Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182101

    Validate with FDA (Live)

    Date Cleared
    2018-11-14

    (103 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute iscoke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The React™ 71 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with a durable polymer outer layer. A lubricious, polytetrafluoroethylene and Engage liner is used to create a structure that has both proximal stiffness and distal flexibility. The React™ 71 Catheter is also designed with an encapsulated radiopaque distal platinum-iridium markerband which is used for visualization under fluoroscopy. The React™ 71 Catheter is introduced into the vasculature through the split y-introducer sheath.

    The proximal end of the React™ 71 Catheter is designed with a thermoplastic elastomer strain relief and a clear hub. The distal end of the React™ 71 Catheter is designed with a hydrophilic coating. The React™ 71 Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion.

    AI/ML Overview

    Based on the provided text, the Riptide™ Aspiration System (React™ 71 Catheter) is a medical device. The acceptance criteria and the studies performed are detailed in the "510(k) Summary: K182101."

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists numerous tests with their acceptance criteria (implicitly, as the results consistently state "met the acceptance criteria") and the reported device performance. Due to the extensive list, a summarized table is provided below, grouping similar types of tests.

    Test CategorySpecific TestAcceptance Criteria (Implicit)Reported Device Performance
    BiocompatibilityCytotoxicityNo evidence of cell lysis/toxicity, Grade < 2 (Mild Reactivity)Test article extract showed Grade 0 (No Reactivity); considered non-cytotoxic.
    SensitizationNo evidence of delayed dermal contact sensitizationNo evidence of delayed dermal contact sensitization; does not elicit a sensitization response.
    IrritationDifference between test article and control extract mean score meets requirementsDifference between test article and control extract mean score was 0.0 and 0.2; considered a non-irritant.
    Acute Systemic ToxicityNo mortality or evidence of systemic toxicityNo mortality or evidence of systemic toxicity; does not indicate signs of toxicity.
    HemolysisHemolytic index within acceptable limits (e.g., <5%)Hemolytic index for direct contact was 0.8%, for extract was 0.0%; considered non-hemolytic.
    Complement ActivationConcentration of SC5b-9 not statistically higher than controlConcentration of SC5b-9 not statistically higher than control; not considered a potential activator of the complement system.
    Thrombogenicity (Canine Model)Lower thrombogenic potential than predicate deviceMean score of 1.7 compared to predicate's 2.0; demonstrates lower thrombogenic potential.
    PyrogenicityNo animal showed temperature rise ≥ 0.5°CNo animal showed a temperature rise ≥ 0.5°C; considered non-pyrogenic.
    MicrobialEthylene Oxide ResidualMet ISO 10993-7 criteriaMet the acceptance criteria.
    Ethylene Chlorohydrin ResidualMet ISO 10993-7 criteriaMet the acceptance criteria.
    Bioburden RecoveryMet ISO 11737-1 criteriaMet the acceptance criteria.
    Bacterial EndotoxinMet ANSI/AAMI ST72 and USP <161> criteriaMet the acceptance criteria.
    BioburdenMet ISO 11737-1 criteriaMet the acceptance criteria.
    PackagingVisual InspectionMet ASTM F1886 criteriaMet the acceptance criteria.
    Bubble LeakMet ASTM F2096 criteriaMet the acceptance criteria.
    Seal StrengthMet ASTM F88 criteriaMet the acceptance criteria.
    Performance (Bench)Visual InspectionMet criteria for x2.5 magnificationMet the acceptance criteria.
    Dimensional MeasurementsMet specified dimensionsMet the acceptance criteria.
    Tip BucklingWithstand maximum compressive forceMet the acceptance criteria.
    Kink ResistanceMet maximum kink diameterMet the acceptance criteria.
    ParticulateMet USP <788> criteriaMet the acceptance criteria.
    Coating LubricityMet average frictional forces criteriaMet the acceptance criteria.
    Tensile StrengthMet ISO 10555-1, Annex B criteriaMet the acceptance criteria at hub and shaft.
    Liquid LeakMet ISO 10555-1, Annex C criteriaMet the acceptance criteria.
    Corrosion ResistanceMet ISO 10555-1, Annex A criteriaMet the acceptance criteria.
    Hub Aspiration ResistanceMet ISO 10555-1, Annex D criteriaMet the acceptance criteria.
    RadiopacityMet criteria for markerband length, wall thickness, and fluoroscopy confirmationMet the acceptance criteria.
    Luer StandardsMet ISO 594-1 and ISO 80369-7 criteriaMet the acceptance criteria.
    Torque to FailureWithstand typical clinical torsional forcesWas able to withstand torsional forces typical of clinical use.
    Dynamic PressureWithstand typical clinical pressuresWas able to withstand pressures typical of clinical use.
    Coating IntegrityRemained coated and lubriciousRemained coated and lubricious.
    System PerformanceRecanalizationMet specified criteria for recanalizationMet the acceptance criteria.
    Vacuum PressureMet specified criteria for vacuum pressureMet the acceptance criteria.
    Suction Flow RateMet specified criteria for suction flow rateMet the acceptance criteria.
    Lumen CollapseRemained patentRemained patent.
    UsabilityMet criteria for maneuverability, flexibility, aspiration, and clot retrievalMet the acceptance criteria.

    2. Sample Size for the Test Set and Data Provenance:

    The document does not explicitly state sample sizes for each bench or animal study. It mentions a "canine model" for the thrombogenicity test and a "porcine model" for the animal performance data. Specific numbers of animals or test articles used in these studies are not provided.

    The data provenance is non-clinical bench and animal testing. There is no mention of human subject data. Therefore, it is entirely retrospective or prospective non-clinical data performed in a laboratory/animal setting. The country of origin of the data is not specified, but the submission is to the U.S. FDA, suggesting the studies were conducted to meet U.S. regulatory standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable as the studies are entirely non-clinical (bench and animal testing). The "ground truth" is based on the objective measurements and observations from the physical and biological tests conducted according to established standards (e.g., ISO, ASTM, USP).

    4. Adjudication Method for the Test Set:

    This information is not applicable as the studies are entirely non-clinical. Adjudication typically refers to expert consensus or review in clinical studies.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The device is a physical medical device (catheter), not an AI software. The studies described are non-clinical bench and animal tests, not an MRMC comparative effectiveness study involving human readers or AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This information is not applicable. The device is a physical medical device, not an algorithm or AI system.

    7. The type of ground truth used:

    The ground truth used for the non-clinical studies is based on:

    • Objective Measurements: Such as dimensional measurements, pressure readings, flow rates, tensile strength, Luer standards, etc.
    • Standardized Test Methods: Adherence to established international (ISO) and national (ASTM, USP, ANSI/AAMI) standards for evaluating materials, sterility, packaging, and performance.
    • Biological Endpoints: In animal models, these include observations for systemic toxicity, sensitization, irritation, pyrogenicity, and direct measurement of thrombogenic potential.
    • Visual Inspection and Functional Assessment: E.g., for coating integrity, recanalization, lumen patency, and usability by trained personnel.

    8. The Sample Size for the Training Set:

    This information is not applicable. This is a submission for a physical medical device. The concepts of "training set" and "testing set" are primarily relevant to machine learning and AI, which are not involved here. The non-clinical studies evaluate the device's inherent physical and biological properties.

    9. How the ground truth for the training set was established:

    This information is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1