Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K081598

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-07-03

    (27 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the KFx Suture Lock Tissue Fixation System is for the fixation of soft tissue to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow. Specifically:
    Shoulder: Bankart lesion repairs, SLAP lesion repairs, acromio-clavicular separation repairs, rotator cuff repairs, capsular shift or capsulolabral reconstructions, biceps tenodesis, deltoid repairs.
    Foot and Ankle: Hallux valgus repairs, medial or lateral instability repairs/reconstructions, Achilles repairs/reconstructions, mid-foot reconstructions, metatarsal ligament repair.
    Knee: Medial collateral ligament repairs, lateral ligament repairs, posterior oblique ligament repairs, lliotibial band tenodesis, patellar tendon repairs.
    Hand, Wrist and Elbow: Scapholunate ligament reconstructions, ulnar or radial collateral ligament reconstructions, tennis elbow repair, biceps tendon reattachment.

    Device Description

    The KFx Tissue Fixation System consists of a Nail Bone Anchor with two suture leads pre-loaded in a single use insertion handle and a Bone Screw Anchor (also pre-loaded in a single use delivery handle) used to capture the sutures, eliminating the need to tie suture knots. The KFx Tissue Fixation System also includes re-usable surgical instruments for the implantation of the Nail Bone and Bone Screw Anchors.
    Devices are provided "STERILE"; sterilization is by Ethylene Oxide (EO) gas and provides a sterility assurance level of 10-6.

    AI/ML Overview

    The provided text describes the KFx Suture Lock Nail Bone Anchor, a medical device for tissue fixation. It refers to "Bench test results" that "indicate that the device is safe and satisfies functional performance requirements" and concludes that the device is "substantially equivalent" to a previously cleared predicate device.

    However, the document does not provide specific acceptance criteria or detailed results of a study in a tabular format as requested. It states that the device is substantially equivalent to a previously cleared device (K072063). The "Biocompatibility and Performance Data" section provides a high-level summary of the study findings, but lacks the granular details needed to fill out all the requested information about a specific study proving it meets acceptance criteria.

    Therefore, much of the requested information cannot be extracted directly from the provided text.

    Here is what can be inferred or stated based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Specific criteria not explicitly stated in the document."Bench test results indicate that the device is safe and satisfies functional performance requirements when used as indicated and do not raise new issues of safety or effectiveness."
    Substantial equivalence to predicate device (K072063) in design, technology, construction, and mechanical performance."The test results demonstrate that the modified KFx Suture Lock Nail Bone Anchor is substantially equivalent to the currently marketed predicate device."
    Biocompatibility"The materials used in the KFx Suture Lock Nail Bone Anchor are biocompatible. The same materials are used in a myriad of legally marketed orthopedic devices."

    Study Details

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample size: Not specified.
      • Data provenance: Not specified (bench tests are typically conducted in a controlled lab environment, not on human data in the context of this device). The document refers to "bench test results."
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable as this is a mechanical device evaluated through bench testing, not image analysis or clinical expert review.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable; this would typically be relevant for studies involving human interpretation or clinical endpoints.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done, as this is a mechanical bone anchor device, not an AI or imaging diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable; this is a physical medical device, not a software algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" (or basis for evaluation) appears to be derived from established engineering and biocompatibility standards for medical devices and comparison to a predicate device. For mechanical performance, this would involve quantitative measurements against specified thresholds, but these thresholds are not provided.
    7. The sample size for the training set:

      • Not applicable, as this is not a machine learning model.
    8. How the ground truth for the training set was established:

      • Not applicable, as this is not a machine learning model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K964935

    Validate with FDA (Live)

    Device Name
    SUTURE LOCK
    Date Cleared
    1997-02-25

    (77 days)

    Product Code
    Regulation Number
    878.4930
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Suture Lock is intended to be use to secure soft tissue to bone, by securing the placement of sutures used in surgical procedures. The Suture Lock is indicated for use in Bankart and Rotator Cuff Repair procedures.

    Device Description

    The Suture Lock is a sterile, single-use device manufactured from polyacetal.

    AI/ML Overview

    This is a medical device 510(k) summary for a "Suture Retention Device" called Suture-Lock. It is a physical device, and therefore the request for information on "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of AI/algorithm performance metrics is not applicable. The provided document does not contain information about an AI/algorithm-based device or its performance studies.

    Therefore, I cannot answer the questions regarding acceptance criteria and performance studies in the requested format. This document describes a traditional medical device and its substantial equivalence to a predicate device based on characteristics like materials, intended use, and indications.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1