 15 Trainee).
    • Data Provenance: Summative usability validation study conducted in a simulated interventional procedural lab setting in the USA (b)(4).
    • Number of Experts for Ground Truth: The study involved participants' self-assessment via questionnaires and a panel of three (3) independent experts who reviewed recorded scans for later analysis.
    • Qualifications of Experts:
      • Expert Participants (15): Capable of independent clinical practice of UGRA, deliver regular clinical care, 14 are members of relevant professional societies, 11 hold advanced further training in UGRA.
      • Trainee Participants (15): Undergoing training for UGRA procedures; 7 deliver regular clinical care, 13 are members of relevant professional societies.
      • Independent Expert Panel (3): Expertise in UGRA, reviewed recorded scans.
    • Adjudication Method: For the expert panel, majority panel view determined the device's performance and safety profile for each scan. For primary and secondary clinical endpoints, it was based on majority view of participants (at least 8/15) for each group (experts, trainees, or combined as 30 participants).
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Yes, in essence. The study involved participants (readers) performing "scans with and without the aid of the device," on two models (cases). This directly assesses the impact of the AI assistance on user performance and perceptions.
      • Effect Size of Human Readers Improve with AI vs. Without AI Assistance:
        • Assistance in obtaining correct ultrasound view: 63% (19/30) participants were assisted.
        • Assistance in identification of anatomical structures: 70% (21/30) participants were assisted.
        • Assistance in supervision and training (for experts): 87% (13/15) experts were assisted.
        • Improved operator confidence: 63% (19/30) participants reported improved confidence.
        • Reduction of risk of mistaking an incorrect view: 70% reduction in this risk reported (presumably from expert panel analysis based on device use in the study).
        • (b)(4) increase in risk (redacted amount) where incorrect highlighting increased the risk.
    • Standalone Performance (Algorithm-only): The expert panel independently evaluated ScanNav Anatomy PNB highlighting on recorded scans, which can be seen as an assessment of the algorithm's standalone performance, albeit within the context of generated images from user interaction. However, the clinical validation study (IU2021 AG 07) is a more direct evaluation of standalone performance (TP, TN, FP, FN rates).
    • Type of Ground Truth:
      • Participant self-assessment/perception: Through questionnaires regarding assistance, confidence, and identification.
      • Expert consensus: The independent panel of three experts reviewed recorded scans and device highlighting, completing the same questionnaire as participants, with their "majority panel view" forming a ground truth for device performance and safety.
    • Sample Size for Training Set & How Ground Truth for Training Set was Established: Not stated in the provided text.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1