Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K113079

    Validate with FDA (Live)

    Date Cleared
    2011-12-13

    (57 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Synthes Hemostatic Bone Putty is indicated for use as a water-soluble implant material and for use in the control of bleeding from bone surfaces.

    Device Description

    Synthes Hemostatic Bone Putty (HBP) stops bone bleeding by establishing a physical barrier along the edges of bones that have been damaged by trauma or cut during a surgical procedure. When applied as directed, HBP forms a mechanical barrier that occludes the vascular openings in the damaged bone. This barrier prevents further bleeding during the surgical procedure and dissolves postoperatively, permitting normal tissue healing and bone regeneration. HBP is a blend of synthetic water soluble polymers that form a ready-to-use hemostatic agent that is substantially eliminated from the defect site in less than 48 hours.

    The constituents of Synthes Hemostatic Bone Putty and Ostene, the predicate, are similar. Ostene is comprised of a proprietary mixture of water soluble alkylene oxide copolymers. HBP is also comprised of water soluble alkylene oxide polymers. The remainder of HBP is a polysaccharide, carboxymethylcellulose (CMC), to improve handling. Ostene does not contain CMC.

    AI/ML Overview

    The provided document is a 510(k) summary for the Synthes Hemostatic Bone Putty. It details the device's description, intended use, and the non-clinical tests performed to demonstrate substantial equivalence to a predicate device. However, it does not describe specific acceptance criteria and detailed device performance results in a format that would allow for the construction of the requested table.

    Specifically, the document lists several non-clinical tests but does not provide:

    • Specific acceptance criteria for each test (e.g., "cytotoxicity level must be below X").
    • Reported device performance in quantitative terms against such criteria.
    • Any information related to clinical studies (MRMC, standalone, human readers, sample sizes for test/training sets, ground truth establishment, expert qualifications, adjudication methods).

    Therefore, I cannot fulfill your request for the detailed table and study information based solely on the provided text. The document focuses on demonstrating substantial equivalence through non-clinical biocompatibility and resorption studies, not on providing performance metrics against predetermined acceptance criteria typically seen in clinical performance studies of diagnostic or AI-powered devices.

    Based on the provided text, the following information is available (and what is not):

    1. Table of acceptance criteria and reported device performance:
    * Not available. The document lists types of non-clinical tests but does not define specific acceptance criteria or report quantitative device performance against them. It states that "Documentation is provided that demonstrates that Synthes Hemostatic Putty is substantially equivalent to other legally marketed devices," implying that the results of these tests met the necessary standards for equivalence, but without providing the specifics.

    2. Sample size used for the test set and the data provenance:
    * Not applicable/Not available. The studies described are non-clinical (in vitro/animal studies), not human clinical trials with test sets derived from human data. The "test set" concept as usually applied to AI/diagnostic devices doesn't fit here.
    * Data provenance: The animal studies mentioned include those in rats and rabbits for resorption rates, and sheep for vertebral body defect evaluation. The specific countries of origin for these animal studies are not stated. All studies appear to be prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    * Not applicable/Not available. As these are non-clinical animal studies, the "ground truth" would likely be established by trained laboratory personnel or veterinary pathologists, not human medical experts in the context of diagnostic interpretation. No specific numbers or qualifications are provided.

    4. Adjudication method for the test set:
    * Not applicable/Not available. Not relevant for the types of non-clinical studies described.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:
    * No. This document is for a hemostatic bone putty, not an AI-powered diagnostic device. An MRMC study is not relevant to this type of medical device's submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
    * No. This is not an algorithm or AI device.

    7. The type of ground truth used:
    * For the animal studies, the "ground truth" would be determined by direct observation, histological analysis, or quantitative measurements (e.g., resorption rates, evaluation of bone healing) in the animal models.

    8. The sample size for the training set:
    * Not applicable/Not available. This is not an AI/machine learning device; therefore, no training set is mentioned or relevant.

    9. How the ground truth for the training set was established:
    * Not applicable/Not available. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103822

    Validate with FDA (Live)

    Date Cleared
    2011-05-19

    (140 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Synthes Hemostatic Bone Putty is indicated for use as a water-soluble implant material and for use in the control of bleeding from bone surfaces.

    Device Description

    Synthes Hemostatic Bone Putty (HBP) stops bone bleeding by establishing a physical barrier along the edges of bones that have been damaged by trauma or cut during the surgical procedure. When applied as directed, HBP forms a mechanical barrier that occludes the vascular openings in the damaged bone. This barrier prevents further bleeding during the surgical procedure and dissolves postoperatively, permitting normal tissue healing and bone regeneration. HBP is a blend of synthetic water soluble polymers that form a ready-to-use hemostatic agent that is substantially eliminated from the defect site in less than 48 hours. The constituents of Synthes Hemostatic Bone Putty and Ostene, the predicate, are similar. Ostene is comprised of a proprietary mixture of water soluble alkylene oxide copolymers. HBP is also comprised of water soluble alkylene oxide polymers. The remainder of HBP is a polysaccharide, carboxymethylcellulose (CMC), to improve handling. Ostene does not contain CMC.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Synthes Hemostatic Bone Putty (HBP) and mentions a variety of non-clinical tests performed for substantial equivalence comparison. However, it does not contain the detailed acceptance criteria or numerical performance data of the device in a structured table, nor does it present a study proving the device meets specific performance criteria.

    The document primarily focuses on establishing "substantial equivalence" of HBP to a predicate device (Ostene CT Soluble Hemostasis Implant Material, K082491), rather than demonstrating direct performance against predefined acceptance criteria. Substantial equivalence in this context means the device is as safe and effective as a legally marketed predicate device.

    Therefore, many of the requested sections regarding acceptance criteria, reported performance, sample sizes, ground truth, and expert involvement cannot be extracted directly from this document.

    Here's what can be inferred or stated based on the provided text, while also explicitly noting what is not available:


    1. Table of acceptance criteria and the reported device performance:

    This information is not provided in the document. The document lists several non-clinical tests that were performed, but it does not specify quantitative acceptance criteria for these tests or present the reported numerical performance of the device against such criteria. The studies listed are:

    • Cytotoxicity Study Using the ISO Elution Method
    • Mouse Peripheral Blood Micronucleus Study
    • ISO Modified Intracutaneous Study, Solution with Measurement
    • Genotoxicity: Bacterial Reverse Mutation Assay
    • Genotoxicity: Mouse Lymphoma Assay
    • ISO Guinea Pig Maximization Sensitization Test-Solution
    • Systemic Toxicity Study
    • An In Situ Study to Determine the HBP Resorption Rate in a Rat Craniotomy Model
    • In Vivo Resorption Rate of a Hemostatic Bone Putty Subcutaneously Implanted in the Rabbit at 2, 4, 7, 14 Days.
    • In Vivo evaluation of Hemostatic Bone Putty in a sheep vertebral body defect at 7 days
    • Evaluation of Hemostatic Bone Putty in a Sheep Vertebral Body Defect
    • An In Vivo Study to Determine Hemostatic Bone Putty Effect on Bone Healing In A Rat Craniotomy Model at 3, 6, and 12 Weeks

    The document states that "Documentation is provided that demonstrates that Synthes Hemostatic Putty is substantially equivalent' to other legally" marketed devices. This implies that the results of these tests were considered satisfactory to demonstrate equivalence, but the specific performance outcomes or the acceptance thresholds are not detailed.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document lists several in vivo and in vitro animal studies.

    • Animal Models Used: Rat, Rabbit, Sheep.
    • Sample sizes for these studies are not specified in the provided text.
    • Data Provenance: Not explicitly stated, though the applicant is Synthes USA Products, LLC, with an address in West Chester, PA, USA. The studies likely originated from contract research organizations or internal labs, but the specific country of origin for the data is not mentioned.
    • Retrospective or Prospective: These would be prospective animal studies, as they involve active experimentation to test the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the evaluation is for a physical medical device (hemostatic bone putty) through non-clinical (animal and lab) studies, not a diagnostic AI device requiring expert consensus or ground truth derived from human interpretation. The "ground truth" would be the biological and physical measurements and observations from the animal models, analyzed by researchers/pathologists. The specific qualifications of those interpreting the animal study results are not provided.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This is not applicable to the non-clinical studies described, which are not based on human-in-the-loop diagnostic interpretations. The evaluation relies on laboratory and in vivo observation methods.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a hemostatic bone putty, not an AI or diagnostic imaging device that would involve human readers or AI assistance.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The device is a hemostatic bone putty, not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For the in vivo studies, the "ground truth" would be established through direct observation, physiological measurements, histology, pathology, and possibly imaging (e.g., to assess bone healing), as performed by qualified personnel in the animal studies.

    • For the resorption rate studies in rats and rabbits, the ground truth would be the measured presence/absence and degradation of the putty.
    • For bone healing studies in rats and sheep, the ground truth would be pathological and histological assessment of bone regeneration and healing.
    • For the hemostasis evaluation in sheep, the ground truth would be direct observation and measurement of bleeding control.
    • For in vitro studies like cytotoxicity, genotoxicity, and sensitization, the ground truth would be laboratory assays according to ISO standards.

    8. The sample size for the training set:

    This is not applicable. The device is a physical medical product, not an AI model requiring a training set. The term "training set" is typically used in the context of machine learning.


    9. How the ground truth for the training set was established:

    This is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1