510(k) Data Aggregation

    K Number
    K111519

    Validate with FDA (Live)

    Date Cleared
    2012-10-03

    (489 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    STERRAD® SEALSURE® Chemical Indicator Tape (PN 14202) is a process indicator (ISO 11140-1:2005) intended for use by healthcare providers to secure non-woven sterilization packs and wraps to be sterilized in the STERRAD® Sterilization Systems. The color of the STERRAD® SEALSURE® Chemical Indicator Tape changes from red to gold (or lighter) when exposed to hydrogen peroxide and is intended to differentiate between processed and unprocessed loads.

    Device Description

    STERRAD® SEALSURE® Chemical Indicator Tape is a process indicator intended for use by healthcare providers to secure non-woven sterilization packs and wraps to be sterilized in the STERRAD® Sterilization Systems. The color of the STERRAD® SEALSURE® Chemical Indicator Tape changes from red to gold (or lighter) when exposed to hydrogen peroxide and is intended to differentiate between processed and unprocessed loads.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the STERRAD® SEALSURE® Chemical Indicator Tape, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Chemical Indicator FunctionalityPassed
    Tape Adhesion StrengthPassed
    End Point / Post Processing Color StabilityPassed
    BiocompatibilityPassed
    Shelf lifePassed

    Study Details

    • Sample Size for Test Set: Not explicitly stated in the provided text.
    • Data Provenance (Test Set): Not explicitly stated in the provided text.
    • Number of Experts for Ground Truth (Test Set): Not applicable. This device is a chemical indicator, not an AI or diagnostic device requiring expert interpretation for ground truth.
    • Qualifications of Experts (Test Set): Not applicable.
    • Adjudication Method: Not applicable.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No, this type of study was not conducted as the device is a chemical indicator, not a diagnostic or AI-assisted tool for human readers.
    • Effect Size of Human Readers with vs. without AI Assistance: Not applicable.
    • Standalone (Algorithm Only) Performance Study: Not applicable. The device's performance is intrinsically linked to its chemical reaction, not an algorithm.
    • Type of Ground Truth Used: The ground truth for the "Chemical Indicator Functionality" and "End Point / Post Processing Color Stability" would be based on the intended color change (red to gold or lighter) upon exposure to a specific sterilization cycle (hydrogen peroxide). "Tape Adhesion Strength" and "Biocompatibility" would have their own established pass/fail criteria from material science and safety testing. "Shelf life" would be determined by remaining functional over time.
    • Sample Size for Training Set: Not applicable. This is a physical chemical indicator, not an AI model that requires a training set.
    • How Ground Truth for Training Set was Established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K103219

    Validate with FDA (Live)

    Date Cleared
    2011-04-01

    (151 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    STERRAD® SEALSURE® Chemical Indicator Tape is a process indicator intended for use by healthcare providers to secure non-woven sterilization packs and wraps to be sterilized in the STERRAD® Sterilization Systems:

    MODELCYCLE
    STERRAD® 100SStandard
    STERRAD® 50Standard
    STERRAD® 200Standard
    STERRAD® NX™Standard
    Advanced
    STERRAD® 100NX™Flex
    EXPRESS

    The color of the STERRAD® SEALSURE® Chemical Indicator Tape changes from red to gold (or lighter) when exposed to hydrogen peroxide and is intended to differentiate between processed and unprocessed loads.

    Device Description

    STERRAD® SEALSURE® Chemical Indicator Tape is a process indicator intended for use by healthcare providers to secure non-woven sterilization packs and wraps to be sterilized in the STERRAD® Sterilization Systems.

    The color of the STERRAD® SEALSURE® Chemical Indicator Tape changes from red to gold (or lighter) when exposed to hydrogen peroxide and is intended to differentiate between processed and unprocessed loads.

    AI/ML Overview

    The provided 510(k) summary describes a chemical indicator tape, not a software device or an AI/ML enabled device. Therefore, the typical acceptance criteria and study designs for AI/ML medical devices, such as those related to reader studies, ground truth establishment by experts, and training/test set details, are not applicable here.

    However, I can extract the information relevant to the performance of this specific device based on the provided text.

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, adapted to the context of a chemical indicator tape:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Study Performed)Reported Device Performance
    Chemical Indicator Functionality PassedPassed
    Tape Adhesion StrengthPassed
    End Point / Post Processing Color StabilityPassed
    BiocompatibilityPassed
    Shelf lifePassed

    Details of the Study for the Chemical Indicator Tape:

    • Device Under Test: STERRAD® SEALSURE® Chemical Indicator Tape
    • Purpose of the Device: To function as a process indicator, changing color from red to gold (or lighter) when exposed to hydrogen peroxide within specific STERRAD® Sterilization Systems, differentiating between processed and unprocessed loads.

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document does not specify sample sizes for the individual tests (e.g., how many tapes were tested for adhesion, how many cycles for functionality).
    • Data Provenance: The studies were conducted internally by Advanced Sterilization Products ("Performance testing was conducted to demonstrate the functionality..."). No information on geographical origin of data or retrospective/prospective nature is provided, though given it's a new product, it would inherently be prospective testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • This is not applicable as the "ground truth" for a chemical indicator is based on physical and chemical reactions (color change indicating exposure to sterilant, adhesion properties, stability) which are objectively measured, not subject to expert interpretation in the way medical images are.

    4. Adjudication Method for the Test Set:

    • Not applicable. The "results" are objective measurements (e.g., color change, adhesion strength), not subjective human interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Not applicable. This type of study is for evaluating human performance, often with AI assistance, in interpreting medical data. This device is a physical indicator.

    6. Standalone Performance (Algorithm Only):

    • Not applicable. This device is not an algorithm. Its "standalone" performance is its intrinsic physical and chemical properties as a process indicator. The nonclinical tests ("Chemical Indicator Functionality," "Tape Adhesion Strength," etc.) essentially describe its standalone performance.

    7. Type of Ground Truth Used:

    • The ground truth is based on the physical and chemical properties of the indicator tape:
      • Functionality: Successful color change from red to gold (or lighter) upon exposure to hydrogen peroxide in specific STERRAD® sterilization cycles.
      • Adhesion: The tape maintaining its adhesive properties.
      • Stability: The color change remaining stable post-processing.
      • Biocompatibility: The material being safe for its intended use (no adverse biological reactions).
      • Shelf Life: The device maintaining its properties over time.
      • These are verified through laboratory testing against established physical and chemical standards/expectations for such indicators.

    8. Sample Size for the Training Set:

    • Not applicable. This device is not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable.

    Overall Conclusion from the Provided Text:

    The STERRAD® SEALSURE® Chemical Indicator Tape underwent a series of nonclinical tests to demonstrate its functionality, adhesion, color stability, biocompatibility, and shelf life. All tests "Passed," indicating that the device met its pre-defined performance requirements. The results support the claim that the device is "as safe and as effective for differentiating processed from unprocessed packages" as its predicate devices, leading to a finding of substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022441

    Validate with FDA (Live)

    Date Cleared
    2002-10-08

    (75 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    STERRAD® SealSure™ Chemical Indicator Tape is a process indicator intended for use by health care providers to secure non-woven sterilization packs and wraps to be sterilized in the STERRAD® Sterilization System.

    The color of the STERRAD® SealSure™ Chemical Indicator Tape changes from red to gold (or lighter) when exposed to hydrogen peroxide and is intended to differentiate between processed and unprocessed loads. Reference the color comparator bars on the package box label.

    Device Description

    STERRAD® SealSure™ Chemical Indicator Tape (Class A per EN867-1 and Class 1 ISO 11140-1) is a through-put process indicator tape to be used with ASP's STERRAD® Sterilization System. STERRAD® SealSure™ Chemical Indicator Tape offers an additional way to verify processing in the sterilization cycle. It should be used in addition to, not in place of, the biological indicator. STERRAD® SealSure™ Chemical Indicator Tape does not signify sterilization; it only indicates that the indicator has been exposed to hydrogen peroxide.

    There are six rolls of STERRAD® SealSure™ Chemical Indicator Tape per box. Each roll contains 55 meters (60 yards) of indicator tape.

    AI/ML Overview

    The STERRAD® SealSure™ Chemical Indicator Tape is a process indicator designed to confirm exposure of medical devices to hydrogen peroxide within the STERRAD® Sterilization System. The device changes color from red to gold (or lighter) when exposed to hydrogen peroxide, differentiating between processed and unprocessed loads.

    Here's an analysis of its acceptance criteria and the studies conducted:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance MetricReported Device Performance
    Functionality: Color change (red to yellow/gold) at half-cycle conditions.All tape samples changed from red to yellow at half-cycle conditions in each STERRAD® Sterilization System.
    Resistance to Chemicals: Adequate resistance to various chemicals.Demonstrated adequate resistance to chemicals, with the exception of organic solvents with high polarity and strong acids.
    Differentiation of Processed vs. Unprocessed Loads: Ability to distinguish.Ability to distinguish between a processed load and a load where the cycle was canceled prior to hydrogen peroxide exposure.
    Shipping Stability: Maintenance of stability in a simulated shipping environment.Maintained stability in a simulated shipping environment.
    Shelf-Life Color Stability (Unopened Package): Pre-processed color and functionality, Post-processed color.Pre-processed samples maintained pre-processing color and functionality at both ambient and accelerated aging conditions for four months. Post-processed tape samples retained their post-processing color after aging in both ambient and accelerated aging conditions for two months. (Note: functionality stability for post-processed tapes not explicitly mentioned here, but indicated for pre-processed).
    Shelf-Life Color and Functionality Stability (Opened Package): Pre-processed color and functionality, Post-processed color.Pre-processed tape samples' color and functionality stability were maintained in both simulated hospital sterilizer room and central supply storage room ambient conditions for one month. Post-processing color was stable for one month. (Note: pre-processed studies will continue for a total of two months).
    Impact on Sterilization Efficacy: No significant effect on the performance of the STERRAD® Sterilization System when using an extreme amount of tape.The effect on sterilization efficacy of each STERRAD® System was insignificant, regardless of using an extreme amount of chemical indicator tape on each validation tray. Complete inactivation of challenge microorganisms was achieved in each system.
    Adhesion Strength and Stability: Meeting minimum adhesion requirements and maintaining them over time.Meets minimum adhesion strength requirements. Stability testing demonstrated maintenance of minimum adhesion strength over three months in accelerated and room temperature aging environments.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific numerical sample size for each test set (e.g., how many rolls of tape were tested for functionality, or how many trays were used for efficacy testing). It generally refers to "all tape samples," "each of the STERRAD® Sterilization Systems," and "an extreme amount of STERRAD® SealSure™ Chemical Indicator Tape on each validation tray."

    The data provenance is not explicitly stated in terms of country of origin. Given the applicant is "Advanced Sterilization Products Division of Ethicon, Inc. 33 Technology Drive Irvine, CA 92618" in the United States, it is highly likely the studies were conducted in the US. The studies appear to be prospective as they describe controlled laboratory-style testing performed to validate the device prior to marketing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    The document does not mention the use of human experts to establish ground truth for the chemical indicator's performance. The ground truth for chemical indicators like this is typically based on the physical and chemical properties of the indicator itself, its interaction with the sterilant, and established industry standards (such as EN867-1 and ISO 11140-1, which are referenced). The "ground truth" here is objective chemical reaction and physical performance, not subjective interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. The device's performance (color change, adhesion, resistance) is based on objective measurements and established chemical/physical criteria, rather than human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically relevant for interpretative devices where human readers assess cases (e.g., radiology images) and compare their performance with and without AI assistance. The STERRAD® SealSure™ Chemical Indicator Tape is an objective process indicator, not an interpretative device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    The studies described are inherently "standalone" in the sense that the device's performance (color change, adhesion, etc.) is tested directly without human interpretation in the loop. The "algorithm" here is the chemical reaction of the indicator itself.

    7. The Type of Ground Truth Used

    The ground truth used for these studies is based on:

    • Objective Chemical Reaction: The known and expected color change in response to hydrogen peroxide exposure.
    • Physical Properties: Measurable characteristics like adhesion strength, stability over time, and resistance to chemical degradation.
    • Microbiological Efficacy (for impact on sterilizer): The complete inactivation of challenge microorganisms, which is a definitive, objective measure of sterilization success.
    • Industry Standards: Compliance with standards like Class A per EN867-1 and Class 1 ISO 11140-1 provides a framework for defining acceptable performance.

    8. The Sample Size for the Training Set

    Not applicable. This device is a passive chemical indicator, not an AI or machine learning algorithm that requires a "training set." Its function is based on a predetermined chemical formulation.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.