Found 4 results

510(k) Data Aggregation

    K Number
    K233804

    Validate with FDA (Live)

    Date Cleared
    2024-02-01

    (64 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Resonic™ Rapid Acoustic Pulse device is indicated for use as an accessory to the 1064 nm Q- Switched laser for black ink tattoo removal in Fitzpatrick Skin Type I-III patients. The Resonic device is also indicated for long-term improvement in the appearance of cellulite as supported by clinical data demonstrating treatment benefits up to one year of observation. In addition, the Resonic device can improve the appearance of skin laxity in conjunction with improvement in the appearance of cellulite.

    Device Description

    The Resonic™ Rapid Acoustic Pulse (RAP) device is designed as an accessory to laser treatments to improve laser tattoo fading efficiency, as well as a standalone device to improve the appearance of cellulite. Resonic uses repeated, rapidly rising acoustic waves, releasing pigment particles from the pigment laden macrophage (PLM) and dissipating the laser-induced whitening. This allows multiple laser passes in a single session, resulting in accelerated tattoo fading and fewer office visits to achieve sufficient tattoo fading. When used for improving the appearance of cellulite, the acoustic waves induce physical effects in the fibrous structures and improve the appearance of skin laxity.

    The Resonic™ Rapid Acoustic Pulse device is composed of three parts: the Console, the Handpiece and the replaceable Cartridge. The Console supplies saline to the Handpiece to enable formation of the shock wave within the acoustic pulse chamber. The Handpiece generates acoustic waves in the saline. The acoustic waves pass through the acoustically transparent window of the Cartridge and acoustic ultrasound gel or similar hydrogel pad, which is placed against the surface of the skin to be treated.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    Device Function / IndicationAcceptance CriteriaReported Device Performance
    Improvement in the appearance of skin laxity in conjunction with improvement in the appearance of celluliteCorrect identification by at least 2 of 3 blinded reviewers of 12-week post-treatment images, with respect to skin laxity from randomized groups of pre- and post-treatment images, should exceed a 60% prespecified threshold.90.2% (46/51) of participants had their post-treatment images (with regards to skin laxity) correctly identified by 2 of 3 blinded reviewers, exceeding the 60% prespecified threshold.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The independent, blinded photo assessment study comprised 56 participants. For the primary effectiveness endpoint on skin laxity, 5 participants were excluded as they were not considered to have skin laxity, resulting in an effective sample size of 51 participants.
    • Data Provenance: The images for the aesthetic study were "collected from the pivotal study that evaluated the Resonic device for improvement in the appearance of cellulite." The document does not explicitly state the country of origin or if the data was retrospective or prospective for the original pivotal study. However, the skin laxity assessment itself was a retrospective analysis of existing images.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Three independent, blinded physicians.
    • Qualifications of Experts: The document states they were "independent, blinded physicians." Specific details regarding their specializations (e.g., dermatologists, plastic surgeons) or years of experience are not provided.

    4. Adjudication Method for the Test Set

    • The adjudication method was a 2 out of 3 consensus. The primary effectiveness endpoint was defined as "correct identification by at least 2 of 3 blinded reviewers of 12-week post-treatment images with respect to skin laxity from randomized groups of pre- and post-treatment images."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not mentioned or conducted. This study focused on assessing the device's direct effect on skin laxity through expert review of treated vs. untreated images.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone (algorithm only) performance study was not conducted or mentioned. The study described relies on human expert assessment of photographic images, not an AI algorithm's performance.

    7. Type of Ground Truth Used

    • The ground truth used was expert consensus (specifically, a 2 out of 3 consensus among independent, blinded physicians) based on the visual assessment of photographic images.

    8. Sample Size for the Training Set

    • The document does not mention a training set as this study does not involve an AI algorithm that would require training. The assessment was based on human expert review of clinical images.

    9. How the Ground Truth for the Training Set Was Established

    • As there was no AI algorithm and therefore no training set, this information is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222629

    Validate with FDA (Live)

    Date Cleared
    2022-09-25

    (25 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Resonic™ Rapid Acoustic Pulse device is indicated for use as an accessory to the 1064 nm Q-Switched laser for black ink tattoo removal in Fitzpatrick Skin Type I-III patients. The Resonic device is also indicated for long-term improvement in the appearance of cellulite as supported by clinical data demonstrating treatment benefits up to 1 year of observation.

    Device Description

    The Resonic™ Rapid Acoustic Pulse (RAP) device is designed as an accessory to laser treatments to improve laser tattoo fading efficiency, as well as a standalone device to improve the appearance of cellulite. Resonic uses repeated, rapidly rising acoustic waves, releasing pigment particles from the pigment laden macrophage (PLM) and dissipating the laser-induced whitening. This allows multiple laser passes in a single session, resulting in accelerated tattoo fading and fewer office visits to achieve sufficient tattoo fading. When used for improving the appearance of cellulite, the acoustic waves induce physical effects in the fibrous structures, such as the fibrous septa in the subcutaneous tissue.

    The Resonic™ Rapid Acoustic Pulse device is composed of three parts: the Console, the Hand Piece and the Cable connecting the Hand Piece to the Console supplies saline to the Hand Piece to enable formation of the shock wave within the acoustic pulse chamber. The Hand Piece generates acoustic waves in the saline. The acoustic waves pass through the acoustically transparent window and acoustic ultrasound gel or similar hydrogel pad, which is placed against the surface of the skin to be treated.

    AI/ML Overview

    The provided text describes the regulatory clearance of the Resonic Rapid Acoustic Pulse device, specifically focusing on a software modification to change the pulse repetition rate for cellulite treatment from 50Hz to 100Hz. The information primarily relates to demonstrating substantial equivalence to a predicate device rather than a comprehensive study for initial device approval.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" for the clinical study. Instead, it describes a "primary effectiveness endpoint" and the results of a comparison study.

    Acceptance Criteria (Inferred Primary Effectiveness Endpoint)Reported Device Performance (100Hz Cellulite Dose Study)
    Blinded independent physician assessment (IPA) panel correctly identifies post-treatment 50 Hz dose and 100 Hz dose photographs.100% correct identification for both 50 Hz and 100 Hz treatment doses.
    No significant difference between the 50 Hz and 100 Hz doses in the Wilcoxon matched pairs signed rank test."No significant difference" found between the two doses in the Wilcoxon matched pairs signed rank test.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: 15 female participants.
    • Data Provenance: The document does not explicitly state the country of origin. It indicates a "bridging study was performed," implying a prospective study. Participants were treated with the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: The document refers to a "blinded independent physician assessment (IPA) panel." The specific number of physicians on this panel is not provided.
    • Qualifications of Experts: Only "physician" is mentioned; no specific qualifications (e.g., specialization, years of experience) are detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • The adjudication method is described as "Correct ID (2 of 3)" in Table 6-2 for the Cellulite Dose Study. This suggests that a decision was considered "correct" if at least two out of three adjudicators agreed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This study compared two treatment settings (50Hz vs. 100Hz) of the same device on patients, with physicians assessing the outcomes after treatment, not with AI assistance. Therefore, there is no information on human reader improvement with/without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not applicable as the device is a physical medical device (acoustic pulse device) and not an AI algorithm. The "performance data" section focuses on hardware safety, software verification, and clinical study outcomes, not an AI algorithm's standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Ground Truth Type: Expert consensus based on "blinded independent physician assessment (IPA)" of before and after photographs of patients. Outcomes data (long-term improvement in cellulite appearance) from previous studies also informed the context.

    8. The sample size for the training set

    • The document describes a clinical study to compare two treatment settings (50Hz vs. 100Hz) of the device. It does not mention a "training set" in the context of machine learning. The device itself is not described as an AI or machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no mention of a machine learning training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212502

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-11-05

    (88 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Soliton Resonic™ Rapid Acoustic Pulse device is indicated for use as an accessory to the 1064 nm Q-Switched laser for black ink tattoo removal in Fitzpatrick Skin Type I-III patients. The Resonic device is also indicated for long-term improvement in the appearance of cellulite as supported by clinical data demonstrating treatment benefits up to 1 year of observation.

    Device Description

    The Soliton Resonic™ Rapid Acoustic Pulse device is designed as an accessory to laser treatments to improve laser tattoo fading efficiency, as well as a standalone device to improve the appearance of cellulite. Resonic uses repeated, rapidly rising acoustic waves, releasing pigment particles from the pigment laden macrophage (PLM) and dissipating the laser-induced whitening. This allows multiple laser passes in a single session, resulting in accelerated tattoo fading and fewer office visits to achieve sufficient tattoo fading. When used for improving the appearance of cellulite, the acoustic waves induce physical effects in the fibrous structures, such as the fibrous septa in the subcutaneous tissue.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Device: Soliton Resonic™ Rapid Acoustic Pulse Device
    Intended Use: Long-term improvement in the appearance of cellulite.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Effectiveness)Reported Device Performance
    Primary Effectiveness Endpoint: A blinded physician panel correctly identifies long-term (>52-weeks) post-treatment photographs. (Specific threshold for correct identification is not explicitly stated as a percentage in the acceptance criteria, but implied by the reported performance).The blinded physician panel correctly identified the long-term (>52-weeks) post-treatment photograph in 95.2% of the cases.
    Mean Cellulite Severity Score Reduction: (Specific target reduction not explicitly stated, but implied as a measure of improvement).The mean cellulite severity score reduction was 1.09.
    Patient Satisfaction (at 52 weeks): Implied, as the study collected patient satisfaction responses and aimed to demonstrate maintained treatment benefits. (Specific percentage target not explicitly stated in acceptance criteria).At 52 weeks, all participants found their treatment areas to appear improved compared to the pre-treatment photos, and 97.6% of participants found there was good improvement in the appearance of cellulite.
    Safety: No significant adverse events, or only mild/moderate expected adverse events. (Implied, as safety is always a key consideration for medical devices).All adverse events observed were categorized as mild or moderate and were expected. No adverse events were related to the device. There were no serious adverse events.
    Long-Term Improvement: Demonstrates no significant reduction in treatment benefits up to 1 year of observation. (This is the overarching acceptance criteria for the new indication).The reported performance data (95.2% correct identification, 1.09 mean severity reduction, high patient satisfaction, and maintained safety profile) collectively demonstrates that the treatment effect with Resonic RAP device in improving the appearance of cellulite is maintained through 1 year after the treatment.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for the Test Set: A total of 67 participants were enrolled.
    • Data Provenance:
      • Country of Origin: United States (4 sites in the United States).
      • Retrospective or Prospective: Prospective multi-site trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: "Blinded independent reviewers" and "blinded physician panel." The exact number of individuals on the panel is not specified in this document.
    • Qualifications of Experts: They were described as "blinded independent reviewers" and "blinded physician panel." While the specific medical specialty (e.g., dermatologist) or years of experience are not explicitly stated, the term "physician panel" implies qualified medical professionals.

    4. Adjudication Method for the Test Set

    The adjudication method is implied as "blinded independent reviewers" or a "blinded physician panel" assessing photographs. The specific scheme (e.g., 2+1, 3+1) for resolving disagreements among multiple reviewers is not explicitly stated. It just mentions they "assessed by blinded independent reviewers to identify pre-treatment images when compared to post treatment images."


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study comparing human readers with vs. without AI assistance was not done.

    This study focused on the effectiveness of the device itself (Resonic RAP) for improving cellulite appearance, as assessed by a blinded physician panel, not on human reader performance assisted by the device or AI. The device is not an AI diagnostic tool that "assists" human readers in making diagnoses. It's a treatment device.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    Not applicable in the conventional sense for an AI device. The Resonic RAP device is a physical acoustic pulse device, not an AI algorithm. Its performance is inherent in its physical operation, not an output from a standalone algorithm. The "standalone" aspect in this context refers to its use as a "standalone device to improve the appearance of cellulite" as opposed to an accessory for tattoo removal, but this is a treatment application, not a diagnostic algorithm.


    7. The Type of Ground Truth Used

    The ground truth for effectiveness was established by:

    • Expert Consensus/Assessment: Assessment of serial clinical photographs by a "blinded independent reviewers" or "blinded physician panel" to identify post-treatment images and evaluate cellulite severity reduction.
    • Patient Reported Outcomes: Patient satisfaction surveys were used to assess participants' perception of improvement.

    8. The Sample Size for the Training Set

    Not applicable/Not provided within this document. This document describes a clinical study for a physical medical device (Resonic RAP), not the development or validation of a machine learning algorithm that requires a separate training set. The "training set" concept is typically relevant for AI/ML devices.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As explained above, this document does not describe the development of an AI/ML algorithm with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210964

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-27

    (27 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Soliton Resonic™ Rapid Acoustic Pulse device is indicated for use as an accessory to the 1064 nm Q-Switched laser for black ink tattoo removal in Fitzpatrick Skin Type I-III patients. The Resonic device is also indicated for short-term improvement in the appearance of cellulite.

    Device Description

    The Soliton Resonic™ Rapid Acoustic Pulse device is composed of three parts: the Console, the Hand Piece and the Cable connecting the Hand Piece to the Console. The Console supplies saline to the Hand Piece to enable formation of the shock wave within the acoustic pulse chamber. The Hand Piece generates acoustic waves in the saline. The acoustic waves pass through the acoustically transparent Window and acoustic ultrasound gel or similar hydrogel pad, which when placed against the surface of the skin to be treated.

    AI/ML Overview

    This document describes the Soliton Resonic™ Rapid Acoustic Pulse device, concentrating on its substantial equivalence to a previously cleared device (K201801). The information provided focuses on the modifications made and the types of testing performed to demonstrate that these changes do not alter the fundamental technology or safety/effectiveness.

    Here's an analysis of the provided text in relation to your request:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state quantitative acceptance criteria or detailed performance metrics. Instead, it relies on demonstrating that the modified device functions "as specified" and is "as safe and effective as its predicate device." The performance data section focuses on ensuring the modifications did not negatively impact safety or fundamental operation.

    Acceptance Criteria (Implied)Reported Device Performance
    Device functions as specifiedVerification testing confirmed the Resonic RAP device functions as specified.
    As safe as predicate deviceElectrical safety and EMC testing repeated an independent lab (IEC 60601-1, IEC 60601-1-2). Biocompatibility established based on predicate (no new patient-contacting materials).
    As effective as predicate deviceVerification testing demonstrated the Resonic RAP device is as effective as its predicate.
    Software changes are acceptableSoftware verification testing conducted and results found acceptable.
    No alteration to fundamental scientific technologyModifications did not alter the fundamental scientific technology (shape, frequency, repetition rate of acoustic waves unchanged).
    No new questions of safety or effectivenessMinor differences do not alter fundamental technological characteristics or present new safety/effectiveness questions.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify a separate sample size for a "test set" in the context of clinical performance. The "verification testing" mentioned would typically refer to engineering and functional tests.
    • Data Provenance: The document does not provide information about the country of origin or whether data是在retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided. The document focuses on engineering verification and equivalence to a predicate, not a clinical study involving expert assessment of ground truth.

    4. Adjudication method for the test set:

    This information is not provided as there is no specific mention of a clinical test set requiring adjudication in this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device described, the Soliton Resonic™ Rapid Acoustic Pulse device, is a physical medical device (an accessory to a laser for tattoo removal and for cellulite improvement), not an AI/software device that would involve human readers or AI assistance in image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable for the same reason as point 5.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Given the context of demonstrating substantial equivalence for hardware modifications, "ground truth" as typically understood in clinical or AI studies (e.g., pathology for disease diagnosis) is not relevant here. The "truth" used for verification is that the device functions according to its engineering specifications and meets safety standards, and that its output (acoustic waves) remains fundamentally unchanged from the predicate device.

    8. The sample size for the training set:

    This is not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    This is not applicable as there is no training set mentioned.

    Summary of what the document does provide regarding the study:

    The document focuses on a design control and risk analysis approach to demonstrate substantial equivalence for minor hardware and software modifications to an already cleared device (K201801). The "study" isn't a traditional clinical efficacy study but rather a series of engineering and functional verification tests.

    • Verification testing: Conducted using the "same methods as used for the predicate device" to show the Resonic RAP functions as specified and is as safe and effective.
    • Electrical safety and EMC testing: Repeated by an independent lab in accordance with IEC 60601-1 and IEC 60601-1-2.
    • Biocompatibility: Established based on the predicate device, as "no new materials are used on the surface of the device that contacts the patient skin."
    • Software verification testing: Conducted, and results were found acceptable for the software changes.

    The core argument is that the "minor differences in the newer device version (the subject device) do not alter the fundamental technological characteristics or present different questions of safety or effectiveness as compared to the predicate device." Therefore, extensive new clinical studies to establish efficacy or clinical ground truth were not deemed necessary for this 510(k) submission.