Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K183486

    Validate with FDA (Live)

    Date Cleared
    2019-07-26

    (221 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RapidVit™ Oocyte: Media for vitrification of human oocytes (MII).
    RapidWarm™ Oocyte: Media for warming of vitrified human oocytes (MII).

    Device Description

    Two sets of media are covered by this 510(k), the RapidVit™ Oocyte for vitrification of oocytes and the RapidWarm™ Oocyte for warming of vitrified oocytes. RapidVit™ Oocyte contains three medium solutions to be used sequentially during oocyte vitrification. RapidWarm™ Oocyte includes four medium solutions to be used sequentially during oocyte warming.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Vitrolife RapidVit™ Oocyte and RapidWarm™ Oocyte devices, based on the provided text:

    Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. Therefore, it primarily presents data to show that the new device is as safe and effective as the existing one, rather than a full, comprehensive study report with all the details typically found in a clinical trial publication.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria provided are mainly for non-clinical performance and a Mouse Embryo Assay (MEA). The clinical performance is reported as observed rates, not explicitly tied to specific numerical acceptance criteria within the document, although they are presented as evidence of successful function.

    Acceptance Criteria CategorySpecific CriterionReported Device Performance / Result
    Non-Clinical Performance
    pH TestingPer USP <791>N/A (Confirmed compliance, values not provided)
    Osmolality TestingPer USP <785>N/A (Confirmed compliance, values not provided)
    Aseptic Filling ValidationPer ISO 11137-1:2006 and ISO 11137-2:2013N/A (Confirmed compliance)
    Bacterial Endotoxins Testing< 0.5 EU/mlN/A (Confirmed compliance, specific value not provided)
    Sterility TestingNo microbial growthN/A (Confirmed compliance)
    Mouse Embryo Assay (MEA)>80% of embryos expand to the blastocyst stage by 96hN/A (Confirmed compliance, specific percentage not provided in summary)
    Stability Testing (Shelf-life & Post-opening)pH, osmolality, sterility, 1-cell MEA, and endotoxin spec. met at end of shelf-life (25 weeks) and 2 weeks after opening.N/A (Confirmed compliance)
    Clinical Performance
    Oocyte Survival Rate(Not explicitly stated as an acceptance criterion in this format, but the observed rate supports device function)94% (555/593)
    Fertilization Rate(Not explicitly stated)78% (434/555)
    Day 5 Blastulation Rate(Not explicitly stated)24% (102/434)
    Day 5/6 Utilization Rate(Not explicitly stated)35% (153/434)
    Clinical Pregnancy Rate (confirmed by fetal heartbeat)(Not explicitly stated)50% (27/54 recipients)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study): 593 oocytes from 64 donors.
    • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It is referred to as "A clinical study was conducted," which typically implies a prospective design, but this is not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. For a medical device like reproductive media, the "ground truth" for clinical performance is generally the observed biological outcomes (survival, fertilization, blastulation, pregnancy), which are objectively measurable laboratory and clinical endpoints rather than subjective expert interpretations of images or data. Therefore, a panel of experts for "ground truth" establishment in the typical sense (e.g., for diagnostic AI) might not be applicable here.


    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (like 2+1 or 3+1) is typically relevant when human experts are assessing a subjective outcome (e.g., classifying an image, making a diagnosis) and their agreement needs to be established or disagreement resolved. For the clinical outcomes measured in this study (oocyte survival, fertilization, pregnancy rate), the outcomes are objective biological events, not subjective interpretations. Therefore, an adjudication method in this context is not applicable and not mentioned.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study assesses how human readers' performance (e.g., radiologists interpreting images) changes with or without AI assistance. The device in question is a media (liquid solutions) used in an in-vitro fertilization process, not an AI diagnostic tool or system that assists human interpretation.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone "algorithm only" performance study was not done. This device is a biological media. Its performance is intrinsically tied to its use in a laboratory setting by human embryologists/clinicians performing the vitrification and warming procedures. The concept of an "algorithm only" performance is not applicable to this type of device.


    7. The Type of Ground Truth Used

    The ground truth used for the clinical study was based on observed biological and clinical outcomes:

    • Oocyte survival post-warming.
    • Fertilization rates.
    • Embryo development (Day 5 blastulation, Day 5/6 utilization).
    • Clinical pregnancy confirmation by fetal heartbeat.

    For the non-clinical tests (pH, osmolality, endotoxins, sterility, MEA), the ground truth was based on established laboratory assay results and conformity to specified criteria.


    8. The Sample Size for the Training Set

    This information is not applicable and not provided. The device is a biological media, not an algorithm or AI model that requires a training set.


    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1