    K Number
    K151471

    Validate with FDA (Live)

    Date Cleared
    2015-09-17

    (108 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Radifocus Glidewire Endoscopic Wire is intended to be used for selective cannulation of the biliary ducts, including but not limited to, the common bile, cystic, right and left hepatic ducts during endoscopic biliary procedures for catheter introduction and exchanges.

    Device Description

    The Radifocus Glidewire Endoscopic Wire is an endoscopic guide wire that is provided sterile and is intended for single use only. It consists of a Nickel Titanium alloy core wire, and a polyurethane containing tungsten and hydrophilic polymer that are applied to the entire wire. There are two shaft configurations: standard and stiff; the stiff shaft has a slightly thicker core wire than that of the standard shaft. The wire distal segment comes in angled or straight configurations and is packaged in a plastic holder that is contained within an individual package. A guide wire inserter is contained within the individual package to assist with the insertion of the wire into an endoscope or catheter.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the Radifocus Glidewire Endoscopic Wire. It is a submission to the FDA demonstrating substantial equivalence to a predicate device, not a study describing acceptance criteria and device performance in the context of an AI-based system. Therefore, most of the requested information regarding AI study specifics like sample sizes for test and training sets, expert consensus, MRMC studies, and standalone performance is not applicable to this document.

    However, I can extract and describe the acceptance criteria and the studies that prove the device meets these criteria, as presented in the document, which are related to the physical performance and safety of the guidewire.

    Acceptance Criteria and Device Performance for the Radifocus Glidewire Endoscopic Wire

    The device is a medical guidewire (non-AI), so the acceptance criteria are based on physical performance, biocompatibility, and sterility, rather than diagnostic accuracy or AI performance metrics.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria & TestStandard / BasisReported Device Performance
    Performance Testing (ISO 11070: 2014)
    RadiodetectabilityISO 11070, Section 4.5All samples tested met the standard applicable to each test.
    Fracture TestISO 11070, Section 8.4All samples tested met the standard applicable to each test.
    Flexing TestISO 11070, Section 8.5All samples tested met the standard applicable to each test.
    Peak Tensile Force of GuidewireISO 11070, Section 8.6All samples tested met the standard applicable to each test.
    Performance Testing (Terumo's Internal Standards)
    Flexibility Test of Distal TipInternal Standards (based on predicate)Meets acceptance criteria
    Torque transmissionInternal Standards (based on predicate)Meets acceptance criteria
    Sliding Resistance/CoatingInternal Standards (based on predicate)Meets acceptance criteria
    Integrity (Product appearance)Internal Standards (based on predicate)Meets acceptance criteria
    Bending StrengthInternal Standards (based on predicate)Meets acceptance criteria
    BiocompatibilityISO 10993-1: 2009Concluded to be biocompatible for its intended use, identical to predicate which has a history of safe and effective use. Classified as Externally Communicating Device, Tissue/bone/dentin communicating, Limited Contact (<24 hrs).
    Sterilization (Sterility Assurance Level - SAL)ISO 11135SAL of 10^-6 (implied by "provide a SAL of 10-9" and general practice, though usually SAL of 10^-6 is specified in FDA guidances for terminally sterilized devices; assuming the document implies meeting current requirements, which would include 10^-6. The document explicitly states 10e-9) achieved.
    Sterilization (Residual Ethylene Oxide/Ethylene Chlorohydrin)ISO 10993-7Residual EO will not exceed 4 mg per device; Residual ECH will not exceed 9 mg per device.
    PyrogenicityUSP <85> Bacterial Endotoxins Test; FDA Guidance June 2012Certified to be non-pyrogenic in the unopened and undamaged package. Confirmed by LAL testing on each lot.
    Risk AnalysisISO 14971: 2007No issues of safety and effectiveness determined taking into account modifications.

    Regarding the specific questions you asked for an AI study, the following are not applicable to this 510(k) submission for a non-AI medical device (guidewire):

    1. Sample sizes used for the test set and the data provenance: Not applicable. Device is a physical guidewire, not an AI system processing data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable. "Ground truth" for this device refers to the physical and chemical properties meeting specified standards.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    However, to address the core of "the study that proves the device meets the acceptance criteria" based on the document:

    The studies that prove the device meets the acceptance criteria are non-clinical bench tests and material assessments. These include:

    • Performance Testing per ISO 11070: 2014: This standard covers "Sterile single-use intravascular introducers, dilators and guidewires." Tests performed included Radiodetectability, Fracture Test, Flexing Test, and Peak Tensile Force of Guidewire. These tests were conducted on both non-aged and accelerated aged samples.
    • Performance Testing per Terumo's Internal Standards: These tests included Flexibility Test of Distal Tip, Torque transmission, Sliding Resistance/Coating, Integrity (Product appearance), and Bending Strength. These were based on the predicate device's expected performance.
    • Biocompatibility Testing: Assessed in accordance with ISO 10993-1, classifying the device and confirming material compatibility through comparison to the predicate device's established safety record and identical material composition.
    • Sterilization Validation: Demonstrated compliance with ISO 11135 for Ethylene Oxide sterilization, ensuring a Sterility Assurance Level (SAL) of 10^-6 (although the document stated 10^-9, which is a higher level of assurance) and meeting residual EO/ECH limits per ISO 10993-7 for limited exposure devices.
    • Pyrogen Testing Validation: Conducted using Limulus Amebocyte Lysate (LAL) testing (Photometric Quantitative Method) in accordance with USP <85> and FDA guidance.
    • Risk Analysis: Performed according to ISO 14971 to identify and mitigate potential risks.

    These non-clinical tests collectively demonstrate that the Radifocus Glidewire Endoscopic Wire is safe and effective and substantially equivalent to the predicate device, fulfilling the regulatory requirements for the 510(k) submission. No clinical trials were conducted or required for this submission.