Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200931

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-24

    (231 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RF Surgical Generator FW-120A is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting and coagulating.

    Device Description

    RF Surgical Generator FW-120A, is a compact source of high power RF energy, enhanced capability of radiosurgery generator, which can be employed for a variety of radiosurgery procedures, such as cut and coagulate living human tissue.

    The device which is consisted of a generator, an IEC Neutral Plate, and a Two Pedal Footswitch, provides 5 modes and corresponding output power ranges that is user/operator selectable based on the surgical procedure undertaken. The action is achieved by front panel selection of waveforms and power level. All selection is affected through push buttons and indicators which give the operator feedback of status. Power level for each mode is indicated by front panel digital displays

    The device is provided with safety features of automatically system initialize self-check after power on and continuously monitored during the device running. The device will provide error information and/ or alerts when detect following faults, such as a) Accessary fail connection, b) Overheating of Main Unit, c) Neutral plates Contact resistance exceed the limit, d) System initialize error. The error monitoring is interlocked with the controls to prevent operation when fault condition happened in the unit.

    The generator is configured for Monopolar and Biopolar output and the final output power could be controlled by pedal foot switch and/or hand switches. Recommended Monopolar Handpieces 510(k) No.: K092634. Recommended IEC Neutral Plate 510(k) No.: K102372.

    AI/ML Overview

    This document is a 510(k) premarket notification for an RF Surgical Generator (Model: FW-120A) and focuses on demonstrating substantial equivalence to a predicate device, rather than providing details on a study to meet specific acceptance criteria for a new device.

    Therefore, many of the requested categories about acceptance criteria, detailed study design, ground truth, and expert involvement are not applicable or not provided in this type of submission. This document primarily reports on verification testing against established standards and a comparison to a predicate device.

    Here's a breakdown based on the information available in the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not outline specific numerical "acceptance criteria" in the way one might see for a diagnostic device (e.g., sensitivity/specificity targets). Instead, it relies on demonstrating compliance with recognized safety and performance standards and showing equivalence to a predicate device.

    CategoryAcceptance Criteria (Implied)Reported Device Performance
    Electrical SafetyCompliance with ANSI AAMI ES60601-1Device evaluated and found in compliance with ANSI AAMI ES60601-1:2005/(R)2012 and associated amendments.
    EMC PerformanceCompliance with IEC 60601-1-2Device evaluated and found in compliance with IEC 60601-1-2 (2014).
    HF Surgical Equipment Specific SafetyCompliance with IEC 60601-2-2Device evaluated and found in compliance with IEC 60601-2-2 (Edition 6.0 2017-03).
    Thermal EffectsPerformance comparable to predicate device for cutting/coagulating tissue with acceptable thermal damage.Thermal damage (length, width, depth of thermal zone) measured in porcine muscle, liver, and kidney in all applicable modes. Performance supports substantial equivalence.
    System Performance/Waveform OutputPerformance comparable to predicate device.System performance and waveform outputs tested. Results support substantial equivalence.
    Software VerificationCompliance with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" for moderate concern level.Software verifications performed to ensure appropriate operation as a moderated concern level software.
    Monopolar Neutral Electrode Monitor ResistanceMAX 1000ΩMAX 1000Ω
    Rated Duty Cycle10s on, 30s off10s on, 30s off
    Output PowerComparable to predicate device (specific values provided in table)Cutting: 120W@500Ω; Blend cutting and coagulation: 80W@500Ω; Hemostasis: 60W@500Ω; Fulguration: 40W@500Ω; Bipolar: 120W@200Ω

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified in terms of number of human subjects or distinct test cases. The testing involved "lab bench testing" and "porcine muscle, liver, and kidney" tissue. The number of samples/tests for each tissue type is not detailed.
    • Data Provenance: The testing was "lab bench testing" using "porcine" tissue. The country of origin for the data is not explicitly stated beyond the applicant being based in China. The data is from prospective testing conducted specifically for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not Applicable / Not Provided: This type of submission (510(k)) for an electrosurgical generator typically relies on objective physical measurements and engineering evaluations against standards, rather than expert-established ground truth like in diagnostic imaging. There is no mention of experts establishing a ground truth for a test set in this context.

    4. Adjudication Method for the Test Set

    • Not Applicable / Not Provided: As there's no mention of expert ground truth establishment, there is no adjudication method described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No: An MRMC comparative effectiveness study is not mentioned as this device is an electrosurgical generator, not an interpretative diagnostic imaging device that involves human readers.

    6. Standalone (Algorithm Only) Performance

    • Yes, implicitly: The "performance data" section (Section 6) describes "System Performance and waveform outputs test," "Thermal effects testing," and "software verifications" which represent the standalone performance of the device itself and its integrated software. These tests involve direct measurement of the device's output and effects, independently of human interpretation of a diagnostic output.

    7. Type of Ground Truth Used

    • Objective Measurements and Physical Science Principles: The "ground truth" for this device's performance is established through objective measurements of electrical output parameters (waveform, power), physical effects on tissue (thermal damage size), and compliance with internationally recognized safety and performance standards (e.g., IEC 60601 series). For software, it's verification against design specifications and guidance documents.

    8. Sample Size for the Training Set

    • Not Applicable / Not Provided: This device, an RF Surgical Generator, does not utilize machine learning or AI models that require a "training set" in the conventional sense. Its functionality is based on established engineering principles and not on learning from a large dataset.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable / Not Provided: As there is no training set for a machine learning model, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1