Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K993983

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-01-27

    (64 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emit® II Plus Phencyclidine Assay is a homogeneous enzyme immunoassay with a 25 ng/mL cutoff (SAMHSA initial test cutoff level). The assay is intended for use in the qualitative and semiquantitative analyses of phencyclidine in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

    The Emit® II Plus Phencyclidine Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

    Device Description

    The Emit® II Plus Phencyclidine Assay is a homogenous enzyme assay intended for use in qualititative and semiquantitative analysis of phencyclidine in human urine. The Emit® II Plus Phencyclidine Assay and has been found to be equivalent to the predicate device: Emit® II Phencyclidine Assay with regard to intended use, assay sample, and overall performance characteristics.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Emit® II Plus Phencyclidine Assay, focusing on acceptance criteria and the supporting study:

    The provided document describes a 510(k) submission for the Emit® II Plus Phencyclidine Assay, establishing its substantial equivalence to a predicate device (Emit® II Phencyclidine Assay). The studies presented are primarily for demonstrating this equivalence.

    Acceptance Criteria and Reported Device Performance

    The document describes several performance characteristics rather than explicit acceptance criteria with pre-defined thresholds for each. However, we can infer the implied acceptance criteria from the reported results and the comparison to the predicate device.

    Performance CharacteristicStated Acceptance Criteria (Inferred)Reported Device Performance (Emit® II Plus)
    Qualitative AnalysisExcellent correlation to the predicate device (Emit® II Phencyclidine Assay).99% agreement with the predicate device using a 25 ng/mL cutoff. One discordant sample (15 ng/mL PCP by GC/MS) was positive by Emit® II Plus and negative by Emit® II.
    Spiked Sample Recovery (Qualitative)Consistently distinguish negative vs. positive: < 25% of cutoff should be negative; > 25% of cutoff should be positive.Spiked levels < 25% of 25 ng/mL cutoff (0 - 18.75 ng/mL) were consistently distinguished as negative. Spiked levels > 25% of 25 ng/mL cutoff (31.25 - 300 ng/mL) were consistently distinguished as positive.
    Spiked Sample Recovery (Semiquantitative)Quantitated results within 10% of nominal concentration for a specified range.Quantitated within 10% of nominal concentration between 8.0 ng/mL and 90 ng/mL.
    Precision (Qualitative - Within-run)Acceptable within-run precision for controls and cutoff (rates).%CV for controls and cutoff (rates) at 0.5%.
    Precision (Qualitative - Total)Acceptable total precision for controls and cutoff (rates).%CVs for controls and cutoff (rates) ranged from 0.5 to 0.6%.
    Precision (Semiquantitative - Within-run)Acceptable within-run precision for controls and cutoff (concentrations).%CV for controls and cutoff (concentrations) ranged from 1.79 to 2.44%.
    Precision (Semiquantitative - Total)Acceptable total precision for controls and cutoff (concentrations).%CV ranged from 2.18 to 2.60%.
    Correlation with GC/MSGood relationship between semiquantitative analyses and GC/MS values. (This is a more general statement of acceptance rather than a specific numeric criterion).Demonstrated a "good relationship" between semiquantitative analyses and GC/MS values for all positive samples and a portion of negative samples (n=20).

    Study Details

    This document describes a performance study conducted to demonstrate substantial equivalence, not a standalone clinical validation study in the traditional sense, especially concerning an AI/ML device.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: 100 specimens for comparative analysis (49 positive, 50 negative by both Emit® II Plus and Emit® II).
      • Data Provenance: Not explicitly stated (e.g., country of origin, demographics). The study appears to be laboratory-based internal testing, likely retrospective using collected urine samples. It is implied the samples are human urine.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is not applicable in the context of this device. The ground truth for drug-of-abuse testing is established by confirmatory methods (like GC/MS) and reference assays (the predicate Emit® II assay), not by human expert opinion.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. When there was a discrepancy between the Emit® II Plus and the predicate Emit® II assay, Gas Chromatography/Mass Spectrometry (GC/MS) was used as the confirmatory (adjudication) method. For example, one discordant sample was confirmed by GC/MS to have 15 ng/mL of PCP.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is not an AI/ML device, nor does it involve human readers interpreting images or data. It is an in vitro diagnostic (IVD) assay that provides quantitative and qualitative results directly.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in the sense that it's an automated assay. The Emit® II Plus Phencyclidine Assay is a homogeneous enzyme immunoassay designed for use with chemistry analyzers, meaning it provides results without direct human interpretation of a visual output where human error in reading is a factor. The "performance" refers to the assay's output itself, which is then interpreted by a human user (e.g., laboratory technician, physician) in the context of the patient. However, it's not "algorithm-only" in the AI/ML sense. It's a chemical assay.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For positive confirmation: Gas Chromatography/Mass Spectrometry (GC/MS) was used as the reference (confirmatory) method.
      • For comparison/equivalence: The predicate device, Emit® II Phencyclidine Assay, served as the comparative method.
      • For "known" samples: Spiked samples with known concentrations of phencyclidine were used.
    7. The sample size for the training set:

      • This information is not provided as this is not an AI/ML device that undergoes a 'training' phase in the conventional sense. The development of such an assay involves reagent formulation and optimization, not machine learning model training on a dataset.
    8. How the ground truth for the training set was established:

      • Not applicable. As stated above, this is not an AI/ML device. The "ground truth" during development would be established through careful analytical chemistry and formulation to ensure the assay reagents react as expected with known concentrations of phencyclidine and its metabolites.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1