Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K142014

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-10-10

    (78 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SOFIA Distal Access Catheter is intended for general intravascular use, including the neuro and peripheral vasculature. It can be used to facilitate the introduction of diagnostic agents. It is not intended for use in coronary arteries.

    Device Description

    The SOFIA™ Distal Access Catheter is a single-lumen, flexible catheter designed with coil and braid reinforcement. The distal segment is steam-shapeable and it has a hydrophilic coating for navigation through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. An introducer sheath and shaping mandrel are also provided.

    AI/ML Overview

    The provided document describes the SOFIA Distal Access Catheter and compares it to a predicate device (K131482). The submission aims to demonstrate substantial equivalence, not to prove clinical effectiveness or diagnostic accuracy in specific use cases as is common for AI/ML devices. Therefore, many of the requested criteria are not applicable, as this is a traditional medical device submission focused on bench testing and biocompatibility.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as pass/fail thresholds in a single table, but rather implied by the "Results" and "Conclusions" for each bench test and biocompatibility assessment. The reported device performance is outlined in the "Results" column.

    Test NameAcceptance Criteria (Implied)Reported Device Performance (Results)
    Bench Testing
    Simulated UseDevice performs as intended under simulated use conditions.Test articles achieved a rating ≥ 3 for preparation/ease of assembly, introducer sheath interaction, introducer peel away, tracking with guidewire/microcatheter, microcatheter/guidewire lockup, lubricity and durability of hydrophilic coating, microcatheter/guidewire removal. Removal/ aspiration of clot, mechanical clot retriever and stent delivery with no particles.
    Equipment InterfaceDevice compatible with recommended accessories commonly used in intravascular procedures.Test articles compatible with 0.035-inch and 0.038-inch guidewires, 6F or larger guide catheter/guiding sheath, common RHVs using insertion tool, stopcocks and ≤ 0.027-inch microcatheters.
    Dimensional and Physical AttributesDevice meets established dimensional and physical specifications.Test articles met the specified dimensional requirements for catheter OD, catheter ID, overall working length, length of distal section, length of distal tip to marker band and total length of hub/strain relief.
    Kink ResistanceDevice resistant to kinking around small radii turns.No kinks at 1 cm, 4 cm, 12 cm and 25 cm from distal tip when wrapped around 0.025-inch and 0.030-inch pin gauges. No kinks noted during simulated use testing.
    Tip ShapeabilityShapeability of distal tip after steam shaping equivalent to competitive devices.Tip angle of test article equivalent to competitive devices after steam shaping using mandrel with an angle of approximately 90°.
    Radio DetectabilityDevice radiopacity equivalent to or better than predicate and competitive devices.Distal marker band visible under fluoroscopy.
    Gauging (ISO 594-2)Device hub meets the requirements of ISO 594-2.Gauging pin and hub align in limit planes.
    Separation Force (ISO 594-2)Device hub meets the requirements of ISO 594-2 (separation force greater than 25 N).Mating parts separation force greater than 25 N.
    Unscrewing Torque (ISO 594-2)Device hub meets the requirements of ISO 594-2 (luer remains attached after applying a torque not less than 0.02 Nm for a minimum of 10 seconds).Test article luer remains attached after applying an unscrewing torque not less than 0.02 Nm for a minimum of 10 seconds.
    Stress Cracking (ISO 594-2)Device hub meets the requirements of ISO 594-2 (no stress cracks).No stress cracks on test article hub.
    Ease of Assembly (ISO 594-2)Device hub meets the requirements of ISO 594-2 (components fit securely with no resistance).Components fit together securely with no resistance observed between test article luer and reference fitting.
    Resistance to Overriding (ISO 594-2)Device hub meets the requirements of ISO 594-2 (luer does not override reference fitting threads).Test article luer does not override reference fitting threads.
    Durability/Lubricity of Hydrophilic CoatingDevice tracks easily with no coating cracking or separation (achieved a rating ≥ 3).Test article achieved a rating of ≥ 3 during simulated use testing for coating durability and lubricity.
    Catheter StiffnessDevice tracks in tortuous anatomy while advancing to target site (stiffness equivalent to predicate and competitive devices).Device stiffness equivalent to predicate and competitive devices.
    Torque StrengthDevice torque strength same as predicate device (no catheter breakage after 50 rotations).No catheter breakage after 50 rotations.
    Catheter Flexural FatigueDevice integrity suitable for intended clinical use (no flexural fatigue).No flexural fatigue following repeated bending during simulated use testing and repeated hoop stress following pressure and air aspiration testing.
    Surface ContaminationDevice integrity suitable for intended clinical use (free from surface contaminants, smooth distal tip, PTFE inner layer not delaminated).Test article free from surface contaminants from uncured coating surface particulates > 0.02 mm², embedded particulates. Distal tip smooth and tapered. PTFE inner layer not delaminated.
    Force at Break (Distal and Hub)Tensile strength test results equivalent to predicate and competitive devices (catheter force at break ≥ 2.25 lbf for distal section and hub/catheter junction).Catheter force at break ≥ 2.25 lbf for distal section and hub/catheter junction.
    Flow RateDevice meets specified requirements for delivery of diagnostic agents (flow rate at 100 psi and 300 psi with diagnostic agents equivalent to predicate device).Flow rate at 100 psi and 300 psi with diagnostic agents (e.g., saline, contrast media) equivalent to predicate device.
    Static Burst PressureDevice integrity suitable for intended clinical use and met requirements of ISO 10555-1 (no damage at 46 psi).No damage of pressurized catheter at 46 psi.
    Fluid Leakage at > 46 psiDevice integrity suitable for intended clinical use and met requirements of ISO 10555-1 (no liquid leakage from hub and catheter shaft at 46 psi for 30 seconds).No liquid leakage from hub and catheter shaft at 46 psi for 30 seconds.
    Air LeakageDevice integrity suitable for intended clinical use and met requirements of ISO 10555-1 (no air leakage at hub into syringe for 15 seconds).No air leakage at hub into syringe for 15 seconds.
    Dynamic BurstDevice met labeled maximum infusion pressure of 300 psi (did not burst at or below 300 psi).Test articles did not burst at or below 300 psi.
    Particulate TestDevice met specifications for maximum allowable particles (less than 25 particles > 10 microns/ml, less than 3 particles > 25 microns/ml, no particles > 70 microns).Less than 25 particles greater than 10 microns per ml volume and less than 3 particles greater than 25 microns per ml volume. No particles greater than 70 microns.
    Biocompatibility
    Cytotoxicity (ISO 10993-5) - MEM elution assayNon-cytotoxic (cell culture treated with test article exhibits only slight reactivity, Grade 1).Cell culture treated with test article exhibited slight reactivity (Grade 1).
    Sensitization/Irritation (ISO 10993-10) - Guinea pig maximization sensitizationWeak allergic potential or sensitizing capacity (extracts elicited no reaction at challenge, 0% sensitization, Grade 1).Extracts of the test article elicited no reaction at the challenge (0% sensitization) following the induction phase (Grade 1).
    Sensitization/Irritation (ISO 10993-10) - Intracutaneous reactivityNon-irritant (extracts did not show a significantly greater biological reaction than control sites).Extracts of the test article did not show a significantly greater biological reaction than the sites injected with the control article.
    Hemocompatibility – Rabbit Blood Direct and Indirect Contact (ISO 10993-4)Non-hemolytic (hemolysis index < acceptance criteria; not explicitly stated but typically < 2%).The hemolysis index was 0.13% (direct contact) and 0.0% (indirect contact).
    Hemocompatibility - Unactivated Partial Thromboplastin Time Assay Direct Contact (ISO 10993-4)No effect on coagulation (no statistically significant difference in UPTT compared to controls).No statistically significant difference found between the Unactivated Partial Thromboplastin Time (UPTT) of the plasma exposed to the test article and that of the plasma exposed to either the negative control or the untreated control.
    Hemocompatibility - Complement Activation Assay (ISO 10993-4)No effect on complement activation (C3a and SC5b-9 levels < negative and untreated controls).C3a and SC5b-9 levels < negative and untreated controls.
    Hemocompatibility - Thrombogenicity Study in Dogs (ISO 10993-4)No significant thrombosis (minimal thrombosis, Grade 0).Minimal thrombosis observed with a Grade 0 in two out of two test sites and two out of two control sites.
    Systemic Toxicity - Systemic Injection Test (ISO 10993-11)No toxic effects (extracts did not induce a significantly greater biological reaction than control extracts).Extracts of test article did not induce a significantly greater biological reaction than the control extracts when injected in Swiss Albino mice.
    Systemic Toxicity - Rabbit Pyrogen Test (ISO 10993-11)Non-pyrogenic (temperature increases < acceptance criteria; typical < 0.5°C).The temperature increases (maximum) was 0.03°C from baseline.

    Regarding the other requested information (2-9), most are not applicable because this is a 510(k) premarket notification for a traditional medical device, not an AI/ML diagnostic or prognostic device. The study presented is a technical verification and validation study against established standards and a predicate device, rather than a clinical effectiveness study.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated for each test, but it would be a reasonable number for bench testing (e.g., 3-10 samples per test, as per common medical device testing practices). The thrombogenicity study in dogs used "two out of two test sites and two out of two control sites" which implies 2 animals for the test and 2 for the control for that specific test.
    • Data Provenance: The document does not specify the country of origin for the test data for the proposed device. For many tests, it explicitly states "(Prior test data from predicate device)" or "Biocompatibility was prior test data from predicate device," indicating the data was generated previously for the predicate device. This is a retrospective use of existing data for the current submission. The origin of those previous tests isn't detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is not a study requiring expert consensus or ground truth in the context of diagnostic accuracy. The "ground truth" for these tests comes from engineering specifications, ISO standards, and comparison to the predicate device's performance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No adjudication method is described as this is not a diagnostic performance study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI/ML device, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable in the traditional sense. The "ground truth" for the tests performed are primarily:
      • Engineering Specifications: Internal design requirements for dimensions, strength, flow rate, etc.
      • International Standards: e.g., ISO 594-2 for luer connectors, ISO 10993 series for biocompatibility, ISO 10555-1 for sterile single-use intravascular catheters. These standards define acceptance criteria.
      • Predicate Device Performance: Direct comparison of the proposed device's performance to that of the predicate device (K131482) to demonstrate substantial equivalence.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/ML device; there is no training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is not an AI/ML device; there is no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131482

    Validate with FDA (Live)

    Date Cleared
    2013-11-22

    (183 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SOFIA Distal Access Catheter is intended for general intravascular use, including The GOT IA Dicture al vasculature. It can be used to facilitate the introduction of the nours and therapeutic agents. It is not intended for use in coronary arteries.

    Device Description

    The SOFIA Distal Access Catheter is a single-lumen, flexible catheter designed with coil and braid reinforcement. The distal segment is steam-shapeable and it has a hydrophilic coating for navigation through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. An introducer sheath and shaping mandrel are also provided.

    AI/ML Overview

    The provided text describes a medical device, the SOFIA™ Distal Access Catheter, and its verification testing for 510(k) clearance. This submission does not include information about AI/ML models, therefore, many of the requested fields related to AI/ML studies cannot be answered. The information pertains to traditional medical device testing.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document presents internal acceptance criteria implicitly through the "Results" and "Conclusions" sections of the verification tests. The "Conclusions" often state that the device "met established specifications" or "performs as intended," implying these were the acceptance criteria.

    Test CategorySpecific TestAcceptance Criteria (Implied)Reported Device Performance
    FunctionalSimulated UseAchieves a rating ≥ 3 for various parameters (preparation, tracking, lubricity, etc.) with no particles after procedures like clot aspiration or stent delivery.Test articles achieved a rating ≥ 3 for preparation/ease of assembly, introducer sheath interaction, introducer peel away, tracking with guidewire/microcatheter, microcatheter/guidewire lockup, lubricity and durability of hydrophilic coating, microcatheter/guidewire removal, removal/aspiration of clot, mechanical clot retriever and stent delivery with no particles.
    Equipment InterfaceCompatible with 0.035-inch and 0.038-inch guidewires, 6F+ guide catheter/sheath, common RHVs, and ≤ 0.027-inch microcatheters.Test articles compatible with 0.035-inch and 0.038-inch guidewires, 6F or larger guide catheter/guiding sheath, common RHVs using insertion tool, stopcocks and ≤ 0.027-inch microcatheters.
    Kink ResistanceNo kinks at specified distances (1, 4, 12, 25 cm) when wrapped around pin gauges (0.025-inch, 0.030-inch) and during simulated use.No kinks at 1 cm, 4 cm, 12 cm and 25 cm from distal tip when wrapped around 0.025-inch and 0.030-inch pin gauges. No kinks noted during simulated use testing.
    Tip ShapeabilityTip angle equivalent to competitive devices after steam shaping with a ~90° mandrel.Tip angle of test article equivalent to competitive devices after steam shaping using mandrel with an angle of approximately 90°.
    RadiodetectabilityDistal marker band visible under fluoroscopy, equivalent to or better than predicate/competitive devices.Distal marker band visible under fluoroscopy.
    Durability/Lubricity of Hydrophilic CoatingAchieves a rating of ≥ 3 during simulated use for coating durability and lubricity.Test article achieved a rating of ≥ 3 during simulated use testing for coating durability and lubricity.
    Catheter StiffnessStiffness equivalent to predicate and competitive devices to track in tortuous anatomy.Device stiffness equivalent to predicate and competitive devices.
    Torque StrengthNo catheter breakage after 50 rotations.No catheter breakage after 50 rotations.
    Catheter Flexural FatigueNo flexural fatigue during repeated bending in simulated use and hoop stress testing.No flexural fatigue following repeated bending during simulated use testing and repeated hoop stress following pressure and air aspiration testing.
    Flow RateFlow rate at 100 psi and 300 psi with diagnostic agents equivalent to or better than competitive devices.Flow rate at 100 psi and 300 psi with diagnostic agents (e.g., saline, contrast media) equivalent to or better than competitive devices.
    Static Burst PressureNo damage at 46 psi, meeting ISO 10555-1.No damage of pressurized catheter at 46 psi.
    Fluid Leakage at > 46 psiNo liquid leakage from hub and catheter shaft at 46 psi for 30 seconds, meeting ISO 10555-1.No liquid leakage from hub and catheter shaft at 46 psi for 30 seconds.
    Air LeakageNo air leakage at hub into syringe for 15 seconds, meeting ISO 10555-1.No air leakage at hub into syringe for 15 seconds.
    Dynamic BurstNo burst at or below 300 psi, meeting labeled maximum infusion pressure.Test articles did not burst at or below 300 psi.
    Dimensional/PhysicalDimensional and Physical AttributesMeets specified dimensional requirements for catheter OD, ID, lengths, etc.Test articles met the specified dimensional requirements for catheter OD, catheter ID, overall working length, length of distal section, length of distal tip to marker band and total length of hub/strain relief.
    Gauging (ISO 594-2)Gauging pin and hub align in limit planes (meets ISO 594-2).Gauging pin and hub align in limit planes.
    Separation Force (ISO 594-2)Mating parts separation force greater than 25 N (meets ISO 594-2).Mating parts separation force greater than 25 N.
    Unscrewing Torque (ISO 594-2)Luer remains attached after applying ≥ 0.02 Nm for ≥ 10 seconds (meets ISO 594-2).Test article luer remains attached after applying an unscrewing torque not less than 0.02 Nm for a minimum of 10 seconds.
    Stress Cracking (ISO 594-2)No stress cracks on test article hub (meets ISO 594-2).No stress cracks on test article hub.
    Ease of Assembly (ISO 594-2)Components fit securely with no resistance (meets ISO 594-2).Components fit together securely with no resistance observed between test article luer and reference fitting.
    Resistance to Overriding (ISO 594-2)Luer does not override reference fitting threads (meets ISO 594-2).Test article luer does not override reference fitting threads.
    Surface ContaminationFree from surface contaminants from uncured coating, particulates > 0.02 mm², embedded particulates. Distal tip smooth and tapered, PTFE not delaminated.Test article free from surface contaminants from uncured coating surface particulates > 0.02 mm², embedded particulates. Distal tip smooth and tapered. PTFE inner layer not delaminated.
    Force at Break (Distal and Hub)Catheter force at break ≥ 2.25 lbf for distal section and hub/catheter junction.Catheter force at break ≥2.25 lbf for distal section and hub/catheter junction.
    Particulate Test< 25 particles > 10 microns/mL, < 3 particles < 25 microns/mL, no particles > 70 microns.Less than 25 particles greater than 10 microns per ml volume and less than 3 particles less than 25 microns per ml volume. No particles greater than 70 microns.
    BiocompatibilityCytotoxicity - MEM Elution Assay (ISO 10993-5)Cell culture treated with test article exhibits slight reactivity (Grade 1) or less (non-cytotoxic).Cell culture treated with test article exhibited slight reactivity (Grade 1).
    Sensitization/Irritation - Kligman Maximization Test (ISO 10993-10)No reaction at challenge (0% sensitization) (weak allergic potential or sensitizing capacity).Extracts of the test article elicited no reaction at the challenge (0% sensitization) following the induction phase (Grade 1).
    Sensitization/Irritation - Intracutaneous Injection Test (ISO 10993-10)Extracts do not show significantly greater biological reaction than control sites (non-irritating).Extracts of the test article did not show a significantly greater biological reaction than the sites injected with the control article.
    Hemocompatibility - Rabbit Blood Direct and Indirect Contact (ISO 10993-4)Hemolysis index low (e.g., < 2%) (non-hemolytic).The hemolysis index was 0.13% (direct contact) and 0.0% (indirect contact).
    Hemocompatibility - Unactivated Partial Thromboplastin Time Assay (ISO 10993-4)No statistically significant difference between test article exposed plasma and controls (no effect on coagulation).No statistically significant difference found between the Unactivated Partial Thromboplastin Time (UPTT) of the plasma exposed to the test article and that of the plasma exposed to either the negative control or the untreated control.
    Hemocompatibility - Complement Activation Assay (ISO 10993-4)C3a and SC5b-9 levels ≤ negative and untreated controls (no effect on complement activation).C3a and SC5b-9 levels ≤ negative and untreated controls.
    Hemocompatibility - Thrombogenicity Study in Dogs (ISO 10993-4)Minimal thrombosis (e.g., Grade 0 or low) (no significant thrombosis).Minimal thrombosis observed with a Grade 0 in two out of two test sites and two out of two control sites.
    Systemic Toxicity - Systemic Injection Test (ISO 10993-11)Extracts do not induce significantly greater biological reaction than control extracts (no toxic effects).Extracts of test article did not induce a significantly greater biological reaction than the control extracts when injected in Swiss Albino mice.
    Systemic Toxicity - Rabbit Pyrogen Test (ISO 10993-11)Temperature increases (maximum) low (e.g., < 0.1°C) (non-pyrogenic).The temperature increases (maximum) was 0.03°C from baseline.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for each test in terms of "N" or "number of unique cases/patients." It refers to "test articles," implying a number of device units were subjected to each test. For some tests, it mentions "two out of two test sites" or "test articles."
    Data provenance is not provided in terms of country of origin or retrospective/prospective clinical data, as these are bench and biocompatibility tests, not clinical studies involving patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    N/A. This is a medical device performance and safety (bench and biocompatibility) study, not an AI/ML diagnostic accuracy study requiring expert ground truth for interpretation.

    4. Adjudication Method for the Test Set

    N/A. Not applicable for bench and biocompatibility testing. Ground truth for these tests is based on physical measurements, chemical analyses, and biological responses according to established standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This document does not mention any MRMC comparative effectiveness study, nor does it involve human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This document is for a physical medical device, not an algorithm or AI model.

    7. The Type of Ground Truth Used

    For the bench testing, the "ground truth" is based on:

    • Physical measurements: E.g., dimensions (OD, ID, length), force at break, unscrewing torque, flow rates, burst pressures. These are measured objectively against defined specifications.
    • Qualitative assessments: E.g., presence/absence of kinks, visibility under fluoroscopy, ease of assembly, presence of stress cracks, evaluation of surface contaminants. These are typically assessed by trained technicians against defined criteria.
    • Performance metrics: E.g., successful tracking, minimal resistance, non-overriding threads. These are observed outcomes during simulated use or specific tests.

    For biocompatibility testing, the "ground truth" is based on:

    • Standardized biological assays: E.g., MEM elution assay for cytotoxicity, Kligman Maximization Test for sensitization, rabbit blood tests for hemocompatibility, rabbit pyrogen test, dog thrombogenicity study. These tests produce quantitative or semi-quantitative results that are interpreted against established acceptance criteria defined in ISO 10993 series of standards.
    • Pathology/Histology: For the thrombogenicity study, "minimal thrombosis observed" would be a histological assessment.

    8. The Sample Size for the Training Set

    N/A. This is not an AI/ML study, so there is no training set.

    9. How the Ground Truth for the Training Set Was Established

    N/A. This is not an AI/ML study, so there is no training set or ground truth for one.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1