510(k) Data Aggregation

    K Number
    K252266

    Validate with FDA (Live)

    Date Cleared
    2025-10-17

    (88 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System may be immediately loaded when good primary stability is achieved and the functional load is appropriate.

    All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components previously cleared K200992, K193096, K170398 and K170392.

    This submission adds Interface Abutments and Healing Abutments to the S.I.N. Dental Implant System. The subject device Interface abutments are provided with MT 16º and CM11.5º connection compatible with the S.I.N. Dental implants.

    Subject Device Interface Abutments

    Subject device Interface Abutments are two-piece abutments for which the second part (or top half) is the ceramic superstructure. The second part is designed and manufactured using CAD-CAM techniques. The titanium component of the Interface Abutments has a cementable platform diameter of 3.5mm, 4.0mm, 4.5mm, or 5.5mm, cementable post-height of 4.0mm or 6.0mm, and a gingival height built into the titanium base component from 0.5mm to 4.0 mm. The two pieces of the Ti-Base are bonded using Panavia Universal Dual-Cure Cement (K150704), as referenced in K190936.

    All subject devices are provided only in straight designs, with no pre-angled conventional abutments or pre-angled CAD-CAM abutments. No subject device abutment is to be customized to create an angled abutment or to correct for angulation, and no superstructure for use with any subject device Interface abutment (CAD-CAM) is to be fabricated to create a final abutment with angulation or to correct for angulation. Therefore, this model has a slight design difference to improve screw insertion and removal. It was purposely designed to optimize fixation and removal of the screw connection. This new presentation allows screw fixation even in cases of anterior aesthetic rehabilitation, avoiding the need to create screw access holes in the sublingual phase. However, it is confirmed that all implant-abutment interfaces are intended exclusively for straight configurations and are not indicated for any angled model. These minor differences in the exact dimensions of the subject device interface abutment as compared to the predicate device and reference devices do not impact safety, effectiveness, or substantial equivalence.

    The subject device interface abutments with the CM11.5º interface are compatible with implants cleared in K200992, K221453, K211921 and K170392 in terms of abutment-implant connection. The subject device with the MT16º interface is compatible with implants cleared in K222231, K193096 and K170392 in terms of abutment-implant connection.

    The design parameters for the CAD-CAM fabrication of patient-specific abutments from CADCAM Abutments are the same for the 11.5° CM implant connection and the MT 16° implant connection. The design limit parameters are:

    • Minimum wall thickness – 0.5 mm
    • Minimum abutment post height for single-unit restoration – 4.0 mm
    • Maximum angle – 0°. Straight only
    • Maximum angle for screw insertion/removal – 20º
    • Maximum gingival height – 5.0mm
    • Minimum Gingival Height – 0.5 mm
    • Maximum allowable abutment Post Height – 6 mm
    • Total Abutment Height – 10mm

    By definition, the abutment post height is considered by FDA to be the "stump" portion above the gingival collar, to which the restorations attach.

    All subject device Interface Abutments are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136 and the same zirconia that is specified for superstructures in K190396, and are to be used with subject abutment screws. All subject devices are provided sterile. All Interface abutments are provided sterile by gamma irradiation; the same sterilization method used in predicate device K190396 and in the additional devices K200992 and K170392.

    All Interface abutments manufactured form titanium alloy are anodized using a standard anodization process is identical to the anodization process used on abutments cleared in K190396.

    The subject device abutments screws are to be used with Interface Abutments present in this submission and they are substantially equivalent to material (titanium alloy) and designs to abutments screws cleared in K170392.

    Subject Device Healing Caps

    The subject device Healing Cap is substantially equivalent to the corresponding design Healing Cap cleared in K170392. The subject device healing cap is designed for the healing period between implant placement and final abutment placement and is not intended to support a temporary prosthesis.

    All Healing Caps are provided sterile by gamma irradiation; the same sterilization method used in predicate device K170392 and additional predicate device K200992.

    All subject device components are provided sterile by gamma irradiation, the same sterilization method used in K190936, K170392 and K200992.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251262

    Validate with FDA (Live)

    Date Cleared
    2025-09-26

    (156 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.- validated milling center for manufacture.

    The Abutment Mini Angled Morse 450 is intended for use with the S.I.N. Dental Implant System Zygomatic implants. S.I.N. Dental Implant System Zygomatic implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System Zygomatic implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K222231, K231127, and K170392.

    This submission includes dental implants Epikut S with a Morse taper (MT) abutment interface and an acid-etched endosseous surface, and Epikut S Plus implants with an endosseous surface produced by acid- etching followed by application of a hydroxyapatite coating (HAnano) and the same described connection. The implant design and endosseous surfaces are nearly identical to the Epikut S and Epikut S Plus implants cleared in K222231, with the exception of the additional body/platform diameter (6.0 /5.2 mm) with lengths of 8.5, 10, 11.5, 13.

    The subject device dental implants are summarized in the following table.

    Implant LinesBody Ø, mmPlatform Ø, mmLengths, mm
    Epikut S6.05.28.5, 10, 11.5, 13
    Epikut S Plus6.05.28.5, 10, 11.5, 13

    This submission includes Abutment Mini Angled Morse 45° in sizes from 1.5, 2.0, 2.5 with a 16° Morse taper connection mating abutments for screw retained, multi-unit prostheses and a prosthetic platform diameter of 4.8 mm, that is for exclusively use with cleared Zygomatic Plus implant. The corresponding abutment screw is PFMAAM 16A.

    This submission includes Interface abutment MT 16° prosthetic components for fabrication of patient-specific abutments that are compatible with implants from S.I.N. Dental Implant System. Subject device Interface Abutments are two-piece abutments for which the second part (or top half) is the ceramic superstructure. The second part is designed and manufactured using CAD-CAM techniques. The titanium component of the Interface Abutments Morse Taper has a cementable platform diameter of 3.5mm, 4.5mm, or 5.5mm, cementable post-height of 4.0mm or 6.0mm, and a gingival height built into the titanium base component from 0.5mm to 4.0mm. The two pieces of the Ti-Base are bonded using Panavia Universal Dual-Cure Cement (K150704), as referenced in K190936. These abutments are used with a fixation screw PFMT 02A.

    After scanning the intraoral area, the S.I.N.-validated milling center will receive the STL file and design the coping/superstructure using computer software (CAD) according to the design parameters below and the space available.

    Design parameters for all Interface CAD-CAM Abutments superstructure are:

    • Minimum wall thickness – 0.5 mm
    • Minimum abutment post height for single-unit restoration – 4.0 mm
    • Maximum angle – 0°. Straight only
    • Maximum gingival height – 5.0 mm
    • Minimum Gingival height – 0 mm
    • Maximum allowable abutment Post Height – 6 mm
    • Total abutment height – 10mm
    • Abutment diameter – depends on available space (for the patient)

    By definition, the abutment post height is considered by FDA to be the "stump" portion above the gingival collar, to which the restorations attach.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251046

    Validate with FDA (Live)

    Date Cleared
    2025-08-13

    (132 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. dental implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    S.I.N. Dental Implant System implants with lengths of 18, 20, 22 or 24mm may be tilted up to 30º. When used in the mandible or maxilla with implants with lengths of 18, 20, 22 or 24 mm at an angulation of 30º, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22 or 24 mm at angulations between 0º and less than 30º, the S.I.N. Dental Implant System implants are only indicated for multiple unit restoration in splinted applications that utilize at least two implants.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes various endosseous implants with 16° Morse tapers interface connection. This submission adds to the S.I.N. Dental Implants System, the Versalis S implants line.

    The Versalis S implant line has the Versalis S and Versalis S Plus implants, and both have a Morse taper connection with an internal 16° cone taper. The Versalis S implants have implant body lengths from 8.5 to 24mm and body diameter Ø from 3.5 to 7.0mm., they are manufactured from unalloyed titanium conforming to ASTM F67, with a double acid-etched surface treatment. The Versalis S Plus implants have implant body lengths from 8.5 to 24mm and body diameter Ø from 3.5 to 7.0mm., they are manufactured from unalloyed titanium conforming to ASTM F67, with a double acid-etched surface treatment plus hydroxyapatite surface coating.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67 Standard Specification for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700), Grade 4. The material used to manufacture the dental implants in this submission (conforming to ASTM F67) is identical to the material used to manufacture the device implants cleared in primary predicate K170392 and in additional predicate K222231.

    AI/ML Overview

    The provided text is a 510(k) clearance letter for a dental implant system. It does not contain any information about acceptance criteria or a study proving that an AI/ML device meets acceptance criteria. The "Performance Data" section discusses non-clinical testing for the dental implant itself, such as sterilization validation, shelf-life testing, and biocompatibility, as well as fatigue testing and MRI environment compatibility. This is for a physical medical device, not a software device that relies on AI/ML.

    Therefore, I cannot provide the requested information as it is not present in the given document.

    To reiterate, the document is about a physical dental implant system and the performance data pertains to its physical properties, sterility, and biocompatibility, not to the performance of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240609

    Validate with FDA (Live)

    Date Cleared
    2024-07-17

    (135 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System Zygomatic implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System Zygomatic implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System. This submission includes zygomatic dental implants and mating abutments for screw-retained, multi-unit prostheses.

    The subject device Zygomatic Plus Morse Taper implants have a body diameter of 4.0 mm that tapers to approximately 2.3 mm at the apex. The implants have a Morse taper ("CM") abutment interface connection, with an internal 16° cone taper. The abutment platform is 4.0 mm. All implants are threaded starting at the apex and extending approximately 17 mm in the coronal direction, with a major thread diameter of 4 mm; the thread tapers over the final 6 mm of the implant to the apex. The implants are provided with overall lengths of 30, 32.5. 35. 37.5. 40. 42.5.45. 47.5. 50. 52.5.57.5. and 60 mm. The surface of all subject device implants is acid-etched from the implant platform to the apex, followed by application of a hydroxyapatite coating (HA nano).

    The subject device abutments are multi-unit, indexed abutments for use only with the subject device Zygomatic Plus Morse Taper implants for maxillary full-arch restorations. The subject device abutments have a prosthetic platform diameter of 4.8 mm, and gingival heights of 1.5, 2.0, and 2.5 mm. The prosthetic post is angled 52° or 60° to the long axis of the implant. These components are used with previously cleared abutment screws.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. All implants have an acid-etched HA100 surface treatment, identical to that cleared in K231127. All subject device abutments are manufactured from titanium alloy (Ti-6A1-4V) conforming to ASTM F136. All subject device implants and abutments are provided sterile to the end user.

    AI/ML Overview

    The provided text is a 510(k) Summary for a dental implant system. It explicitly states that no clinical data were included in this submission to support substantial equivalence (see "No clinical data were included in this submission." on page 6 of 9).

    Therefore, it is not possible to provide specific details about acceptance criteria, device performance, sample sizes, expert ground truth establishment, adjudication methods, or MRMC studies, as这些 types of studies were not conducted or reported in this 510(k) submission.

    The "Performance Data" section (pages 6-7 of 9) refers to non-clinical data only:

    • mechanical testing
    • gamma irradiation sterilization
    • bacterial endotoxin testing
    • MR environment evaluation
    • sterile barrier shelf life data
    • biocompatibility data

    These tests are typically conducted to assess the physical and biological characteristics of the device, not its clinical performance in patients as would be measured in an AI/MRMC study.

    In summary, based on the provided document, no study was conducted to prove the device meets acceptance criteria related to AI/MRMC performance as outlined in your request. The submission relies on substantial equivalence to predicate devices, supported by non-clinical performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232821

    Validate with FDA (Live)

    Date Cleared
    2024-05-31

    (261 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Unitite Slim, Unitite, Unitite Compact, Strong SW CM S.I.N Dental Implant System implants with lengths less than 7 mm are intended for delayed loading only.

    Epikut CM, Epikut Plus CM, Epikut S, Epikut S Plus S.I.N. Dental Implant System implants with lengths of 18, 20, 22, or 24 mm may be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    Epikut CM, Epikut Plus CM, Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add labeling claims for a surface treatment designated HA™®, which is applied to implants of the S.I.N. Dental Implant System, as previously cleared in K170392. K193096. K200992, K211921, K221453, and K222231. The new labeling claims and the information supporting the claims are shown below in Performance Data.

    The following information is a summary of the designs of the subject device implants, as they have been previously cleared with the HA"" surface treatment. Note that there are no changes to any aspect of the design, nor to the HA" surface treatment, as described in the respective 510(k) Premarket Notifications referenced above.

    The previously cleared implants that are the subject of this submission includes eleven (11) product lines from six (6) 510(k)s. Implant body diameters, platform diameters and lengths for each product line are shown below in Table 10.1 Summary of Subject Device Designs and Sizes.

    K170392: Unitite Slim, Unitite, Unitite Compact K193096: Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus K200992: Strong SW Plus K211921: Epikut Plus CM, Epikut Plus HE K221453: Epikut Plus CM (additional lengths) K22231: Epikut S Plus

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The HA1000 surface treatment is identical to that cleared in K211921.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the U.S. FDA for the S.I.N. Dental Implant System. It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices by comparing their indications for use, designs, materials, and manufacturing processes, with an emphasis on adding new labeling claims related to a surface treatment.

    Crucially, this document states: "No clinical data were included in this submission." This means that the device's performance was not proven through a study involving human subjects or artificial intelligence (AI). The document relies on non-clinical data such as in vitro testing, animal testing referenced from published literature, sterilization, endotoxin, shelf-life, biological evaluation, MR compatibility, and fatigue testing to support the claims and substantial equivalence.

    Therefore, most of the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, and specific performance metrics for an AI/device study cannot be extracted from this document.

    However, I can extract the information related to the non-clinical performance data presented to meet the criteria for substantial equivalence to predicates.

    Here's a breakdown of what can be extracted and an explanation of why other requested information is not present:

    Information that CANNOT be extracted from this document (and why):

    • A table of acceptance criteria and the reported device performance (for clinical/AI studies): Not applicable as no clinical or AI study was performed. The acceptance criteria relate to demonstrating substantial equivalence through non-clinical data, not specific performance metrics against a ground truth for a diagnostic claim.
    • Sample sizes used for the test set and the data provenance: Not applicable.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • Adjudication method for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable (not an AI device).
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for test set: Not applicable as no clinical study or diagnostic AI was evaluated.
    • The sample size for the training set: Not applicable (not an AI device).
    • How the ground truth for the training set was established: Not applicable (not an AI device).

    What CAN be extracted/inferred regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of this 510(k) submission:

    The "acceptance criteria" for a 510(k) are typically met by demonstrating substantial equivalence to existing legally marketed devices (predicates). The "study that proves the device meets the acceptance criteria" in this case refers to the non-clinical tests and comparisons performed to show this equivalence and support the new labeling claims.

    1. A table of acceptance criteria and the reported "device performance" (in terms of non-clinical criteria):

    The acceptance criteria for this 510(k) submission are implicitly met by demonstrating that the subject device is as safe and effective as the predicate devices and that the new labeling claims are supported by scientific evidence. The "performance" here refers to meeting specific engineering, biological, and material standards, not diagnostic accuracy.

    Acceptance Criterion (Implicitly required for 510(k) & specific claims)Reported Device "Performance" (Evidence provided)
    Equivalence in Intended UseSubject device's Indications for Use (IFUS) are identical or substantially similar to those of predicate and reference devices (K170392, K193096, K200992, K211921, K221453, K222231). This includes general placement, loading conditions, specific length limitations for delayed loading, and angulation requirements for longer implants.
    Equivalence in Design/Physical CharacteristicsThe design (body/platform diameters, lengths, implant/abutment interface) of the subject device implants (including the new HA™® surface) is identical to those previously cleared in predicate/reference 510(k)s (K170392, K193096, K200992, K211921, K221453, K222231).
    Equivalence in MaterialsSubject device is manufactured from unalloyed titanium conforming to ASTM F67, identical to predicate devices. The HA™® surface treatment is identical to that cleared in K211921.
    Sterilization EfficacyDemonstrated by gamma irradiation sterilization to a sterility assurance level of 10^-6 by VDmax25 method, according to ISO 11137-1 and ISO 11137-2 (referenced from K211921).
    Absence of Pyrogens/EndotoxinsBacterial endotoxin testing (LAL test) according to ANSI/AAMI ST72 on manufacturing water (weekly) and sterilized product (quarterly) demonstrated all sterile product meets a limit of < 20 EU/device (referenced from K211921).
    Shelf-Life StabilityShelf life testing, including testing of samples after 4 years of real-time aging according to ASTM F1929 and F88/F88M (packaging sterile barrier) and sterility testing of product (referenced from K211921).
    BiocompatibilityBiological evaluation performed according to ISO 10993-1, with test results leveraged from K170392 for ISO 10993-3 (genotoxicity), ISO 10993-5 (cytotoxicity), ISO 10993-6 (implantation), ISO 10993-10 (sensitization, irritation), and ISO 10993-11 (systemic toxicity) to support biocompatibility.
    MRI Safety and CompatibilityNon-clinical analysis and testing according to ASTM F2052 (magnetically induced displacement force), ASTM F2213 (magnetically induced torque), ASTM F2182 (RF induced heating), and ASTM F2119 (image artifact), and the FDA guidance document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" (issued May 2021). No specific results are provided in the summary.
    Mechanical Strength/FatigueFatigue testing according to ISO 14801 referenced from K170392, K193096, K200992, K211921, K221453, and K222231 is applicable to the subject device (the specific results or "performance" against a benchmark are not provided in the summary, just that such testing was leveraged).
    Support for HAnano Surface ClaimsSurface Wettability (Hydrophilicity): Referenced Bezerra et al., 2017. Page 6: "Water contact angle was significantly less on HAnano treated titanium surface than on dual acid-etched titanium surface."Osteoblast Behavior/Differentiation: Referenced Bezerra et al., 2017 (Pages 4, 10) and da Silva et al., 2020 (Page 7). Claims include greater transition to differentiation-related signaling, changes in osteogenic gene marker expression, and spreading morphology different from machined/hydrophilic dual acid-etched surfaces. Results of in vitro testing are not necessarily predictive of human clinical performance.

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not applicable. The "test set" here would refer to the samples used in the non-clinical tests (e.g., in vitro cell cultures, physical test samples, animal models where applicable for referenced studies). Exact sample sizes for these referenced non-clinical studies are not provided in this summary.
    • Data Provenance:
      • Country of Origin: Not specified for the referenced studies (Bezerra et al., da Silva et al.). The manufacturing country for the device is Brazil (S.I.N. - Sistema de Implante Nacional S.A., São Paulo, Brazil).
      • Retrospective or Prospective: Not applicable as these were non-clinical, in vitro, or animal studies, not human clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is a non-clinical submission for a dental implant, not an AI/diagnostic device requiring expert ground truth for imaging interpretation. The "ground truth" for the non-clinical properties is established by the methods, standards, and scientific literature cited (e.g., ASTM, ISO standards, and published research papers).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. There was no expert reader study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This is not an AI-enabled device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was not done. This is not an AI-enabled device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the claims made are based on non-clinical in vitro and animal studies, and adherence to recognized standards for material properties, sterilization, biocompatibility, and mechanical performance. Specifically for the HAnano claims, the ground truth is derived from the experimental results reported in the referenced scientific publications (e.g., measurements of water contact angle, observation of cell morphology, assessment of gene marker expression).

    8. The sample size for the training set:

    • Not applicable. This is not an AI device trained on a dataset.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K231127

    Validate with FDA (Live)

    Date Cleared
    2023-12-07

    (231 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System Zygomatic implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System Zygomatic implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System. This submission includes zygomatic dental implants, mating abutments for screw-retained, multi-unit prostheses, and the corresponding abutment screws.

    The subject device Zygomatic Plus Morse Taper implants have a body diameter of 4.0 mm that tapers to 1.65 mm at the apex. The implants have a Morse taper ("CM") abutment interface connection, with an internal 16° cone taper. The abutment platform is 4.0 mm. All implants are threaded starting at the apex and extending approximately 17 mm in the coronal direction, with a major thread diameter of 4 mm; the thread tapers over the final 6 mm of the implant to the apex. The implants are provided with overall lengths of 30, 32.5, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5, 57.5, and 60 mm. The surface of all subject device implants is acid-etched from the implant platform to the apex, followed by application of a hydroxyapatite coating (HA nano).

    The subject device abutments are multi-unit, indexed abutments for use only with the subject device Zygomatic Plus Morse Taper implants for maxillary full-arch restorations. The subject device abutments have a prosthetic platform diameter of 4.8 mm, and gingival heights of 2 mm and 3 mm. The prosthetic post is angled 45° to the long axis of the implant. These components are used with subject device abutment and prosthetic screws, or previously cleared screws.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. All implants have an acid-etched HA1200 surface treatment, identical to that cleared in K200992. All subject device abutments are manufactured from titanium alloy (Ti-6A1-4V) conforming to ASTM F136. All subject device implants, abutments, and abutment screws are provided sterile to the end user.

    AI/ML Overview

    I am sorry, but based on the provided text, I cannot provide a comprehensive description of the acceptance criteria and the study that proves the device meets them, as requested. The document is an FDA 510(k) summary for a dental implant system. While it mentions performance data, it primarily focuses on establishing substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a full study report of the subject device's performance against those criteria.

    Here's why I cannot fulfill each requested point:

    1. A table of acceptance criteria and the reported device performance: The document does not explicitly state acceptance criteria for the new device in a quantifiable way beyond citing compliance with standards like ISO 11137-1, ISO 11137-2, ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119, ASTM F67, and ASTM F136. It mentions mechanical testing was done "to demonstrate that the subject device abutments... have sufficient strength for the intended use," but it doesn't provide specific numerical acceptance limits or the reported performance values against them.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): The text states "Non-clinical data provided in this submission...mechanical testing conducted using a method modified from ISO 14801 and engineering analysis," but it does not specify the sample size used for this mechanical testing. There is no information about data provenance (country of origin, retrospective/prospective). No clinical data was included in this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): This information is not relevant to the type of non-clinical, mechanical, and sterilization testing described in the document. No expert review of a "test set" in the context of clinical interpretation or diagnosis is mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, as no expert "test set" for clinical evaluation is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is a dental implant system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device is a dental implant system, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" here would be the physical properties and performance characteristics determined by established engineering and sterilization standards (e.g., sterility assurance level, mechanical strength, biocompatibility, MR compatibility). These are not expert consensus in a clinical sense, pathology, or outcomes data.

    8. The sample size for the training set: Not applicable. This is a physical medical device, not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established: Not applicable, for the same reason as point 8.

    The document focuses on demonstrating that the S.I.N. Dental Implant System (specifically the new zygomatic implants and abutments) is substantially equivalent to legally marketed predicate devices through non-clinical data, including:

    • Gamma irradiation sterilization validation.
    • Mechanical testing (modified from ISO 14801).
    • MR compatibility testing (referenced from K222231 and standards ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119).
    • Sterile barrier shelf life data (referenced from K203725).
    • Biocompatibility data (referenced from K200992 and standards ASTM F67, ASTM F136).

    The substantial equivalence argument highlights similarities in indications for use, design principles, materials, manufacturing, sterilization, and shelf life to various predicate and reference devices. However, it does not provide detailed acceptance criteria and specific performance results for the subject device beyond qualitative statements like "sufficient strength."

    Ask a Question

    Ask a specific question about this device

    K Number
    K230069

    Validate with FDA (Live)

    Date Cleared
    2023-04-06

    (86 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    All digitally-designed custom abutments for use with Pre-Milled CAD-CAM Abutments are to be sent to a S.I.N.-validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K221453, K222231, K211921, K193096, and K170392.

    This submission adds Pre-Milled CAD-CAM Abutments to the S.I.N. Dental Implant System. The subject device abutments are provided with 11.5° CM and 16° CM connections compatible with the above S.I.N. dental implants.

    The design parameters for the CAD-CAM fabrication of patient-specific abutments from Pre-Milled CAD-CAM Abutments are the same for the 11.5° CM implant connection and the 16° CM implant connection. The design limit parameters are:

    Minimum wall thickness - 0.55 mm Minimum post height for single-unit restoration - 4.0 mm Minimum gingival height - 0.55 mm Maximum gingival height - 3.5 mm Minimum prosthetic platform diameter - 3.5 mm Maximum prosthetic post height - 6 mm Maximum angulation - 30°

    The subject device abutments are compatible with previously cleared S.I.N. Dental Implant bodies summarized in the following table.

    All subject device abutments are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136, and are to be used with compatible abutment screws previously cleared in K170392. All subject device abutments are provided non-sterile and are to be moist heat sterilized by the end user.

    AI/ML Overview

    This document describes the regulatory clearance for the S.I.N. Dental Implant System, specifically focusing on the addition of Pre-Milled CAD-CAM Abutments. This is a 510(k) submission, which means the device is seeking clearance by demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials.

    Therefore, the information you're requesting regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of clinical performance, sample sizes for test sets, ground truth establishment by experts, MRMC studies, and training sets, is not applicable to this type of regulatory submission.

    Here's why and what information is provided in the document related to "acceptance criteria" from a substantial equivalence perspective:

    In a 510(k) submission for a device like this, "acceptance criteria" are primarily met through non-clinical performance data demonstrating that the new device (the subject device with the added components) is as safe and effective as the predicate device(s). The "study" proving this typically involves engineering analyses and bench testing, comparing the subject device to the predicate device(s) or established standards.

    Detailed breakdown based on the provided document:

    1. Table of Acceptance Criteria (from a Substantial Equivalence Perspective) and Reported Device Performance:

    The "acceptance criteria" for a 510(k) are essentially demonstrating that the new device has the same intended use and similar technological characteristics as the predicates and that any differences do not raise new questions of safety or effectiveness. The reported performance is shown through the non-clinical tests.

    Acceptance Criteria (Demonstration of Substantial Equivalence)Reported Device Performance/Evidence from Document
    Intended Use EquivalenceS.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading. All digitally-designed custom abutments for use with Pre-Milled CAD-CAM Abutments are to be sent to a S.I.N.-validated milling center for manufacture. This statement is identical to the primary predicate device (K193096) for the core indications, and similar to reference devices with minor differences that are addressed (e.g., specific implant lengths not relevant to the subject device, or predicate not explicitly mentioning CAD-CAM abutments which is addressed by K193096).
    Technological Characteristics EquivalenceAbutment Design Parameters: The design parameters for the CAD-CAM fabrication of patient-specific abutments from Pre-Milled CAD-CAM Abutments are provided (Minimum wall thickness - 0.55 mm, Minimum post height for single-unit restoration - 4.0 mm, Minimum gingival height - 0.55 mm, Maximum gingival height - 3.5 mm, Minimum prosthetic platform diameter - 3.5 mm, Maximum prosthetic post height - 6 mm, Maximum angulation - 30°). These parameters are compared to those of predicate abutments (e.g., Abutment Cemented Angled Indexed SIT with CM 11.5° interface and Abutment Cemented Morse Angled with CM 16° interface). Material: Titanium alloy (Ti-6Al-4V) conforming to ASTM F136, same as predicate devices. Sterilization Method: Non-sterile, to be moist heat sterilized by end-user. Validation performed to a sterility assurance level of 10^-5 by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2. Similar to some predicate devices. Biocompatibility: Referenced from K170398 for abutment material ASTM F136 (ISO 10993-5 cytotoxicity). MR Safety: Referenced from K221453 (ASTM F2052, F2213, F2182, F2119, and FDA guidance). Mechanical Performance: Engineering analysis provided to demonstrate that the subject device abutments, in combination with compatible previously cleared dental implants, do not create a new worst-case construct in terms of mechanical testing according to ISO 14801. Mechanical testing data were referenced from K200992 conducted according to ISO 14801.
    Safety and Effectiveness Equivalence (absence of new risks)An engineering rationale was provided to demonstrate that the use of the subject device abutments in combination with the compatible implants raise no new risks and that mechanical testing data provided in a prior submission (K200992) are applicable to the subject abutments. The non-clinical performance data (sterilization, biocompatibility, MR safety, mechanical) supports this.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: Not applicable in the context of clinical performance evaluation for this 510(k). The "test set" here refers to the samples of the device and implant components used for bench testing (e.g., mechanical testing, sterility validation). The document indicates mechanical testing was done per ISO 14801, which would have defined the number of samples required for that specific test, but the exact number isn't provided in this summary.
    • Data Provenance: The manufacturing entity, S.I.N. - Sistema de Implante Nacional S.A., is based in São Paulo, Brazil. The testing would typically be performed either in-house by the manufacturer or by contract research organizations (CROs) in a relevant country. The document does not specify the country where the non-clinical tests (e.g., mechanical, sterilization validation) were explicitly conducted. These non-clinical tests are prospective, meaning they are performed specifically to support the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable. This device clearance is based on non-clinical performance data (bench testing, engineering analysis) and substantial equivalence to predicate devices, not on human interpretation or diagnosis. Therefore, there is no "ground truth" established by medical experts in the way it would be for an AI diagnostic device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. As no expert review or clinical trial data requiring adjudication was performed, this concept is irrelevant to this 510(k) submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI/software as a medical device (SaMD) and no MRMC study was conducted or is required for this type of implant device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical dental implant system, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For this 510(k) submission, the "ground truth" for demonstrating device performance is typically established by:
      • International Standards: e.g., ISO 14801 for mechanical testing of dental implants, ISO 10993 for biocompatibility, ISO 17665 for sterilization. These standards provide accepted methodologies and performance requirements.
      • Engineering Principles and Analyses: Demonstrating that the new components, when combined with existing cleared components, do not introduce new failure modes or decrease mechanical integrity compared to the predicate.
      • Comparison to Predicate Devices: Showing that the subject device's characteristics (design, material, intended use) are sufficiently similar to legally marketed devices.

    8. The sample size for the training set:

    • Not Applicable. This is not a machine learning/AI device, so there is no "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set, this question is not relevant.

    In summary: The provided document is a 510(k) clearance letter for a dental implant system, a physical medical device. The "acceptance criteria" and "proof" are demonstrated through extensive non-clinical (bench) testing and engineering analyses to show substantial equivalence to predicate devices and adherence to relevant industry standards, rather than clinical trials involving human subjects or AI algorithm performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221453

    Validate with FDA (Live)

    Date Cleared
    2022-12-15

    (210 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    S.I.N. Dental Implant System implants with lengths of 18, 20, 22, or 24 mm may be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implant System implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K211921, K203725, K200992, K193096, K170398, and K051859.

    This submission includes dental implants Epikut CM with a Morse taper (CM) abutment interface and an acid-etched endosseous surface, and Epikut Plus CM implants with an endosseous surface produced by acid-etching followed by application of a hydroxyapatite coating (HA ""0"). The implant design and endosseous surfaces are identical to those cleared in K211921, with the exception of the additional body/platform diameter (4.0 mm) and the longer lengths (18, 20, 22, and 24 mm).

    This submission includes Multifunctional Abutments with Morse taper connections (16°, 11.5°, and 4°), and protectors and temporary cylinders for these abutments; the subject device Multifunctional Abutments are compatible with subject device implants (CM 11.5°) and previously-cleared implants (CM 16° and 4°). This submission also includes Multifunctional Abutment components (cylinders) manufactured from Co-Cr-Mo alloy compatible with the subject Multifunctional Abutments and with abutments cleared in K170392; CAD-CAM abutment components (called "Interface") manufactured from Ti-6A1-4V alloy and Co-Cr-Mo alloy, compatible with the subject Multifunctional Abutments and with abutments cleared in K170392; and screws compatible with subject device components and previously cleared components with the 11.5° Morse taper connection. All subject device abutments are straight, with no angulation allowed.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The acid etching procedure is applied to all subject device dental implants. The acid etching process in this submission is identical to the process used to manufacture the dental implants cleared in K211921, and the HAM00 surface treatment is identical to that cleared in K211921.

    The subject device implants are compatible with abutments and prosthetic components in this submission and components cleared previously in K200992, K193096, K170392, and K051859.

    The subject device abutments and prosthetic components are compatible with implants and components in this submission and components cleared previously in K211921. K200992. K193096, and K170392.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the S.I.N. Dental Implant System. It does not describe an acceptance criteria or a study proving device performance against an acceptance criteria in the way typically found for AI/ML-driven medical devices. Instead, it demonstrates substantial equivalence to previously cleared predicate devices through non-clinical performance data and detailed comparisons of technological characteristics.

    Therefore, the requested information elements related to acceptance criteria, specific study designs (test sets, training sets, expert review, MRMC studies, standalone performance), and ground truth establishment are not applicable to this type of submission.

    The document focuses on showing that the new components (implants and abutments) of the S.I.N. Dental Implant System are as safe and effective as existing legally marketed predicate devices. This involves demonstrating equivalency in:

    • Intended Use: The indications for use are substantially equivalent to the predicate devices.
    • Technological Characteristics: Materials, manufacturing processes, design principles, sterilization methods, and performance data (non-clinical) are comparable or identical.

    Here's a breakdown of the requested information based on the provided document, highlighting why most are not directly present:

    1. Table of Acceptance Criteria and Reported Device Performance

    This type of table, with specific numerical acceptance criteria and a corresponding performance metric for an AI device, is not present because this is a 510(k) for a physical medical device (dental implants and components) and not an AI/ML software. The "performance data" provided are non-clinical tests demonstrating material properties, sterilization efficacy, biocompatibility, and mechanical integrity, which are described qualitatively or by reference to established standards (e.g., ISO, ASTM).

    Here's an illustrative table based on the provided "PERFORMANCE DATA" section, but it does not represent AI/ML acceptance criteria:

    Acceptance Criterion (Implicitly based on standards)Reported Device Performance (as summarized)
    Sterility Assurance Level (SAL) of 10⁻⁶ (ISO 11137-1 & 11137-2 for gamma; ANSI/AAMI/ISO 17665-1 & TIR 17665-2 for moist heat)Demonstrated by selecting/substantiating a 25 kGy dose (gamma) and by overkill method (moist heat), referenced from predicate devices.
    Bacterial Endotoxin Limit < 20 EU/device (ANSI/AAMI ST72)Demonstrated by LAL test on water samples (weekly) and sterilized product samples (quarterly), referenced from predicate devices.
    Shelf Life (packaging sterile barrier integrity and product sterility) after 4 years (ASTM F1929, F88/F88M)Demonstrated by testing samples after 4 years of real-time aging, referenced from predicate devices.
    Biocompatibility (ISO 10993-5, -3, -6, -10, -11)Demonstrated for implant (ASTM F67), abutment (ASTM F136, F1537), and zirconia coping (ISO 13356) materials, and for HAnano coating, referenced from predicate devices.
    HAnano Coating Characterization (SEM, XPS, TEM, XRD, adherence)Characterization performed and leveraged from predicate devices.
    Implant Acid-Etched Surface Characterization (SEM)Characterization performed and leveraged from predicate devices.
    MR Environment Safety (ASTM F2052, F2213, F2182, F2119 and FDA guidance)Non-clinical analysis and testing performed to evaluate metallic subject devices in MR environment.
    Mechanical Integrity (Implants with Abutments) (ISO 14801)Engineering analysis demonstrated subject devices do not create a new worst-case construct; previous mechanical testing applicable.

    2. Sample size used for the test set and the data provenance: Not applicable (no AI/ML test set mentioned). The document refers to non-clinical testing of physical device components.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable (no AI/ML ground truth mentioned).

    4. Adjudication method for the test set: Not applicable (no AI/ML test set mentioned).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable (no AI/ML or human reader study mentioned).

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable (no AI/ML algorithm mentioned).

    7. The type of ground truth used: Not applicable (no AI/ML ground truth mentioned). The "ground truth" for this submission are the established performance characteristics and safety profiles of the predicate devices and general standards for dental implants.

    8. The sample size for the training set: Not applicable (no AI/ML training set mentioned).

    9. How the ground truth for the training set was established: Not applicable (no AI/ML training set mentioned).

    Ask a Question

    Ask a specific question about this device

    K Number
    K222231

    Validate with FDA (Live)

    Date Cleared
    2022-10-27

    (94 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    S.I.N. Dental Implants with lengths of 18, 20, 22, or 24 mm be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K211921, K200992, and K170392.

    This submission includes dental implants Epikut S with a Morse taper (CM) abutment interface and an acid-etched endosseous surface, and Epikut S Plus implants with an endosseous surface produced by acidetching followed by application of a hydroxyapatite coating (HA™M). The implant design and endosseous surfaces are nearly identical to the Epikut Plus CM implants cleared in K211921, with the exception of the implant-abutment connection (subject device 16° Morse taper, K211921 11.5° Morse taper), the additional body/platform diameter (4.0 mm), and the longer lengths (18, 20, 22, and 24 mm).

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The acid etching procedure is applied to all subject device dental implants. The acid etching process in this submission is identical to the process used to manufacture the dental implants cleared in K211921, and the HA100 surface treatment is identical to that cleared in K211921.

    AI/ML Overview

    This FDA 510(k) summary (K222231) describes a dental implant system and focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than on presenting a study with specific acceptance criteria, test sets, or performance metrics in the way these terms are typically used for AI/ML-based diagnostic devices.

    The document is a regulatory submission for a medical device (S.I.N. Dental Implant System), not a study evaluating algorithm performance. Therefore, it does not contain information about:

    • A table of acceptance criteria and reported device performance in the context of an algorithm's diagnostic accuracy.
    • Sample sizes, data provenance, number of experts, adjudication methods for a test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
    • Standalone algorithm performance.
    • Sample size for training sets or how ground truth for training sets was established.

    Instead, the document focuses on non-clinical performance data to show that the new components of the S.I.N. Dental Implant System (subject device) are as safe and effective as existing legally marketed devices.

    Here's an breakdown of the acceptance criteria (or rather, the demonstration of equivalence) and the studies (non-clinical tests and analyses) that prove the device meets these criteria, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are based on demonstrating that the subject device's design, materials, manufacturing, sterilization, and performance characteristics are substantially equivalent to predicate devices already cleared by the FDA. The performance data presented are non-clinical and relate to the physical and biological properties of the implant.

    Acceptance Criteria (Demonstrated Equivalence to Predicate Devices)Reported Device Performance (as demonstrated by non-clinical studies/analysis)
    Intended Use (Functional and esthetic rehabilitation of edentulous mandible or maxilla)Equivalent to K211921, K203725, K170392, and K050406. The subject device's Indications for Use statement is largely identical or similar to the predicate devices, with specific language related to tilting angles and required number of implants/splinting justified by K050406 and K203725.
    Material Composition (Unalloyed titanium, ASTM F67)Equivalent to K211921, K203725, K170392. All subject device implants are manufactured from the same material.
    Endosseous Surface Treatment (Acid-etched; HA™ for Epikut S Plus)Equivalent to K211921 and K170392 for acid-etching. The HA™ treatment is identical to that cleared in K211921. Characterization (SEM, XPS, TEM, XRD, adherence) from K211921 leveraged.
    Sterilization Method & Sterility Assurance Level (SAL) (Gamma irradiation to 10⁻⁶ SAL at 25 kGy)Equivalent to K211921 and K203725. Sterilization method selected and substantiated according to ISO 11137-1 and ISO 11137-2.
    Bacterial Endotoxin Limit (< 20 EU/device)Equivalent to K211921. Demonstrated via Limulus amebocyte lysate (LAL) testing according to ANSI/AAMI ST72 on samples from manufacturing water (weekly) and sterilized product (quarterly).
    Shelf Life/Packaging Integrity (4 years real-time aging)Equivalent to K211921. Testing of samples after 4 years of real-time aging according to ASTM F1929 and F88/F88M (packaging sterile barrier) and sterility testing of product.
    BiocompatibilityEquivalent to K211921. Biological evaluation performed according to ISO 10993-1, with test results leveraged from K211921.
    MR Safety (Magnetically induced displacement force, torque, RF heating, image artifact)Non-clinical analysis and testing performed according to ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119, and FDA guidance. Results demonstrate safety in the MR environment.
    Mechanical Performance / Implant-Abutment Connection (Engineering analysis of new components)Equivalent to K200992 (for mechanical testing applicability). Engineering analysis demonstrated that the subject device implants, with compatible previously-cleared abutments, do not create a new worst-case construct, thus making previous mechanical testing applicable. (Note: Specific mechanical test results/values are not provided in this summary, but the applicability of prior testing is the demonstration of equivalence here).
    Physical Dimensions (Body/platform diameters and lengths, including new 4.0mm diameter and 18-24mm lengths)Equivalent or within the range of previously cleared devices. New 4.0mm diameter is within the range of K170392. 18-24mm lengths are similar to K203725.

    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable as this document describes a 510(k) submission for a physical medical device, not an algorithm's test set. The "test sets" here refer to batches of physical devices or materials undergoing non-clinical laboratory testing. The provenance of the data is the manufacturer (S.I.N. - Sistema de Implante Nacional S.A.) or its testing facilities, and the studies are non-clinical evaluations.

    3. Number of Experts Used to Establish the Ground Truth and Qualifications

    This is not applicable for a non-clinical device submission. Ground truth, in this context, would relate to objective measurements against established standards (e.g., ASTM, ISO standards) rather than expert consensus on medical images.

    4. Adjudication Method

    Not applicable. Adjudication methods are typically associated with resolving discrepancies in expert interpretations during clinical or algorithm performance studies. The non-clinical tests described have defined protocols and objective pass/fail criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This type of study is relevant for evaluating the impact of an AI algorithm on human reader performance, which is not the subject of this 510(k) submission for a dental implant.

    6. Standalone Performance Study

    No. This refers to the performance of an algorithm without human intervention. This document details the performance characteristics of a physical implant device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests is established by compliance with recognized consensus standards (e.g., ISO, ASTM, AAMI) and the device's adherence to its design specifications, materials testing, and performance against these standards. For example:

    • Sterility: Achieved according to ISO 11137-1 and ISO 11137-2.
    • Bacterial Endotoxin: Meets limits (< 20 EU/device) as per ANSI/AAMI ST72.
    • Packaging: Integrity verified against ASTM F1929 and F88/F88M.
    • Biocompatibility: Evaluated against ISO 10993-1.
    • MR Safety: Assessed against ASTM F2052, F2213, F2182, F2119.
    • Mechanical Performance: Demonstrated through engineering analysis applying previous mechanical testing (K200992).
    • Material Composition: Conforms to ASTM F67.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device.

    In summary, this 510(k) pertains to a physical dental implant and uses non-clinical testing and engineering analysis to establish substantial equivalence to previously cleared predicate devices, rather than performing clinical studies or AI/ML algorithm validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211921

    Validate with FDA (Live)

    Date Cleared
    2021-10-20

    (121 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System components cleared in K200992, K193096, and K170398.

    This submission includes dental implants Epikut CM with a Morse taper (CM) abutment interface and a double acid-etched endosseous surface, and Epikut Plus CM implants with an endosseous surface produced by double acid-etching followed by application of a hydroxyapatite coating (HA™Y). These implant endosseous surfaces are identical to those cleared in K200992.

    This submission also includes dental implants Epikut HE with an external hexagon (HE) abutment interface and a double acid-etched endosseous surface, and Epikut Plus HE implants with an endosseous surface produced by double acid-etching followed by application of a hydroxyapatite coating (HA1000). These implant endosseous surfaces also are identical to those cleared in K200992.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The acid etching procedure is applied to all subject device dental implants. The acid etching process in this submission is identical to the process used to manufacture the dental implants cleared in K200992. K193096, and K170398. The Epikut Plus CM and Epikut Plus HE implants have the identical HA™ surface treatment cleared in K200992 and K193096.

    The subject device dental implants are compatible with abutments and prosthetic components cleared previously in K200992. K193096. K170398. and K051859. Epikut HE and Epikut Plus HE dental implants are not for use with angled abutments.

    AI/ML Overview

    The provided document (K211921 510(k) Summary for S.I.N. Dental Implant System) is a 510(k) premarket notification summary submitted to the FDA. It does not describe an AI medical device or a clinical study intended to prove its performance against acceptance criteria for such a device.

    Instead, this document describes a dental implant system and its substantial equivalence to previously cleared predicate devices. The performance data presented are non-clinical bench tests (sterilization, bacterial endotoxin, shelf-life, biocompatibility, and coating characterization) to demonstrate the device's physical and biological properties.

    Therefore, I cannot directly answer your prompt using the provided text, as the information regarding acceptance criteria, AI device performance, sample sizes for test sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details are not applicable to the content of this document.

    The document explicitly states: "No clinical data were included in this submission." and focuses on demonstrating substantial equivalence through non-clinical data and comparisons to predicate devices.