Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223022

    Validate with FDA (Live)

    Date Cleared
    2023-02-15

    (139 days)

    Product Code
    Regulation Number
    870.5100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SC HONK YTONK PTCA Balloon Dilatation Catheter is indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.

    Device Description

    The SC HONKYTONK™ PTCA Balloon Dilatation Catheter is a rapid exchange catheter, including a semi-compliance balloon close to the distal tip. The catheter's distal portion is dual lumen and coaxial. The dual lumen includes an outer lumen for balloon inflation, and an inner lumen to allow a ≤0.014 in (0.36 mm) quidewire to facilitate catheter advancement to and through the stenosis that is to be dilated. The proximal portion of the catheter is a single-lumen, stainless steel hypotube that includes a single luer hub for balloon inflation /deflation. The balloon design provides for an inflatable segment of both known diameter and length at recommended pressures. A balloon protective sheath placed over the balloon maintains a low profile. A mandrel is positioned inside the inner lumen to protect catheter patency. The catheter's tapered tip aids advancement to and through the stenosis. The effective length of the catheter is 145 cm. Radiopaque marker bands, in conjunction with fluoroscopy, are aid for the placement of the catheter's balloon. The 1.50 mm device size has one radiopaque marker band, while all other sizes have two radiopaque marker bands. The entire balloon has a hydrophobic coating. The shaft has a hydrophilic coating from the guidewire port to just distal of the balloon.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the SC HONKYTONK™ PTCA Dilatation Catheter. It focuses on demonstrating substantial equivalence to a predicate device, the Emerge™ Monorail PTCA Dilatation Catheter (K163174), through non-clinical testing. This type of submission generally relies on bench testing and does not typically involve clinical studies with human participants or AI performance evaluations as described in your prompt.

    Therefore, many of the requested categories are not applicable to this type of regulatory submission and the provided document.

    Here's an breakdown based on the information provided and standard 510(k) practices for this type of device:


    Acceptance Criteria and Device Performance for SC HONKYTONK™ PTCA Dilatation Catheter

    This submission relies on non-clinical (bench) testing to demonstrate substantial equivalence to a predicate device. The "acceptance criteria" are implied by the successful completion and meeting of specifications for each test. The "reported device performance" refers to the outcomes of these non-clinical tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Mechanical PerformanceDevice meets specified engineering and functional requirements."Met all acceptance criteria and performed similarly to the predicate device."
    Dimensional VerificationWithin specified tolerances.Successful
    Simulated UseFunctions correctly under simulated clinical conditions.Successful
    Balloon FatigueMaintains integrity after multiple inflation/deflation cycles.Successful
    Inflation/Deflation TimeWithin specified time limits.Successful
    Flexibility/KinkMaintains patency and structural integrity during manipulation.Successful
    RadiopacityClearly visible under fluoroscopy.Successful
    ParticulateMeets limits for particulate matter release.Successful
    Component Dimensional CompatibilityComponents fit and function together as designed.Successful
    Rated Burst PressureWithstands specified pressure without bursting prematurely.Successful (Specific pressures listed in Table 5-1 match expectations)
    ComplianceBalloon expands to expected diameter at rated pressure.Successful
    Bond Strength (including Tip Pull)Connections remain secure under stress.Successful
    TorqueTransmits rotational force effectively.Successful
    Coating IntegrityCoatings remain intact and functional.Successful
    BiocompatibilityMaterials are biocompatible and do not elicit adverse biological responses."SC HONKYTONK™ PTCA Balloon Dilatation Catheter to be biocompatible."
    CytotoxicityNon-toxic to cells.Successful
    Intracutaneous ReactivityNo significant irritation.Successful
    Complement Activation (SC5b-9)Low or acceptable level of complement activation.Successful
    Material Mediated PyrogenicityNo pyrogenic response.Successful
    Partial Thromboplastin TimeNo or acceptable impact on blood clotting time.Successful
    SensitizationNo allergic/sensitization response.Successful
    Hemolysis (Direct and Extract)Acceptable level of red blood cell lysis.Successful
    ThrombogenicityLow thrombogenic potential.Successful
    Acute Systemic ToxicityNo short-term systemic toxic effects.Successful
    Heparinized Blood Platelet and Leukocyte Count AssayAcceptable impact on blood cell counts.Successful

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated as a number of "patients" or "images" since this is a non-clinical bench study. The "sample size" refers to the number of devices or components tested for each specific bench test (e.g., a certain number of balloons for burst pressure, a certain number of catheters for simulated use). These numbers are not detailed in the summary document but would be part of the full testing report submitted to the FDA.
    • Data Provenance: Non-clinical (bench) laboratory testing performed by the manufacturer, Sino Medical Sciences Technology Inc. The data is generated in a controlled laboratory environment, not from human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable (N/A). This is a non-clinical bench study for a PTCA catheter. "Ground truth" in the clinical sense (e.g., expert diagnosis of an image) is not established. The "truth" is based on engineering specifications and validated test methods.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable (N/A). Adjudication methods like 2+1 or 3+1 are used for clinical studies involving multiple readers (typically human experts reviewing images/data). This submission is based on automated or semi-automated bench test results against predefined criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a non-clinical 510(k) submission for a medical device (catheter), not an AI/CADe/CADx device that would involve human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is a non-clinical 510(k) submission for a medical device (catheter) with no AI component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Engineering Specifications and Industry Standards: The "ground truth" for non-clinical testing is based on the device's design specifications, relevant ISO/ASTM standards (e.g., for material properties, biocompatibility), and comparison to the predicate device's established performance parameters. For example, a balloon's rated burst pressure is a design specification, and the test verifies it meets that "ground truth."

    8. The sample size for the training set

    • Not Applicable (N/A). There is no "training set" as this is not an AI/machine learning device. The device itself is manufactured, and specific units are selected for verification and validation (V&V) testing.

    9. How the ground truth for the training set was established

    • Not Applicable (N/A). As there is no training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1