Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K150627

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-05-27

    (77 days)

    Product Code
    Regulation Number
    890.5710
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rapid Aid Infant Heel Warmer is primarily used in hospitals, doctor's offices, and other healthcare facilities. It is an instant warm pack intended to be used on an infant's heel to increase blood circulation to the area to aid in the drawing of blood for analysis. It is a single use, non-toxic, non-sterile, disposable warmer.

    Device Description

    The Rapid Aid Infant Heel Warmer is a self contained unit comprised of a flexible, poly/nylon outer pouch containing:

    • a) A metal disc that can be flexed during activation,
    • b) Liquid solution of food grade sodium acetate and water,
    • c) Minute crystals of sodium acetate.
      An adhesive tape is attached to the top of the unit is activated by squeezing firmly on the inner fluid pouch, this will cause the inner perforated pouch to activate. Rapid crystallization occurs when the liquid contents are exposed to the minute crystals of sodium acetate contained within the poly/nylon outer pouch. This exothermnic reaction causes the unit to heat up to 104ºF. The adhesive tape strip is used to hold the warmer in place if desired.
    AI/ML Overview

    Here's an analysis of the provided text regarding the Rapid Aid Instant Disposable Infant Heel Warmer, focusing on acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The main performance characteristic tested for this device is its temperature profile.

    Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance (New Device)Reported Device Performance (Predicate Device)
    Maximum TemperatureWithin specified range102.6 ± 0.2 °F102.2 ± 0.4 °F
    Dwell Time (within 101-104 °F)Within specified range8.5 ± 1.0 min8.5 ± 0.8 min

    Note: The exact numerical "specified range" for maximum temperature and dwell time is not explicitly stated as a formal acceptance criterion but is implied by the comparison to the predicate device and the statement "both within the specified range for this device."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document mentions "The Infant Heel Warmer with Disc was tested against the predicate device," implying a test set, but does not provide the number of units tested.
    • Data Provenance: Not explicitly stated, but it's a bench test conducted by the manufacturer, Rapid Aid Corp. (Canada). This would be prospective data generated for the 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of device and study. The testing described is bench testing of physical properties (temperature, dwell time, material properties), not diagnostic accuracy requiring expert interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable as the study involves objective physical measurements, not subjective evaluations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not conducted. This device is a physical warmer, not an AI or diagnostic tool that would involve human readers interpreting cases.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is a physical medical device, not an algorithm, so "standalone performance" in the context of AI does not apply. The performance evaluation was essentially a standalone bench test of the device's physical properties.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluation appears to be the established performance characteristics of the legally marketed predicate device (Rapid Aid Instant Disposable Infant Heel Warmer, K040856) and the "specified range for this device" (though this range is not explicitly defined in the document). In essence, the new device's performance was compared to that of a previously cleared, substantially equivalent device.

    For material properties, the ground truth is based on ASTM standards (for poly/nylon pouch material) and toxicological acceptability (for sodium acetate).

    8. The Sample Size for the Training Set

    This information is not applicable. This is a physical medical device, not an AI model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040856

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-05-03

    (31 days)

    Product Code
    Regulation Number
    890.5710
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rapid Aid Infant Heel Warmer is primarily used in hospitals, Doctor's offices, and other healthcare facilities. It is an instant warm pack intended to be used on an infant's heel to increase blood circulation to the area to aid in the drawing of blood for analysis. It is a single use, non-toxic, non-sterile, disposable warmer.

    Device Description

    The Rapid Aid Infant Heel Warmer is a self contained unit comprised of a flexible, poly/nylon outer pouch containing:

    • a) A flexible, perforated, polyethylene/polyester inner pouch that holds the liquid solution.
    • b) Liquid solution of food grade sodium acetate and water contained in a).
    • c) Minute crystals of sodium acetate.
      An adhesive tape is attached to the top of the unit. The unit is activated by squeezing firmly on the inner fluid pouch, this will cause the inner perforated pouch to activate. Rapid crystallization occurs when the liquid contents are exposed to the minute crystals of sodium acetate contained within the poly/nylon outer pouch. This exothermic reaction causes the unit to heat up to 105 degrees F. The adhesive tape strip is used to hold the warmer in place on the infant's heel.
    AI/ML Overview

    This 510(k) summary does not contain sufficient detail to fully answer all aspects of the request, particularly regarding specific numerical acceptance criteria and a detailed study design for device performance. However, based on the provided text, here's an analysis of the acceptance criteria and the study conducted for the Rapid Aid Infant Heel Warmer:

    The primary "acceptance criterion" for this device, as presented in the 510(k), is substantial equivalence to predicate devices, specifically concerning temperature characteristics and intended use. The study focuses on demonstrating that the Rapid Aid Infant Heel Warmer performs "very similarly" to and is "substantially equivalent" to predicate infant heel warmers.


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document does not explicitly state numerical acceptance criteria for temperature output beyond "within the range of the predicate devices" and "reaching a maximum temperature around 105 F."

    Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance
    Intended UseSame as predicate devices (aid in drawing blood from infant heel)"intended use is the same"
    Performance AttributesSame as predicate devices"performance attributes are the same"
    Temperature OutputWithin the range of predicate devices; reach max ~105°F"temperature output... is within the range of the predicate devices"; "performed very similarly with temperatures within the same ranges, reaching a maximum temperature around 105 F within the first minutes and then steadily decreasing in temperature."
    Chemical CompositionPrimarily food-grade sodium acetate and water, same as predicate"chemical composition is primarily the same... food grade sodium acetate and water"
    Material SafetyLatex-free, non-sensitizing, non-toxic, food-grade chemicals"outer poly/nylon pouch material... is latex free and non-sensitizing"; "sodium acetate... is a nontoxic, food grade chemical and has been found to be toxicologically acceptable for it's intended use."
    Physical PropertiesMeets ASTM thickness, tensile strength, seal width, seal/burst integrity"tested for: thickness following ASTM D1203, tensile strength following ASTM D-882 and seal width"; "subject to incoming inspection for width/length, scal integrity and burst strength."
    Peak TemperatureConsistent with production batch results and predicate performance"Summary of peak temperature from batch production results is attached [not provided in extract]"; performs "very similarly" to predicates reaching ~105°F.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the performance testing against predicate devices. It refers to "batch production results" for peak temperature but does not specify how many units or batches were tested.

    • Sample Size: Not specified.
    • Data Provenance: Not explicitly stated, but assumed to be internal testing by Rapid Aid Ltd. There is no mention of country of origin for data or whether it was retrospective or prospective, although performance testing is generally prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This type of device (a simple warming pack) does not typically require expert-established ground truth in the way diagnostic or prognostic AI devices do. The "truth" here is objective physical measurements of temperature and material properties. Therefore:

    • Number of Experts: Not applicable/not specified for ground truth establishment.
    • Qualifications of Experts: N/A. The "ground truth" is based on instrumental measurements and adherence to ASTM standards.

    4. Adjudication Method for the Test Set

    Not applicable. This device does not involve subjective interpretations requiring adjudication (e.g., medical image reading). Performance is evaluated through objective measurements and comparisons to predicate device characteristics.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device (an infant heel warmer), not an AI diagnostic or prognostic tool. Therefore, MRMC studies and "human reader improvement with AI" are not relevant.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.


    7. The Type of Ground Truth Used

    The ground truth for this device's performance is based on:

    • Objective Physical Measurements: Temperature (peak and duration), material properties (thickness, tensile strength, seal integrity, burst strength), and chemical composition analysis.
    • Predicate Device Performance: The primary "ground truth" for comparison is the established performance characteristics and safety profile of the legally marketed predicate infant heel warmers.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires a "training set." The device's function is based on a chemical reaction, not a learned algorithm.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1