510(k) Data Aggregation

    K Number
    K190976

    Validate with FDA (Live)

    Date Cleared
    2019-07-12

    (88 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Radial Medical Cirvo Compression System is indicated for use in:

    • Preventing deep vein thrombosis (DVT)
    • Enhancing blood circulation
    • Diminishing post-operative pain and swelling
    • Reducing wound healing time
    • Treatment and assistance in healing: stasis dermatitis; venous stasis ulcers; arterial and diabetic leg ulcers
    • Treatment of chronic venous insufficiency
    • Reducing edema
    Device Description

    The Cirvo Compression System is a light-weight, mobile, sequential compression device (SCD) placed around the calf that sequentially compresses the calf to augment blood flow. SCD's are used in both the hospital and home setting. most commonly to prevent deep vein thrombosis (DVT), heal venous leg ulcers (VLU) and reduce swelling. The Cirvo Compression System includes three components: the Cirvo sequential compression device, an AC charger and a smart device app.
    The Cirvo compression unit generates compression cvcles using a small motor to sequentially cycle the compression unit on the calf. The Cirvo sequential compression device is placed around the lower leg using a detachable shin guard with self-aligning magnets, with the padded compression unit aligned on the calf. Proper fitting is achieved by adjusting a pair of hook and loop straps. The Cirvo compression system continuously monitors and maintains the prescribed compression therapy, records the duration of therapy session as well as compliance with the prescribed therapy. Therapy can be initiated and stopped using either the app interface on a smart device or using the capacitive touch sensor on the compression unit. Patients can track their progress with the app and can opt in to have text reminders sent to reinforce compliance with their therapy and notifications sent to their physician. All data is saved and transmitted using high-level, AES-128 encryption.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study information for the "Cirvo Compression System," structured according to your request.

    Please note: The provided document is a 510(k) summary for a medical device. This type of document primarily focuses on demonstrating substantial equivalence to a predicate device, rather than detailed clinical efficacy trials often described in scientific publications. As such, information regarding clinical studies, AI performance metrics, ground truth establishment, and specific multi-reader multi-case studies might be limited or absent.


    Description of Acceptance Criteria and Study to Prove Device Meets Criteria

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary for the Cirvo Compression System K190976 states that "All testing met pre-established acceptance criteria." However, it does not explicitly detail the quantitative acceptance criteria for each test performed. It lists the types of tests conducted and concludes that they were successful. Therefore, the table below reflects what is implied as meeting the acceptance criteria based on the narrative.

    Acceptance Criteria Category (Implied)Reported Device Performance
    Bench Testing
    Device strap tensile strengthSuccessfully passed
    Device buckle to cord tensile strengthSuccessfully passed
    Simulated use performanceSuccessfully passed
    Reliability performanceSuccessfully passed
    Packaging integritySuccessfully passed
    Software Testing
    Battery power supervisor softwareCompliant with FDA/IEC 62304 "moderate level of concern"
    Mobile application client softwareCompliant with FDA/IEC 62304 "moderate level of concern"
    Signal board softwareCompliant with FDA/IEC 62304 "moderate level of concern"
    Security verificationSuccessfully verified
    Electrical Safety Testing
    IEC 60601-1:2006 (Basic safety)Compliant
    IEC/EN 60601-1-2:2007 (EMC)Compliant
    FCC Part 15 & Part 18Compliant
    Industry Canada ICES-003 Issue 5Compliant
    Coexistence with intentional radiatorsSuccessfully demonstrated
    Therapy Parameters
    Pressure Delivered (40-60 mmHg)Maintained "prescribed compression therapy" (Implied)
    Therapy Duration MonitoringContinuously monitored (Implied as functional)
    Compliance with Therapy MonitoringContinuously monitored (Implied as functional)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific sample sizes for the "test set" in the context of clinical data. The performance testing described (bench, software, electrical safety) uses equipment and methods, but doesn't refer to a "test set" of patients or data in the way an AI/clinical study might. The data provenance is not applicable in the context of device performance testing; it's about the device itself and its components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The testing described is primarily engineering and regulatory compliance, not typically involving human expert-established ground truth on patient data.

    4. Adjudication Method for the Test Set

    This information is not provided and is not applicable to the type of performance testing documented here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No MRMC comparative effectiveness study was done or reported in this 510(k) summary. This document focuses on demonstrating substantial equivalence to a predicate device based on technical characteristics and performance testing, not on comparative clinical effectiveness with or without AI assistance. The device is a "sequential compression device," not an AI diagnostic or assistive tool in the MRMC sense.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device itself is a standalone physical compression system with software components. The software is part of the control system and user interface. The performance testing "without human-in-the-loop" would essentially be the bench testing and software verification, which were indeed performed. However, this is not a diagnostic algorithm in the typical sense of standalone AI performance.

    7. Type of Ground Truth Used

    For the engineering tests, the "ground truth" would be established by:

    • Industry Standards: IEC, FCC, Industry Canada standards for electrical safety and electromagnetic compatibility.
    • Design Specifications: Internal design requirements for tensile strength, pressure delivery, duration tracking, etc.
    • FDA Guidance: "General Principles of Software Validation; Final Guidance for Industry and FDA Staff" and IEC 62304 for software.

    There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for the performance testing described.

    8. The Sample Size for the Training Set

    This information is not applicable as the document does not describe the development of a machine learning or AI model that would require a "training set" of data in the common sense. The software is described as being developed in accordance with validation principles, but not trained on a data set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no described training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181651

    Validate with FDA (Live)

    Date Cleared
    2018-10-12

    (112 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Radial Medical Compression System is indicated for use in:

    • Preventing deep vein thrombosis (DVT)
    • Enhancing blood circulation
    • Diminishing post-operative pain and swelling
    • Reducing wound healing time
    • Treatment and assistance in healing: stasis dermatitis; venous stasis ulcers; arterial and diabetic leg ulcers
    • Treatment of chronic venous insufficiency
    • Reducing edema
    Device Description

    The Radial Medical Compression System (RMCS) is a light weight, mobile, sequential compression device (SCD) placed around the lower leg. The device is designed to provide sequential compression to the calf to augment blood flow. The system includes three components: the Cirvo sequential compression device, an inductive charging cradle and a smart device app (i.e. tablet or smartphone). The RMCS records the amount and duration of compression received during a therapy session as well as compliance with the prescribed therapy. Identified patient data, and compiled, de-identified data will be shared with the treating physician to facilitate clinical decision making.
    The Cirvo uses a mechanical system to sequentially compress the leg utilizing four tension zones tensioned in series. A circumferential sleeve of polyester material is placed around the calf, held in place using hook and loop straps. During the compression cycle, a reduction in the effective circumference of the sleeve exerts a radially inward force on the calf. The device has a compression cycle time of 6 seconds and a physician-determined dwell time between compression cycles. The range of compression cycle pressures of the Cirvo is 40-60 mmHg.

    AI/ML Overview

    The provided text describes the Radial Medical Compression System (RMCS) and its clearance via a 510(k) pathway, demonstrating substantial equivalence to a predicate device. However, the information provided does not include the details typically found in a study proving an AI/device meets acceptance criteria using a test set with ground truth established by experts and adjudication methods, nor does it describe a multi-reader multi-case (MRMC) comparative effectiveness study with human readers assisted by AI.

    The document outlines performance testing for a medical device (a sequential compression system) for therapeutic purposes, not an AI or diagnostic device that would require the establishment of ground truth for a test set, expert consensus, or specific acceptance criteria around diagnostic performance metrics like sensitivity, specificity, or AUC.

    Therefore, I cannot directly answer your prompt's specific requirements regarding acceptance criteria and a study proving an AI device meets those criteria, as the provided text describes a mechanical medical device and its validation for FDA clearance, which focuses on device performance, safety, and substantial equivalence to a predicate.

    However, I can extract the closest analogous information from the provided text regarding the device's performance testing and clinical evaluation.

    Based on the provided text, here's an attempt to extract relevant details, acknowledging that it's for a mechanical device and not an AI/diagnostic system as implied by the prompt's structure:


    The Radial Medical Compression System (RMCS) is a light-weight, mobile, sequential compression device (SCD) designed to provide sequential compression to the calf to augment blood flow. Its clearance was based on demonstrating substantial equivalence to a legally marketed predicate device (Vasculaire Compression System, K122609).

    Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" for diagnostic performance (as would be typical for an AI device) but rather functional and safety performance specifications. The "performance" assessment is primarily comparative and related to its intended physiological effect.

    Table 1: Key Performance Attributes and Comparison (Extracted from "Summary Comparison to Predicate" and "Clinical Performance Testing")

    AttributeAcceptance Criteria (Implied / Compared to Predicate)Reported Device Performance (Radial Medical Compression System)
    Pressure DeliveredWithin specified physiological range and comparable to predicate. The predicate delivers 50-150 mmHg.40 – 60 mmHg
    Therapy Delivery ModeSequential compression of calf from caudal to cranial to augment venous flow. Similar to predicate.Mechanical system, four tension zones, tensioned in series. Delivered as sequential compression from caudal to cranial.
    Clinical EffectAugment venous flow (measured as peak venous velocity); comparable or superior to baseline and predicate.- Average 3.5-fold femoral venous flow augmentation (high setting).- Radial device (low/high setting) resulted in higher average peak venous velocities compared to baseline and IPC device (p<0.05).
    SafetyNo serious safety events or adverse events.No Adverse Events or safety issues reported in the course of the study.
    BiocompatibilityMaterials in patient contact are biocompatible (per ISO 10993-1:2009).Cytotoxicity, irritation, and sensitization testing performed and successfully completed.
    Electrical SafetyCompliance with relevant IEC standards (IEC 60601-1, IEC 60601-1-2, etc.).Successfully demonstrated compliance.
    Software ValidationClassified as "moderate level of concern," prepared in accordance with FDA guidance and IEC 62304.Compliant with FDA guidance "General Principles of Software Validation" and IEC 62304.
    Human FactorsUsability and safety evaluated, no serious safety events due to use/misuse.Formative usability testing performed, device meets requirements. FMEA revealed no serious safety events related to use/misuse.

    Study Details (Clinical Performance Testing)

    The "Clinical Performance Testing" section describes a feasibility study rather than a full-scale clinical trial used for a new AI diagnostic device to establish standalone or human-in-the-loop performance. This study's purpose was to evaluate venous flow augmentation.

    1. Sample Size and Data Provenance:
    * Test Set Sample Size: 10 subjects.
    * Data Provenance: The study was conducted on recruited subjects. The text mentions "an IRB approved study," suggesting prospective data collection, likely within the US, given the FDA submission context. No explicit country of origin is stated beyond the general context of US regulatory approval.

    2. Number of Experts and Qualifications:
    * This type of study (measuring physiological parameters like peak venous velocity) typically involves trained medical professionals performing the measurements (e.g., ultrasound technicians, physicians). However, the document does not specify the number or qualifications of experts used for data collection or analysis within this feasibility study. There's no "ground truth" adjudicated by multiple experts in the sense of image interpretation for AI.

    3. Adjudication Method:
    * Not applicable to this type of study. The measurements (peak venous velocity) are quantitative.

    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
    * No, an MRMC study was not done. This study was a physiological measurement study on a medical device, not an AI diagnostic tool involving multiple readers interpreting cases. Therefore, there's no effect size for human readers improving with AI assistance.

    5. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):
    * Not applicable in the typical sense for an AI device. The "performance" described is the device's direct physiological effect (venous flow augmentation) when worn by a subject, which is an "algorithm only" type of assessment in that it evaluates the device's physical function. However, it's not an AI algorithm generating a diagnostic output.

    6. Type of Ground Truth Used:
    * The "ground truth" in this context is the physiologically measured peak venous velocity (cm/s) at specific anatomical sites (popliteal and femoral veins) using an objective measurement method (implied to be Doppler ultrasound, common for such measurements). This serves as the quantitative outcome data.
    * It also compared to a baseline measurement and a commercially available portable intermittent pneumatic compression (IPC) device.

    7. Sample Size for Training Set:
    * Not applicable. This document describes a medical device, not an AI system that undergoes a separate training phase with a distinct training dataset.

    8. How the Ground Truth for the Training Set Was Established:
    * Not applicable. As above, there's no AI training set or associated ground truth establishment process described.