510(k) Data Aggregation

    K Number
    K232533

    Validate with FDA (Live)

    Date Cleared
    2023-11-16

    (87 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ROSA® Partial Knee System, for use with the ROSA® RECON platform, is indicated as a stereotaxic instrumentation system for Partial Knee replacement (PKA) surgery. It is to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of the knee implant components.

    The robotic arm placement is performed relative to anatomical landmarks as recorded using the system intraoperatively, and based on a three-dimensional representation of the bone structures determined preoperatively using compatible X-ray or MRI based imaging technologies.

    It includes a robotic arm, an optical sensor navigation system and accessories, software system, surgical instruments and accessories.

    The ROSA® Partial Knee System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population that is suitable for the implants compatible with the ROSA® Partial Knee System.

    The ROSA® Partial Knee System is to be used with Persona Partial Knee (PPK) fixed bearing knee replacement system in accordance with its indications and contraindications.

    Device Description

    The ROSA® Partial Knee System for use with the ROSA® RECON Platform is used to assist surgeons in performing Partial Knee Arthroplasty (PKA) with features to assist with the bone resections as well as assessing the state of the soft tissues to facilitate implant positions intraoperatively.

    The system uses a Non-Device Medical Device Data System (MDDS) called the Zimmer Biomet Drive Portal which manages the creation and tracking of the surgical cases. The cases resides on the portal until it is uploaded to the ROSA® RECON Platform before surgeries.

    If the case is image-based, a 3D virtual bone model is generated pre-operatively by the PSI systems (X-PSI Knee System or CAS PSI Knee System) to create a 3D model of the patient's femur/tibia and allows the preparation of a pre-operative ROSA® Total Knee System (TKA) surgical plan. However, the pre-operative surgical plan is not provided in the ROSA® Partial Knee System and is only made available if a switch is performed intra-operatively from ROSA® Partial Knee System to the ROSA® Knee System. Landmarks taken intra-operatively on the patient's bony anatomy are used to create the intraoperative surgical plan.

    An image-less option is also available where landmarks taken intra-operatively on the patient's bony anatomy are used to create the surgical plan.

    Accuracy of resections, knee state evaluation, and soft tissue assessment are the same between image-based and imageless options as they are always based on intra-operative landmarks.

    The intra-operative workflow and surgical concepts implemented in the system remain close to the conventional PKA workflow. As such, at the time of the surgery, the system mainly assists the surgeon in (1) determining reference alignment axes in relation to anatomical landmarks, (2) planning the orthopedic implants location based on these reference alignment axes and orthopedic implant geometry (planning optionally based on using pre-operative imaging), and (3) precisely positioning the cut guide relative to the planned orthopedic implant location by using a robotic arm and (4) using the Zimmer Biomet Persona Partial Knee (PPK) spacer block to perform the femoral distal cut

    AI/ML Overview

    The provided text describes the ROSA® Partial Knee System's 510(k) submission, confirming its substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria or a detailed study proving the device meets said criteria. The document mentions performance data was provided, but the specifics of that data are not included in the excerpt.

    Therefore, the following information cannot be extracted from the provided text:

    • A table of acceptance criteria and the reported device performance
    • Sample sizes used for the test set
    • Data provenance for the test set
    • Number of experts used to establish ground truth for the test set and their qualifications
    • Adjudication method for the test set
    • Whether an MRMC comparative effectiveness study was done, or its effect size
    • Whether a standalone (algorithm only) performance study was done
    • The type of ground truth used
    • The sample size for the training set
    • How the ground truth for the training set was established

    What can be extracted is:

    Summary of Performance Data (Nonclinical and/or Clinical) - General Statements:

    The submission states that the following performance data was provided in support of the substantial equivalence determination (though the details of the results are not given):

    1. Biocompatibility Testing: No additional testing was deemed necessary. Existing testing for the Persona Partial Knee instrumentation and ROSA® Partial Knee instruments remains valid.
    2. Electrical Safety and Electromagnetic Compatibility (EMC): No additional testing was deemed necessary on the ROSA® RECON Platform. Existing EMC testing is still applicable.
    3. Device Performance Testing:
      • Physical/Performance Tests: To ensure the performance of implemented features and verify related design inputs.
      • Engineering Analysis: To ensure the performance of implemented features and verify related design inputs.
      • Validation Lab: Performed to validate that using the ROSA® Partial Knee System is equivalently safe and effective, and that the performances of the system are acceptable under full simulated use on cadaveric specimens.
    4. Software Verification and Validation Testing:
      • Tests were conducted to satisfy requirements of the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices and IEC 62304 (Medical Device Software- Life Cycle Process).
      • The software was considered a "major" level of concern.
      • The documentation provided suggests that a failure of any device software function could present a hazardous situation with a probable risk of death or serious injury.
      • The testing demonstrates that the ROSA® Partial Knee System does not raise any new issues of safety and effectiveness as compared to the predicate device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210121

    Validate with FDA (Live)

    Date Cleared
    2021-04-19

    (90 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ROSA® Partial Knee System, for use with the ROSA® RECON platform, is indicated as a stereotaxic instrumentation system for Partial Knee replacement (PKA) surgery. It is to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of the knee implant components.

    The robotic arm placement is performed relative to anatomical landmarks as recorded using the system intraoperatively, and optionally based on a three-dimensional representation of the bone structures determined preoperatively using compatible X-ray or MRI based imaging technologies.

    It includes a robotic arm, an optical sensor navigation system and accessories, software system, surgical instruments and accessories.

    The ROSA® Partial Knee System is designed for use on a skeletally mature patient population has the same characteristics as the population that is suitable for the implants compatible with the ROSA® Partial Knee System.

    The ROSA® Partial Knee System is to be used with Persona Partial Knee (PPK) fixed bearing knee replacement system in accordance with its indications and contraindications.

    Device Description

    The ROSA® Partial Knee System for use with the ROSA® RECON Platform is used to assist surgeons in performing Partial Knee Arthroplasty (PKA) on the medial compartment with features to assist with the bone resections as well as assessing the state of the soft tissues to facilitate implant positioning intraoperatively.

    The ROSA® Partial Knee System uses a Non-Device Medical Device Data System (MDDS) called the Zimmer Biomet Drive Portal which manages the creation and tracking of the surgical cases. The cases reside on the portal until it is uploaded to the ROSA® RECON Platform before surgeries.

    If the case is image-based, a 3D virtual bone model is generated pre-operatively by the PSI systems (X-PSI Knee System or CAS PSI Knee System) to create a 3D model of the patient's femur/tibia and allows the preparation of a pre-operative ROSA Total Knee System (TKA) surgical plan. However, the pre-operative surgical plan is not provided in the ROSA® Partial Knee System and is only made available if a switch is performed intra-operatively from ROSA® Partial Knee System to the ROSA® Knee System. Landmarks taken intra-operatively on the patient's bony anatomy are used to create the intraoperative surgical plan.

    An image-free option is also available where landmarks taken intra-operatively on the patient's bony anatomy are used to create the surgical plan.

    Accuracy of resections, knee state evaluation, and soft tissue assessment are the same between image-based and image-free options as they are always based on intraoperative landmarks.

    The intra-operative workflow and surgical concepts implemented in the system remain close to the conventional PKA workflow. As such, at the time of the surgery, the system mainly assists the surgeon in (1) determining reference alignment axes in relation to anatomical landmarks, (2) planning the orthopedic implants location based on these reference alignment axes and orthopedic implant geometry (planning optionally based on using pre-operative imaging), and (3) precisely positioning the cut guide relative to the planned orthopedic implant location by using a robotic arm.

    AI/ML Overview

    The provided text is a 510(k) summary for the ROSA® Partial Knee System. While it describes the device, its intended use, and general performance testing (biocompatibility, electrical safety, software verification and validation), it does not contain the specific details required to answer all parts of your request regarding acceptance criteria and a study proving the device meets those criteria.

    The document states:

    • "Verification and Validation Testing for ROSA® Partial was conducted with the following aspects: Physical/Performance Tests- to ensure the performance of the implemented features and verify related design inputs"
    • "Validation Lab- performed to validate that using ROSA® Partial Knee System is safe and effective and that the performances of the system are acceptable under full simulated use on cadaveric specimens"

    However, it does not provide:

    • A table of specific acceptance criteria (e.g., accuracy thresholds for bone cuts, alignment targets) and the quantitative reported device performance against these criteria.
    • The sample size for a test set used for performance validation.
    • Data provenance (country of origin, retrospective/prospective).
    • Details on expert involvement for ground truth (number, qualifications, adjudication method).
    • Whether an MRMC comparative effectiveness study was done involving human readers and AI assistance. Robotic surgical systems often focus on robotic accuracy and surgeon assistance, not typically AI-driven image interpretation in the same way an AI diagnostic tool would.
    • Specific standalone algorithm performance (as it's a robotic system with a human-in-the-loop, not a purely AI diagnostic algorithm).
    • The type of specific ground truth used for quantifying the performance (e.g., CT measurements post-cut, direct measurements, etc.).
    • The sample size for the training set (if any machine learning was explicitly involved in training a model for decision-making, separate from software verification).
    • How ground truth for the training set was established.

    Based on the provided text, I cannot generate the table or provide detailed answers to most of your questions. The document focuses on regulatory compliance and substantial equivalence, not the detailed results of a performance study with specific quantitative acceptance criteria and outcomes.

    Therefore, many parts of your request will be answered with "Not available in the provided text."


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Metric (Hypothetical, as not provided)Acceptance Threshold (Hypothetical, as not provided)Reported Device Performance (Not available in text)
    Robotic Arm AccuracyPositional Accuracye.g., ± 0.5 mmNot available in provided text
    Angular Accuracye.g., ± 1.0 degreeNot available in provided text
    Bone Resection PrecisionCut Depth Accuracye.g., ± 0.5 mmNot available in provided text
    Cut Plane Orientatione.g., ± 1.0 degreeNot available in provided text
    Navigation SystemRegistration Accuracye.g., < 1.0 mm RMSNot available in provided text
    Overall System EfficacySuccessful Implant Placement RateNot explicitly stated (implied safe & effective)Not available in provided text

    The available text broadly states that "performances of the system are acceptable under full simulated use on cadaveric specimens" but does not provide specific quantitative acceptance criteria or reported values.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not available in the provided text. The text mentions "cadaveric specimens" for validation.
    • Data Provenance: Not available in the provided text. It does not specify the country of origin of the data or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not available in the provided text.
    • Qualifications of Experts: Not available in the provided text. The evaluation method mentioned is "Validation Lab- performed to validate that using ROSA® Partial Knee System is safe and effective and that the performances of the system are acceptable under full simulated use on cadaveric specimens." This implies surgical or engineering expertise, but details are absent.

    4. Adjudication method for the test set

    • Adjudication Method: Not available in the provided text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: Not available in the provided text. Robotic surgical systems like ROSA primarily assist surgeons in performing physical tasks (bone cuts, implant placement) with precision based on pre-operative plans and intra-operative navigation, rather than assisting human "readers" in interpreting images for diagnosis like an AI diagnostic algorithm would. Therefore, a typical MRMC comparative effectiveness study as conceived for diagnostic AI is unlikely to be applicable or reported here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not explicitly detailed or quantified in the provided text. The ROSA® Partial Knee System is described as assisting the surgeon and involving a robotic arm, an optical sensor navigation system, and software. It is inherently a "human-in-the-loop" system. While the individual components (e.g., software calculations) would have undergone internal standalone verification, the ultimate performance metric is the system's ability to assist in surgery, which is human-dependent.

    7. The type of ground truth used

    • Type of Ground Truth: The text indicates that "Validation Lab- performed to validate that using ROSA® Partial Knee System is safe and effective and that the performances of the system are acceptable under full simulated use on cadaveric specimens." This suggests that the ground truth would involve direct measurements on the cadaveric bones after robotic intervention (e.g., measuring cut dimensions, alignment, and implant position) to assess accuracy and compare them to the pre-operative plan or ideal surgical outcomes. Specific details (e.g., using CT scans post-surgery on cadavers, direct physical measurement) are not provided.

    8. The sample size for the training set

    • Sample Size for Training Set: Not available in the provided text. This document is a 510(k) summary for a robotic surgical system, not a deep learning model for image analysis, so a "training set" in the common AI sense might not be explicitly relevant or defined in the same way. The software is verified through "Software Verification and Validation Testing" and is considered "major" level of concern.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set Establishment: Not applicable/Not available in the provided text for the reasons stated above (see point 8).