Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K123149

    Validate with FDA (Live)

    Date Cleared
    2014-06-20

    (623 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SERVO i Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants, and adults with respiratory failure or respiratory insufficiency. SERVO-i is a ventilator system to be used only by healtheare providers in hospitals or healthcare facilities and for in-hospital transport.

    The added indications for use of the NAVA option is when the electrical signal from the diaphragm is intact; NA VA will improve synchrony between the ventilator and patients with no contraindication for insertion/exchange of a Naso-Gastric tube.

    The SERVO-i Ventilator is classified as MR Conditional for 1.T. 1.5T and 3T MR scanners. This means that it is safe to use in the MR environment if the MR Environment Declaration for SERVO-i are met.

    The SERVO-i Ventilator System with Heliox option is indicated for use with the delivery of Air, Oxygen, or Heliox (a mixture of Helium and Oxygen).

    The SERVO-s Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants, and adults with respiratory failure or respiratory insufficiency. SERVO-s is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities and for in-hospital transport.

    Device Description

    The SERVO-i Ventilator System (here after called SERVO-i) is intended to provide continuous ventilation for neonate to adult patients in the weight range 0.5-250 kg and with tidal volumes from 2 mL to 4000 mL. SERVO-i consists of a Patient Unit where gases are mixed and administered, and a User Interface where the settings are made and ventilation is monitored. The ventilator delivers controlled or supported breaths to the patient, with either constant flow or constant pressure, using a set oxygen concentration. SERVO-i will produce visual and audible alarms if vital parameters vary beyond pre-set, or default, limits. The system contains provisions for at least two battery modules to supply the system in the case of mains power failure or during in-hospital transport. The ventilator functionality is controlled by software. The SERVO-i Ventilator System is available in three software versions. Infant. Adult and Universal.

    The NAVA (Neurally Adjusted Ventilatory Assist) option is a supported mode for SERVO-i that uses the Edi signal (the electrical activity of the diaphragm) as an addition to the flow/pressure trigger to synchronize the patient efforts with the onset and cycle off. The NAVA option is available in invasive and non-invasive mode.

    SERVO-i is MR conditional. The SERVO-i ventilator with MR option have been tested with 1.0, 1.5, 3.0 T scanners without impairing its performance or the image quality of the scanner. Each scanner and its environment form an individual device. The MR Environment Declaration describes how a SERVO-i with MR option can be qualified to be used with an MR scanner forming a safe Medical System. All vital parts of the ventilator have been tested for performance in excessive magnetic fields.

    The SERVO-i with Heliox option requires a different mechanical adaptor on the air supply inlet to allow a mixture of Helium and Oxygen to be connected. Furthermore is the software updated to allow safe delivery and monitoring of the Heliox gas mixture.

    Accessories for CO2-monitoring, nebulization and flow monitoring at the Y -piece (Y-sensor) are integrated as options in the SERVO-i and the drivers are controlled by the software in the ventilator.

    This 510(k) submission for the SERVO-i include changes to receive a new baseline based on compatibility to the third edition standard package of AAMVANSI 60601-1 :2005 and its collateral and particular standards for intensive care ventilators. The submission does also include modifications of the software and hardware to update existing functionalities since the last submission (K073149).

    The SERVO-s ventilator system (here after called SERVO-s) is based on the SERVO-i ventilator family platform. SERVO-s ventilation system is a downscaled version based on the SERVO-i ventilator system notified in K041223.

    The SERVO-s Ventilator System is intended to provide continuous ventilation for neonate to adult patients in the weight range 2-250 kg and with tidal volumes from 10 mL to 2000 mL. The SERVO-s Ventilator System consists of a Patient Unit where gases are mixed and administered, and a User Interface where the settings are made and ventilation is monitored. The ventilator delivers controlled or supported breaths to the patient, with either constant flow or constant pressure, using a set oxygen concentration. SERVO-s Ventilator System will produce visual and audible alarms if vital parameters vary beyond preset, or default, limits. The system contains two internal batteries to supply the system with power in the case of mains power failure or during inhospital transport. The ventilator functionality is controlled by software. The SERVO-s Ventilator System is available in two software versions, Infant and Adult.

    This 510(k) submission for the SERVO-s include changes to receive a new baseline based on compatibility to the third edition standard package of IEC 60601-1 :2005 and its collateral and particular standards for intensive care ventilators. The submission does also include addition of the Infant option, patient weight range 2-10 kg, with tidal volumes from 10 mL to 350 mL and modifications of the software and hardware to update existing functionalities.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies mentioned in the provided 510(k) summary for the GETINGE GROUP SERVO-i and SERVO-s Ventilator Systems, organized according to your requested format.

    It's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. Therefore, the "studies" described are primarily verification and validation activities rather than formal clinical trials designed to prove efficacy from scratch. The acceptance criteria are largely implied by compliance with standards and successful performance in these verification and validation tests.


    Acceptance Criteria and Device Performance

    This section synthesizes the implicit acceptance criteria from the various verification and validation activities described and attempts to align them with the reported "performance" based on the conclusions drawn in the document.

    Acceptance Criteria (Implicit from V&V and predicate equivalence)Reported Device Performance (as stated or implied)
    General Device Functionality:
    Compliance with IEC 60601-1 (2005) and collateral/particular standards (e.g., ISO 80601-2-12, ISO 80601-2-55, ISO 5356-1, CGA V-5)Design verification and validation demonstrated performance within specifications and applied standards for both SERVO-i and SERVO-s.
    Software functionalityCode review, static code analysis, and regression testing performed on software changes. Free User Testing (FUT) conducted.
    Hardware functionalityDesign verification and validation performed.
    Proper functioning of User InterfaceTested as part of overall system and FUT.
    Accuracy Statements (Ventilation Parameters):Performance is the same as predicate, with updated measurement and presentation methods to comply with new standards.
    Noise Level:Performance is the same as predicate, with updated measurement and presentation methods to comply with new standards.
    Ingress Protection (IP21):Improved Ingress Protection to IP21 by chassis improvements.
    Mechanical Stability (Transport):Brakes on all four wheels (vs. previous two) to improve stability and meet new transport requirements.
    Gas Inlet Pressure Specification:Max pressure lowered to comply with required testing.
    Pre-Use Check (Patient Circuit Resistance & Compliance):Added measurement of patient circuit resistance to Pre-Use Check to comply with ISO 80601-2-12.
    NIV NAVA functionality (SERVO-i only):Added ventilation mode, leveraging existing cleared NIV and NAVA modes. Implied successful integration and functionality.
    Second RS232 port functionality (SERVO-i only):Added, identical in function to the first port. Implied proper functionality.
    Built-in Nebulizer (Aeroneb) functionality (SERVO-i only):Nebulization performance is equal to the predicate's SUN nebulizer.
    Stress Index (SI) Calculation Accuracy (SERVO-i only):<10% discrepancy when compared with an existing system used in published articles, which was within acceptance criteria.
    Nuisance Alarm Reduction (RR and MV in NAVA) (SERVO-i only):New algorithms effectively calculated RR and MV while reducing nuisance RR and MV alarms, as evaluated at MAQUET.
    Apnea Ventilation and Alarm Behavior (NAVA) (SERVO-i only):New algorithm for switching between NAVA and backup ventilation is acceptable and reduces nuisance alarms, meeting intended use and user needs.
    SERVO-s specific changes (e.g., enclosure, power supply, weight/tidal volume ranges):Adapted to new smaller enclosure, 2 internal rechargeable battery modules (vs. 2-6), 4 updated PC boards. Performance within specifications for its intended patient range (2-250 kg, 10-2000 mL tidal volume).
    Substantial Equivalence:Demonstrated through design verification, validation, and comparison to predicate devices, showing no new functions requiring clinical investigation for basic safety/efficacy.

    Study Information

    The document describes verification and validation activities rather than a single, formal "study" with a defined test set sample size for regulatory approval of this specific 510(k) submission's changes. The clinical activities mentioned are more akin to post-market evaluations or validation of specific features, built upon the foundation of the predicate devices.

    1. Sample sized used for the test set and the data provenance:

    • Test Set (General Verification & Validation): Not explicitly stated as a uniform "test set" in terms of number of patients or specific data points for the entire device. The document refers to "existing and new test cases at the system and subsystem level" and "regression tests."
    • Data Provenance (General): Not specified. Likely a combination of lab simulations, engineering tests, and potentially previous datasets from predicate device development.
    • Specific Clinical Validation Activities (SERVO-i):
      • Nuisance Alarm Reduction (NAVA RR/MV): 22 patient treatments recorded across 6 sites. Provenance: Post-market evaluation, specifics (e.g., country) not stated but likely within markets where MAQUET operates.
      • Stress Index (SI): 10 adult patients with ALI or ARDS. Provenance: Not stated, likely clinical settings (hospitals).
      • Apnea Ventilation/Alarm Behavior (NAVA): Patients in 3 sites during four weeks. 25 clinicians involved. Provenance: Post-market evaluation, likely clinical settings.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • General V&V: "Software testers and clinicians" were involved in "Free User Testing (FUT)" in MAQUET's test laboratory. Specific number and qualifications not detailed.
    • Specific Clinical Validation Activities (SERVO-i):
      • Nuisance Alarm Reduction (NAVA RR/MV): "Evaluation at MAQUET showed that the new algorithms did effectively calculate RR and MV while reducing the occurrence of nuisance RR and MV alarms." This implies internal experts at MAQUET (engineers, potentially medical staff) evaluated the data. Qualifications not specified.
      • Stress Index (SI): Comparison against "an existing system used in multiple published articles." This "existing system" acts as the de facto ground truth reference. The evaluation of results was likely done by MAQUET's internal team.
      • Apnea Ventilation/Alarm Behavior (NAVA): 25 clinicians filled in questionnaires. Their responses, evaluated by MAQUET, established the "ground truth" (or acceptance) that the new algorithm was acceptable and reduced nuisance alarms. Their specific qualifications (e.g., years of experience, specialty) are not detailed beyond "clinicians."

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • No formal adjudication method like "2+1" or "3+1" is described. Evaluations were made by MAQUET internally or based on comparison with existing systems/feedback from clinicians.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No, an MRMC comparative effectiveness study was not done. The device is a ventilator, not an AI diagnostic tool where human readers interpret output.

    • Effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as it's not an AI-assisted diagnostic tool for human readers. The "AI" (software algorithms) are intrinsic to the device's function.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The extensive "design verification and validation" for both SERVO-i and SERVO-s, including requirement verification, regression testing, code review, static code analysis, and compliance with product standards, represents standalone testing of the algorithm (software) and hardware. The "Free User Testing" involves human interaction, but the core functionality tests are against specifications.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • General V&V: Primarily established by device specifications, regulatory requirements (standards), and comparison to predicate device performance.
    • Specific Clinical Validation Activities (SERVO-i):
      • Nuisance Alarm Reduction: Reduced frequency of nuisance alarms (an objective metric) and effective calculation of RR/MV (comparison to expected values).
      • Stress Index: Outputs from "an existing system used in multiple published articles."
      • Apnea Ventilation/Alarm Behavior: Clinician feedback via questionnaires, representing user acceptance and perceived reduction in nuisance alarms.

    7. The sample size for the training set:

    • Not applicable. This device is a medical ventilator (hardware and software) and not a machine learning model that undergoes a distinct "training phase." The "training" for the device's development involves engineering design, coding, and iterative testing, not data-driven machine learning in the conventional sense.

    8. How the ground truth for the training set was established:

    • Not applicable, as there is no specific "training set" in the context of a machine learning model. The "ground truth" for the device's design and functionality is established through engineering principles, adherence to medical device standards, clinical requirements, and performance of existing predicate devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K073179

    Validate with FDA (Live)

    Date Cleared
    2008-06-09

    (209 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SERVO-i Ventilator System with Heliox option is intended for treatment and monitoring of patients in the range of neonates, infants, and adults with respiratory failure or respiratory insufficiency. SERVO-i is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities. The SERVO-i Ventilator System with Heliox option is also MR conditional.

    The SERVO-i Ventilator System with Heliox option is indicated for use with the delivery of Air, Oxygen, or Heliox (a mixture of Helium and Oxygen).

    Device Description

    The predicate device SERVO-i Ventilator System is a ventilator, which gives, ventilation (Invasive and Non Invasive) to critical care patients in the weight range 0.5 to 250 Kg's. The modified device adds an option, called "Heliox option", which is a mechanical adaptor along with software for SERVO-i Ventilator System integration.

    AI/ML Overview

    The provided text is a 510(k) summary for the Maquet SERVO-i Ventilator System with Heliox option. It focuses on demonstrating substantial equivalence to predicate devices and describes the device's intended use and technological characteristics. Crucially, this document does not contain information about acceptance criteria, device performance metrics, or study designs used to prove the device meets specific performance criteria beyond establishing substantial equivalence for regulatory purposes.

    Therefore, I cannot extract the requested information regarding acceptance criteria, reported device performance, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, or ground truth details. These types of studies are typically found in more detailed technical documentation or clinical trial reports, which are not part of this 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063404

    Validate with FDA (Live)

    Date Cleared
    2007-02-15

    (98 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SERVO-i Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency. SERVO-i is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities and for in-hospital transport.

    The SERVO-i Ventilator with MR environment option is classified as MR Conditional for 1.T. 1.5T and 37 MR scanners. This means that it is safe to use in the MR environment if the conditions in the MR Environment Declaration for SERVO-i are met.

    Device Description

    The ventilator is a platform with several selectable ventilation modes, which monitor patients whom need respiratory assistance. The MR environment option makes it possible to qualify the Servo-i ventilator to be used in MR environment

    AI/ML Overview

    The SERVO-i Ventilator System's acceptance criteria and testing are focused on its "MR environment option" which allows it to be used safely in conjunction with MRI scanners. The provided document is a 510(k) summary, which typically provides an overview of testing rather than a detailed study report.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria (e.g., specific thresholds for magnetic field strength, RF interference limits, or image artifacts) or precise, quantitative reported device performance values. Instead, it describes general claims of compatibility and safety.

    Acceptance Criteria CategoryReported Device Performance
    MR Compatibility StatementThe SERVO-i Ventilator with MR environment option is classified as MR Conditional for 1T, 1.5T, and 3T MR scanners. This implies that it is safe to use in the MR environment if specific conditions are met.
    Ventilator Performance in MR EnvironmentThe SERVO-i ventilator with MR option has been tested with 1.0, 1.5, and 3.0 Tesla (T) scanners without impairing its performance.
    Impact on MR Image QualityThe SERVO-i ventilator with MR option has been tested with 1.0, 1.5, and 3.0 T scanners without impairing the image quality of the scanner.
    Safety MarginsVital parts of the ventilator have been tested for performance in excessive magnetic fields to evaluate the safety margins. (No specific margin values reported)
    Electromagnetic Emission ReductionThe emission from the SERVO-i has been further decreased with the MR option to reduce the probability for artifacts on the scanner images.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a numerical sample size for the devices tested. It refers to "the SERVO-i ventilator with MR option" in general terms.
    • Data Provenance: The document implies the testing was conducted by Maquet Critical Care AB, which is a Swedish company. The nature of the testing (e.g., lab testing, in-house verification) suggests it's prospective data generated specifically for this submission. There is no indication of patient data or retrospective studies for this specific MR compatibility claim.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish a "ground truth" in the typical sense for clinical performance. The testing described appears to be engineering and compatibility testing to ensure the device functions correctly in an MR environment and does not interfere with the MRI scanner. Therefore, no information on the number or qualifications of clinical experts for ground truth is provided.

    4. Adjudication Method for the Test Set

    Not applicable. The testing described is technical verification and validation of MR compatibility, not clinical performance requiring adjudication of patient cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on the technical compatibility and safety of the ventilator in an MR environment, not on the comparative clinical effectiveness or improvement of human readers with AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, in a sense. The testing performed on the SERVO-i ventilator with the MR option can be considered "standalone" as it evaluates the device's technical performance and its interaction with an MRI system, without direct human-in-the-loop performance being a primary outcome measure for the MR compatibility. The ventilator itself is a standalone medical device.

    7. The Type of Ground Truth Used

    The ground truth used for this testing is based on:

    • Engineering specifications and regulatory standards for MR compatibility: The device must maintain its intended ventilatory function.
    • MRI image quality assessment: The device must not cause unacceptable artifacts in the MR images.
    • Magnetic field strength measurements: To ensure safety margins.

    This is a technical ground truth rather than a clinical ground truth derived from expert consensus or pathology in a diagnostic context.

    8. The Sample Size for the Training Set

    Not applicable. This submission concerns a physical medical device (ventilator) and its MR compatibility, not an AI/ML algorithm that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no AI/ML algorithm or training set discussed in the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062793

    Validate with FDA (Live)

    Date Cleared
    2007-02-07

    (142 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    1 - 120
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Servo-i ventilator with NAVA option is intended for treatment and monitoring to improve synchromy between the ventilator and patients. The Servo-i with NAVA option is suitable for patient ranges of neonates, infants and adults with respiratory failure or respiratory insufficiency. Servo-i is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities and for in-hospital transport. The added indications for use of the NAVA option is when the brain to the brain to the diaphragm is intact; NAVA will improve synchrony between the ventilator and patients with no contraindication for insertion/exchange of a Naso Gastric tube.

    Device Description

    The predicate device Servo-i is a ventilator, which gives, ventilation (Invasive and Non Invasive) to critical care patients in the weight range 0.5 to 250 Kg's The modified device is an option, called NAVA option (Neurally Adjusted Ventilatory Assist) which is a HW module with SW for Servo-i integration, which are added to the predicate device. The Nava option uses an amplifier which in conjunction with a nasogastric feeding tube with microelectrodes detects signals to the diaphragm (Edi). The Edi is used as an additional detector to improve the synchrony between the patient and the ventilator and to give the patient corresponding ventilatory support.

    AI/ML Overview

    This document is a 510(k) summary for the Maquet Servo-i ventilator with the NAVA (Neurally Adjusted Ventilatory Assist) option. It focuses on demonstrating substantial equivalence to predicate devices and does not contain detailed acceptance criteria and performance data for a clinical study in the format requested.

    The document indicates that "Clinical and non clinical data is submitted to verify that the safe performance is substantial equivalent to the Servo-i ventilator with NAVA option" (Section-Page 13 (55)), but it does not provide the acceptance criteria or results of a specific study to prove the device meets those criteria.

    Therefore, I cannot populate the table or answer most of the questions as the information is not present in the provided text.

    Here's what can be extracted based on the limitations:

    1. Table of acceptance criteria and the reported device performance:
    This information is not provided in the document. The document states that clinical and non-clinical data were submitted to verify safe performance and substantial equivalence, but it does not detail specific acceptance criteria or performance metrics and results.

    2. Sample size used for the test set and the data provenance:
    This information is not provided in the document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not provided in the document.

    4. Adjudication method for the test set:
    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No information about an MRMC study or AI assistance is provided. The NAVA option is described as a module to improve synchrony between the patient and the ventilator by detecting signals to the diaphragm (Edi), not an AI-assisted diagnostic or interpretive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
    The NAVA option is an integral part of a ventilator system, assisting in ventilation. It's not a standalone algorithm in the typical sense of a diagnostic tool that produces an output without human intervention for review. Its function is to modulate the ventilator's output based on physiological signals, which implies continuous operation within the human-in-the-loop context of patient ventilation.

    7. The type of ground truth used:
    This information is not provided in the document. The general claim is about verifying "safe performance" and "substantial equivalence," but the specific ground truth for any underlying studies is not detailed.

    8. The sample size for the training set:
    This information is not provided in the document.

    9. How the ground truth for the training set was established:
    This information is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040221

    Validate with FDA (Live)

    Date Cleared
    2004-02-20

    (18 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Servo' Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency. Servo' is a ventilator system to be used only by health care providers in hospitals or health care facilities and for in-hospital transport.

    Device Description

    The ventilator is a ventilator with several selectable modes to individually monitor and treat patients whom needs respiratory assistance. The ventilator is the same as described in the notification K010925 and K022132 (addition of BiVent mode and CO2 module).

    AI/ML Overview

    The provided text is a 510(k) summary for a Special 510(k) submission for the Servo-i Ventilator System. It focuses on a minor modification to the device, specifically changing the low minute volume alarm for infants. Crucially, the document does not contain information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert involvement as requested.

    The 510(k) submission is for a modification to an existing device (Servo-i Ventilator System) and claims substantial equivalence to previously cleared devices (K010925 and K022132 for Servo-i, and K970839 for Servo Ventilator 300 A). The core of this particular submission is that the low minute volume alarm for infants has been changed from 0.10 l/min to 0.06 l/min, to match that of the predicate device, the SV300 ventilator.

    Therefore, I cannot provide the requested information from the given text.

    This document is a regulatory submission for a minor device change, not a detailed study report on novel device performance or AI/algorithm effectiveness. Such a document would typically not contain the robust clinical study data requested.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1