Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K963542

    Validate with FDA (Live)

    Date Cleared
    1996-09-27

    (23 days)

    Product Code
    Regulation Number
    862.1665
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The safety and effectiveness of Sigma Diagnostics Electrolyte Reference Reagent, Procedure Number E3891, are demonstrated by its substantial equivalency to Beckman Electrolyte Reference Reagent Kit, Part No. 443315. Both electrolyte reference reagents are used to measure sodium, chloride, potassium, and carbon dioxide concentrations in serum or plasma, sodium, potassium, and chloride concentrations in urine, and chloride concentrations in CSF on the SYNCHRON CX®3 System, and the reaction principles for both reagents are identical.

    Device Description

    The Sigma Diagnostics methods use ion selective electrodes for determining sodium, potassium, and chloride and rate of pH change for determining carbon dioxide on the SYNCHRON CX®3 System.

    AI/ML Overview

    The provided text describes the performance of the Sigma Diagnostics Electrolyte Reference Reagent (Procedure Number E3891) in comparison to the Beckman Electrolyte Reference Reagent Kit (Part No. 443315) on the SYNCHRON CX®3 System. The study aims to demonstrate substantial equivalence.

    Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" as defined thresholds for correlation coefficients or regression equations. Instead, it presents the achieved performance metrics as evidence of substantial equivalence to the predicate device. The implied acceptance is that the correlation and regression fall within an acceptable range for a substantially equivalent device. Similarly, for precision, the stated %CVs are the achieved performance, implying these are considered acceptable.

    Metric (Analyte, Sample Type)Reported Sigma Diagnostics PerformanceImplied Acceptance Criteria (relative to predicate)
    Correlation Coefficient
    Sodium (Serum)0.963High correlation (implied >~0.95 given the values)
    Potassium (Serum)0.999High correlation
    Chloride (Serum)0.973High correlation
    Total CO2 (Serum)0.992High correlation
    Sodium (Urine)0.997High correlation
    Potassium (Urine)0.998High correlation
    Chloride (Urine)0.997High correlation
    Chloride (CSF)0.891High correlation (lower but still presented as acceptable)
    Regression Equation (y = Sigma, x = Beckman)
    Sodium (Serum)y = 0.90x + 13.48Slope close to 1, intercept close to 0 (indicating good agreement)
    Potassium (Serum)y = 0.98x + 0.08Slope close to 1, intercept close to 0
    Chloride (Serum)y = 0.91x + 10.52Slope close to 1, intercept close to 0
    Total CO2 (Serum)y = 0.95x + 1.54Slope close to 1, intercept close to 0
    Sodium (Urine)y = 0.95x - 0.10Slope close to 1, intercept close to 0
    Potassium (Urine)y = 0.97x + 0.49Slope close to 1, intercept close to 0
    Chloride (Urine)y = 0.97x + 2.06Slope close to 1, intercept close to 0
    Chloride (CSF)y = 0.98x + 4.00Slope close to 1, intercept close to 0
    Precision (%CV)
    Sodium (Serum)< 1.1% (within-run & total)Low %CV (implied acceptable precision)
    Sodium (Urine)< 1.2% (within-run & total)Low %CV
    Potassium (Serum)< 1.0% (within-run & total)Low %CV
    Potassium (Urine)< 1.6% (within-run & total)Low %CV
    Chloride (Serum)< 1.3% (within-run & total)Low %CV
    Chloride (Urine)< 1.9% (within-run & total)Low %CV
    Chloride (CSF)< 1.8% (within-run & total)Low %CV
    Linearity Range (mmol/L)
    Sodium (Serum)100.0 - 200.0Wide appropriate range
    Sodium (Urine)10 - 200Wide appropriate range
    Potassium (Serum)1.00 - 15.00Wide appropriate range
    Potassium (Urine)2 - 200Wide appropriate range
    Chloride (Serum)50.0 - 200.0Wide appropriate range
    Chloride (Urine)15 - 300Wide appropriate range
    Chloride (CSF)50 - 200Wide appropriate range
    Total CO2 (Serum)5.0 - 40.0Wide appropriate range

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the sample size used for the comparison studies. It refers to "comparison studies" and "serum samples," "urine samples," and "CSF samples" in a general sense, but no numerical count is provided.

    The data provenance is also not explicitly stated. It's presented as data obtained in "comparison studies" without mentioning the country of origin or whether the data was retrospective or prospective. Given the nature of a 510(k) submission, it would likely be prospective data collected specifically for the submission, but this is not confirmed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    This information is not applicable to this type of device and study. The "ground truth" here is the measurement obtained by the predicate device (Beckman Electrolyte Reference Reagent Kit), not an expert interpretation. This is a direct analytical comparison.

    4. Adjudication Method for the Test Set:

    This information is not applicable. There is no "adjudication" in the sense of resolving discrepancies between human readers or between a device and human readers. The comparison is between two analytical devices.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. This document describes an in-vitro diagnostic (IVD) device for measuring electrolytes, not an AI-assisted diagnostic imaging or interpretation system that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, the study described is essentially a standalone performance evaluation of the Sigma Diagnostics Electrolyte Reference Reagent. It's comparing the device's measurements directly against those of a predicate device, without human intervention in the measurement process itself.

    7. The type of ground truth used:

    The "ground truth" in this context is the measurements provided by the predicate device, the Beckman Electrolyte Reference Reagent Kit, Part No. 443315, on the SYNCHRON CX®3 System. The study's purpose is to show consistency with this established method.

    8. The Sample Size for the Training Set:

    This information is not applicable. The device is a chemical reagent kit used for analytical measurement, not a machine learning model that requires a "training set" in the conventional sense.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1