Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K063305

    Validate with FDA (Live)

    Date Cleared
    2006-11-29

    (28 days)

    Product Code
    Regulation Number
    874.1820
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used with both the Medtronic Nerve Integrity Monitor (NIM) and XPS Drill Systems, the Stimulating Bur Guard is intended to stimulate cranial and peripheral motor nerves, including spinal nerve roots, with a standard bur / blade for location and identification during the incision and removal of soft and hard tissue or bone.

    The Stimulating Bur Guard is indicated for nerve monitoring during the incision and removal of soft and hard tissue or bone with a standard bur / blade during otology, neurotology, sinus, laryngeal, nasopharyngeal, head and neck, general and plastic, and orthopedic surgical procedures, including spinal applications.

    Device Description

    The Stimulating Bur Guard is a sterile, single use device used in conjunction with the XPS 3000 BF Powered Drill System and the Medtronic Xomed Nerve Monitoring System. It locks onto a handpiece and has an opening on the anterior end to accept a standard bur / blade. The primary purpose of the Stimulating Bur Guard is to deliver the stimulating current from the Nerve Monitoring console to the standard bur / blade on the drill handpiece. The second purpose of the Stimulating Bur Guard is to provide support for the extended length burs.

    AI/ML Overview

    Here is an analysis of the provided 510(k) summary regarding the Stimulating Bur Guard, focusing on acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Generalized from text)Reported Device Performance
    BiocompatibilityCompliant with ISO 10993-1:2003 and FDA G95-1 for external communicating devices with tissue/bone contact < 24 hrs.The device is considered biocompatible and "appropriate tests were performed on all patient contacting materials." Specific results not detailed.
    Shelf Life AgingThree (3) months shelf life required for limited market release; minimum one (1) year at full market release; ultimately achieve four (4) year shelf life."Accelerated aging studies for the Stimulating Guard will ensure three (3) months shelf life prior to a limited market release." (Future projections for full market release and beyond.)
    Operating LifeThe device must meet acceptance criteria for worst-case use during a surgical procedure for single-use."The device met the acceptance criteria for worst case use during a surgical procedure."
    Bench TestingSpecific criteria for: Reliability and Useful life, Bur/blade Compatibility, Current delivery, Current delivery compared to 2.3mm ball tip probe, Noise, Temperature, Current drop out, Electrical Isolation."The device met the acceptance criteria for: Reliability and Useful life, Bur / blade Compatibility, Current delivery, Current delivery compared to 2.3mm ball tip probe, Noise, Temperature, Current drop out, Electrical Isolation." Specific values/results not detailed.
    EMC & Electrical SafetySystem (Stimulating Guard + nerve monitoring + drill system consoles) compliance with EN 60601-1-2 (EMC) and EN / IEC 60601-1 (Electrical Safety)."The devices are evaluated for EMC as a system... in accordance with EN 60601-1-2. Electrical safety testing is conducted in accordance with EN / IEC 60601-1." "Both the XPS 3000 and NIM systems are BF rated."

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify a sample size for a test set in the context of clinical or performance data from human subjects. The studies described are primarily bench testing and biocompatibility testing.

    • Bench Testing: No sample size is explicitly stated for the number of devices or tests performed in areas like reliability, current delivery, or noise.
    • Biocompatibility: No sample size of materials or tests is explicitly stated, beyond "all patient contacting materials."
    • Data Provenance: The studies are described as bench testing and laboratory studies (biocompatibility, shelf life). There's no mention of human subject data, retrospective or prospective, or country of origin for such data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable and not provided in the document. The device is a surgical accessory, and the studies described are technical performance evaluations, not clinical studies requiring expert ground truth for interpretation of outcomes or images.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. Adjudication methods are typically used in clinical studies, especially those involving expert interpretation of data (e.g., imaging studies), which is not the case here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done and is not mentioned. Such studies are generally conducted for diagnostic devices (e.g., AI for medical imaging) to assess the impact of AI on human reader performance, which doesn't apply to this surgical stimulating bur guard.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are essentially standalone performance evaluations of the device's physical and electrical properties. There is no "algorithm" in the sense of AI or software interpreting data for human users. The device itself is an accessory to existing systems (Nerve Integrity Monitor and XPS Drill Systems). The various bench tests directly assess the device's intended function and safety characteristics in a standalone manner (though in conjunction with the specified predicate systems for electrical tests).

    7. The Type of Ground Truth Used

    The "ground truth" for the various tests described would be based on:

    • Bench Test Specifications: Pre-defined engineering specifications, electrical standards, and performance limits for parameters like current delivery, noise, temperature, and reliability.
    • Regulatory Standards: Compliance with ISO and FDA biocompatibility standards.
    • Manufacturer's Internal Standards: For shelf life and operating life tests, likely based on internal validation protocols.

    Essentially, the ground truth is derived from objective engineering and regulatory standards, rather than clinical outcomes, pathology, or expert consensus on clinical data.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. This device does not use an "algorithm" or "AI" that requires a training set. The descriptions relate to traditional device verification and validation.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable and not provided as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1