510(k) Data Aggregation

    K Number
    K143220

    Validate with FDA (Live)

    Date Cleared
    2015-01-08

    (59 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Wright's SIDEKICK® EZ FRAME™ External Fixation System is intended for:

    • · Triple Arthrodesis
    • · Isolated Rearfoot Arthrodesis
    • · Midfoot Arthrodesis
    • Comminuted Trauma
    • · Diabetic Charcot Reconstruction
    • · Most foot pathology not requiring fixation above the ankle
    Device Description

    The SIDEKICK® EZ FRAME™ External Fixation System uses a series of pins and wires for compression or distraction in the foot and uses a boot to secure the tibia. The system is adjustable to accommodate variations in patient anatomy. The SIDEKICK® EZ FRAMETM External Fixation System is compatible with the SIDEKICK® FREEDOM External Fixation System (K043289) components, and these components provide additional fixation options in the system.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the SIDEKICK® EZ FRAME™ External Fixation System. It asserts substantial equivalence to a predicate device, rather than providing a study demonstrating its own performance against acceptance criteria. Therefore, most of the requested information regarding acceptance criteria and performance studies for the device in question cannot be extracted from this document.

    Here's what can be extracted based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide specific acceptance criteria or reported device performance in terms of metrics. Instead, it relies on demonstrating substantial equivalence to a predicate device.

    Acceptance CriteriaReported Device Performance
    Not applicable. The submission focuses on demonstrating substantial equivalence to a predicate device, not on meeting specific performance criteria for the new device.Not applicable. Device performance metrics against specific acceptance criteria are not presented.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. The submission states "Substantial Equivalence - Non-Clinical Evidence" and "Through technological comparison, the subject system does not represent a new worst-case." and "Substantial Equivalence - Clinical Evidence N/A". This indicates that no clinical or performance test set was used for the sidekick EZ FRAME in this submission. The evaluation is based on technological comparison and material/design similarities to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No test set was used.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a mechanical external fixation system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a mechanical external fixation system, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    Not applicable. Ground truth is not relevant in a substantial equivalence filing for a mechanical device based on technological comparison.

    8. The sample size for the training set

    Not applicable. There is no mention of a training set for an algorithm.

    9. How the ground truth for the training set was established

    Not applicable. There is no mention of a training set or ground truth for it.

    Summary based on the document:

    The document asserts that the SIDEKICK® EZ FRAME™ External Fixation System is substantially equivalent to a legally marketed predicate device (K130044: SIDEKICK® EZ FRAME™ External Fixation System) based on technological characteristics comparison. The submission explicitly states:

    • Non-Clinical Evidence: "Through technological comparison, the subject system does not represent a new worst-case."
    • Clinical Evidence: "N/A" (Not Applicable)
    • Conclusion: "The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate device."

    This means that the "study" proving the device meets criteria is primarily a comparative analysis of its design, materials, and intended use against an already approved predicate device, demonstrating that it does not introduce new safety or effectiveness concerns. No new performance studies with acceptance criteria, test sets, or ground truth establishment were conducted or reported for this 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130044

    Validate with FDA (Live)

    Date Cleared
    2013-04-05

    (87 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Triple arthrodesis .
    • lsolated rearfoot arthrodesis .
    • Midfoot arthrodesis
    • . Comminuted trauma
    • . Diabetic Charcot reconstruction
    • Most foot pathology not requiring fixation above the ankle .
    Device Description

    The SIDEKICK® EZ FRAME™ External Fixation System uses a series of pins and wires for compression or distraction in the foot and uses a boot to secure the tibia. The system is adjustable to accommodate variations in patient anatomy. The SIDEKICK® EZ FRAME™ External Fixation System is compatible with the SIDEKICK® FREEDOM External Fixation System (K043289) components, and these components provide additional fixation options in the system.

    AI/ML Overview

    The provided documentation describes a 510(k) submission for the SIDEKICK® EZ FRAME™ External Fixation System, seeking substantial equivalence to a predicate device. This type of submission focuses on demonstrating that a new device is as safe and effective as a legally marketed predicate device, rather than providing extensive clinical studies to prove effectiveness against predefined acceptance criteria for a novel device.

    Therefore, many of the requested details regarding acceptance criteria, extensive study design, ground truth, and reader studies are not applicable or explicitly stated in this type of submission for this device.

    Here's an analysis based on the information provided:


    Analysis of Acceptance Criteria and Study for SIDEKICK® EZ FRAME™ External Fixation System

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is a 510(k) submission for an external fixation system, the "acceptance criteria" are primarily established by demonstrating substantial equivalence to a predicate device through mechanical and design comparisons, rather than clinical performance metrics like sensitivity, specificity, or accuracy.

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical Equivalence: Does the new device perform mechanically at least as well as the predicate device, without representing a new worst-case scenario?The submission states: "Through mechanical component analysis and comparison the subject system does not represent a new worst-case." This indicates that mechanical performance was evaluated against the predicate and deemed equivalent or better.
    Design and Materials Equivalence: Are the design characteristics, materials, and method of operation similar enough to the predicate device?The submission states: "The subject device operates in the same method as the predicate" and "The design characteristics of the subject system do not raise new types of questions of safety or effectiveness." This implies that the design and method of operation are demonstrably similar to the predicate.
    Intended Use Equivalence: Does the new device have the same or similar intended uses as the predicate device?The intended uses listed for the SIDEKICK® EZ FRAME™ (Triple arthrodesis, Isolated rearfoot arthrodesis, Midfoot arthrodesis, Comminuted trauma, Diabetic Charcot reconstruction, Most foot pathology not requiring fixation above the ankle) are implicitly considered equivalent or acceptable in scope compared to the predicate device, as FDA cleared the device for these indications.
    Safety and Effectiveness: Does the device perform at least as well as the predicate systems without raising new questions of safety or effectiveness?The submission concludes: "From the evidence submitted in this 510(k), the subject device system can be expected to perform at least as well as the predicate systems." The FDA's clearance letter confirms this conclusion regarding substantial equivalence.

    2. Sample Size Used for the Test Set and the Data Provenance:

    • Not Applicable. This submission relies on substantial equivalence through mechanical and design comparison, not a "test set" of patient data in the typical sense of a clinical trial for a novel diagnostic or therapeutic device. The "test set" would primarily consist of the device components themselves undergoing mechanical testing. The document does not specify a sample size for such mechanical tests, nor does it detail data provenance in terms of patient population or origin.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • Not Applicable. Ground truth, in the context of clinical studies involving expert interpretation, is not relevant here. The evaluation is against the performance and characteristics of a predicate device, which itself has established safety and effectiveness.

    4. Adjudication Method for the Test Set:

    • Not Applicable. As there is no "test set" involving human interpretation or clinical endpoints requiring adjudication, no adjudication method is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    • No. This device is an external fixation system (a hardware medical device), not an AI-powered diagnostic or interpretive tool. Therefore, an MRMC study comparing human reader performance with or without AI assistance is not relevant and was not conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • No. This is a physical medical device, not a software algorithm. Therefore, "standalone" performance in the context of an algorithm is not applicable.

    7. The Type of Ground Truth Used:

    • Not Applicable / Predicate Device Performance. For this type of submission, the "ground truth" is effectively the established safety and effectiveness profile, mechanical performance, and design characteristics of the legally marketed predicate device. The new device's performance is gauged against this established benchmark.

    8. The Sample Size for the Training Set:

    • Not Applicable. As this is not an AI/ML device, there is no "training set" of data as understood in that context. The device's design and manufacturing rely on established engineering principles and materials science, not machine learning algorithms trained on data.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. Since there is no training set, there is no ground truth to establish for it.