510(k) Data Aggregation

    K Number
    K042196

    Validate with FDA (Live)

    Device Name
    SAVEDENT BLUE
    Date Cleared
    2004-11-01

    (80 days)

    Product Code
    Regulation Number
    872.6070
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Light source for the activation of light cured resin polymers and glass ionomer materials used in dentistry as pit and fissure sealants, restorative materials, and bonding materials.

    Device Description

    SAVEDENT The BLUE is a LED based curing light intended to polymerise and therefore cure, harden or set resin or glass ionomer based dental materials used for restorative, sealing or cementing procedures. It is a lightweight handheld unit delivering up to 1000mW/cm2 for set times between 10 and 40 seconds.

    AI/ML Overview

    The provided text describes a 510(k) submission for a dental light-curing unit called SAVEDENT BLUE. The submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study with acceptance criteria in the typical sense of a clinical trial for diagnostic performance.

    Here's an analysis based on the provided text, addressing the requested points where information is available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The text does not explicitly state formal "acceptance criteria" in a quantitative, pre-defined manner often seen for diagnostic devices (e.g., sensitivity, specificity thresholds). Instead, the demonstration of equivalence is based on meeting or exceeding the performance of the predicate device for "cure depth."

    Acceptance Criteria (Implied)Reported Device Performance
    Same or better performance in terms of cure depth for a given exposure time compared to the predicate device"Comparative tests have demonstrated that SAVEDENTTM BLUE provides the same or better performance in terms of cure depth for a given exposure time as the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    The text does not specify the sample size used for the comparative tests or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions "Comparative tests."

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    This information is not applicable or available from the provided text. The device is a light-curing unit, and its performance is measured by objective physical properties (cure depth) rather than a diagnostic interpretation requiring expert judgment.

    4. Adjudication Method for the Test Set

    This information is not applicable or available from the provided text, as the "ground truth" (cure depth) is based on physical measurement, not expert interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices where human readers interpret medical images or data. The SAVEDENT BLUE is a therapeutic device (light-curing unit).

    6. If a Standalone Performance (Algorithm Only) Study Was Done

    No, a standalone performance study in the context of an algorithm or AI is not applicable. The SAVEDENT BLUE is a hardware device (light-curing unit), not an algorithm. The "standalone performance" mentioned implicitly refers to its physical output parameters and ability to cure materials as designed.

    7. The Type of Ground Truth Used

    The ground truth used is the measured "cure depth" of the dental materials. This is an objective, quantitative physical property, not based on expert consensus, pathology, or outcomes data in a clinical sense.

    8. The Sample Size for the Training Set

    This information is not applicable for this device. As a hardware light-curing unit, there is no "training set" in the context of machine learning or AI algorithms. The development involves engineering and testing, not algorithmic training.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no "training set" for this type of device.