Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K952879

    Validate with FDA (Live)

    Date Cleared
    1996-04-22

    (304 days)

    Product Code
    Regulation Number
    870.4360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 3M™ Sarns™ Centrifugal Pumps with Duraflo® II Treatment are indicated as extracorporeal pumps for use in cardiopulmonary bypass procedures only and for use exclusively with Sams™ centrifugal control systems.

    Device Description

    The Sarns™ Centrifygal Pump is a centrifugal pump which imparts velocity to blood through action of a rotating impeller. The pump is comprised of two chambers. The pump chamber contains the impeller with vanes, seals components, inlet port and outlet port for connection to tubing. The rear chamber contains bearings used to support and align the magnet rotor. The rear chamber is sealed off from the pump chamber. The rear chamber is isolated from atmosphere via a hermetic seal of the housing components forming this chamber. The Duraflo® II Treatment provides a heparin-based blood path containing up to 400 USP units of heparin. This treatment improves the blood compatibility of non-biological surfaces in the extracorporeal circuit.

    AI/ML Overview

    This document describes a 510(k) summary for the 3M™ Sarns™ Centrifugal Pump with Duraflo® II Treatment. It does not describe an AI/ML device or its performance characteristics, but rather a medical device and its physical and biological performance. Therefore, many of the requested fields are not applicable.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance
    Blood CompatibilityHemolysisNot explicitly stated as a numerical threshold, but implied to be within acceptable ranges based on predicate and previous testing."some change noted in hemolysis between the two groups" but "results are in line with results from previous centrifugal pump testing at worst case conditions."
    Platelet DepletionNo statistical difference between coated and uncoated units."no statistical difference in %PLT depletion...between the coated and uncoated units."
    White Blood Cell DepletionNo statistical difference between coated and uncoated units."no statistical difference...in average % WBC depletion between the coated and uncoated units."
    Heparin LeachingNot explicitly stated, but implied to be within safe/effective levels.Results were satisfactory (implicit, as no adverse findings are noted).
    Heparin QuantitationNot explicitly stated, but implied to be within desired range for the "up to 400 USP units of heparin" treatment.Results were satisfactory (implicit, as no adverse findings are noted).
    Functional Performance12-hour Life TestDevice must pass the test (remain functional)."All Duraflo® II treated pumps passed these tests."
    Burst TestDevice must pass the test (withstand over-pressurization without failure)."All Duraflo® II treated pumps passed these tests."

    2. Sample size used for the test set and the data provenance

    • Hemolysis, Platelet Depletion, White Blood Cells, Heparin Leaching, Heparin Quantitation: The specific sample sizes for these in vitro tests are not explicitly stated. The text mentions a comparison to "a study that was conducted in early 1995 on the predicate device (K915363)," suggesting the data is retrospective in the sense of comparing against historical predicate data. The provenance is likely internal testing by the manufacturer.
    • 12-hour Life Test: n=16
    • Burst Test: n=16
    • Data Provenance: The tests are described as in vitro studies conducted by the manufacturer. No country of origin is specified for the data, but it's generated by the company (Baxter Healthcare Corporation, Bentley Division, Irvine, California).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical device performance study, not an AI/ML study requiring expert ground truth establishment for a diagnostic output. The ground truth for this device's performance is derived from physical and biological measurements against predefined test protocols.

    4. Adjudication method for the test set

    Not applicable. This is a medical device performance study, not an AI/ML study requiring human adjudication of results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device performance study for a physical pump, not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device performance study for a physical pump.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this device's performance is established through physical and biological measurements obtained from in vitro testing, following established test methods based on the proposed "Special Control Protocol for Evaluation of Blood Trauma Caused by Extracorporeal Centrifugal Pumps." For the functional tests (Life and Burst), the ground truth is simply whether the device passed the defined operational requirements.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device; there is no training set in the context of machine learning.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI/ML model, there is no ground truth established for a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1