Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K152134

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-01-15

    (168 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The reprocessed vessel sealer/dividers are bipolar electrosurgical instruments intended for use with the ForceTriad™ Energy Platform in general, minimally invasive and open surgical procedures where ligation and division of vessels and lymphatics is desired. The instrument creates a seal by application of RF electrosurgical energy to vascular structures (vessels and lymphatics) interposed between the iaws of the instrument. A blade within the instrument is surgeon actuated to divide tissue.

    Indications for use include general open and minimally invasive procedures including urologic, vascular, thoracic and thoracoscopic, and gynecologic procedures where ligation and division of the vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectomy, Nissen fundoplication, colectomy, adhesiolysis, oophorectomy, etc. The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

    The reprocessed vessel sealer/dividers can be used on vessels and lymphatics up to and including 7 mm, and tissue bundles.

    Device Description

    The reprocessed vessel sealer is a sterile, hand-held bipolar RF electrosurgical instrument designed exclusively for use with the ForceTriad™ energy platform (generator) to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics clamped between the jaws, grasping tissue, and blunt dissection during general surgical procedures (as indicated), laparoscopic or open, using radio frequency (RF) energy. A hand actuated lever allows the user to open and close the instrument jaws without having to latch the lever, which includes a clicking mechanism that indicates to the user that the jaws are in the grasping zone, a button (switch) to activate the LigaSureTM mode by closing the handle against the button (switch) for vessel sealing, and a trigger to actuate an independent cutting blade.

    AI/ML Overview

    This document describes the regulatory clearance for reprocessed vessel sealers (device name: Reprocessed Vessel Sealers, models LF1737 and LF1723) which are electrosurgical instruments. The submission aims to demonstrate substantial equivalence to predicate devices (Covidien LigaSure™ 5 mm, Maryland Jaw Sealer/Dividers, K133338).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance values in the way one might expect for an AI/CADe device. Instead, the acceptance criteria are implicitly defined by a series of non-clinical tests designed to demonstrate that the reprocessed devices perform as intended and are safe and effective, similar to the original predicate devices. The reported device performance is generally stated as achieving compliance with these tests.

    However, based on the "Summary of Non-Clinical Tests Conducted" section, we can infer the following:

    Acceptance Criteria (Inferred from Tests)Reported Device Performance (Summary)
    Cleaning Validation:Validated cleaning procedures.
    Sterilization Verification:Validated sterilization procedures.
    Ethylene Oxide Residual Testing:Compliant with ISO 10993-7.
    Packaging Validation:Compliant with ASTM D4169, ASTM F88, ASTM F2096.
    Shelf-Life Validation:Compliant with ASTM 1980-07.
    Visual Inspection:Part of manufacturing process and bench testing to demonstrate appropriate functional characteristics.
    Fatigue Testing:Performed as part of functional performance (bench testing).
    Electrical Safety:Testing performed.
    Device Functionality:Testing performed.
    Vessel Seal Burst (Static and Burst Pressure):Testing performed. Performance shows reprocessed vessel sealers perform as originally intended.
    Vessel Seal Thermal Spread:Testing performed.
    EMC:Testing performed.
    Mold Stress:Testing performed.
    Drop:Testing performed.
    Spillage:Testing performed.
    Push:Testing performed.
    Biocompatibility:Testing performed for Cytotoxicity, Pyrogen, Thrombogenicity, Irritation, Acute Toxicity, Immune Response, Hemolysis, Sensitization.

    Overall Conclusion: "Performance testing shows the reprocessed vessel sealers to perform as originally intended." and "Sterilmed concludes that the reprocessed vessel sealers are safe, effective, and substantially equivalent to the predicate devices..."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document mentions that "Representative samples of reprocessed devices were tested" for various functional and safety characteristics. However, specific sample sizes (e.g., number of devices, number of seals, number of tests for each parameter) for each test are not provided.

    The data provenance is not explicitly stated in terms of country of origin. The study appears to be bench and laboratory testing performed by the manufacturer/reprocessor (Sterilmed, Inc., located in Minnesota, USA). The nature of these tests (cleaning, sterilization, packaging, functional performance, biocompatibility) indicates they are prospective tests performed on manufactured devices, not retrospective data analysis from clinical cases.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable as the device is a reprocessed surgical instrument, not an AI/CADe device or diagnostic tool that requires expert human interpretation to establish ground truth for a test set. The "ground truth" for this device's performance is established by objective engineering and safety standards (e.g., burst pressure thresholds, sterility levels, biocompatibility profiles) and comparison to the performance of the original device.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This section is not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in diagnostic studies involving human readers or AI outputs. The tests described are objective engineering and laboratory tests where the outcome is determined by adherence to a predefined standard or measurement, not by expert consensus or adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating diagnostic performance, especially for AI-assisted image interpretation. The device in question is a reprocessed surgical instrument, and its performance is evaluated through bench testing and safety assessments, not through human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable. The device is not an algorithm or AI system. It is a physical medical instrument. Therefore, the concept of "standalone (algorithm only)" performance does not apply.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is based on objective engineering standards, validated laboratory testing protocols, and comparison to the established performance of the original, new device. For example:

    • Sterility: Absence of microorganisms, verified by sterilization indicators and microbial testing.
    • Ethylene Oxide Residuals: Concentrations below a specified limit (ISO 10993-7).
    • Mechanical Functionality: Ability to open, close, grasp, seal, and cut within specified force ranges and dimensions, often compared to the original device's specifications.
    • Vessel Seal Burst Strength: Ability to withstand a certain pressure without leaking (e.g., specified kPa or mmHg).
    • Electrical Safety: Compliance with electrical safety standards (e.g., leakage current, insulation).
    • Biocompatibility: Absence of adverse biological reactions (cytotoxicity, irritation, etc.) validated through in vitro and in vivo tests.

    Essentially, the ground truth is established by meeting predefined scientific and engineering thresholds that demonstrate the device is safe and performs identically or equivalently to the original device.

    8. The sample size for the training set

    This section is not applicable. The device is a physical reprocessed medical instrument, not an AI model. Therefore, there is no "training set" in the context of machine learning. The "training" for the reprocessing procedures involves process validation and adherence to established protocols for cleaning, sterilization, and functional testing.

    9. How the ground truth for the training set was established

    This section is not applicable as there is no "training set" in the machine learning sense. However, if one were to loosely interpret "training set" as the data that informed the reprocessing procedures and quality checks, then the "ground truth" for establishing these procedures would be derived from:

    • Regulatory standards: Follow FDA guidelines and relevant international standards (e.g., ISO for sterilization, biocompatibility).
    • Manufacturer's specifications: Understanding the original device's design, materials, and performance requirements.
    • Scientific and engineering principles: Applying established methods for cleaning, disinfection, sterilization, and material compatibility.
    • Validation studies: Performing rigorous studies to prove that the reprocessing methods consistently achieve the desired outcomes (e.g., sterility, functionality).

    The "ground truth" in this context is the successful demonstration that the reprocessed device can consistently meet the same safety and performance characteristics as the new, original device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123096

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-01-25

    (115 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The reprocessed vessel sealers are indicated for use in general and gynecological, laparoscopic surgical prooedures where ligation of vasculature is desired. The reprocessed vessel sealers can be used to seal to seal to seal to seal and including 7 imm, lymphatics, and tissue bundles. The reprocessed vessel scalers can also be used to seal pulmonary vasculature when used with the ForceTriad™ energy platform.

    Device Description

    The vessel sealers are electrosurgical instruments for use with the ForceTriad™ energy platform when performing laparoscopic procedures. The vessel sealers are capable of sealing vessels and lymphatics, grasping tissue and dissection. The vessel sealers have a 5 mm shaft diameter, 37 cm shaft length and a shaft rotation of 159 degrees. The distal end mechanism may have a blunt or dolphin nose tip.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Reprocessed Vessel Sealers:

    Based on the provided information, this submission focuses on the substantial equivalence of a reprocessed medical device to its original predicate devices, rather than a novel device requiring a typical clinical study with acceptance criteria for efficacy. The "acceptance criteria" here therefore refers to demonstrating that the reprocessed device performs identically to the predicate device and meets established safety and functional standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance
    Functional EquivalencePerformance identical to predicate devices in design, materials, clinical applications, patient population, performance specifications, and method of operation."The reprocessed vessel sealers are identical to the predicate devices in design, materials of construction, and intended use. There are no changes to the clinical applications, patient population, performance specifications, or method of operation." "Performance testing shows the reprocessed vessel sealers to perform as originally intended."
    Cleaning ValidationValidation of cleaning procedures."Process validation testing was performed to validate the cleaning... procedures." Specific details of acceptance metrics for cleaning are not provided, but it's stated as validated.
    Sterilization ValidationValidation of sterilization procedures (ISO 11135, USP <71>)."Process validation testing was performed to validate the ... sterilization procedures." Adherence to ISO 11135 and USP <71> implies meeting their respective sterility assurance level (SAL) requirements.
    BiocompatibilityAssurance of biocompatibility (ISO 10993)."Biocompatibility testing (ISO 10993)" was performed. Adherence to ISO 10993 implies meeting its requirements for biological safety.
    Ethylene Oxide ResidualsTesting for ethylene oxide residuals (ISO 10993-7)."Ethylene oxide residual testing (ISO 10993-7)" was performed. Adherence to ISO 10993-7 implies meeting its requirements for acceptable residual levels.
    Packaging ValidationValidation of packaging (ASTM D.4169, ASTM F 88, ASTM F 2096)."Packaging validation (ASTM D.4169, ASTM F 88, ASTM F 2096)" was performed. Adherence to these ASTM standards implies meeting requirements for package integrity and sterile barrier maintenance.
    Shelf Life ValidationValidation of shelf life (ASTM 1980-07)."Shelf life validation (ASTM 1980-07)" was performed. Adherence to ASTM 1980-07 implies meeting requirements for maintaining sterility and functionality over the declared shelf life.
    Functional PerformanceDemonstrated appropriate functional characteristics; bench testing through simulated use, visual inspection, fatigue testing, and function testing."Representative samples of reprocessed vessel sealers were tested to demonstrate appropriate functional characteristics." "validation of functional performance (bench testing) was performed through simulated use, visual inspection, fatigue testing, and function testing." This indicates the device met pre-defined functional requirements, though specific metrics are not detailed.
    Manufacturing ProcessIncludes visual and validated functional testing of all products produced."the manufacturing process includes visual and validated functional testing of all products produced." This is an ongoing quality control measure to ensure individual units meet standards.

    2. Sample Size and Data Provenance

    The document does not explicitly state specific sample sizes for each non-clinical test (e.g., how many reprocessed devices were tested for cleaning validation). It refers to "Representative samples" for functional testing.

    • Test Set Sample Size: Not explicitly stated for specific tests, but general reference to "representative samples."
    • Data Provenance: The studies are described as internal non-clinical tests performed by Sterilmed, Inc. (submitter). The country of origin for the data is implicitly the USA, where Sterilmed is located and where the submission was made to the FDA. The data is prospective in the sense that these tests were conducted by Sterilmed for the purpose of this 510(k) submission to demonstrate equivalence.

    3. Number of Experts and Qualifications for Ground Truth

    This type of submission (reprocessed device) primarily relies on engineering and laboratory test standards rather than expert interpretation of clinical data or images. Therefore, the concept of "ground truth established by experts" as seen in diagnostic AI is not directly applicable here. The "experts" would be the engineers, microbiologists, and other technical specialists who conducted and validated the non-clinical tests according to established international and national standards (e.g., ISO, ASTM, USP). Their qualifications would be expertise in those particular testing methodologies and standards.

    4. Adjudication Method

    Not applicable. This is not a study involving human reader interpretation or clinical outcomes requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This is a submission for a reprocessed medical device demonstrating equivalence through non-clinical testing. It does not involve AI, human reader performance, or clinical comparative effectiveness studies in the traditional sense of a novel diagnostic or therapeutic device.

    6. Standalone Performance Study

    Yes, in a sense. The "standalone performance" is demonstrated through the various non-clinical tests (cleaning, sterilization, biocompatibility, functional performance, etc.) where the reprocessed device is tested in isolation against defined criteria and regulatory standards. The purpose is to show the device performs on its own, as originally intended, after reprocessing.

    7. Type of Ground Truth Used

    The ground truth used in this context is predominantly:

    • Engineering/Performance Standards: Adherence to established performance specifications (e.g., sealing capability, fatigue resistance, visual inspection criteria, shaft rotation).
    • Regulatory/Industry Standards: Compliance with international and national standards for sterility (ISO 11135, USP <71>), biocompatibility (ISO 10993), packaging (ASTM D.4169, ASTM F 88, ASTM F 2096), shelf life (ASTM 1980-07), and residual chemicals (ISO 10993-7).
    • Predicate Device Performance: The original functional specifications and performance of the predicate Covidien LigaSure™ Laparoscopic Instruments serve as the benchmark for "as originally intended" performance.

    8. Sample Size for the Training Set

    Not applicable. This is a reprocessed device submission based on non-clinical testing, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI model in this submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1