Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173200

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-06-11

    (252 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SOFIA Plus Aspiration Catheter with the Gomco 405 Aspiration Tubing Kit is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA therapy are candidates for treatment.

    Device Description

    The SOFIA Plus Aspiration Catheter is a single lumen catheter designed to be introduced over a steerable guidewire to access the neurovasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. A single radiopaque marker at the distal end facilitates fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic coating to reduce friction during navigation in the vasculature. A luer fitting on the microcatheter hub is used for the attachment of accessories. The strain relief at the hub provides kink resistance for the proximal end. A steam shaping mandrel and introducer sheath are also packaged with the catheter. The SOFIA Plus Aspiration Catheter is used to remove thrombus/embolus from the neurovasculature using aspiration tubing and pump.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the SOFIA Plus Aspiration Catheter. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials. Therefore, the information provided does not align with the typical structure of an AI/ML medical device study demonstrating algorithmic performance against acceptance criteria.

    The document primarily details the pre-clinical testing on the device itself and an animal study to evaluate its aspiration performance and safety. It does not describe an AI/ML algorithm or a study of its performance.

    However, I can extract and structure the information about the acceptance criteria and performance of the physical device based on the provided text, while acknowledging that this is not an AI/ML study.

    Here's an interpretation of the request using the available information:


    The SOFIA Plus Aspiration Catheter underwent extensive pre-clinical testing to demonstrate its performance and substantial equivalence to a predicate device. The acceptance criteria for these tests were typically defined by established specifications, industry standards (e.g., ISO), or direct comparison to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Biocompatibility
    Cytotoxicity (ISO 10993-5)Non-cytotoxicCell culture treated with test article exhibited slight reactivity (Grade 1), concluded as non-cytotoxic. PASSED
    Sensitization/Irritation (ISO 10993-10) - Guinea pig maximization sensitizationWeak allergic potential or sensitizing capacityExtracts of the test article elicited no reaction at the challenge (0% sensitization) following the induction phase (Grade 1). PASSED
    Sensitization/Irritation (ISO 10993-10) - Intracutaneous reactivityNon-irritantExtracts of the test article did not show a significantly greater biological reaction than the sites injected with the control article. PASSED
    Hemocompatibility – Rabbit Blood Direct and Indirect Contact (ISO 10993-4)Non-hemolyticHemolysis index was 0.13% (direct contact) and 0.0% (indirect contact). PASSED
    Hemocompatibility – Unactivated Partial Thromboplastin Time Assay Direct Contact (ISO 10993-4)No effect on coagulationNo statistically significant difference found between the Unactivated Partial Thromboplastin Time (UPTT) of the plasma exposed to the test article and that of the plasma exposed to either the negative control or the untreated control. PASSED
    Hemocompatibility - Complement Activation Assay (ISO 10993-4)No effect on complement activationC3a and SC5b-9 levels < negative and untreated controls. PASSED
    Hemocompatibility - Thrombogenicity Study in Dogs (ISO 10993-4)No significant thrombosisMinimal thrombosis observed with a Grade 0 in two out of two test sites and two out of two control sites. PASSED
    Systemic Toxicity - Systemic Injection Test (ISO 10993-11)No toxic effectsExtracts of test article did not induce a significantly greater biological reaction than the control extracts when injected in Swiss Albino mice. PASSED
    Systemic Toxicity - Rabbit Pyrogen Test (ISO 10993-11)Non-pyrogenicThe temperature increases (maximum) was 0.03°C from baseline. PASSED
    Bench Testing
    Simulated UseDevice performs as intended, demonstrates equivalency to predicate deviceAchieved a rating $\geq$ 3 for preparation/ease of assembly, introducer sheath interaction, introducer peel away, tracking with guidewire/microcatheter, microcatheter/guidewire lockup, lubricity and durability of hydrophilic coating, microcatheter/guidewire removal, removal/aspiration of clot. PASSED
    Dimensional and Physical AttributesDevice meets specified dimensional requirements, substantially equivalent to predicate deviceMet specified dimensional requirements for catheter OD, catheter ID, overall working length, length of distal section, length of distal tip to marker band and total length of hub/strain relief. PASSED
    Kink ResistanceDevice resistant to kinking around small radii turns, same as predicate deviceNo kinks when wrapped around 0.030 and 0.040-inch pin gauges. No kinks noted during simulated use testing. PASSED
    Durability/Lubricity of Hydrophilic CoatingDevice tracks easily with no coating cracking or separationAchieved a rating of $\ge$ 3 during simulated use testing for coating durability and lubricity. PASSED
    Catheter StiffnessDevice tracks in tortuous anatomy while advancing to target site, equivalent to predicate and competitive devicesDevice stiffness equivalent to predicate and competitive devices. PASSED
    Catheter Flexural FatigueDevice integrity suitable for intended clinical useNo flexural fatigue following repeated bending during simulated use testing and repeated hoop stress following pressure and air aspiration testing. PASSED
    Force at Break (Shaft and hub)Tensile strength test results meet acceptance criteria and equivalent to predicate and competitive devicesCatheter force at break $\ge$ 15N for shaft section and hub/catheter junction. PASSED
    Force at Break (After tip shaping)Device meets acceptance criteria and no difference between pre- and post- force at breakCatheter force at break $\ge$ 15N for distal section assessed pre- and post-tip shaping. PASSED
    Static Burst PressureDevice integrity suitable for intended clinical use and met requirements of ISO 10555-1No damage of pressurized catheter at 46 psi. PASSED
    Fluid Leakage at > 46 psiDevice integrity suitable for intended clinical use and met requirements of ISO 10555-1No liquid leakage from hub and catheter shaft at 46 psi for 30 seconds. PASSED
    Air LeakageDevice integrity suitable for intended clinical use and met requirements of ISO 10555-1No air leakage at hub into syringe for 15 seconds. PASSED
    Dynamic BurstDevice met labeled maximum infusion pressure of 300 psiTest articles did not burst at or below 300 psi. PASSED
    Catheter CollapseNo observation of device collapse, equivalent to predicate deviceTest articles were pressurized for a duration of 6 minutes and no observation of catheter collapse. PASSED
    Vacuum PressureNo difference was measured between device and predicateVacuum pressure testing of catheter at distal tip was compared to vacuum pressure of source. No difference detected. PASSED
    Torque ResponseEquivalent or better than predicateCatheter torque response was assessed in vascular model and met acceptance criteria. PASSED
    Torque StrengthMet acceptance criteriaCatheter was torqued and did not break. PASSED
    Advance/RetractMet acceptance criteria, equivalent or better than predicateCatheter was subjected to advance/retraction test in a vascular model and was equivalent or better than predicate. PASSED
    RadiopacityMarker band is visible under fluoroscopyMarker band was detectable under fluoroscopy. PASSED
    Corrosion ResistanceCatheter is corrosion resistantMet ISO 10555-1. PASSED
    Particulate TestingCatheter does not generate particulateMet USP <788> criteria. PASSED

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for each bench test. Instead, it refers to "test articles" and general results (e.g., "Test articles achieved a rating...") and in some cases, "two out of two test sites" for the animal study.

    • Test Set Provenance: The tests are pre-clinical bench tests performed by the manufacturer (MicroVention, Inc., Tustin, California, USA) and an animal study (swine model). The information does not specify the country of origin of data in terms of retrospective/prospective human patient data, as this is not a human clinical trial. All tests would have been performed prospectively as part of the device development and submission.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This section is not applicable as the document describes pre-clinical physical device testing and an animal study, not an AI/ML algorithm study that requires expert adjudication of image-based ground truth. For the simulated-use tests, the performance ratings (e.g., $\geq3$) would be based on qualitative assessment by trained personnel, but they are not referred to as "experts" in the context of clinical ground truth establishment.

    4. Adjudication Method for the Test Set

    Not applicable. As this is not an AI/ML algorithm study involving interpretation of clinical data by multiple readers, there is no need for an adjudication method like 2+1 or 3+1.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, this was not done. This document pertains to the 510(k) clearance of a physical medical device. MRMC studies are typically performed for AI-driven diagnostic or assistive technologies to assess their impact on human reader performance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, this was not done. This is not an AI/ML algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through:

    • Engineering Specifications and Standards: For dimensional, mechanical, and safety properties (e.g., ISO standards, specified psi limits, force at break criteria).
    • Comparative Performance to Predicate Devices: Performance attributes (e.g., stiffness, vacuum pressure, torque response) were compared to the legally marketed predicate device (Penumbra Reperfusion Catheter ACE64 and ACE68).
    • Animal Study Observations: For aspiration performance and histopathological evaluations in a swine model.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML study, therefore there is no training set for an algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no AI/ML algorithm training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1