    K Number
    K130478

    Validate with FDA (Live)

    Date Cleared
    2013-07-26

    (151 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a lumbar intervertebral body fusion device, SUSTAIN® and SUSTAIN® Radiolucent (SUSTAIN®R) Spacers are intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The SUSTAIN® and SUSTAIN® R Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the REVERE®, REVOLVE™ or BEACON® Stabilization Systems.

    When used as a cervical intervertebral body fusion device, the SUSTAIN® and SUSTAIN®R Spacers are intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The SUSTAIN® and SUSTAIN®R Spacers are to be filled with autogenous bone graft material. These devices are intended to be used with supplemental fixation, such as the ASSURE®, PROVIDENCE® or XTEND® Anterior Cervical Plate System.

    When used as a vertebral body replacement device, SUSTAIN® and SUSTAIN® R Spacers are intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The spacers are intended to be used with supplemental spinal fixation systems that have been labeled for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems and anterior screw and rod systems). The interior of the spacer can be packed with bone grafting material. SUSTAIN® and SUSTAIN® R Spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period.

    Device Description

    SUSTAIN® and SUSTAIN® R Spacers are devices that can be used as intervertebral fusion devices or as vertebral body replacement devices. These spacers are available in different shapes and heights to accommodate various surgical approaches and anatomical needs. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to resist expulsion. Each spacer has an axial hole to allow grafting material to be packed inside the spacer.

    These spacers are used to provide structural stability in skeletally mature individuals following discectomy, corpectomy, or vertebrectomy (including partial). Lumbar spacers may be inserted using a posterior, transforaminal, anterior, anterolateral, or lateral lumbar approach. Cervical spacers are inserted using an anterior cervical approach.

    The SUSTAIN® Spacers are made from commercially pure titanium or titanium alloy as specified in ASTM F67, F136, and F1295.

    The SUSTAIN® R Spacers are made from radiolucent peek polymer with titanium alloy or tantalum markers as specified in ASTM F136, F560, F1295, and F2026.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the SUSTAIN® and SUSTAIN® Radiolucent Spacers, which are intervertebral body fusion devices and vertebral body replacement devices. The submission focuses on adding intervertebral indications and making sterile spacers available.

    Here's an analysis of the acceptance criteria and the study based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance ClaimReported Device Performance
    Material Composition (SUSTAIN® Spacers)Made from commercially pure titanium or titanium alloy (ASTM F67, F136, F1295).
    Material Composition (SUSTAIN® R Spacers)Made from radiolucent PEEK polymer with titanium alloy or tantalum markers (ASTM F136, F560, F1295, F2026).
    Lumbar Intervertebral Body Fusion:For DDD at L2-S1, skeletally mature patients, minimum 6 months nonoperative treatment, up to Grade 1 spondylolisthesis/retrolisthesis. Used with autogenous bone graft and supplemental fixation (REVERE®, REVOLVE™, BEACON®).
    Cervical Intervertebral Body Fusion:For DDD at C2-T1 (one level), skeletally mature patients, minimum 6 weeks non-operative treatment. Used with autogenous bone graft and supplemental fixation (ASSURE®, PROVIDENCE®, XTEND®).
    Vertebral Body Replacement:For T1-L5 to replace collapsed/damaged/unstable vertebral body due to tumor/trauma. Used with supplemental spinal fixation systems and bone grafting material. Designed for anterior spinal column support even without fusion.
    Mechanical Testing:Static and dynamic compression, static and dynamic torsion, static and dynamic compression shear, and subsidence. Conducted in accordance with "Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Intervertebral Fusion Device," June 12, 2007.
    Substantial Equivalence:Substantially equivalent to predicate SUSTAIN® and SUSTAIN® R Spacers for design, technical characteristics, performance, and vertebral body replacement indications.
    Equivalent to predicate PATRIOT® Spacers for intervertebral body fusion device indications.
    Safety and Effectiveness:As safe, as effective, and perform as well as or better than predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a clinical study on human subjects with a "test set" for the device's performance. Instead, it describes mechanical testing and a claim of substantial equivalence to predicate devices. Therefore, the concepts of "sample size used for the test set" and "data provenance" in the context of clinical data do not apply here. The "test set" for mechanical performance would be the specific number of devices tested in each mechanical test, but this detail is not provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the submission describes mechanical testing and a comparison to predicate devices, not a study involving human subjects or expert assessment for ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reasons as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This submission is for an implantable spinal device, not an AI-assisted diagnostic or imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is an implantable medical device, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    For the mechanical testing, the "ground truth" would be established by the engineering standards and specifications outlined in the referenced guidance document ("Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Intervertebral Fusion Device, June 12, 2007") and the ASTM standards mentioned for material composition (e.g., ASTM F67, F136). The performance of the device against these objective engineering standards serves as its "ground truth."

    For substantial equivalence, the ground truth is the performance and characteristics of the legally marketed predicate devices (PATRIOT® Spacers, SUSTAIN® Spacers, CALIBER™ Spacers, COALITION® Spacers, BAK/Cervical Interbody Fusion System).

    8. The Sample Size for the Training Set

    This is not applicable, as this is not a machine learning or AI device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reasons as point 8.