Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K232075

    Validate with FDA (Live)

    Date Cleared
    2024-02-23

    (226 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood, and neonatal heel stick specimens throughout all hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy.

    The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonatal heel stick.

    It is not intended for use with neonate cord blood specimens.

    It is not intended for the screening or diabetes mellitus but is indicated for use in determining dysglycemia.

    The StatStrip Glucose Hospital Meter System includes the following components:

    • StatStrip Glucose Hospital Meter
    • StatStrip Glucose Test Strips
    Device Description

    StatStrip Glucose Hospital Meter System:
    The StatStrip Glucose Hospital Meter System is a hand-held testing device that works in conjunction with the StatStrip Glucose Test Strips to measure glucose in a whole blood sample. Meter operation is selfprompting using an illuminated color touch-screen Graphical User Interface (GUI).

    StatStrip Glucose Test Strips:
    The test strips contain a reaction layer that contains a glucose-enzyme (greater than 1.0 IU) and ferricyanide as a mediator. The test strip is touched to a drop of blood to initiate the test process. The strip is designed such that when a drop of blood is touched to the strip, the blood is drawn into the reaction space via capillary action. A simple one-step provides a blood glucose result. Test strips will be sold in vials of 25 strips.

    StatStrip Glucose Control Solutions:
    The control solutions are aqueous assayed solutions containing buffered D-Glucose, viscosity-adjusting agent, preservatives and other non-reactive ingredients (dye). They contain no products of human origin. There are three levels of controls, (Level 1, Level 2 and Level 3). These solutions will be offered for sale separately from the meter.

    StatStrip Glucose Linearity Solutions:
    There are 5 levels of Linearity solutions containing buffered D-Glucose, viscosity-adjusting agent, preservatives and other non-reactive ingredients (dve). They contain no products of human origin. These solutions are offered separately from the system for users to verify the performance of the system.

    Charging (Docking) Station:
    The meter charging station is a stationary accessory used to recharge the meter. The charging station has one slot for the meter to be placed and charged wirelessly. The charging station should be located central to the patient care area being served by the meter (e.g., a nursing station). The data charging station must remain plugged in to a wall outlet for power.

    The system still allows the charging station to be used to transfer data from the meter to a central workstation and allow meter setup information to be downloaded from the central workstation to the meter.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the StatStrip Glucose Hospital Meter System (K232075), based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from the FDA Guidance for POC Device Acceptance Criteria (Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use, September 29, 2020). The performance of both the proposed and predicate devices is reported against these criteria.

    Acceptance Criteria (FDA Guidance)Proposed Device Performance (< 75 mg/dL)Predicate Device Performance (< 75 mg/dL)Proposed Device Performance (≥ 75 mg/dL)Predicate Device Performance (≥ 75 mg/dL)
    Within ± 5 mg/dL33/36 (91.7%)32/36 (88.9%)--
    Within ± 10 mg/dL36/36 (100%)36/36 (100%)--
    Within ± 12 mg/dL36/36 (100%)36/36 (100%)--
    Within ± 15 mg/dL36/36 (100%)36/36 (100%)--
    Within ± 5 %--205/264 (77.3%)211/264 (80.7%)
    Within ± 10 %--263/264 (99.6%)263/264 (99.6%)
    Within ± 12 %--264/264 (100%)264/264 (100%)
    Within ± 15 %--264/264 (100%)264/264 (100%)

    Linear Regression Values:

    ConditionSlopeInterceptCorrelation Coefficient (r)
    Proposed Device vs. YSI0.97995.23570.9977
    Predicate Device vs. YSI0.98485.21640.9984
    Proposed Device vs. Predicate Device0.99350.35710.9978

    Precision Data (Linearity Solutions):

    Linearity LevelsPredicate Device Mean (SD/%CV)Proposed Device Mean (SD/%CV)
    Level 120.2 (0.96)20.4 (0.85)
    Level 262.2 (1.61)62.4 (1.76)
    Level 3116.4 (2.43)116 (2.58)
    Level 4300.8 (2.71)301.5 (2.40)
    Level 5505.6 (3.04)508.1 (3.78)

    Precision Data (Venous Whole Blood Samples):

    YSI MeanPredicate Device Mean (SD/%CV)Proposed Device Mean (SD/%CV)
    4949.7 (2.75)49.1 (2.32)
    8890.1 (2.79)88.8 (3.36)
    156154.4 (3.66)156.1 (3.23)
    263263.9 (3.05)265.9 (3.09)
    371364.0 (3.04)363.0 (2.69)
    458456.5 (2.80)453.0 (2.01)
    531535.9 (2.37)532.1 (1.73)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Accuracy Study): 100 deidentified venous whole blood samples with glucose concentrations ranging from 20-600 mg/dL.
    • Sample Size (Method Comparison and Precision Study - Venous Whole Blood): Seven (7) levels of deidentified venous whole blood samples, each tested 60 times (implied N for mean calculations in the table).
    • Sample Size (Method Comparison and Precision Study - Linearity Solutions): Five (5) levels of linearity solutions, each tested 60 times.
    • Data Provenance: "deidentified venous whole blood samples from consented donors." The country of origin is not specified, but it implies a prospective collection for the purpose of the study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth was established using a "reference analyzer" (YSI), not by human experts for this device. Therefore, this question is not applicable in the context of the provided document.


    4. Adjudication Method for the Test Set

    Not applicable. The ground truth was established by a reference analyzer (YSI), not through human adjudication.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a blood glucose meter, not an AI-assisted diagnostic imaging device or a system requiring human interpretation. The study focuses on the accuracy and precision of the meter itself compared to a reference method and a predicate device.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies conducted (Accuracy, Method Comparison, and Precision) are standalone performance evaluations of the device (meter and test strips) and its algorithm. The device measures glucose levels directly from blood samples and outputs a quantitative result without human-in-the-loop interpretation.


    7. The Type of Ground Truth Used

    The ground truth for the glucose measurements was established using a reference analyzer, specifically referred to as "YSI" in the tables and descriptions.


    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning. The studies described are performance validation studies for a medical device. Therefore, information about a training set for an AI algorithm is not provided as it's not relevant to this type of device submission.


    9. How the Ground Truth for the Training Set Was Established

    As noted above, a "training set" for an AI algorithm is not discussed in the document for this device. The ground truth for the performance validation studies was established using a reference analyzer (YSI).

    Ask a Question

    Ask a specific question about this device

    K Number
    K181043

    Validate with FDA (Live)

    Date Cleared
    2018-07-12

    (84 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, neonate arterial whole blood, and neonate heel stick specimens throughout all hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy.

    The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.

    It is not intended for use with neonate cord blood specimens.

    It is not intended for the screening or diabetes mellitus but is indicated for use in determining dysglycemia.

    Device Description

    The StatStrip Glucose Hospital Meter System was previously cleared under K150281. The purpose of this dual 510(k) and CLIA Waiver submission for the StatStrip Glucose Hospital Meter System is to expand the indication for use to include the use of capillary specimens on patients receiving intensive medical intervention/therapy and support a CLA Waived categorization for this expanded indication for use to include the quantitative determination of glucose in capillary finger stick throughout all hospital and all professional healthcare settings. No changes have been made to the software, the test strips, controls or linearity solutions.

    The system contains the following:

      1. StatStrip Meter, with integrated Wi-Fi connection and antenna option
      1. Charging Station
      1. Vial of StatStrip Test Strips
      1. Battery (3.7 V Lithium)
      1. Quick Reference Guide
      1. Instructions for Use Manual
      1. StatStrip Log Book

    Offered separately:

    • . StatStrip Test Strips
    • Quality Control Solutions, Level 1, 2, 3
    • Linearity Solutions, Levels 1, 2, 3, 4, 5.

    Meter: The StatStrip Glucose Hospital Meter is a hand-held testing device that works in conjunction with Nova StatStrip Glucose test strips to measure glucose in a whole blood sample. Meter operation is selfprompting using an illuminated touch-screen Graphical User Interface (GUI). In addition to measuring glucose, the meter also stores patient test data, quality control test data and other information relating to the patient, patient sample, operator, reagents and meter.

    The self-prompting menu system is navigated by means of a combination of touch-panel on-screen keys, on-screen soft keyboard. The operator can use a finger or a PDA stylus to select options for the onscreen display. It offers audible feedback for user inputs, and audible and/or visual feedback for prompts and user alerts. The Meter also has a barcode scanner that automates data entry.

    The rechargeable batteries provide sufficient power to operate for 8 hours before requiring recharging. A "battery fuel gauge" bitmap constantly informs the user as to the current state of charge on the battery. Battery charge state information is available on the "meter Welcome screen". The meter will shutoff (Sleep) after 90 seconds of inactivity. Test data and meter setup information will be stored in a nonvolatile form to prevent data loss.

    Charging (Docking) Station: The meter charging station is a stationary accessory used to recharge the meter. The charging station has one slot for the meter to be placed for charging and a slot for an extra battery to be charged.

    The charging station should be located central to the patient care area being served by the meter (e.g. a nursing station). The data charging station must remain plugged in to a wall outlet for power.

    The system still allows the charging station to be used to transfer data from the meter to a central workstation, and allow meter setup information to be downloaded from the central workstation to the meter.

    Test Strips: The test strips contain a reaction layer that contains a glucose-enzyme (greater than 1.0 IU) and ferricvanide as a mediator. The test strip is touched to a drop of blood to initiate the test process. The strip is designed such that when a drop of blood is touched to the strip, the blood is drawn into the reaction space via capillary action. A simple one-step process provides a blood glucose result. Test strips will be sold in vials of 25 strips.

    Control Solutions: The control solutions are aqueous assayed solutions, containing buffered D-Glucose, viscosity-adjusting agent, preservatives and other non-reactive ingredients (dye). They contain no products of human origin. There are three levels of controls, (Level 1, Level 2 and Level 3). These solutions will be offered for sale separately from the meter.

    Linearity Solutions: There are 5 levels of Linearity solutions containing buffered D-Glucose, viscosity-adjusting agent, preservatives and other non-reactive ingredients (dye). They contain no products of human origin. These solutions are offered separately from the system for users to verify the performance of the system.

    Replacement batteries will be offered separately.

    Wi-Fi Option: The StatStrip Glucose Hospital Meter provides a Wi-Fi communication method with a healthcare facility's network system. The wireless radio chipset utilizing the IEEE 802.11 alb/g communication protocols (Wi-Fi) is used as an optional means to transmit and receive test results, previously cleared under K150281.

    AI/ML Overview

    The provided text describes the Nova Biomedical Corporation's StatStrip Glucose Hospital Meter System. Here's a breakdown of the acceptance criteria and study information:


    Acceptance Criteria and Device Performance for StatStrip Glucose Hospital Meter System

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for glucose meters, particularly those for critical care, typically align with ISO 15197 standards or internal company-defined standards that demonstrate clinical agreement. While explicit "acceptance criteria" are not given in quantitative terms like "must achieve X% within Y mg/dL," the clinical study performance data serves as the reported performance against FDA's substantial equivalence requirements. Therefore, the "acceptance criteria" can be inferred from the performance metrics demonstrated.

    Interpreted Acceptance Criteria (Based on Study Performance):

    • For glucose concentrations < 75 mg/dL: A high percentage of results should be within ±15 mg/dL of the comparator method.
    • For glucose concentrations ≥ 75 mg/dL: A high percentage of results should be within ±20% of the comparator method.

    Reported Device Performance (from Clinical Study Performance section):

    Glucose Concentration RangePerformance MetricStudy #1 (Prospective)Study #2 & #3 (Combined Real-World)
    < 75 mg/dLWithin ± 5 mg/dL100% (1/1)47.9% (907/1894)
    Within ± 10 mg/dL100% (1/1)77.6% (1470/1894)
    Within ± 12 mg/dL100% (1/1)85.2% (1614/1894)
    Within ± 15 mg/dL100% (1/1)91.7% (1737/1894)
    Exceeds ± 15 mg/dL0% (0/1)8.3% (157/1894)
    ≥ 75 mg/dLWithin ± 5%48.9% (277/567)50.2% (7473/14884)
    Within ± 10%79.4% (450/567)79.3% (11807/14884)
    Within ± 12%85.4% (484/567)86.0% (12799/14884)
    Within ± 15%91.0% (516/567)92.1% (13712/14884)
    Within ± 20%96.8% (549/567)96.4% (14350/14884)
    Exceeds ± 20%3.2% (18/567)3.6% (534/14884)

    2. Sample size used for the test set and the data provenance:

    • Prospective Clinical Study (Site #1):
      • Sample Size: 568 critical care patients.
      • Data Provenance: Prospective study, conducted at a single leading medical center in Minnesota, U.S. (one of the top-ranked hospitals in Diabetes & Endocrinology).
    • Real-World Evidence Studies (Site #2 & #3):
      • Sample Size:
        • Study #2: 2,133 critical care patients.
        • Study #3: 14,645 critical care patients.
        • Combined for reporting: 1894 (for <75 mg/dL) and 14884 (for >=75 mg/dL). This suggests that after filtering or specific data points, these are the totals used for analysis.
      • Data Provenance: Retrospective data mining ("real world evidence") from two leading medical centers in the U.S. (one in Minnesota and one in Maryland). These were described as CLIA Waived retrospective studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The ground truth was established by a central laboratory comparator method.

    4. Adjudication method for the test set:

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a glucose meter, not an imaging device or AI diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Yes, the studies assessed the performance of the StatStrip Glucose Hospital Meter System (algorithm-driven device) by comparing its readings to a central laboratory comparator method. This is essentially a standalone performance evaluation of the device itself. The "CLIA Waived operators" used the device, but the performance being measured is of the device's accuracy compared to a gold standard, not the operator's diagnostic ability.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth was established using plasma glucose specimens obtained from either an arterial or venous specimen as measured on a central laboratory comparator method, specifically an IDMS traceable hexokinase comparator method (Roche Cobas System, Roche Diagnostics, Indianapolis, IN). This is considered a highly accurate and standardized laboratory method for glucose determination.

    8. The sample size for the training set:

    This information is not provided in the document. The document describes clinical studies for performance evaluation for a 510(k) submission, not specifically developmental or training data for the device's underlying technology. The device itself was previously cleared (K150281) and this submission is an expansion of its indication for use, implying the core technology was already developed and validated.

    9. How the ground truth for the training set was established:

    This information is not provided in the document, as the document focuses on the expanded indication's validation rather than the initial device development.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132121

    Validate with FDA (Live)

    Date Cleared
    2014-09-24

    (441 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood and neonate heel stick specimens.
    The StatStrip Glucose Hospital Meter System is also intended for use in the quantitative determination of glucose in venous whole blood, arterial whole blood, neonatal heel stick and neonatal arterial whole blood samples throughout all hospital and all professional healthcare settings.
    The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.
    It is not intended for use with neonate cord blood specimens.
    It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.

    Device Description

    The StatStrip Glucose Hospital Meter is a hand-held testing device that works in conjunction with the StatStrip Glucose Test Strips to measure Glucose in a whole blood sample. Meter operation is self-prompting using an illuminated color touch-screen Graphical User Interface (GUI). In addition to measuring Glucose, the meter also stores patient test data, quality control test data, and other information relating to the patient sample, operator, reagents, and meter. Meter setup options relating to authorized operators, reagent lots, QC preferences, and other operational settings are customizable. Data is transferred bidirectionally between the meter, data docking station, and separate data management system each time a meter is placed in to a data docking station.
    Operator ID, patient ID, reagent lot number, test strip code, and sample accession number can be scanned into the meter using an optional built-in barcode scanner. The meter can store up to 1,000 patient test results and 200 quality control test results. The user can recall and sort or search the meter test results database to find specific test information. A rechargeable battery provides power to operate. A battery low warning will alert the user to recharge the battery. Battery charge state information is available on the "meter status screen". A System Manager selectable auto shutoff option is provided to conserve power when the meter is not in use. All test data is stored in a non-volatile form to prevent data loss.
    The Test Strip is designed with an electrode that measures Glucose in the blood sample mixes with reagent on the test strip that produces an electric current. The amount of current that is produced depends on how much Glucose is in the blood.
    The strip is designed such that when a drop of blood is touched to the end of the strip, the blood is drawn into the reaction space via capillary action. A simple one-step process provides a blood glucose result. Test strips will be sold in vials of 50 strips.
    Control Solutions are aqueous solutions that contain no products of human origin. The controls solutions are for use with the StatStrip family of meters. There are three (3) levels of control solutions (1, 2, and 3)
    The StatStrip Glucose Linearity Sclutions are aqueous materials with a known concentration of glucose intended to verify performance of the Nova Biomedical Analyzers. Assay values for expected ranges are included on every bottle of linearity standards. If the results obtained are outside the expected range. the system may not be performing correctly. The combined linearity solutions are for use with the StatStrip family of meters. There are five (5) Levels of Linearity Standards.
    The meter Data Docking Station is a stationary accessory used to recharge the battery in the meter. The charging station has one slot for the meter to be placed for charging and additional slots for extra batteries to be charged.
    The charging station is typically located central to the patient care area being served by the meter (e.g. a nursing station). The Data Docking Station must remain plugged in to a wall outlet for power.
    The Data Docking Station is used to charge the meter and can be used with a USB cable to transfer data from the meter to a central workstation (i.e. Computer Database or a Data Capture System).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the StatStrip Glucose Hospital Meter System. The submission is for a correction to a previous letter and primarily focuses on an expanded indication for use for an already cleared device, K063821 (Nova StatStrip Glucose Hospital Meter System). The manufacturer states that the device itself has not changed, only the labeling to allow for its use with arterial or venous samples in all hospital and professional healthcare settings.

    Given this context, the acceptance criteria and supporting study described are geared towards demonstrating that the existing device performs equivalently for these expanded indications as it did for its previous indications, and equally to central laboratory methods.

    Here's the breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state quantitative acceptance criteria in terms of specific performance metrics (e.g., accuracy percentages within certain glucose ranges) for this particular submission. Instead, it relies on the established performance of the predicate device and demonstrates that the current device, with expanded indications, performs substantially equivalently to central laboratory methods.

    The "Comparison of Predicate and Proposed device" table (Pages 6-7) highlights the characteristics that are identical between the predicate and proposed device, implying that the established performance for these characteristics remains valid.

    CharacteristicPredicate - Nova StatStrip Glucose Hospital Meter System (K063821)Proposed - StatStrip Glucose Hospital Meter SystemPerformance Assessment (Implicit from text)
    Test MeasuredGlucoseGlucoseIdentical
    Operating PrincipleCoulometric Electro-chemical SensorCoulometric Electro-chemical SensorIdentical
    GLU Measuring Range10-600 mg/dL10-600 mg/dLIdentical
    Hematocrit Range25% to 60%25% to 60%Identical
    GLU Sample Size1.2 µL1.2 µLIdentical
    Glucose Unitsmg/dLmg/dLIdentical
    Sample type (Expanded)Whole Blood: Capillary, Venous Arterial and NeonateCapillary whole blood (finger stick), venous whole blood, arterial whole blood, neonate heel stick, and neonate arterial whole blood specimens.Venous whole blood, arterial whole blood, neonatal heel stick, and neonatal arterial whole blood samples throughout all hospital and all professional healthcare settings.Performance shown to be substantially equivalent to central laboratory methods for expanded types.
    GLU Analysis Time6 seconds6 secondsIdentical
    GLU Test Strips Active ReagentGlucose OxidaseGlucose OxidaseIdentical

    2. Sample size used for the test set and the data provenance:

    The document states: "The laboratory and clinical performance data for the StatStrip Glucose Hospital Meter System supports the expanded use of the system for arterial and venous specimens in all hospital and all professional healthcare settings." However, specific sample sizes for the test set are not provided within the given text. The provenance of the data (e.g., country of origin, retrospective or prospective) is also not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    The nature of the device (a blood glucose meter) suggests that ground truth would be established through a central laboratory reference method, not by expert clinicians or radiologists. The document states "Results of clinical performance and laboratory testing demonstrate that the StatStrip Glucose Hospital Meter System produces results that are substantially equivalent to central laboratory methods." Thus, the "experts" facilitating the ground truth would be the personnel operating the central laboratory equipment, but no specific number or qualifications of these individuals are mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    As the ground truth is established by a central laboratory reference method, adjudication by multiple experts is not applicable in this context.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    An MRMC study is not applicable here as this is a blood glucose meter, not an imaging device typically evaluated with human readers or AI assistance. The performance is assessed against a laboratory reference method.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device itself is a standalone measurement system (the meter and test strips). Its performance is evaluated intrinsically against a reference method, so the concept of "standalone (algorithm only)" is applicable to the device itself. The clinical and laboratory performance data discussed in the document would represent this standalone performance, but without specific raw data, the exact standalone performance metrics aren't detailed beyond the statement of substantial equivalence.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth used for the performance studies was established by central laboratory methods for glucose determination.

    8. The sample size for the training set:

    The document explicitly states: "No changes have been made to the meter, the software, the test strips, the controls or the linearity solutions." This implies that the device software and algorithms were already developed and "trained" prior to this submission (likely for the predicate device K063821). The submission itself is not about developing a new algorithm or device, but extending the indication for an existing one. Therefore, no new training set information is provided or relevant to this specific submission.

    9. How the ground truth for the training set was established:

    Given that no new training set is described for this submission, this question is not applicable here. The ground truth for the original development of the predicate device's algorithms would similarly have been established by comparison to reference laboratory methods.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1