    Intended Use

    The Safety pen needle is intended for use with pen injector devices for subcutaneous injection of drugs. It has sharps protection feature of the patient-end and cartridge-end. For Model SD, the needle tube tip is 3 bevel. For Model SDW, the needle tube tip is 5 bevel.

    Device Description

    Safety Pen Needle, could be divided into Model SD and Model SDW.
    Model SD/SDW consists of needle hub,shield sleeve,rear spring,front spring, slider, inner shell,protective jacket,needle tube,outer container and seal(dialysis paper).

    Safety Pen Needle,could be divided into two models according to the needle tube tip is 3 or 5 bevel; the needle tube is thin-wall. Although the product of different model have different needle tube tip, but the same is that a double-ended needle tube is fixed with needle hub, the needle tip is protected by a protective jacket, the product is sealed with medical dialysis paper and needle container. The device is sterilized by EO gas and for single use. The shelf life is 5 years. The device is non-toxic and non-pyrogenic.

    Needle tube is fixed in the center of the needle hub with UV glue, the needle tip is protected with protective jacket before use, medical dialysis paper is covered and sealed needle container to maintain sterile of the device. The cartridge end of needle can be inserted into the rubber of insulin pen, meanwhile the needle hub is connected to insulin pen with screw thread to provide sterile fluid path for injection of insulin during use. The patient end and the cartridge end of the needle tube are lubricated with silicone oil for ease of injection and rubber penetration.

    Safety Pen Needle has sharps injury protection feature that it can reduce the occurrence of accidental needlesticks from the patient end and cartridge end of the needle due to the effect of safe-lock guard.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K212514

    Validate with FDA (Live)

    Date Cleared
    2022-04-29

    (262 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Safety Pen Needle for Single Use is intended for use with pen injector devices for the subcutaneous injection of insulin.

    The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.

    Device Description

    The devices are provided in two types of configurations: Safety Pen Needle for Single Use and Insulin Pen Needle.

    The Safety Pen Needle for Single Use is intended for use with pen injector devices for the subcutaneous injection of insulin. It consists of a needle cap, needle hub, safety protective cover, self-destruction seat and sealed paper. The safety protective cover is manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.

    The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin. It consists of protective cap, needle tube, needle protective cover and sealed paper.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a medical device, specifically Safety Pen Needles for Single Use and Insulin Pen Needles. It focuses on demonstrating substantial equivalence to a predicate device (K181447) rather than presenting a study for general acceptance criteria alone.

    Therefore, the response will be structured to address the acceptance criteria as derived from the comparative and standalone testing performed to establish substantial equivalence.

    Here's an analysis based on the provided text:


    1. Table of acceptance criteria and the reported device performance

    The document lists numerous non-clinical tests and standards that the proposed device was tested against. The acceptance criteria are implicit in compliance with these standards and the comparisons to the predicate device. The performance is stated as "complied with" or "met the requirements".

    Acceptance Criteria (Standard / Test)Reference Clause/StandardReported Device Performance
    Biocompatibility
    CytotoxicityISO 10993-5: 2009No cytotoxicity
    Irritation/Intracutaneous ReactivityISO 10993-10: 2010No intracutaneous reactivity
    SensitizationISO 10993-10: 2010No skin sensitization
    Systemic ToxicityISO 10993-11: 2017No systemic toxicity
    HemolysisASTM F756-17No Hemolysis
    PyrogenUSP <151>No Pyrogen
    Subacute ToxicityNot explicitly listed in table, but mentioned in textNo adverse effect
    Sterilization & Endotoxin
    EO Residue (EO, ECH)ISO 10993-7:2008Did not exceed limit
    Bacterial Endotoxins LimitUSP <85>Did not exceed 20 EU/device
    Sterility Assurance Level (SAL)Not explicitly listed in table, but mentioned in text10^-6
    Packaging & Shelf Life
    Seal Integrity (Visual)ASTM F1886 / F1886M-16Deemed acceptable
    Seal StrengthASTM F88/F88M-15Complied with requirements
    Dye Penetration (Seal Leaks)ASTM F1929-15Complied with requirements
    Shelf Life (5 years)Physical, Mechanical, Chemical, Package TestsMaintain its performance
    Physical & Mechanical (General Needles)
    DimensionsISO 11608-2:2012 Cl. 4.2, ISO 9626:2016 Cl. 5.6Complied with requirements
    Flow RateISO 11608-2:2012 Cl. 4.3Complied with requirements
    Bond (Hub & Needle Tube)ISO 11608-2:2012 Cl. 4.4, ISO 7864:2016 Cl. 4.12Complied with requirements
    Needle PointsISO 11608-2:2012 Cl. 4.5, ISO 7864:2016 Cl. 4.11Complied with requirements
    Freedom from DefectsISO 11608-2:2012 Cl. 4.6Complied with requirements
    LubricationISO 11608-2:2012 Cl. 4.7Complied with requirements
    Dislocation of Measuring PointISO 11608-2:2012 Cl. 4.8Complied with requirements
    Functional Compatibility with Pen SystemsISO 11608-2:2012 Cl. 4.9Complied with requirements
    Ease of Assembly/DisassemblyISO 11608-2:2012 Cl. 4.10Complied with requirements
    CleanlinessISO 7864:2016 Cl. 4.3, ISO 9626:2016 Cl. 5.3Complied with requirements
    Acidity/AlkalinityISO 7864:2016 Cl. 4.4, ISO 9626:2016 Cl. 5.4Complied with requirements
    Extractable MetalsISO 7864:2016 Cl. 4.5Complied with requirements
    Size DesignationISO 7864:2016 Cl. 4.6, ISO 9626:2016 Cl. 5.5Complied with requirements
    Color CodingISO 7864:2016 Cl. 4.7Complied with requirements
    Needle HubISO 7864:2016 Cl. 4.8Complied with requirements
    Needle CapISO 7864:2016 Cl. 4.9Complied with requirements
    Needle TubeISO 7864:2016 Cl. 4.10Complied with requirements
    Lumen PatencyISO 7864:2016 Cl. 4.13Complied with requirements
    Surface Finish & AppearanceISO 9626:2016 Cl. 5.2Complied with requirements
    StiffnessISO 9626:2016 Cl. 5.8Complied with requirements
    Resistance to BreakageISO 9626:2016 Cl. 5.9Complied with requirements
    Resistance to CorrosionISO 9626:2016 Cl. 5.10Complied with requirements
    Particulate TestingUSP <788>Complied with requirements
    Specific Safety Pen Needle Features
    Safety Feature PerformanceISO 23908:2011 & FDA GuidanceMet pre-established criteria
    Safety Feature Activation Forces(Compared to Predicate)Average at 3.71N (Predicate: 3.73N)

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each test in the acceptance criteria. It broadly states "Non-clinical tests were conducted..." and "Physical, Mechanical, and Chemical testing listed in following table were performed on the proposed device."
    Regarding data provenance, the manufacturer is Berpu Medical Technology Co., Ltd. in Wenzhou, Zhejiang, China. Therefore, the data provenance is China, and the data is prospective as it involves direct testing of the proposed devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable as the document describes non-clinical laboratory and performance testing for a medical device (pen needles), not a diagnostic algorithm requiring expert ground truth for interpretation of images or other data. The "ground truth" here is defined by compliance with established international and ASTM standards.

    4. Adjudication method for the test set

    This section is not applicable for the same reasons as point 3. The evaluation is based on objective measurements against specified standard criteria, not subjective expert judgment requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The device is a physical medical device (pen needle), not an AI-powered diagnostic tool, and therefore, no MRMC study or AI assistance evaluation was performed or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable as the device is a physical medical instrument, not an algorithm or software. The non-clinical tests described are "standalone" in the sense that they evaluate the device's intrinsic properties against standards, without human interaction influencing the core performance metrics being measured (e.g., flow rate, bond strength).

    7. The type of ground truth used

    The "ground truth" for the non-clinical tests and the product's performance is established by international and national standards (e.g., ISO, ASTM, USP) and the specifications derived from these standards. For example, the criteria for cytotoxicity or sterility assurance level (SAL) are defined by the respective ISO and USP standards. The "pre-established criteria" mentioned for the simulated clinical study and safety feature test would also fall under this category, likely derived from the relevant standards (ISO 23908:2011) and FDA guidance.

    8. The sample size for the training set

    This section is not applicable. The document describes the testing of a physical medical device. There is no "training set" in the context of an AI/ML algorithm. The training in device manufacturing typically refers to process validation and quality control, not data training.

    9. How the ground truth for the training set was established

    This section is not applicable for the same reason as point 8; there is no training set mentioned in the context of this device's approval process.


    Ask a Question

    Ask a specific question about this device

    K Number
    K210864

    Validate with FDA (Live)

    Date Cleared
    2022-02-17

    (331 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The safety pen needle is intended for use with pen injector devices for subcutaneous injection of fluids and FDA-approved drugs.

    HE Model: Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection.

    HS Model: Additionally, the product has two safety shields, which lock in place after use (patient end) and upon removal of the needle from the pen (pen connection end). The locked shields help reduce the occurrence of needle sticks from both ends of the needle.

    Device Description

    The proposed device, Safety Pen Needle, is a sterile, non-pyrogenic, single-use device, which is intended for use with pen injector devices for subcutaneous injection of fluids and FDA-approved drugs. It incorporates a sharps injury prevention feature, which can reduce the occurrence of accidental needle sticks.

    The proposed device is provided with two models, i.e. HE Model and HS Model. HE Model has an attached safety shield, which can automatically lock in place and reduce the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection. HS Model has two safety shields, which lock in place after use (patient end) and upon removal of the needle from the pen (pen connection-end). The locked shields help reduce the occurrence of accidental needle sticks from both ends of the needle.

    Both of HE Model and HS Model contain main body, needle tube, front shield, rear shield, big spring, turning tube, container (outer cap) and peel tab (sealing paper). Both ends of the needle tube (i.e. the patient end and the pen connection end) is lubricated by silicone oil. The needle tube and the needle hub are bonded by glue. The container sealed with the peel tab constitutes the sterile barrier system. The common components of both models are almost the same in materials and structure and size, and the only difference is the structure of the rear shield. The biggest difference between the two models is that HS Model contains a small spring additionally and has a relatively more complex structure.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Safety Pen Needle). It describes the device, its intended use, comparison to a predicate device, and non-clinical testing conducted to demonstrate substantial equivalence.

    However, the text does not contain the detailed information required to answer the specific questions about acceptance criteria and a study proving the device meets those criteria, particularly for a software-based AI/ML device. The document pertains to a physical medical device (pen needle) and relies on standard performance testing and simulated clinical use for safety features, not AI/ML performance metrics.

    Therefore, I cannot extract the requested information regarding:

    • A table of acceptance criteria and reported device performance (for AI/ML metrics like sensitivity, specificity, AUC). The document lists "Requirements" (standards) and "Results" (Pass/Fail) for various physical and biological tests, but these are not the AI/ML performance metrics you asked for.
    • Sample size used for the test set and data provenance (for AI/ML). The document mentions 500 subject devices for a "simulated clinical use study" of the safety feature, but this is not an AI/ML test set.
    • Number of experts used to establish ground truth and their qualifications. This is not relevant to a physical pen needle's mechanical testing.
    • Adjudication method for the test set. Not relevant.
    • If a multi-reader, multi-case (MRMC) comparative effectiveness study was done. Not relevant, as this is for a physical device, not an AI assisting human readers.
    • If a standalone (algorithm only) performance study was done. Not relevant.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.). The "ground truth" for the physical device's safety feature was a lack of needle-stick injuries and protective feature failure in a simulated environment.
    • The sample size for the training set. Not applicable for a physical device.
    • How the ground truth for the training set was established. Not applicable.

    In summary, the provided document describes the regulatory submission for a physical medical device and does not contain the information requested about the acceptance criteria and study for an AI/ML powered device.