Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K964850

    Validate with FDA (Live)

    Date Cleared
    1997-02-25

    (84 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Not Found

    Device Description

    The subject of this submission is the Millipore 0.22 micron filter which Baxter intends to incorporate into various configurations of solution sets for the administration of intravenous solutions. The filter is manufactured by Millipore and has been recently cleared for marketing in Abbott solution administration sets covered by K960466. Baxter will purchase the 0.22 micron filter from Millipore and incorporate it into currently marketed solution administration sets. Baxter is making no changes to the design, components or materials of the Millipore filter.

    AI/ML Overview

    This 510(k) summary describes a new solution administration set incorporating a 0.22 micron filter manufactured by Millipore, intended for use with intravenous solutions. The submission focuses on the filter itself, as Baxter will be integrating it into their existing solution administration set designs. The predicate devices are Abbott and Baxter solution sets, both incorporating 0.22 micron filters.

    Here's an analysis of the provided information based on your requested criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance
    Functional PerformanceMet or exceeded all functional requirements.
    Reverse Pressure Filter IntegrityMet (implied by "meets or exceeds all functional requirements").
    Unit VentingMet (implied by "meets or exceeds all functional requirements").
    Gravity FlowMet (implied by "meets or exceeds all functional requirements").
    Inlet Axial StressMet (implied by "meets or exceeds all functional requirements").
    Housing Integrity/Female Luer FitmentMet (implied by "meets or exceeds all functional requirements").
    Forward Pressure Filter IntegrityMet (implied by "meets or exceeds all functional requirements").
    Downstream Particle CountMet (implied by "meets or exceeds all functional requirements").
    Accelerated EnduranceMet (implied by "meets or exceeds all functional requirements").
    Outlet Deflection StressMet (implied by "meets or exceeds all functional requirements").
    In-Line Filter WettabilityMet (implied by "meets or exceeds all functional requirements").
    BiocompatibilityMet or exceeded all functional requirements (this usually encompasses biocompatibility testing).
    Bacterial EndotoxinsMet (implied by "functional, microbiological, and drug compatibility data indicate that the proposed filter meets or exceeds all functional requirements").
    Bioburden EvaluationMet (implied by "functional, microbiological, and drug compatibility data indicate that the proposed filter meets or exceeds all functional requirements").
    Structural IntegrityMet or exceeded all functional requirements.
    Burst StrengthMet (implied by "meets or exceeds all functional requirements").
    MVI ResistanceMet (implied by "meets or exceeds all functional requirements").
    Drug CompatibilityMet (explicitly stated: "drug compatibility data indicate that the proposed filter meets or exceeds all functional requirements").

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The summary mentions "Data regarding the functional performance [...] were generated by Millipore," but does not specify the number of units tested for each criterion.
    • Data Provenance: The data was generated by Millipore, the manufacturer of the 0.22 micron filter. It is not specified whether the data is retrospective or prospective, nor the country of origin, though given the context of a US 510(k) submission, it's highly likely to be considered within a US regulatory framework.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of medical device submission. This refers to clinical studies where expert consensus or diagnostic interpretation is required for ground truth. For a filter, the "ground truth" is typically defined by standardized physical, chemical, and microbiological tests.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical trials or diagnostic studies to resolve discrepancies between multiple expert readers. For a filter, objective measurements are taken, and results are compared against predefined acceptance limits.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This submission is for a physical medical device (a filter), not an AI-powered diagnostic tool. Therefore, MRMC studies and the concept of human reader improvement with AI assistance are irrelevant here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. As mentioned above, this is a physical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on objective, quantitative measurements obtained from established test methods for filter performance, biocompatibility, and structural integrity. This includes:

    • Physical measurements (e.g., burst strength, flow rates).
    • Chemical purity assessments (e.g., bacterial endotoxins, particle counts).
    • Microbiological testing (e.g., bioburden evaluation) demonstrating sterile filtration capabilities.
    • Drug compatibility studies.

    These "ground truths" are derived from validated test procedures and industry standards relevant to medical filters.

    8. The Sample Size for the Training Set

    This is not applicable. This device is a physical product, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as there is no training set for a physical filter device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1