510(k) Data Aggregation

    K Number
    K201626

    Validate with FDA (Live)

    Date Cleared
    2021-03-24

    (281 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Safety Subcutaneous Infusion Set is indicated for subcutaneous infusion of medication administered by an external infusion pump or syringe.

    Device Description

    The Safety Subcutaneous Infusion Set is an intravascular administration set used to administer medication subcutaneously. The device consists of subcutaneous stainless-steel needle mounted to a wing stabilizer at 90° at the distal end of the infusion set. The subject device has closing type wings to allow the user to insert the needle into the subcutaneous tissue and also secure the needles on the infusion site. A female luer connector is provided at the proximal end to allow connection to external devices. Each set consists of the administration set and standard medical dressing to secure the needles on the infusion site. The subject device is provided in various needle gauges and lengths. Each needle is equipped with an open end slide clamp on the tubing to facilitate stop flow when required during priming of individual needles.

    AI/ML Overview

    This document is a 510(k) summary for the "Safety Subcutaneous Infusion Set" and does not contain information related to an AI/ML powered device. As such, it does not include acceptance criteria directly applicable to AI/ML performance, nor a study proving that an AI/ML device meets such criteria.

    The information provided pertains to the substantial equivalence determination for a medical device that facilitates subcutaneous infusion of medication. Therefore, I cannot extract the requested information regarding AI device acceptance criteria or studies from this document.

    However, I can provide the acceptance criteria and performance data for the physical device described in this document, which are based on non-clinical tests:

    Acceptance Criteria and Device Performance (Non-Clinical for Physical Device):

    Acceptance Criteria CategorySpecific Test/StandardReported Device Performance/Conclusion
    BiocompatibilityISO 10993-1:2018Biocompatibility studies performed
    Particulate MatterUSP <788>Particulate matter testing performed
    SterilizationETO, SAL of 10-6Sterilization validation performed, achieved SAL of 10-6 (Same as predicate)
    Ethylene Oxide ResidualsISO 10993-7:2008Ethylene oxide residuals testing performed
    Shelf LifeASTM F1980-16Shelf life testing performed
    Sharps ProtectionISO 23908:2011Sharps protection compliant to ISO 23908:2011 (Same as predicate)
    Tensile Strength & Water TightnessISO 8536-8:2015Tensile strength and water tightness testing performed
    Luer Lock ConnectionISO 80369-7Luer lock connection test performed
    Comparison to Predicate (Technological Characteristics)(See detailed comparison table in document)Found to have similar design, intended use, proximal interface, needle lengths, number of needles, needle safety feature, wing type, and sterilization method. Differences in protective cap design, priming volume, and materials did not raise new safety/effectiveness questions.

    The following requested information is not applicable to this document as it describes a physical medical device, not an AI/ML product:

    • Sample size used for the test set and the data provenance
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    • Adjudication method for the test set
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done
    • If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    • The sample size for the training set
    • How the ground truth for the training set was established
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1