Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K023290

    Validate with FDA (Live)

    Date Cleared
    2003-04-03

    (183 days)

    Product Code
    Regulation Number
    880.6860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Premarket Notification is to expand the Indications for Use for the STERRAD 50 Sterilizer (K981625) and the STERRAD 100S Sterilizer (K991999) to include titanium as a compatible material and to expand the lumen sizes to include medical devices with only a single stainless steel lumen in the following configurations:

    • An inside diameter of 1 mm or larger and a length of 125 mm or shorter
    • An inside diameter of 2 mm or larger and a length of 250 mm or shorter .

    The validation testing for these two new lumen sizes was conducted using a maximum of 10 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing.

    For information pertaining to previously cleared materials and lumen sizes, please reference K981625 and K991999.

    Device Description

    The STERRAD® 50 & 100S Sterilizers are self-contained stand-alone systems of hardware and software designed to sterilize medical instruments and devices using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer bowl where the solution is heated and transformed into a vapor, introducing the vapor into the process chamber under negative pressure and transforming the vapor into a gas plasma with radio frequency (RF) electrical energy.

    The equipment (hardware and software) for the STERRAD® 50 & 100S Sterilizers is the same as that of the predicate devices. The hardware consists of a sterilization chamber onto which is mounted a variety of instruments and components, housed in a covered frame. The system also uses accessories such as disposable sterilant cassettes, reusable instrument trays, printer paper and ink cartridges.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies conducted for the STERRAD® 50 and STERRAD® 100S Sterilization Systems, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Goal)Reported Device Performance (STERRAD® 50 & 100S)
    Sterilizing medical devices with titanium surfacesAchieved a Sterility Assurance Level (SAL) of at least 10^-6 for medical device surface sterilization in the complete sterilization process.
    Sterilizing stainless steel lumens:
    - 1 mm inside diameter (ID) or larger and 125 mm or shorterAchieved an SAL of 10^-6 in a full sterilization cycle. SLRs for simulated use (unwashed/washed) were 6.1/6.0 (STERRAD 50) and 6.1/6.2 (STERRAD 100S).
    - 2 mm inside diameter (ID) or larger and 250 mm or shorterAchieved an SAL of 10^-6 in a full sterilization cycle. SLRs for simulated use (unwashed/washed) were 6.9/6.9 (STERRAD 50) and 7.0/6.9 (STERRAD 100S).
    Sterilizing lumens in Tyvek-Mylar pouchesAchieved an SAL of 10^-6 in a full sterilization cycle for stainless steel lumens (1 mm ID x 125 mm long and 2 mm ID x 250 mm long) packaged in STERRAD Sterilization pouches, validating the use of these pouches.
    Sterilization efficacy with organic/inorganic challenge (simulated use)The process was not affected by the presence of an organic and inorganic challenge, as indicated by the achieved SALs and SLR (Sterility Log Reduction) values (e.g., SLR 6.1 to 7.0 across various lumen sizes and conditions, demonstrating effective sterilization even with soil).
                                                                       The In-Use testing demonstrated that the STERRAD® 50 and 100S Sterilization Systems were effective sterilizers of stainless steel lumen devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
    

    | No cytotoxicity or toxic residuals on processed materials | Cytotoxicity test results indicated that titanium materials processed did not induce in vitro cytotoxicity. Residues testing showed no toxic sterilant residuals. |

    Note: SAL of 10^-6 means there is a one in a million chance of a non-sterile unit. SLR values of 6.0 or higher indicate a 6-log reduction in microbial population, which is a common standard for sterilization.

    2. Sample Size and Data Provenance

    • Test Set Sample Sizes:
      • Titanium Surface Sterilization: Not explicitly stated, but performed with "titanium materials representative of the materials commonly used in re-usable medical devices."
      • Lumen Sterilization: Not explicitly stated, but for the validation loads, a "maximum of 10 lumens per load" was used for the 1mm x 125mm and 2mm x 250mm lumens. The half-cycle validation tests used sufficient materials to demonstrate the SAL.
      • Tyvek-Mylar Pouched Device Sterilization: Not explicitly stated, but conducted with 1mm x 125mm and 2mm x 250mm stainless steel lumens in pouches.
      • Simulated Use Testing: Not explicitly stated how many devices, but used "devices representative of lumen claims."
      • In-Use Sterility Testing: Not explicitly stated how many devices, but used "Devices representative of lumen claims for the STERRAD® 50 Sterilization System" that were "used in routine surgeries at local hospitals."
    • Data Provenance: Retrospective and prospective. The In-Use Sterility Testing involved devices "used in routine surgeries at local hospitals," suggesting data collected from real-world usage scenarios (retrospective element) which were then transported for prospective testing with the STERRAD systems. The other validation testing appears to be prospectively designed lab studies. No specific country of origin is mentioned beyond "local hospitals" for the in-use testing, implying a US-based context given the FDA submission.

    3. Number of Experts and Qualifications for Ground Truth

    • This information is not provided in the summary. The ground truth for sterilization is typically established through microbiological testing (e.g., demonstrating the death of a highly resistant test organism), rather than expert interpretation of images or clinical outcomes.

    4. Adjudication Method for the Test Set

    • This information is not applicable as the studies are primarily microbiological challenge tests for sterilization efficacy, not human interpretation tasks requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was mentioned or performed. The device is a sterilizer, meaning its performance is assessed by its ability to render medical devices sterile, not by human interpretation or diagnosis. Therefore, the concept of human readers improving with or without AI assistance is not relevant to this type of device.

    6. Standalone Performance Study

    • Yes, standalone performance studies were done. All the validation tests described (Surface Sterilization, Lumen Sterilization, Tyvek-Mylar Pouched Device Sterilization, Simulated Use Testing, In-Use Sterility Testing, Toxicity Testing) are assessing the algorithm/system-only performance of the STERRAD® 50 and 100S sterilizers without human intervention in the sterilization process itself. The "overkill" approach used in the validation testing directly assesses the system's ability to achieve a robust sterility assurance level.

    7. Type of Ground Truth Used

    • The primary ground truth used is microbiological evidence of sterility, specifically:
      • Inactivation of B. stearothermophilus endospores: This is a highly resistant bacterial spore used as a biological indicator to challenge sterilization processes. Its inactivation to a specific Sterility Assurance Level (SAL) (e.g., 10^-6) or log reduction (e.g., 6-log reduction) serves as the ground truth for effective sterilization.
      • Sterility Assessment: Direct assessment of the absence of viable microorganisms after processing.
      • Toxicity testing: Used to establish ground truth that no cytotoxic effects or toxic residuals remain.

    8. Sample Size for the Training Set

    • This information is not applicable. The STERRAD® Sterilization System is a physical-chemical sterilization device, not an AI/ML algorithm that is "trained" on data in the traditional sense. Its operation is based on pre-defined scientific principles of hydrogen peroxide gas plasma sterilization, not on learning from a dataset.

    9. How Ground Truth for the Training Set Was Established

    • This information is not applicable for the reasons stated above. The system's "operating parameters" or "settings" are developed through scientific and engineering principles, not through a "training set" with established ground truth. The validation studies described in the summary are designed to prove the device meets performance criteria, not to train it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K981625

    Validate with FDA (Live)

    Date Cleared
    1999-01-11

    (249 days)

    Product Code
    Regulation Number
    880.6860
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STERRAD® 50 Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture. (See list of recommended Materials in Section 3 of the Operator's Manual.)

    The STERRAD® 50 Sterilizer can sterilize instruments which have diffusionrestricted spaces, such as the hinged portion of forceps and scissors.

    Metal and nonmetal lumened instruments with inside diameters of 6 mm (¼ inch) or larger and lengths of 310 mm (12 inches) or shorter can be processed in the STERRAD® 50 Sterilizer. (See list of recommended Materials in Section 3 of the Operator's Manual.)

    Additionally, medical devices with only a single stainless steel lumen which has an inside diameter of 3 mm (1/8 inch) or larger and a length of 400 mm (15 ¾ inches) or shorter can be processed in the STERRAD® 50 Sterilizer.

    Device Description

    The STERRAD® 50 Sterilizer is a self-contained stand-alone system of hardware and software designed to sterilize certain medical instruments and devices, using a patented hydrogen peroxide gas plasma process. Sterilization is accomplished by injecting aqueous hydrogen peroxide into the vaporizer bowl where the solution is heated and transformed into a vapor, introducing this vapor into the process chamber under negative pressure and transforming the vapor into a gas plasma with RF electrical energy. The technology is particularly suited to the sterilization of heat and moisture sensitive instruments.

    The hardware consists of a sterilization chamber onto which is mounted a variety of instruments and components, housed in a covered frame. The system also uses accessories such as disposable sterilant cassettes, reusable instrument trays, single-use biological and chemical indicators, printer paper and ink cartridges.

    The sterilizer sits on a cart. The cart is designed for transportation of the sterilizer to its installation site, for mobility during servicing and as a structural support device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the STERRAD® 50 Sterilizer, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for a sterilizer are generally defined by achieving a certain Sterility Assurance Level (SAL), typically 10⁻⁶, meaning a 1 in a million chance of a non-sterile unit. The studies consistently aimed to demonstrate this SAL.

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Dose-Response Relationship (Hydrogen Peroxide)Achieve SAL of 10⁻⁶ for various materials.Achieved SAL of 10⁻⁶ for all materials tested with 360 µL or greater injection volume under half-cycle conditions.
    Surface Sterilization (Medical Device Materials)Achieve SAL of 10⁻⁶ for various substrate materials.Achieved SAL of at least 10⁻⁶ for medical device surface sterilization with 720 µL and 1080 µL injection volumes under less than half-cycle conditions.
    Mated Surfaces SterilizationAchieve SAL of 10⁻⁶ for mated surfaces.Achieved SAL of 10⁻⁶ for mated surface sterilization with no spore survivors observed under half-cycle conditions.
    Lumen SterilizationAchieve SAL of 10⁻⁶ for lumens (3 mm x 400 mm).Achieved SAL of 10⁻⁶ for lumens with no spore survivors after half-cycle and modified Total-kill Endpoint tests.
    Tyvek-Mylar Pouched Device SterilizationAchieve SAL of 10⁻⁶ for pouched devices with lumens.Achieved SAL of 10⁻⁶ with no spore survivors observed under half-cycle conditions for 3 mm x 400 mm stainless steel lumens in Tyvek-Mylar pouches.
    Bacteriostasis Testing (Processed Carriers)No bacteriostatic effect from processed carriers on B. stearothermophilus outgrowth.No bacteriostatic effect indicated; all test carriers/materials demonstrated desired outgrowth within 14 days.
    Sporicidal Microbiological TestingDemonstrate sporicidal activity against B. subtilis and Clostridium sporogenes.No growth observed on any carriers (silk suture loops, penicylinders) for B. subtilis and Clostridium sporogenes.
    In-Use Sterility TestingSuccessfully sterilize actual surgical instruments.Successfully sterilized actual surgical instruments (open surfaces, mated/hinged surfaces, 3mm x 400mm lumens) from clinical use.
    In-Use Bacteriostasis and Fungistasis TestNo bacteriostatic or fungistatic effects on processed instruments.No bacteriostatic or fungistatic effects seen on stainless steel open, mated/hinged, and 3mm x 400mm lumened instruments when challenged with Cl. sporogenes, Candida albicans, or B. subtilis.
    Simulated Use Testing (Organic/Inorganic Challenge)Maintain effectiveness even with improper washing; achieve 6.1 log reduction with proper washing.Sterilizer minimally affected by organic and inorganic challenge; effective at inactivating highly resistant spores in diffusion-restricted environments even without proper washing. Achieved 6.1 log reduction with proper washing.
    Toxicity Testing of Processed MaterialsNo toxic sterilant residuals on processed materials.Cytotoxicity and in vivo biocompatibility testing showed no toxic sterilant residuals.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Test Organism: B. stearothermophilus spores (at least 1 x 10⁶ endospores for most tests).
      • Materials: Various materials representative of medical devices (e.g., metallic, polymer carriers, stainless steel). Specific quantities are not provided, but tests involved "various materials," "inoculated spore carriers," "mated biological indicator carriers," "stainless steel lumens," and "Tyvek pouched stainless steel lumens (Ten)."
      • Data Provenance: The studies were conducted by Advanced Sterilization Products (ASP), a division of Johnson & Johnson Medical, Inc., in Irvine, CA, USA, as part of their 510(k) submission. This indicates a retrospective evaluation of the device's performance through laboratory testing. "In-Use Sterility Testing" involved devices used in routine surgeries at a local hospital, then transported to ASP for sterilization, making this aspect prospective for the sterilization step with used instruments.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The ground truth for all sterilization studies is established by microbiological culture for the presence or absence of viable spores. The text does not mention the number or specific qualifications of human experts (e.g., microbiologists) used to perform or interpret these culture results. However, the methodology aligns with standard microbiological validation practices for sterilizers.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable in the conventional sense for a medical imaging or diagnostic device. The "ground truth" is determined by microbiological growth/no growth, which is an objective measurement based on established protocols. There is no mention of expert consensus or adjudication for these microbiological results.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a sterilization device, not an AI-assisted diagnostic or imaging system. Therefore, MRMC studies and AI-related effect sizes are not applicable.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable as this is a sterilization device without an AI algorithm component. The performance evaluated is that of the sterilizer itself.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The primary ground truth used is microbiological culture data (presence/absence of B. stearothermophilus growth or other specified organisms). This is a direct measurement of sterility assurance. In the case of toxicity testing, "cytotoxicity and in vivo biocompatibility testing" provided the ground truth for residual toxicity.
    7. The sample size for the training set:

      • Not applicable. This is a physical device (sterilizer), not a machine learning model. There is no concept of a "training set" in the context of validating its sterilization efficacy.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this type of device validation.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1