Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132138

    Validate with FDA (Live)

    Date Cleared
    2014-04-11

    (274 days)

    Product Code
    Regulation Number
    874.1820
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rhythmlink Disposable Concentric Stimulating Probe is used to perform localized stimulation of neural tissue and to locate, identify and monitor cranial motor nerves, peripheral nerve and spinal nerve roots during surgery.

    The Rhythmlink Disposable Concentric Stimulating Probe is a single patient use device.

    Device Description

    target nerves and to locally stimulate them in order to provide a measurable response. The stimulus will have a very small current spread to reduce the innervation of the surrounding nerves.

    The concentric design has two main parts, an inner stainless steel wire and an outer cannula also insulated from reading un-intended signals.

    The Inner wire is isolated from the outer stainless steel cannula using a biocompatible heat shrink tubing. This Inner Wire acts as the stimulator and is surrounded by the outer cannula. The Outer Cannula is isolated with a biocompatible heat shrink which isolates the outer cannula allowing contact in a localized area of the intended nerves of interest.

    The inner wire, the cathode, is stimulated using EMG/EP electroneurodiagnostic equipment cleared for the stimulation of nerve tissue and recording muscle activity during surgical procedures (not part of this 510(k) submission) The outer cannula acts as the anode or reference.

    The Inner wire and outer cannula are isolated from each other using a dielectric heat shrink and are connected to a color coded pair of leadwires which are terminated by two DIN 42 802 touch proof connectors. The concentric stimulators are terminated on the distal end inside of a plastic handle and are independently connected to the leadwires.

    AI/ML Overview

    This document describes a 510(k) submission for the Rhythmlink Disposable Concentric Stimulation Probe. It outlines the safety and effectiveness testing to demonstrate substantial equivalence to a predicate device, not the performance of an AI model. Therefore, many of the requested fields related to AI model acceptance criteria and study design are not applicable.

    Here's an analysis based on the provided text, focusing on the available information:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    DimensionalProbe diameter and lengths within specified ranges (comparable to predicate)Conducted (specific ranges not explicitly stated, but implied to be acceptable based on comparison to predicate in "Technological Characteristics" table)
    Performance - Continuity0.5 OhmsAchieved 0.5 Ohms
    Performance - Hi PotNo insulator breakdownPassed (testing of insulator breakdown conducted and analyzed)
    Performance - Stimulation DeliveryEffective stimulation delivery (comparable to predicate)Conducted and analyzed (implied to be effective for intended use)
    Performance - Pull-off StrengthAdequate strength (specific value not stated, but implied to be acceptable)Conducted and analyzed
    Performance - Leadwire StrengthAdequate strength (specific value not stated, but implied to be acceptable)Conducted and analyzed
    SterilizationCurrent, validated EO Sterilization Cycle 93007 met; acceptable residual EtO and ECH levelsSterilized under Cycle 93007; residuals at "lowest possible limits"
    Biocompatibility - ISO BET Get Clot TestingNo interference with lysate reaction; no inhibition or enhancement; no clotting at neat concentration; geometric mean endpoint concentration < 0.06 EU/mL and < 0.9 EU/Device of bacterial endotoxin.The system did not interfere with the lysate reaction, and no inhibition or enhancement was present. The test articles did not clot at the neat concentration. The geometric mean endpoint concentration of each test article was <0.06 EU/mL and each contained <0.9 EU/Device of bacterial endotoxin. Results acceptable.
    Biocompatibility - USP Inhibition and Enhancement Testing, Gel Clot MethodNo inhibition or enhancement of Bacterial Endotoxin Test according to USP guidelines.The test Concentric Stimulation Probe does not inhibit or enhance the Bacterial Endotoxin Test according to the USP guidelines.
    Biocompatibility - ISO MEM Elution TestNo biological reactivity (Grade 0) of cells exposed to test article extract. Positive and negative controls confirm suitability.There was no biological reactivity (Grade 0) of the cells exposed to the test article extract. The response obtained from the positive and negative control article extracts confirmed the suitability of the test system.
    Biocompatibility - ISO Intracutaneous Reactivity TestTest article sites do not show significantly greater biological reaction than control article sites.The test article sites did not show a significantly greater biological reaction than the sites injected with the control article.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each non-clinical test conducted. It states "Non-clinical bench testing was comprised of dimensional measurements and performance tests." and "biocompatibility tests on the device". The data provenance is internal to the manufacturer (Rhythmlink International, LLC) and likely collected in a laboratory setting as part of product development and regulatory submission. This is not patient data; it's device performance data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a submission for a medical device (stimulating probe), not an AI algorithm. Ground truth, in the context of an AI model, refers to the verified correct output for a given input. For this device, "ground truth" is established through engineering specifications, validated test methods, and compliance with industry standards (e.g., ISO, USP). There are no "experts" establishing ground truth in the sense of clinical annotations for AI.

    4. Adjudication method for the test set

    Not applicable. This is a submission for a medical device (stimulating probe), not an AI algorithm. Adjudication methods are relevant for resolving discrepancies in expert opinions used to create labels for AI training/test sets.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a submission for a medical device (stimulating probe), not an AI algorithm. An MRMC study pertains to the evaluation of diagnostic performance involving multiple human readers and an AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a submission for a medical device (stimulating probe), not an AI algorithm.

    7. The type of ground truth used

    For the non-clinical tests:

    • Dimensional: Engineering specifications and comparisons to the predicate device.
    • Performance Tests (Continuity, Hi Pot, Stimulation Delivery, Pull-off, Leadwire Strength): Engineering specifications and functional requirements based on the intended use and predicate device's performance.
    • Sterilization: Regulatory standards for ethylene oxide sterilization and validated processes.
    • Biocompatibility: Established international (ISO) and national (USP) standards for biological evaluation of medical devices, with pass/fail criteria defined by these standards.

    8. The sample size for the training set

    Not applicable. This is a submission for a medical device (stimulating probe). There is no "training set" in the context of artificial intelligence. The device design and manufacturing processes are validated through engineering and biocompatibility tests.

    9. How the ground truth for the training set was established

    Not applicable. As there is no AI training set, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1