Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K101409

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-08-02

    (75 days)

    Product Code
    Regulation Number
    880.6885
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heatsensitive medical devices in healthcare facilities.

    The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 46 to 55°C, and rinses the load with extensively treated* potable water. The processed load should be used immediately.

    The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

    • The extensive treatment of EPA potable water consists of: *
        1. Pre-filtration through two pre-filters:
        • Pre-filter 1 is a gross depth filter that removes approximately 5 micron . or larger particles/contaminants.
        • Pre-filter 2 is a surface filter that removes particles/contaminants > . ().1 micron.
        1. UV Irradiation:
        • During transit through the UV water treatment chamber, a UV dose . sufficient to achieve a ≥ 6-log reduction of MS2 virus is delivered to the water.
        1. 0.1 micron filtration:
        • The water prepared by pre-filtration and UV irradiation is filtered . through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa > 0.1 micron.
    Device Description

    The SYSTEM IE Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible, heat sensitive, flexible and rigid endoscopes and their accessories, and microsurgical instruments. The system consists of the SYSTEM 1E Processor and the S40 Sterilant Concentrate, interchangeable processing trays/containers and Quick Connects.

    The SYSTEM 1E Processor is an automated, self-contained device which creates and maintains the conditions necessary for liquid chemical sterilization in 6 minutes. Following the processing, the liquid chemically sterilized articles are rinsed with extensively treated water produced by passing EPA potable tap water through pre-filters, an ultraviolet light treatment subsystem, and then through two 0.1 micron filter membranes. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.

    S40 Sterilant Concentrate is a single use chemical sterilant concentrate developed exclusively for use in the SYSTEM 1E Processor. The active ingredient in S40, peracetic acid, is combined with inert ingredients to form a use dilution which inhibits corrosion of metals, polymers and other materials.

    The interchangeable processing trays/containers are made to accommodate a variety of instrument types, models and procedure specific sets. Each container is designed of maintain instruments in appropriate position while specific Quick Connects ensure thorough circulation of the sterilant solution and rinses.

    AI/ML Overview

    This document is a 510(k) Summary for the STERIS SYSTEM 1E Liquid Chemical Sterilant Processing System, focusing on a modification involving an alternative UV water treatment subsystem. It does not contain information about an AI/ML device or a study comparing AI performance against acceptance criteria. Therefore, I cannot provide the requested information. The document details the device, its intended use, and its substantial equivalence to a predicate device based on this specific change.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1