Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K171041

    Validate with FDA (Live)

    Device Name
    SilverSeal
    Date Cleared
    2017-08-22

    (137 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For Over-the-Counter Use: SilverSeal Hydrogel Wound Dressing made with X-Static is indicated for the topical management of minor superficial wounds, minor lacerations, minor abrasions and minor first degree burns.

    Device Description

    Hydrogel Wound Dressings with X-Static are designed to intimately contact the wound as a primary dressing and permit the passage of fluids. Hydrogel Wound Dressings with X-Static are made of flexible sterile nonadherent hydrogel consisting of water and polymer with nylon fibers coated with metallic silver suspended in the hydrogel.

    AI/ML Overview

    The provided document is a 510(k) summary for the SilverSeal® Hydrogel Wound Dressing, a medical device. It does not contain the information requested regarding acceptance criteria and a study proving device performance in the context of an AI/algorithm-based medical device.

    The document states:
    "Hydrogel Wound Dressing with X-Static is the same product that was cleared in K040019 as the X-Static SilverSeal Hydrogel Wound Dressing. Therefore, the performance data provided in K040019 (section 6.2 pages 12-13 of K040019) is applicable to the 510(k) and no new performance data is provided."

    This means that the current submission (K171041) relies on the performance data of a previously cleared device (K040019) and does not include any new studies or performance data. Furthermore, the device is a wound dressing, not an AI or algorithm-based device.

    Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment, as these concepts are not applicable to the type of device and submission described in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063190

    Validate with FDA (Live)

    Date Cleared
    2007-08-21

    (305 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Silverseal® Wound Packing Strips with X-Static® are intended for the control of local wound bleeding and nasal hemorrhage, and to encourage draining by wicking fluids from a body cavity, infected area, or abscess, and to help remove necrotic tissue from ulcers or other infected wounds when used as a wet-to-dry packing.

    Device Description

    SILVERSEAL® Wound Packing Strip with X-Static® is a flexible, non-adherent, protective dressing consisting of nylon fibers with a metallic silver surface. It is made of flexible, sterile, non-adherent fabric consisting of a knitted continuous nylon fiber substrate with a metallic silver surface containing approximately 1.5% silver oxide.

    AI/ML Overview

    The provided text is a 510(k) summary for the SILVERSEAL® Wound Packing Strips with X-Static®. This type of document is for clearance of a medical device, and therefore focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria in a detailed clinical study with numerical thresholds.

    As such, many of the requested categories for a rigorous study design (e.g., sample size for test set, number of experts, adjudication methods, multi-reader multi-case studies, effect size, training set details) are not applicable or not present in this type of submission. The primary "study" for a 510(k) is a comparison to a legally marketed predicate device.

    Here's an analysis based on the information provided:


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission focused on substantial equivalence, there are no explicit numerical acceptance criteria described in the same way one might find in a clinical trial protocol (e.g., "sensitivity must be >90%"). Instead, the "acceptance criteria" are implied by demonstrating that the new device shares fundamental technological characteristics, intended use, and similar performance to the predicate devices.

    Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance
    Functional Design (Similar to predicates)Flexible, non-adherent, protective dressing. Consists of nylon fibers with a metallic silver surface.
    Performance (Similar to predicates)Designed to intimately contact the wound as a primary dressing and permit the passage of fluids. Silver provides effective protection of the dressing against microbial contamination.
    Intended Use (Similar to predicates)Control of local wound bleeding and nasal hemorrhage; encourages draining by wicking fluids from a body cavity, infected area, or abscess; helps remove necrotic tissue from ulcers or other infected wounds when used as a wet-to-dry packing.
    Biocompatibility (Safe for intended use)Passed standard in vivo biocompatibility tests: cytotoxicity, sensitization, and acute intracutaneous reactivity. Performed in accordance with ISO 10993.
    Technological CharacteristicsFlexible primary contact wound dressing, permeability to oxygen and fluids, and protection against microbial contamination of the dressing. Substantially equivalent to cited predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable. There was no specific "test set" of cases or patients in the sense of a clinical performance study with statistical endpoints. The evaluation was primarily based on material characterization and in vivo biocompatibility testing.
    • Data Provenance: The biocompatibility tests were in vivo studies, likely conducted in animal models, but specific details on the "country of origin" or "retrospective/prospective" nature of patient data (which is not applicable here) are not provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable. There was no "ground truth" established by experts for a test set of clinical cases. The evaluation involved laboratory testing and comparison to existing device classifications and intended uses.

    4. Adjudication Method for the Test Set

    • Not applicable. No adjudication method was used, as there was no clinical test set requiring expert review or consensus.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • Not applicable. This device is a wound packing strip, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study is irrelevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used

    • The "ground truth" for the biocompatibility assessment was derived from standardized in vivo biological evaluations based on ISO 10993, rather than expert consensus, pathology, or outcomes data from human patients for performance claims.
    • The "ground truth" for substantial equivalence relies on the established intended use, technological characteristics, and safety profile of the legally marketed predicate devices.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI or machine learning device requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K050726

    Validate with FDA (Live)

    Date Cleared
    2005-12-05

    (262 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For over-the-counter use, SILVERSEAL® Adhesive Strips with X-Static® may be used for: First aid to help in minor abrasions, minor cuts, lacerations, scrapes, minor scalds and burns. Under the supervision of a health care professional, SILVERSEAL® Adhesive Strips may be used for the management of: SILVERSEAL® Adhesive Strips with X-Static® are sterile, non-adherent dressings intended for local management of partial thickness burns, incisions, skin grafts, donor sites, lacerations, abrasions, and Stage I-IV dermal ulcers (vascular, venous, pressure and diabetic).

    Device Description

    SILVERSEAL® Adhesive Strips with X-Station are designed to intimately contact the wound as a primary dressing and permit the passage of fluids. The silver provides effective protection of the dressing abainst microbial contamination. The nylon fabric permits the passage of exygen and fluids to and from the wound. The surface of the nylon fibers in SILVERSEAL® Adhesive Strips with X-Static; consists or a thin layer of approximately 22. metallic silver containing approximately 1.5 silver oxide that provides effective protection of the dressing against microbial contamination. SILVERSEAL: Adhesive Strips are made of flexible, sterile, non-adherent fabric consisting of a knitted continuous nylon fiber substrate with 22% metallic silver surface containing approximately 1.58 silver oxide.

    AI/ML Overview

    This 510(k) premarket notification for the SILVERSEAL® Adhesive Strip with X-Static® does not include acceptance criteria or a study proving that the device meets specific acceptance criteria in the way a clinical trial or performance study for a diagnostic device would.

    Instead, the submission focuses on substantial equivalence to existing predicate devices. The performance data provided are biocompatibility tests, which are standard for medical devices that come into contact with the body.

    Here's an breakdown based on the information provided, highlighting why some sections of your request cannot be fully addressed:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The acceptance criteria for this type of device (biocompatibility) are generally qualitative (e.g., "safe," "non-sensitizing") rather than quantitative performance metrics like sensitivity/specificity for a diagnostic device. The document states all tests were performed in accordance with ISO 10993 and indicated safety.

    Feature / Test CategoryAcceptance Criteria (Implied by ISO 10993)Reported Device Performance
    BiocompatibilityNo cytotoxicityMet (indicated safe)
    No sensitizationMet (indicated safe)
    No acute intracutaneous reactivityMet (indicated safe)

    Explanation of Device Performance:

    • Biocompatibility: The device was subjected to standard in vivo biocompatibility tests, including cytotoxicity, sensitization, and acute intracutaneous reactivity.
    • Results: All tests were performed in accordance with ISO 10993 (Biological Evaluation of Medical Devices) by Nittin American Science Associates, Inc. (NAMSA). The studies indicated that SILVERSEAL® Adhesive Strips with X-Station are safe for their intended use.
    • Microbial Contamination Protection (of dressing): The silver and silver oxide in the dressing provide "effective protection of the dressing against microbial contamination." This claim is based on "In Vitro studies" and has not been studied in a clinical setting. This is a functional claim for the dressing itself, not a direct patient outcome claim.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Biocompatibility): Not specified in the provided document. Biocompatibility tests usually involve animal models (in vivo) or cell cultures (in vitro), with a specific number of subjects/samples depending on the ISO 10993 part being followed. The document only mentions "standard in vivo biocompatibility tests."
    • Data Provenance: Not explicitly stated, but "Nittin American Science Associates, Inc. (NAMSA)" is a contract research organization (CRO) with a global presence, often performing tests in various locations.
    • Retrospective or Prospective: Biocompatibility tests are typically prospective studies performed specifically for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable in this context. For biocompatibility testing, the "ground truth" is established by adherence to standardized protocols (ISO 10993) and established scientific methods, rather than expert consensus on complex diagnostic interpretations. The CRO (NAMSA) is implicitly qualified to perform these tests.

    4. Adjudication Method for the Test Set

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used for expert image review or clinical outcome assessment in studies where human judgment is a primary input. Biocompatibility testing relies on objective, laboratory-based measurements and observations according to validated protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. An MRMC study is not relevant for this device. These studies are typically performed for diagnostic imaging devices to evaluate how AI affects human reader performance.

    6. If a Standalone Study Was Done

    • Yes, in essence. The biocompatibility studies described are "standalone" in the sense that they evaluate the device's inherent safety properties without human intervention in the loop. The "In Vitro studies" for microbial protection are also standalone algorithm/device performance, though the details are not provided.
    • "Algorithm only without human-in-the loop performance": This phrasing is more applicable to AI/software as a medical device. For a wound dressing, the "algorithm" is the physical/chemical properties of the dressing. Its performance (e.g., biocompatibility or antimicrobial action against the dressing itself) is assessed directly.

    7. The Type of Ground Truth Used

    • Biocompatibility: The "ground truth" is established by standardized biological responses observed in validated in vivo and in vitro models, as defined by ISO 10993. This is based on established scientific principles regarding material interaction with biological systems.
    • Microbial Protection: The ground truth for the dressing's protection against microbial contamination is based on in vitro microbiological assays, which quantitatively measure microbial growth on the dressing material under controlled conditions. This is a form of laboratory-based evidence.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a physical wound dressing, not an AI algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, there is no "training set" for this type of device.

    In summary, the 510(k) submission for the SILVERSEAL® Adhesive Strip with X-Static® relies on demonstrating substantial equivalence to predicate devices and proving biocompatibility through standard in vivo tests. It also mentions in vitro studies for protection against microbial contamination of the dressing itself. The performance data presented are not in the context of diagnostic AI algorithms, so many of the requested categories (like sample size for test/training sets, expert ground truth, adjudication, or MRMC studies) are not directly applicable or available in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051256

    Validate with FDA (Live)

    Date Cleared
    2005-07-21

    (66 days)

    Product Code
    Regulation Number
    890.3420
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SILVERSEAL® Orthotic Components with X-Static® are used as components on various orthotic devices intended for medical purposes that can be worn as a cervical collar or brace. The component comprises a knitted stockinette which is made of 90% cotton or polyester, and 10% X-Static®, which is 1.5% silver oxide. The component provides the skin with an antimicrobial barrier.

    Device Description

    SILVERSEAL® Orthotic Components with X-Static® are made of flexible, non-adherent fabric consisting of a knitted or made or in fabric which is 90% cotton or polyester, and 10% of a continuous nylon fiber substrate with a metallic silver surface containing approximately 1.5% silver oxide.

    AI/ML Overview

    The Noble Fiber Technologies, Inc. 510(k) submission for the SILVERSEAL® Orthotic Components with X-Static® does not include acceptance criteria or a study that evaluates device performance against specific efficacy endpoints in the same way a diagnostic or therapeutic device might. Instead, this submission focuses on biocompatibility and substantial equivalence to a predicate device.

    Here's an breakdown of the relevant information provided:

    1. A table of acceptance criteria and the reported device performance:

    This document does not present quantitative acceptance criteria or a performance table for efficacy. The "performance" assessment is based on biocompatibility and antimicrobial properties, which are discussed qualitatively.

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    BiocompatibilityAbsence of sensitization, irritation, and acute in vivo blood incompatibility reactions"All tests were performed in accordance with [ISO] Standard 10995." "The studies indicated that SILVERSEAL® Orthotic Components with X-Static® are safe for their intended use."
    Antimicrobial PropertyProvides an antimicrobial barrier to the skin."The technological characteristics of the device, aids in the protection against microbial contamination of the skin that are substantially equivalent to the predicate devices cited." (This refers to the inherent property of X-Static® silver.)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    The document states that the device was "subjected to standard in vivo biocompatibility" tests, including sensitization, irritation, and acute in vivo blood incompatibility. However, it does not specify the sample size for these tests. It also does not explicitly state the country of origin of the data or whether the tests were prospective or retrospective, although biocompatibility testing is typically prospective. The testing was performed by "North American Science Associates, Inc. (NAMSA)."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This information is not provided. Biocompatibility testing typically involves laboratory professionals and toxicologists, but no details on "experts" or their qualifications for establishing ground truth are given in this K051256 summary.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not provided. Adjudication methods are not typically relevant for standard biocompatibility testing. The assessment of biocompatibility usually relies on established laboratory protocols and interpretation of results against predetermined biological endpoints.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. This device is a component of an orthotic, and its performance evaluation is based on biocompatibility and antimicrobial properties, not on diagnostic or therapeutic effectiveness requiring human reader-dependent interpretation or AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable. This device is not an algorithm or AI system. Its performance relates to material properties and biological interaction.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth for biocompatibility testing is based on established biological endpoints and toxicological principles, as defined by international standards (ISO 10995, likely a typo for ISO 10993 which is for Biological Evaluation of Medical Devices). This involves evaluating for adverse biological reactions such as irritation, sensitization, and specific toxic effects, usually observed in animal models (in vivo) or in vitro cell cultures.

    8. The sample size for the training set:

    This is not applicable. This device is not an AI/ML algorithm that requires a "training set."

    9. How the ground truth for the training set was established:

    This is not applicable. As mentioned above, there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050649

    Validate with FDA (Live)

    Date Cleared
    2005-06-13

    (91 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For over-the-counter use, Silverseal® Wound Contact Dressing with X-Static® may be used for first aid management of minor abrasions, cuts, scrapes, scalds and burns.

    Under the supervision of a health care professional, Silverseal® Contact Wound Dressing may be used for the management of partial thickness burns, incisions, skin grafts, donor sites, lacerations, abrasions, and Stage I-IV dermal ulcers (vascular, venous, pressure and diabetic).

    Device Description

    SILVERSEAL® Burn Contact Wound Dressings with X-Static® are designed to intimately contact the wound as a primary dressing and permit the passage of fluids. The silver provides effective protection of the dressing against microbial contamination.

    The nylon fabric permits the passage of oxygen and fluids to and from the wound. The surface of the nylon fibers in SILVERSEAL® Contact Wound Dressings with X-Static® consists of a thin layer of metallic silver containing approximately 1.5% silver oxide that provides effective protection of the dressing against microbial contamination.

    SILVERSEAL® Contact Wound Dressings are made of flexible, sterile, non-adherent fabric consisting of a knitted continuous nylon fiber substrate with a metallic silver surface containing approximately 1.5% silver oxide.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the SILVERSEAL® Burn Glove with X-Static® and related dressings:

    No acceptance criteria or study demonstrating performance against such criteria are explicitly provided in this 510(k) summary.

    The document primarily focuses on establishing substantial equivalence to predicate devices based on technological characteristics and biocompatibility.

    Here's a breakdown of the requested information based on the provided text, highlighting what is present and what is not:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does NOT explicitly state any acceptance criteria for device performance (e.g., specific percentages for infection reduction, healing rates, or other clinical outcomes).

    Instead, the "performance" discussed is related to biocompatibility and protection against microbial contamination of the dressing itself.

    Acceptance Criteria (Explicitly Stated)Reported Device Performance
    N/A (No specific numerical acceptance criteria for clinical efficacy are provided.)- Biocompatibility: Passed standard in vivo tests (cytotoxicity, sensitization, acute intracutaneous reactivity) in accordance with ISO 10993. - Protection Against Microbial Contamination (of the dressing): The silver content provides effective protection of the dressing against microbial contamination (based on in vitro studies).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set for Biocompatibility: The sample size for the in vivo biocompatibility tests (cytotoxicity, sensitization, acute intracutaneous reactivity) is not specified.
    • Data Provenance: The biocompatibility tests were performed by North American Science Associates, Inc. (NAMSA), which is a common contract research organization for medical device testing. The country of origin for the data is implicitly the USA, where NAMSA operates. These tests are typically prospective in nature, conducted specifically for the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable.

    • For biocompatibility tests, "ground truth" is typically established by laboratory standards and protocols (e.g., observing for specific cellular reactions, skin reactions) rather than expert consensus on a clinical outcome.
    • There is no mention of a human-centric "test set" requiring expert consensus for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable.

    As there is no "test set" requiring human interpretation or clinical judgment in the traditional sense for efficacy or diagnostic performance, no adjudication method is mentioned. Biocompatibility results are typically objective measurements or observations against established standards.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was NOT done.

    The document explicitly states: "All claims are the result of In Vitro studies and have not been studied in a clinical setting." This confirms that no clinical studies comparing the device's effectiveness, either with or without AI assistance, were performed.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not Applicable, and no such study was performed.

    This is a medical dressing, not an algorithm or AI device. Therefore, the concept of "standalone algorithm performance" does not apply. The performance evaluation was focused on materials, biocompatibility, and antimicrobial properties of the dressing itself.


    7. The Type of Ground Truth Used

    • Biocompatibility: The "ground truth" for biocompatibility was established through performance against standardized laboratory protocols and observations for cytotoxicity, sensitization, and acute intracutaneous reactivity, as defined by ISO 10993.
    • Antimicrobial properties (of the dressing): The "ground truth" for protection against microbial contamination of the dressing was established through in vitro studies, which involve lab-based testing of the material's ability to inhibit microbial growth.

    8. The Sample Size for the Training Set

    Not applicable.

    This product is a medical dressing, not an AI/machine learning algorithm. Therefore, there is no "training set" in the context of AI model development.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    As there is no AI algorithm or "training set," there is no ground truth established in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043526

    Validate with FDA (Live)

    Date Cleared
    2005-03-10

    (79 days)

    Product Code
    Regulation Number
    888.5940
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K033587

    Validate with FDA (Live)

    Device Name
    SILVERSEAL
    Date Cleared
    2004-08-05

    (266 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    X-Static® SILVERSEAL™ Contact Wound Dressings are sterile, non-adherent dressings intended for local management of partial thickness burns, incisions, skin grafts, donor sites, lacerations, abrasions, and Stage I-IV dermal ulcers (vascular, venous, pressure, and diabetic).

    Device Description

    X-Static® SILVERSEAL™ Contact Wound Dressings are designed to intimately contact the wound as a primary dressing and permit the passage of fluids. The silver provides effective protection of the dressing against microbial contamination. The nylon fabric permits the passage of oxygen and fluids to and from the wound. The surface of the nylon fibers in X-Static® SILVERSEAL™ Contact Wound Dressing consists of a thin layer of metallic silver containing approximately 1% silver oxide that provides effective protection of the dressing against microbial contamination. X-Static® SILVERSEAL™ Contact Wound Dressings are made of flexible, sterile, non-adherent fabric consisting of 1 or 4 layers of a knitted continuous nylon fiber substrate with a metallic silver surface containing approximately 1% silver oxide.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a wound dressing, not a study evaluating an AI/ML medical device. Therefore, the information requested in your prompt (e.g., acceptance criteria for device performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types) is not present in the document.

    The document describes the device, its intended use, and claims substantial equivalence to predicate devices based on technological characteristics and biocompatibility testing.

    Here's what I can extract related to performance in the context of this specific regulatory submission:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Implied for Biocompatibility)Reported Device Performance
    Non-cytotoxicPassed biocompatibility tests
    Non-sensitizingPassed biocompatibility tests
    Non-acutely intracutaneous reactivePassed biocompatibility tests

    Explanation of Device Performance:

    The document explicitly states: "X-Static® SILVERSEAL™ 1-Layer and 4-Layer Contact Wound Dressings were subjected to standard in vivo biocompatibility tests including cytotoxicity, sensitization, and acute intracutaneous reactivity. All tests were performed in accordance with Part-10993 of the International Standard Organization (ISO) Standard (Biological Evaluation of Medical Devices) by North American Science Associates, Inc. (NAMSA). All claims are the result of In Vitro studies and have not been studied in a clinical setting." and "The studies indicated that X-Static® SILVERSEAL™ 1 Layer and 4 Layer Contact Wound Dressings are safe for their intended use."

    Regarding the other points in your request:

    • 2. Sample sized used for the test set and the data provenance: Not applicable. Biocompatibility tests typically involve animal models (in vivo) and cell cultures (in vitro), but specific sample sizes and provenance for these are not detailed in this summary. The "claims are the result of In Vitro studies" suggesting cell culture or non-living material tests, but no detail on sample size is provided.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Biocompatibility testing relies on standardized protocols and laboratory analysis, not expert consensus on image/diagnostic interpretation.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.
    • 7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For biocompatibility, the "ground truth" is defined by the results of standardized biological tests (e.g., cell viability, immune response indicators) against established safety thresholds.
    • 8. The sample size for the training set: Not applicable. There is no training set mentioned as this is not an AI/ML device.
    • 9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K040019

    Validate with FDA (Live)

    Device Name
    SILVERSEAL
    Date Cleared
    2004-07-23

    (199 days)

    Product Code
    Regulation Number
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X-Static® Silverseal™ Hydrogel Wound Dressing is intended for the management of wounds and to provide an antimicrobial barrier. The dressing is indicated for use on partial and full thickness dermal ulcers, leg ulcers, superficial wounds, abrasions, first and second degree burns, donor sites and over debrided and grafted partial thickness wounds.

    Device Description

    X-Static® SILVERSEAL™ Hydrogel Wound Dressings are designed to intimately contact the wound as a primary dressing and permit the passage of fluids. The silver provides effective protection of the dressing against microbial contamination. The nylon fabric permits the passage of oxygen and fluids to and from the wound. The surface of the nylon fibers in X-Static® SILVERSEAL™ Hydrogel Wound Dressing consists of a thin layer of metallic silver containing approximately 1.5% silver oxide that provides effective protection of the dressing against microbial contamination. X-Static® SILVERSEAL™ ilydrogel Wound Dressings are made of flexible, sterile, non-adherent fabric consisting of 1 layer of a knitted continuous nylon fiber substrate with metallic silver surface containing approximately 1% silver oxide.

    AI/ML Overview

    The provided text describes the X-Static® SILVERSEAL™ Hydrogel Wound Dressing, its intended use, and its substantial equivalence to predicate devices, but it does not contain information about acceptance criteria or a study proving that the device meets those criteria in terms of performance metrics like sensitivity, specificity, or improvement with AI assistance.

    The "Assessment of Performance Data" section primarily focuses on safety testing (cytotoxicity, sensitization, intracutaneous, and acute reactivity) performed in accordance with ISO 10993 standards. It explicitly states, "All antibiotic claims are the result of In Vitro studies and that the effect of the antibiotic has not been studied in a clinical setting." This means there is no clinical data presented to support performance related to antimicrobial efficacy on actual wounds or other clinical outcomes.

    Therefore, I cannot fill out the requested table or answer the questions related to performance criteria, sample sizes for test/training sets, expert involvement, or AI-related metrics because this information is not present in the provided document.

    The document discusses:

    • Safety Assessments: The device underwent standard tests for cytotoxicity, sensitization, intracutaneous, and acute reactivity as per ISO 10993.
    • Antimicrobial Claims Basis: Antimicrobial claims are based solely on in vitro studies, and the effect has not been studied in a clinical setting.

    Without further information describing performance criteria and a study that assesses the device against those criteria (beyond basic safety), a complete answer to your request cannot be generated from the given text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1