510(k) Data Aggregation

    K Number
    K160216

    Validate with FDA (Live)

    Date Cleared
    2016-04-22

    (84 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Revolution™ Spinal Fixation System is intended for pedicle screw fixation of the noncervical, posterior spine in order to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) spondylolisthesis, (3) trauma (i.e., fracture or dislocation), (4) spinal stenosis, (5) deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), (6) tumor, (7) pseudoarthrosis, (8) failed previous fusion.

    Device Description

    The Intelligent Implant Systems' Revolution™ Spinal Fixation System consists of monoaxial bone screws of various lengths and diameters and a series of connector assemblies for connecting the bone screws. This 510(k) submission adds four types of components to the system: 1) Transition Connector, 2) Transition Bone Screws, 3) Two-Post Connector, and 4) Two-Post Bone Screws. These components, when utilized with the existing standard Revolution™ connectors and standard bone screws, allow stabilization of multiple levels of the spine. Like the existing bone screws, the Transition and Two-Post bone screws are available in diameters of 4.5 mm - 7.5 mm, and in lengths from 25 mm to 55 mm. All screw diameter sizes except for the 4.5mm are cannulated for use with a 1.4 mm k-wire. The Transition Bone Screws and Two-Post Bone Screws have a standard thread for interfacing bone and a partially threaded post designed to engage the Revolution™ Connectors. The Transition Connector and Two-Post Connector effectively create two threaded posts from a single bone screw. These threaded posts are utilized to connect to the Revolution™ Connectors. The standard Revolution™ Connectors are provided in various lengths and are used to connect the pedicle screws and create a rigid structure. To allow for variation in screw placement, each end of the connector allows for angular compensation. One end allows for full polyaxial angulation, similar to a typical polyaxial screw. The other end has a pivoting-slide that allows for angulation in only one plane along the long axis of the Connector while also sliding within an opening in the Connector. As the distance between two screws is always variable, the slide allows the Connector to compensate for this difference, minimizing inventory and eliminating the need to cut spinal rods. In addition, the Connectors have a threaded locking nut built into each end and each nut fits over the top of a bone screw post. These locking nuts are turned clockwise to tighten. A calibrated torque wrench provides the correct setting for locking the bone screw post with sufficient force. The system can be unlocked, if necessary, by turning the locking nuts counterclockwise. All implant components of the Revolution™ Spinal Fixation System. including the new multi-level components, are manufactured from Ti-6Al-4V ELI alloy, conforming to ASTM F136.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Revolution™ Spinal Fixation System, which is a medical device and not an AI/ML regulated device. Thus, the provided information does not contain the requested details on acceptance criteria and a study that proves a device meets them in the context of AI/ML performance.

    Therefore, I cannot provide the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142939

    Validate with FDA (Live)

    Date Cleared
    2015-01-13

    (96 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Revolution™ Spinal Fixation System is intended for pedicle screw fixation of the noncervical posterior spine in order to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) spondylolisthesis, (3) trauma (i.e., fracture or dislocation), (4) spinal stenosis, (5) deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), (6) tumor, (7) pseudoarthrosis, (8) failed previous fusion.

    Device Description

    The Intelligent Implant Systems' Revolution™ Spinal Fixation System consists of monoaxial bone screws of various lengths and diameters and a series of connector assemblies for connecting the bone screws. The Revolution™ Bone Screws have a standard thread for interfacing bone and a partially threaded post designed to engage the Revolution™ Connectors. Screws are available in 4.5mm major diameter and lengths from 25mm to 50mm and 5.5mm, 6.5mm, and 7.5mm major diameters and lengths from 25mm to 55mm. All screw diameter sizes except for the 4.5mm are cannulated for use with a 1.4 mm k-wire. To allow the surgeon to determine the height of the connector relative to the bone, screws with extended non-threaded collars are provided. These extended collars effectively provide built in spacers that prevent the bone screw from being driven too deep and control the overall height of the completed assembly. The Revolution™ Connectors are provided in various lengths and are used to connect the pedicle screws and create a rigid structure. To allow for variation in screw placement, each end of the connector allows for angular compensation. One end allows for full polyaxial angulation, similar to a typical polyaxial screw. The other end has a pivoting-slide that allows for angulation in only one plane along the long axis of the Connector while also sliding within an opening in the Connector. As the distance between two screws is always variable, the slide allows the Connector to compensate for this difference, minimizing inventory and eliminating the need to cut spinal rods. In addition, the Connectors have a threaded locking nut built into each end and each nut fits over the top of a bone screw post. These locking nuts are turned clockwise to tighten. A calibrated torque wrench provides the correct setting for locking the nuts on the bone screw post with sufficient force. The system can be unlocked, if necessary, by turning the locking nuts counterclockwise. All implant components of the Revolution™ Spinal Fixation System are manufactured from Ti-6Al-4V ELI alloy, conforming to ASTM F136.

    AI/ML Overview

    This FDA 510(k) summary describes a spinal fixation system, not a device with AI components. Therefore, the questions related to AI performance, ground truth, experts, and training/test sets are not applicable.

    The document focuses on the Revolution™ Spinal Fixation System, a mechanical implant. For such a device, acceptance criteria and testing revolve around its physical properties, material compatibility, and structural integrity under simulated physiological conditions.

    Here's an analysis of the provided information, addressing the applicable questions and indicating when a question is not relevant to this type of device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the testing was conducted "according to methods defined in ASTM F 1798-13, 'Standard Test Method for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants', and ASTM F 1717-14, 'Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model'."

    Since specific numerical acceptance criteria (e.g., minimum strength in Newtons) are not explicitly stated in this summary, the table below will reflect the type of tests performed and the general conclusion given in the document. The implicit acceptance criterion is that the device meets the requirements of these ASTM standards and performs comparably to predicate devices.

    Acceptance Criterion (Implicit based on ASTM standards)Reported Device Performance
    ASTM F1798-13 Tests:(Implied to meet standard, comparable to predicate)
    Static Flexion-Extension StrengthComparable to published values on plate and rod systems
    Axial Gripping StrengthComparable to published values on plate and rod systems
    Static A-P Pullout StrengthComparable to published values on plate and rod systems
    ASTM F1717-14 Tests:(Implied to meet standard, comparable to predicate)
    Static Compression BendingComparable to published values on plate and rod systems
    Static Torsional BendingComparable to published values on plate and rod systems
    Dynamic Compression BendingComparable to published values on plate and rod systems
    Material Compatibility:Complies with ASTM F-136.
    Material: Ti-6Al-4V ELI

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "sample size used for the test set" in terms of patient data, as this is a mechanical device. Instead, it refers to mechanical testing. The sample size for mechanical testing (i.e., the number of devices or components tested) is not explicitly stated in this summary. The data provenance is mechanical testing conducted in a laboratory setting to validate the device's physical properties.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is not an AI/diagnostic system. Ground truth in this context refers to the physical properties measured against established engineering standards (ASTM).

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods are typically for clinical studies or AI algorithm outputs where multiple human reviewers assess data. Here, the "truth" is determined by standardized mechanical testing procedures.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This is a mechanical spinal fixation system, not an AI or diagnostic device that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device does not involve an algorithm.

    7. The Type of Ground Truth Used

    For this mechanical device, the "ground truth" is defined by established engineering and material standards, specifically:

    • ASTM F 1798-13: "Standard Test Method for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants"
    • ASTM F 1717-14: "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model"
    • ASTM F-136: Standard for the Ti-6Al-4V ELI material.

    The device's performance is compared against the requirements and typical performance values for devices conforming to these standards.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This device does not have a "training set" in the context of AI.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032219

    Validate with FDA (Live)

    Date Cleared
    2003-12-19

    (151 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spinal Innovations REVOLUTION ™ Spinal Fixation System, when used as a pedicle screw fixation system, is indicated for use in patients: a) having severe spondylolisthesis (Grade 3 and 4) at the LS-S1 joint; b) who are receiving fusion using autogenous graft only; c) who are having the device fixed or attached to the lumbar or sacral spine; and d) who are having the device removed after the development of a solid fusion mass. When used for this indication, the fusion mass may not go above the L5-S1 joint, the levels of pedicle screw fixation may span from L3 to the sacrum.

    The Spinal Innovations REVOLUTION ™ Spinal Fixation System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

    When used as a hook and sacral screw system (other than pedicle screw fixation system for high grade spondylolisthesis), the Spinal Innovations REVOLUTION ™ Spinal Fixation System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with the degencration of the disc confirmed by a history and radiographic studies), idiopathic scoliosis, spondylolisthesis, kyphotic deformity of the spine, loss of stability due to tumors, spinal stenosis, vertebral fracture or dislocation, pseudoarthrosis, and previous failed spinal surgery/fusion. When used for this indication, screws of the Spinal Innovations REVOLUTION ™ Spinal Fixation System are intended for sacral/iliac attachment only. Hooks and transverse connectors of the system are intended for posterior thoracic and/or lumbar use only. The levels of use for hook and sacral screw fixation of this system are TI to the sacrum.

    Device Description

    The Spinal Innovations REVOLUTION ™ Spinal Fixation System is a temporary implant system used to correct spinal deformity and facilitate the biological process of spinal fusion. This system is intended for posterior use in the thoracic, lumbar and sacral areas of the spine. Implants of this system consist of hooks and/or screws connected to rods and are intended to be removed after solid fusion has occurred. This system includes fixed and polyaxial screws of varying diameters and lengths, and hooks in varying designs and lengths, curved and straight rods and transverse connectors.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Spinal Innovations REVOLUTION™ Spinal Fixation System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance
    Biomechanical PerformanceInterconnection testing of individual system componentsPer ASTM F 1798-97"The testing results of the various system components show that the data compares directly to predicate testing and meets or exceeds only predicate device data standards."
    Biomechanical PerformanceStatic flexion/extension testingNot explicitly stated, implied by comparison to predicate devices"The testing results of the various system components show that the data compares directly to predicate testing and meets or exceeds only predicate device data standards."
    Biomechanical PerformanceFatigue flexion/extension testingNot explicitly stated, implied by comparison to predicate devices"The testing results of the various system components show that the data compares directly to predicate testing and meets or exceeds only predicate device data standards."
    Overall EquivalenceMaterial, Design, and FunctionSubstantially equivalent to predicate devices"The Spinal Innovations REVOLUTION™ Spinal Fixation System is substantially equivalent in material, design and function to the predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document refers to "various system components" being tested, implying a range of components were subjected to the biomechanical tests.
    • Data Provenance: Not explicitly stated. However, given that it's a 510(k) submission for a spinal fixation system, the testing would be conducted in a laboratory setting, likely in the country of origin of the manufacturer (Tennessee, USA). The study is prospective in the sense that the device was manufactured and then tested to demonstrate its performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This is not applicable to this type of study. The ground truth for this device's acceptance is established through objective, quantitative biomechanical testing against established ASTM standards and comparison to predicate devices, not through expert consensus on qualitative data.

    4. Adjudication Method for the Test Set

    • Not applicable. See point 3. Biomechanical testing results are objective measurements and do not require adjudication in the same way clinical or imaging studies might.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. This document describes a biomechanical performance study of a physical medical device (spinal fixation system), not an AI-powered diagnostic or assistive tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This device is a physical implant, not a software algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • The ground truth used for this device's acceptance is based on objective biomechanical measurements and established engineering standards (ASTM F 1798-97). The device's performance is compared directly against the performance of legally marketed predicate devices, implying that their established performance serves as the "ground truth" for what is considered safe and effective in this context.

    8. The Sample Size for the Training Set

    • Not applicable. This is a biomechanical study, not a machine learning study. There is no "training set" in the context of device performance testing through physical measurements.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. See point 8.