510(k) Data Aggregation

    K Number
    K242173

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-12

    (111 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.

    Device Description

    The ReTrace Ureteral Access Sheath is a sterile, single use device that is inserted over a guidewire into the upper or lower urinary tract of patients requiring ureteroscopic procedure sin a hospital environment. The ReTrace Ureteral Access Sheath consists of:

    • Reinforced tube/sheath
    • Introducer/dilator
    • Connector
    • Clip

    The reinforced polymeric introducer sheath includes a hydrophilic coating on the exterior and a lubricious inner surface. The sheath is reinforced to provide kink resistance. The distal tip of the sheath is fitted with a radiopaque ring. A white connector is fitted on the proximal end of the sheath and an orange connector is on the proximal end of the introducer to allow the two components to be attached.

    The dilator is fitted with a Luer connector on the proximal end and has a hydrophilic coating on the distal end. A guidewire entry and exit eye and exit holes for fluid delivery are located at the distal end of the introducer. The dilator protrudes out the distal end and allows for a gradual enlargement of the passage if needed and improves trackability along the guidewire. Once in the desired location, the dilator is removed to establish a hollow conduit for insertion of endoscopes and other instruments into and out of the urinary tract.

    The ReTrace Ureteral Access Sheaths are available in two diameters 10/12 Fr and 12/14 Fr. Lengths range between 28 and 55 cm.

    This submission is for modifications to device materials, labeling, packaging, and connector design.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based solely on the provided text.

    Crucial Note: The provided document is a 510(k) summary for a medical device (Ureteral Access Sheath) that underwent modifications. This means the primary argument for clearance is substantial equivalence to a previously cleared predicate device, rather than proving efficacy from scratch with a new device. Therefore, the details requested in your prompt regarding AI/ML models (e.g., training/test sets, human reader studies, ground truth establishment by experts) are generally not applicable to this type of device clearance. The performance data focuses on demonstrating that the modifications do not negatively impact the device's basic safety and function compared to the original cleared version.


    Acceptance Criteria and Reported Device Performance

    Given this is a 510(k) for a modified device, the "acceptance criteria" are implicitly that the modified device performs comparably to the predicate device in relevant mechanical, biocompatibility, and sterilization tests, and the modifications do not raise new questions of safety or effectiveness. The reported performance is that the device met these criteria.

    Acceptance Criteria (Implied for Device Modification)Reported Device Performance (Summary)
    Biocompatibility: Meets established standards (ISO 10993-1)."Biocompatibility testing was conducted based upon ISO 10993-1 (2018)... Meets Standard."
    Mechanical Performance: Maintains integrity and functionality as per design specifications."Mechanical testing was completed in support of the substantial equivalence determination and changes to the product line." This includes: Visual Inspection, Dimensional Inspection, Viscous fluid test, Simulated use test, Compatibility with accessories, Compatibility with Solutions, Peelability test, Tensile strength on introducer, Tensile strength between introducer and luer, Tensile strength between connector and sheath, Folding resistance on sheath, Radiopacity test on introducer, Friction test on sheath, Friction test on introducer tip, No perforation test, Tip flexibility, Nitinol tube compression / position.
    Sterilization Efficacy: Achieves a Sterility Assurance Level (SAL) of 10⁻⁶."The ReTrace Ureteral Sheaths are sterilized using ethylene oxide in a validated cycle demonstrating a microbial assurance level of 10⁻⁶."
    Packaging Integrity: Maintains sterility and product integrity through shelf life."Packaging testing included dye penetration testing, peel strength, peelability and visual inspection including through five year simulated accelerated aging."
    Shelf Life: Supports a 5-year shelf life."Shelf life testing to support 5 years." (Across various tests)

    Study Details (Based on Provided Text)

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated as a number of devices for each test. The document lists types of mechanical tests performed (e.g., tensile strength, friction tests) which implicitly involve testing multiple samples, but the exact quantity is not given.
      • Data Provenance: Not specified regarding country of origin. The studies are non-clinical testing (mechanical, biocompatibility, sterilization, packaging) and not human subject data. Thus, the concept of "retrospective or prospective" human data does not apply here.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. This document describes the clearance of a physical medical device (Ureteral Access Sheath) through non-clinical performance testing for substantial equivalence after modifications. It is not an AI/ML device that requires expert-established ground truth for a test set. The "ground truth" for these tests is defined by the technical specifications, standards (like ISO 10993-1), and acceptable performance parameters for the device.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not Applicable. Again, this pertains to expert review of data, typically in AI/ML performance studies. Here, the "adjudication" is whether the device passed specific engineering and performance tests, which would be determined by laboratory results and quality system procedures, not expert consensus on ambiguous data.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This type of study (MRMC, human readers with/without AI) is specifically for AI-powered diagnostic/interpretive devices. The Retrace Ureteral Access Sheath is a physical, mechanical medical device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. This is a physical medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this device's performance testing is defined by engineering specifications, industry standards (e.g., ISO 10993-1), and established test methodologies to ensure the device's physical and functional properties meet design requirements and maintain equivalency to the predicate device.
    7. The sample size for the training set:

      • Not Applicable. This is not an AI/ML device with a training set.
    8. How the ground truth for the training set was established:

      • Not Applicable. As above, no training set for an AI/ML model.

    Summary of why AI/ML related questions are not addressed:

    The provided text is a 510(k) summary for a physical, mechanical medical device (a Ureteral Access Sheath) that underwent modifications. The FDA clearance pathway for such devices, especially when arguing substantial equivalence, relies heavily on non-clinical performance testing (biocompatibility, mechanical integrity, sterilization validation, packaging) to demonstrate that the device is as safe and effective as a legally marketed predicate device. It does not involve AI/ML components, diagnostic interpretation, or human reader studies. Therefore, many of the questions asked in your prompt are not relevant to the content of this specific FDA submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181811

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-09-07

    (63 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ReTrace Ureteral Access Sheath is used to establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.

    Device Description

    The ReTrace Ureteral Access Sheath is comprised of the following components: Reinforced tube/sheath, Introducer/dilator, Connector, Clip. The reinforced polymeric introducer sheath includes a hydrophilic coating on the exterior and a lubricious inner surface. The sheath is reinforced to provide kink resistance. The distal tip of the sheath is fitted with a radiopaque ring. A white connector and orange clip is fitted on the proximal end of the sheath. The introducer is fitted with a Luer connector on the proximal end and has a hydrophilic coating on the distal end. A guidewire entry and exit eye and exit holes for fluid delivery are located at the distal end of the introducer. This submission expands the product line to include 28 cm long and 55 cm long sheaths with inner diameters of 10 and 12 Fr.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the ReTrace Ureteral Access Sheath. This specific submission is for an expansion of the product line, including new lengths and diameters of the existing device.

    Based on the information provided, this document does not contain the details of a study that proves the device meets specific acceptance criteria in the manner requested by the prompt. The submission is a "Special 510(k) Summary" which aims to demonstrate substantial equivalence to previously cleared predicate devices (K140523 and K123675) for new sizes of the ReTrace Ureteral Access Sheath, rather than presenting a de novo clinical study with specific acceptance criteria related to accuracy, sensitivity, or specificity.

    Therefore, many of the requested sections about acceptance criteria, device performance, sample sizes, ground truth establishment, expert involvement, and MRMC studies are not applicable to the content of this document.

    However, I can extract the information related to the device's mechanical and sterilization testing which serves as "performance data" in the context of this 510(k) submission for demonstrating substantial equivalence.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in the format of a clinical trial (e.g., sensitivity > X%, specificity > Y%). Instead, it lists mechanical tests performed to ensure the new device variations maintain equivalence to the predicate devices. The "reported device performance" is essentially that these tests were completed successfully and support substantial equivalence.

    Test CategorySpecific Test / CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Mechanical TestingVisualConformance to visual specificationsCompleted (implies conformance)
    Dimensional verificationConformance to specified dimensions (10/12 Fr ID, 28/55 cm L)Completed (implies conformance)
    Guidewire pullout forceMaintain appropriate force for safe useCompleted (implies satisfactory performance)
    Device integrity and functionalityDevice remains intact and functions as intendedCompleted (implies satisfactory performance)
    Simulated usePerforms as expected in simulated surgical environmentCompleted (implies satisfactory performance)
    Viscous fluidMaintains patency and allows passage of viscous fluidCompleted (implies satisfactory performance)
    Kink resistanceResists kinking during useCompleted (implies satisfactory performance)
    Shelf life testing to support 5 yearsMaintains mechanical and functional properties for 5 yearsCompleted (supports 5 years shelf life)
    BiocompatibilityBiocompatibilityDevice materials are biocompatible (no additional testing needed due to predicate)No additional testing provided, relies on predicate's data
    SterilizationSterilization (Ethylene Oxide - Microbial Assurance)Sterility Assurance Level (SAL) of 10⁻⁶Demonstrated SAL of 10⁻⁶ in a validated cycle

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document does not specify sample sizes for the mechanical or sterilization testing. These types of tests typically involve a defined number of units per batch or configuration, but the specific numbers are not disclosed.
    • Data Provenance: The tests are conducted internally by the manufacturer (Coloplast A/S, based in Denmark, with a US contact in Minneapolis) as part of their design control and quality system. It is laboratory-based testing, not clinical data from patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • This is not applicable as the document describes mechanical and sterilization testing, not studies requiring expert review of patient data to establish "ground truth." The "ground truth" for these tests are objective measurements against engineering specifications.

    4. Adjudication Method for the Test Set:

    • This is not applicable for mechanical and sterilization testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No animal studies or clinical testing were provided to support substantial equivalence between the subject and predicate devices." This is a regulatory pathway (510(k)) that, for this specific device and modification, relies on non-clinical data to demonstrate substantial equivalence, not comparative effectiveness.

    6. If a Standalone (Algorithm Only Without Human-in-the-loop Performance) Was Done:

    • This is not applicable. The device is a physical medical instrument (ureteral access sheath), not an algorithm or AI software. Therefore, there's no "standalone algorithm" performance to assess.

    7. The Type of Ground Truth Used:

    • The "ground truth" for the mechanical and sterilization testing referenced is based on engineering specifications, international standards, and validated methods for device performance and sterility. For example, a "guidewire pullout force" test would have an engineering specification for the minimum acceptable force, and the test result confirms it meets that specification.

    8. The Sample Size for the Training Set:

    • This is not applicable as there is no "training set" in the context of an AI/algorithm. The device is a physical medical instrument.

    9. How the Ground Truth for the Training Set Was Established:

    • This is not applicable as there is no "training set."

    Summary of Key Findings from the Document:

    The document is a regulatory submission demonstrating substantial equivalence for new sizes of an existing medical device. It relies on non-clinical (mechanical and sterilization) testing to show that the new sizes perform similarly to the predicate devices and do not raise new questions of safety or effectiveness. It explicitly states that no animal or clinical studies were conducted for this specific submission to support substantial equivalence.