Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K963738

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-05-23

    (248 days)

    Product Code
    Regulation Number
    870.3680
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ELA Medical endocardial leads are designed to be used with implantable cardiac pacemakers. ELA Medical Stela ™ Model BJ44 and BJ45 bipolar pacing leads are intended for atrial implantation only.

    Device Description

    Stela™ BJ44 and BJ45 models are silicone rubber, tined, J-shaped, bipolar transvenous leads that provide a permanent electrical pathway between a pacemaker and the atrium. They are similar in design and construction to other such leads in commercial distribution.

    AI/ML Overview

    The provided text describes the Stela™ Models BJ44 and BJ45 pacing leads. Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Test/Performance MetricPass/Fail CriterionReported Device Performance
    In-Vitro Functional TestingStylet Insertion/RemovalMet established criterionAll test results met the established pass/fail criterion.
    Stylet BottomingMet established criterionAll test results met the established pass/fail criterion.
    Electrical ResistanceMet established criterionAll test results met the established pass/fail criterion.
    IS-1 Connector ConformityMet established criterionAll test results met the established pass/fail criterion.
    Leak ResistanceMet established criterionAll test results met the established pass/fail criterion.
    Distal Tip Electrode/Conductor Coil Tensile StrengthMet established criterionAll test results met the established pass/fail criterion.
    Insulation IntegrityMet established criterionAll test results met the established pass/fail criterion.
    Conductor Coil FatigueMet established criterionAll test results met the established pass/fail criterion.
    Tip FlexibilityMet established criterionAll test results met the established pass/fail criterion.
    BiocompatibilityBiocompatibilityNot applicable (successful history of materials)Not performed due to successful history with same materials in other lead products.
    Sterilization & PackagingSterilization TestingMet established criterionPerformed, and all test results met the established pass/fail criterion.
    Mechanical/Environmental Packaging ValidationMet established criterionPerformed, and all test results met the established pass/fail criterion.
    Clinical PerformancePacing CharacteristicsAppropriate pacing characteristicsExhibit appropriate pacing characteristics.
    Sensing CharacteristicsAppropriate sensing characteristicsExhibit appropriate sensing characteristics.
    ComplicationsNo complications observedNo complications were observed.

    2. Sample Size Used for the Test Set and Data Provenance:

    • In-vitro functional testing, sterilization, and packaging validation: The text states "All test results demonstrated that the established pass / fail criterion was met in all cases." However, it does not specify the sample size for these in-vitro tests.
    • Clinical study: Described as a "limited clinical study." The exact sample size is not provided, only that it involved the Model BJ45 lead.
    • Data Provenance:
      • In-vitro tests: Likely conducted in a lab environment by the manufacturer (ELA Medical).
      • Clinical study: The text indicates the device was "recently introduced into commercial distribution outside the U.S." and that "No unanticipated adverse device effects have been reported for these leads" from this OUS experience. The limited clinical study mentioned is likely related to this OUS experience or a precursor to it. It is a prospective study for the clinical data from the "limited clinical study."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Not Applicable. The described studies are primarily in-vitro engineering tests and a limited clinical study evaluating device performance, not diagnostic accuracy requiring expert panel review for ground truth.

    4. Adjudication Method:

    • Not Applicable. See point 3.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. An MRMC study was not described. This device is a medical lead, not an imaging or diagnostic AI tool that would typically undergo an MRMC study.

    6. Standalone Performance:

    • Partially. The in-vitro functional testing, sterilization, and packaging validation can be considered standalone performance assessments of the device itself (algorithm only is not applicable here as it's a hardware device).
    • The "limited clinical study" also assesses the device's performance in patients in a standalone manner (without another human-in-the-loop comparison).

    7. Type of Ground Truth Used:

    • In-vitro functional testing: The ground truth is based on engineering specifications and established pass/fail criteria for physical and electrical properties.
    • Clinical study: The ground truth for pacing and sensing characteristics would be derived from physiological measurements and clinical observations in patients. The absence of complications serves as a negative ground truth.

    8. Sample Size for the Training Set:

    • Not Applicable. The description does not involve a "training set" in the context of machine learning or AI. This is a hardware device.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1