510(k) Data Aggregation

    K Number
    K992823

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-05-12

    (263 days)

    Product Code
    Regulation Number
    868.1840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spirovit SP-2 / PneumoCheck ® II can be used to measure the following spirometric parameters within the Healthcare Facility:

    Forced Vital Capacity (FVC), Forced Expiratory Volume in one second (FEV1), FEV 1.0 / FVC, FEF 0.2-1.2, FEF 25-75%, FEF 75-85%, PEF, FEF 25%, FEF 50%, FEF 75%, FIVC, FIV 1.0, FIV 1.0 / FIVC, PIF, FIF 50%, SVC, ERV, IRV, TV, MVV, MV, RR, TV, in patients 6 years of age or older.

    Device Description

    Not Found

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the SPIROVIT SP-2 / PneumoCheck® II device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not outline specific numerical acceptance criteria for the spirometric parameters measured by the device. Instead, it relies on demonstrating equivalence to a predicate device and compliance with established standards. The reported "performance" for the new device is largely asserted as being "the same" as the predicate device or having passed relevant electrical and software standards.

    Feature / CriteriaPredicate Device (SP-250) "Acceptance"SPIROVIT SP-2 / PneumoCheck® II Reported Performance
    Dimensions
    Flow Sensor7.5 cm85x190x27 mm
    Handle10 cmn.a. (not applicable, implies no separate handle)
    MouthpieceDisposableSame (Disposable)
    Weight180 g275 g
    Environmental Conditions
    Operating Temp.+10° - 40° CSame
    Storage Temp.-10° - +50° CSame
    Relative Humidity25% - 95% (non-condensing)Same
    Electrical Safety / EMC
    ESD (Functional)Below 4 kV (Open Air)Same
    ESD (No damage)Below 8 kV (Open Air)Same
    Radiated EmissionsLess than 30 dB MicrovoltsSame
    Radiated Immunity3 Volts per meterSame
    Data CommunicationRS-232 interfaceIrDA
    Spirometry StandardsNot explicitly stated for predicate in comparison table, but ATS compliance is mentioned for new device.ATS Spirometry Statement 1994 (All tests passed)
    Electrical Safety StandardsNot explicitly stated for predicate in comparison table.IEC 601-1-1 (All tests passed)
    EMC StandardsNot explicitly stated for predicate in comparison table.IEC 601-1-2 (All tests passed)
    Software Quality StandardsNot explicitly stated for predicate in comparison table.IEC 601-1-4 (All tests passed)

    Discussion of Differences: The document explicitly states, "None of the above differences (1, 2 or 3) can be considered as safety relevant differences," and concludes that the submitted device is "as safe and effective as the Predicate (SP-250) device."

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states "B2 Clinical Tests n.a." (not applicable). This indicates that no clinical test set was used to validate the performance of the device against specific spirometric parameters in a patient population. The device's performance is demonstrated through compliance with standards and comparison to the predicate device's specifications.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable, as no clinical test set with ground truth established by experts was performed. The "ground truth" for the device's technical specifications and compliance was established through rigorous testing against international and industry standards (ATS, IEC).

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring adjudication was performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a spirometer, a medical measurement device, not an AI-powered diagnostic imaging or interpretation tool.Therefore, no MRMC study or AI assistance comparison was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable in the context of AI. The device itself is a standalone measurement system. Its performance relies on its hardware and software implementing spirometry standards correctly, rather than an AI algorithm's standalone performance.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests:

    • Technical Specifications: The ground truth for device dimensions, weight, environmental conditions, and data communication was the design specifications of the device itself and its predicate.
    • Electrical Safety and EMC: The ground truth was the requirements and test methodologies specified in IEC 601-1-1, IEC 601-1-2.
    • Software Quality: The ground truth was the requirements and guidelines from IEC 601-1-4 and the "reviewer guidance for computer controlled medical devices undergoing 510(k) review."
    • Spirometry Performance: The implicit ground truth for accurate spirometric measurements is the ATS Spirometry Statement (November 11th, 1994). The document states that the device was tested to be in accordance with this standard and "All tests are passed."

    8. The Sample Size for the Training Set

    Not applicable, as this device does not involve machine learning or AI models that require a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.