Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K994081

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-01-18

    (46 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A surgeon's glove is intended to be worn on the hands of operating room personnel to prevent contamination between such personnel and the patient during a surgical procedure.

    Device Description

    Safeskin Tactylon® PF powder-free surgical gloves are sterile disposable devices intended to be worn by operating room personnel to prevent contamination between such personnel and the patient during a surgical procedure. Safeskin® Tactylon PF powder-free surgical gloves possess the following technological characteristics (as compared to ASTM or equivalent standards): Dimensions Meets ASTM D 3577, Physical Properties Meets ASTM D 3577, Freedom from pinholes Meets ASTM D 3577 Meets ASTM D 5151, Powder-Free Meets ASTM D 6124 2 mg/glove maximum.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device, specifically a surgical glove. It does not describe an AI/ML powered device, nor does it detail a clinical study with an AI component. Therefore, much of the requested information regarding AI acceptance criteria, sample sizes for AI training/test sets, expert ground truth, MRMC studies, or standalone algorithm performance is not applicable.

    However, I can extract information related to the acceptance criteria and the study (or lack thereof) to prove the device meets these criteria as presented in the document for the physical device.


    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsAcceptance Criteria (Standards)Reported Device Performance
    DimensionsMeets ASTM D 3577Meets ASTM D 3577
    Physical PropertiesMeets ASTM D 3577Meets ASTM D 3577
    Freedom from pinholesMeets ASTM D 3577 & Meets ASTM D 5151Meets ASTM D 3577 & Meets ASTM D 5151
    Powder-FreeMeets ASTM D 6124 (2 mg/glove maximum)Meets ASTM D 6124 (2 mg/glove maximum)
    Primary Skin IrritationPasses in Rabbits (Biocompatibility)Passes
    Guinea Pig SensitizationPasses (Biocompatibility)Passes

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for the individual tests (e.g., how many gloves were tested for pinholes, or how many rabbits/guinea pigs were used). It implicitly refers to compliance with ASTM standards, which would dictate appropriate sampling methods.

    The data provenance is not explicitly stated as "country of origin" for testing, but the company is based in San Diego, CA, USA, and the submission is to the US FDA. The testing described is intrinsic to the product's physical and biocompatibility characteristics, not clinical data provenance in the sense of patient records.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable as this is a physical device (surgical glove) meeting manufacturing standards, not a diagnostic or AI device requiring expert-established ground truth from medical images or data. The "ground truth" here is the pass/fail criteria defined by ASTM and biocompatibility standards.

    4. Adjudication method for the test set

    Not applicable. The tests are based on objective measurements against engineering and biological standards, not subjective expert assessment requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical device submission and does not involve AI or human readers for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical device submission and does not involve an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance is established by:

    • ASTM (American Society for Testing and Materials) Standards: These are consensus standards for material properties and testing methods (e.g., D 3577 for rubber surgical gloves, D 5151 for detecting pinholes, D 6124 for powder residue).
    • Biocompatibility Standards: These standards (implied by "Passes" for skin irritation and sensitization) ensure the material is safe for contact with human tissue.

    8. The sample size for the training set

    Not applicable. This is a physical device and the concept of a "training set" for an AI model does not apply. The manufacturing process is analogous to "training" in the sense of quality control and process optimization, but not in the AI context.

    9. How the ground truth for the training set was established

    Not applicable. As noted above, there is no AI training set. The "ground truth" for the device's acceptable quality is defined by adherence to established, external manufacturing and material standards (ASTM, biocompatibility).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1