Found 2 results

510(k) Data Aggregation

    K Number
    K112983

    Validate with FDA (Live)

    Device Name
    SPACER-S
    Manufacturer
    Date Cleared
    2011-12-12

    (67 days)

    Product Code
    Regulation Number
    888.3690
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Interspace Shoulder is indicated for temporary use (maximum of 180 days) as a shoulder replacement (SR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. The head and stem are inserted into the glenoideal cavity and the humeral medullary canal, respectively, following removal of the existing implant and radical debridement. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection). Interspace Shoulder is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.).

    Device Description

    The Interspace Shoulder devices is a temporary device composed of fully formed PMMA bone cement with gentamicin and an inner stainless steel metal core. The design mimics a hemi-shoulder prosthesis.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Tecres Interspace Shoulder Temporary Shoulder Prosthesis:

    Based on the provided 510(k) summary (K112983), this submission is for a modified medical device, not a new AI/software-as-a-medical-device (SaMD) or diagnostic device. As such, the typical "acceptance criteria" and "study" questions you'd ask for an AI model's performance metrics (like sensitivity, specificity, AUC) are not directly applicable in this context.

    Instead, the "acceptance criteria" here refer to production specifications and performance adequate for in vivo application under temporary conditions of use, and the "study" is a performance testing conducted to demonstrate substantial equivalence to a previously cleared device.

    Here's the breakdown of the information as requested, translated into the context of this medical device submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance
    Material PropertiesMechanical properties matching predicate device and suitable for use.Evaluated and found to support substantial equivalence.
    Drug Release ProfileGentamicin release consistent with intended therapeutic effect.Evaluated and found to support substantial equivalence.
    StabilityDevice stability for temporary use (up to 180 days).Evaluated and found to support substantial equivalence.
    Manufacturing StandardsMeeting production specifications.Evaluated and found to support substantial equivalence.
    BiocompatibilitySuitable for in vivo application.Implicitly covered by substantial equivalence to predicate.
    Indications for UseSame as predicate device (temporary shoulder replacement for septic process, up to 180 days).Confirmed as the same, thus meeting this "performance" criterion.

    Note: The document states that the performance testing was to verify that the implant performance "continues to meet the production specifications and be adequate for in vivo application under the temporary conditions of use." The primary "acceptance criteria" seem to be demonstrating that the modified device is substantially equivalent to its predicate (K060535) despite the addition of a metal reinforcing structure and a new material supplier.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated. The "performance testing" described is likely mechanical and chemical testing of the device itself and its components, rather than a clinical trial with patient data. Therefore, the "test set" would refer to the number of devices or material samples subjected to laboratory tests (e.g., fatigue testing, tensile strength, gentamicin elution studies). Without further documentation, the exact number cannot be determined from this summary.
    • Data Provenance: Not applicable in the sense of patient data. The testing would have been conducted in a laboratory setting by the manufacturer (Tecres S.p.A.).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not applicable. For material and mechanical performance testing, "ground truth" is established through standardized engineering and chemical testing methods, not expert consensus on clinical cases.
    • Qualifications of Experts: Not applicable. The testing would be performed by qualified engineers, chemists, and technicians following established protocols.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. This type of testing involves objective measurements against predefined specifications, not human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI diagnostic tool on human reader performance, which is not the nature of this device.

    • Effect Size: Not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Standalone Performance: Not applicable. This device is an implanted prosthesis, not an algorithm or a software device.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: For the performance testing, the "ground truth" would be established by:
      • Engineering specifications and standards: For mechanical properties (e.g., ISO standards for implant materials, internal design specifications for strength and durability).
      • Chemical analysis standards: For gentamicin release and stability (e.g., pharmacopeial methods, validated analytical chemistry techniques).
      • Comparison to predicate device's established performance: To demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This is not an AI/machine learning device. The device itself is the product. The term "training set" is not relevant here.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as there is no "training set."

    Summary of the 510(k) Submission's Core Argument:

    The K112983 submission for the Interspace Shoulder Temporary Shoulder Prosthesis argues for substantial equivalence to its own previously cleared version (K060535). The key modifications are the introduction of a metal reinforcing structure and a new material supplier. The manufacturer conducted performance testing (mechanical properties, gentamicin release, and stability data) to demonstrate that these changes did not negatively impact the device's ability to meet its production specifications and remain adequate for its temporary in vivo application. The FDA's clearance indicates that they found this demonstration of substantial equivalence sufficient without requiring complex clinical trials or new performance metrics beyond what was presented for the predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060535

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-05-24

    (85 days)

    Product Code
    Regulation Number
    888.3690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spacer-S is intended for use as a temporary (maximum 180 days) shoulder replacement in patients undergoing a two-stage procedure due to a septic process.

    The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection) as defined by physician and/or surgeon.

    Spacer-S is indicated for temporary use (maximum 180 days) as an adjunct to total shoulder replacement and hemi shoulder replacement procedures in skeletally mature patients undergoing a two-stage revision procedure due to a septic process. Spacer-S is only indicated for an implantation period of 180 days or less.

    Device Description

    Spacer-S is composed of fully formed gentamicin/polymethylmethacrylate (PMMA) bone cement. The one piece design mimics a hemi-shoulder prosthesis.

    AI/ML Overview

    The K060535 filing describes the Tecres Spacer-S, a temporary shoulder prosthesis. The information provided primarily focuses on its device description, intended use, and a demonstration of substantial equivalence to a predicate device through performance testing, rather than an AI/ML device and its associated acceptance criteria and study data.

    Therefore, many of the requested categories (e.g., sample size for test/training sets, experts for ground truth, MRMC studies, standalone performance, etc.) are not applicable or cannot be extracted from the provided text, as this is a medical device approval for a physical implant, not a software or AI/ML diagnostic tool.

    However, I can extract the relevant performance data that was used to establish substantial equivalence.

    Here's the information that can be extracted or inferred from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Fatigue strengthEvaluated and found adequate for anticipated in vivo load applications under temporary conditions of use.
    Static strengthEvaluated and found adequate for anticipated in vivo load applications under temporary conditions of use.
    Antibiotic releaseEvaluated and found to support substantial equivalence.
    Overall Goal:Substantial equivalence to the predicate Equinoxe Shoulder System device (#K042021).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The performance testing refers to mechanical and material evaluations, not clinical trials with human subjects in the typical sense of test sets for AI/ML.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. Ground truth, in the context of this traditional medical device, is established through engineering and laboratory testing (e.g., measuring physical properties like strength and antibiotic release), not through expert consensus on diagnostic images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. Adjudication methods are relevant for studies involving human interpretation or clinical endpoints, which is not the focus of the performance testing described for this device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This is a physical medical implant, not an AI or imaging diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is a physical medical implant, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the performance data, the "ground truth" would be established through engineering standards and laboratory measurements of the device's mechanical properties (fatigue strength, static strength) and its antibiotic release characteristics. There is no mention of clinical pathology or outcomes data being used to define the "ground truth" for these specific performance tests.

    8. The sample size for the training set

    This information is not applicable. There is no "training set" in the context of this traditional medical device approval based on performance testing of the device itself.

    9. How the ground truth for the training set was established

    This information is not applicable. As there is no "training set," there is no ground truth to establish for it.