510(k) Data Aggregation

    K Number
    K131920

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-05-01

    (309 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The I.B.S.™ osteosynthesis screws are intended for:

    • The fixation of arthrodeses, osteotomies or fractures of long or short bones of the upper and lower limbs
    • Osteosyntheses requiring a mono or bicortical compression
      The size of the chosen screw should be adapted to the specific indications.
    Device Description

    The I.B.S.TM osteosynthesis screws are cannulated screws, available in a compression and a neutralization design. The cannulation of the screws provides a helpful feature during surgery, as a wire is used to guide insertion of the screw. The compression design has a non-threaded shaft, allowing optimal compression between the two bone fragments, which may enhance bone osteosynthesis. The neutralization design is fully threaded and may be preferred when stabilization of the bone fragments without strong compression is required, as for example in some fractures in multiple fragments, or osteoporotic bone. Both designs are self-drilling and self-tapping screws, which enables introduction of the screw without any preparation of the hole (using a drill and /or a tap) in most cases.
    Sizes: The I.B.S.TM osteosynthesis screws are available in 2.5mm, 3.0mm, 3.5mm, 4.5mm, 6.0mm, 6.5mm, and 8.0mm diameters, in length ranging from 20mm to 160mm.
    Material: The I.B.S.TM osteosynthesis screws are manufactured from titanium alloy Ti6A14Vas per ISO 5832-3 and ASTM F136. They do not have any coatings.
    Single use: The I.B.S.TM osteosynthesis screws are designed for single use only.
    Sterilization: The I.B.S.TM osteosynthesis screws are supplied sterile, using gamma irradiation.
    Place of use: The I.B.S.TM osteosynthesis screws are indicated for use in a hospital, or outpatient surgery center where sterile fields may be created and maintained. They are prescription devices.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the In2Bones I.B.S.™ osteosynthesis screws, focusing on acceptance criteria and supporting studies:

    This submission is for a medical device (osteosynthesis screws), not an AI/ML powered device. Therefore, many of the typical acceptance criteria and study components related to AI devices (such as sample size for test/training sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable.

    The approval for this device is based on substantial equivalence to existing predicate devices, meaning it has similar intended use, materials, dimensions, and designs, and does not raise new questions of safety or effectiveness. The study demonstrating this is primarily a comparative engineering and dimensional analysis rather than a clinical trial or AI performance study.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Demonstrated Equivalence)Reported Device Performance (as demonstrated by comparison to predicates)
    Intended Use: Fixation of arthrodeses, osteotomies or fractures of long or short bones of upper and lower limbs; osteosyntheses requiring mono or bicortical compression.Identical. The I.B.S.™ osteosynthesis screws have the same indications for use as the predicate devices. This is stated as: "The I.B.S.™ osteosynthesis screws are intended for: - The fixation of arthrodeses, osteotomies or fractures of long or short bones of the upper and lower limbs - Osteosyntheses requiring a mono or bicortical compression."
    Technological Characteristics:Similar/Equivalent. The device's technological characteristics are deemed the same as the predicate devices. Specifically: - Cannulation: The I.B.S.™ screws and all predicates are cannulated. - Material: The I.B.S.™ screws are made from Titanium alloy Ti6A14V (ISO 5832-3 and ASTM F136), similar to most predicates (except SBI® AutoFix which uses Stainless Steel 316L). No new materials are introduced. - Compression Design: The I.B.S.™ Compression screw has a non-threaded part for compression, similar to all compression predicates. - Neutralization Design: The I.B.S.™ Neutralization screw is fully threaded, similar to all neutralization predicates. - Size Range: The I.B.S.™ screws (2.5mm to 8.0mm diameters, 20mm to 160mm lengths) are covered by the size ranges of the predicate devices.
    Material/Biocompatibility: Titanium alloy Ti6A14V (ISO 5833-3, ASTM F136).Compliant/Equivalent. Manufactured from titanium alloy Ti6A14Vas per ISO 5832-3 and ASTM F136, which are established medical device standards and consistent with predicate devices.
    Sterilization Method: Gamma irradiation.Compliant/Equivalent. Supplied sterile via gamma irradiation, a standard sterilization method for medical devices.
    Safety and Effectiveness: Demonstrate no new questions of safety or efficacy.Demonstrated by Substantial Equivalence. Based on the identical intended use and similar technological characteristics (materials, design features, size range), the device is deemed as safe and effective as the predicate devices. An "engineering / dimensional comparison to the predicates was performed to demonstrate substantial equivalence."

    2. Sample size used for the test set and the data provenance

    • Not Applicable. This is a hardware medical device seeking substantial equivalence, not an AI/ML device requiring a test set of data. The "test" here refers to the comparison of the physical device to its predicates.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. No ground truth in the context of expert labeling or adjudication is required for this type of device submission. The "ground truth" for substantial equivalence is the established safety and efficacy of the predicate devices, which has been previously reviewed and cleared by the FDA based on their respective data.

    4. Adjudication method for the test set

    • Not Applicable. No test set or expert adjudication process as would be used for an AI/ML device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI-assisted diagnostic or therapeutic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" in this context is the regulatory clearance of the predicate devices by the FDA, which implicitly confirms their safety and effectiveness for their stated indications. The I.B.S.™ screws are compared against these already-cleared devices. No new clinical "ground truth" (e.g., pathology or outcome data specific to this device) was required for this 510(k) submission.

    8. The sample size for the training set

    • Not Applicable. This is a hardware medical device, not an AI/ML device. The "training" for such a device involves engineering design, material selection, and manufacturing processes, not data training.

    9. How the ground truth for the training set was established

    • Not Applicable. As above, no training set in the context of AI/ML. The design and manufacturing of the device would have followed established engineering principles and medical device standards (e.g., ISO, ASTM), and in-house testing to ensure it meets its specifications, but this is not referred to as a "training set" with "ground truth" in the AI sense.
    Ask a Question

    Ask a specific question about this device

    K Number
    K050346

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-03-17

    (34 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STABILIZATION SCREW is indicated for fixation of bone fractures or for bone reconstruction. Examples include:

    • Mono or Bi-cortical osteotomies in the foot or hand (including Hallux Valgus treatment)
    • Fractures management in the foot and hand
    • Fixation of bone fragments in long bones or small bones fractures
    • Arthrodesis in hand, foot or ankle surgery
      The size of the chosen screw should be adapted to the specific indication.
    Device Description

    The STABILIZATION SCREW is a cannulated compression screw with a nonthreaded shaft, allowing optimal compression. It also has a self-tapping screw tip. It is provided in diameters 3.0 mm and 4.3 mm and in length from 10 mm to 34 mm for the 3 mm and from 22 mm to 60 mm for the 4.3 m. The STABILIZATION SCREW is made from Titanium alloy (Ti-6A1-4V ELI).

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "STABILIZATION SCREW." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing extensive clinical study data with detailed acceptance criteria and performance metrics typically seen in a PMA (Premarket Approval) submission.

    Therefore, the document does not contain the level of detail requested for acceptance criteria, device performance, or a comprehensive study plan with statistical outcomes, sample sizes, and expert qualifications in the way a clinical trial report would.

    However, I can extract information related to the type of studies performed and their conclusion regarding equivalence.

    Here's a summary based on the provided text, addressing the points you requested where information is available:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria or specific numerical performance metrics for the STABILIZATION SCREW. Instead, it states that the mechanical properties were shown to be "equivalent" to predicate devices.

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical properties equivalent to predicate devices as per ASTM F543-02 standardTorsional, pullout, and 3-point bending tests showed mechanical properties are equivalent to predicate devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "Torsional and pullout strength tests" and "3-point bending tests" were carried out. It does not specify the sample sizes used for these mechanical tests or the provenance of the data (e.g., country, retrospective/prospective). These are typically bench testing results.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the studies described are mechanical bench tests, not clinical studies involving human observers or "ground truth" as typically understood in a diagnostic accuracy study.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for mechanical bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is for a medical implant (bone fixation screw), not an AI-powered diagnostic device. No human reader involvement or AI assistance is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a mechanical implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the mechanical tests, the "ground truth" would be the physical measurements obtained from testing the screws against the specified ASTM standard (ASTM F543-02) and comparing them to predicate devices. There is no biological or diagnostic "ground truth" in this context.

    8. The sample size for the training set

    Not applicable. There is no concept of a "training set" for mechanical bench testing of this type of device.

    9. How the ground truth for the training set was established

    Not applicable.


    Summary of Study (from Section I. Summary of Studies):

    • Study Type: Mechanical testing.
    • Tests Performed: Torsional strength, pullout strength, and 3-point bending tests.
    • Standard Followed: ASTM F543-02 (Standard Specification and Test Methods for Metallic Medical Bone Screws).
    • Conclusion: "Results have shown that mechanical properties of the STABILIZATION SCREW are equivalent to the predicate devices."

    The core of this 510(k) submission relies on demonstrating substantial equivalence in terms of design, material, indications for use, dimensions, and comparable mechanical performance to already legally marketed devices.