Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K243139

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-01-22

    (114 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reverso Pro System is a non-invasive device intended to be used by aesthetic-related physicians or dermatologists. The Reverso Pro System utilizing the Reverso Applicator is a non-invasive device intended for use in dermatological procedures requiring ablation and resurfacing of the skin. The Reverso Pro System utilizing the Quadro RF Applicator is indicated to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The massage device is intended to provide a temporary reduction in the appearance of cellulite.

    Device Description

    The Reverso Pro System, positioned on a rolling cart, is a computerized system comprised of a console (main unit) and two Applicators: The Reverso Applicator with detachable single-use tips and the Quadro RF Applicator. The system using the Reverso Applicator delivers RF energy in a fractional manner to the skin of a treatment area for ablation and resurfacing of the skin, or using the Quadro RF Applicator in a continuous wave manner to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The Reverso Pro System provides individual adjustments of energy settings and treatment time, and, in the Reverso mode, it also enables the adjustment of tip pattern and pulse profile, to achieve appropriate efficiency and safety for each eligible patient. The system console includes a power supply unit, controller card and LCD screen with touch panel. The Reverso Pro System Applicators are designed for use in dermatological procedures, they are hand-held and ergonomically designed for the treatment requirements. The applicators are connected to the system via a separate cable, each. When not in use, the applicators can be placed within the applicator holder, where the Reverso Applicator is positioned on the system console and the Quadro RF Applicator is positioned on the rolling cart. The Reverso Applicator is compatible with five types of fractional RF tip heads: 44, 80, 88, 160 pin tip head and 176 pin tip head. The Quadro RF Applicator is emitting continues RF energy through a set of two pairs of stainless steel, RF electrodes. The applicator is hand-held and ergonomically designed to induce heat of the skin and beneath tissues for topical heating sensation.

    AI/ML Overview

    This document is a 510(k) summary for the Indiba S.A.U. Reverso Pro System (K243139). It details the device's technical specifications, indications for use, and a comparison to predicate devices to establish substantial equivalence.

    Based on the provided text, the device in question is an electrosurgical cutting and coagulation device. The information provided outlines the non-clinical (bench) performance data for the device, and a qualitative assessment of its similarities to predicate devices to establish substantial equivalence. However, there is no detailed clinical study or acceptance criteria table in the format requested, nor is there information about AI/ML algorithm performance.

    The document states:

    • "Clinical Performance Data: Not Applicable" (Page 8)
    • The "Non-Clinical (Bench) Performance Data" section primarily discusses the device's RF performance specifications and thermal effects. It does not describe a study involving human or expert evaluation for determining accuracy or clinical efficacy in the context of an AI/ML device.
    • The entire submission focuses on establishing substantial equivalence to predicate devices based on technological characteristics and safety standards, rather than proving the device meets specific performance criteria through a rigorous clinical trial or AI/ML validation study.

    Therefore, I cannot provide a detailed answer to your request regarding acceptance criteria and a study proving the device meets them, specifically for an AI/ML algorithm with human-in-the-loop performance, ground truth establishment, or sample sizes for test/training sets, because this submission is for a medical device that does not appear to involve an AI/ML component or a clinical performance study in the way typically required for AI/ML-driven devices.

    The information provided relates to the physical device's electrical and mechanical safety, electromagnetic compatibility, and thermal performance.

    Based on the provided text, the device is a radiofrequency aesthetic device, not an AI/ML-driven diagnostic or therapeutic device. As such, the types of studies and acceptance criteria you've asked about (e.g., sample size for test set, number of experts for ground truth, MRMC studies, standalone algorithm performance) are not applicable to the information presented in this 510(k) summary. The "study" mentioned for performance is a skin temperature test on 3 human volunteers for the Quadro RF Applicator, which is a very different type of study than an AI/ML validation.

    Therefore, providing a specific table and detailed answers to your questions on AI/ML performance would be misleading as the document does not contain this information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212107

    Validate with FDA (Live)

    Device Name
    Reverso
    Date Cleared
    2021-11-10

    (127 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reverso is a non-invasive device intended for use in dermatological procedures requiring ablation and resurfacing of the skin.

    Device Description

    The Reverso device is a tabletop computerized system comprised a console (main unit) and Applicator with detachable single-use tips. The system delivers RF energy in a fractional manner to the skin of a treatment area for ablation and resurfacing of the skin. The Reverso device provides individual adjustments of energy settings, tip pattern, pulse profile to achieve appropriate efficiency and safety for each eligible patient. The device console includes a power supply unit, controller card and LCD screen with touch panel. The Reverso device Applicator is designed for use in dermatological procedures requiring ablation and resurfacing of the skin. The applicator is hand- held and ergonomically designed for the treatment requiring ablation and resurfacing of the skin. The applicator is connected to the system via a cable. When not in use, the applicator can be placed within the applicator holder positioned on the system console. The Reverso Applicator is compatible with two types of fractional RF tip heads; 80 pin tip head and 160 pin tip head.

    AI/ML Overview

    The provided text is a 510(k) summary for the Reverso device, which is an electrosurgical device for dermatological ablation and resurfacing. It aims to prove substantial equivalence to a predicate device, the Venus Viva MD.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided document:

    Key Takeaway: This document is not about an AI/ML device where acceptance criteria would involve performance metrics like sensitivity, specificity, or AUC against a ground truth established by experts. Instead, it's a traditional medical device 510(k) focused on demonstrating substantial equivalence to an existing cleared device through engineering and performance testing, primarily with bench and ex-vivo studies. Therefore, many of the typical AI/ML-specific questions (number of experts, adjudication methods, MRMC studies, standalone performance metrics) are not applicable to this type of submission.


    Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" in this context are not presented as a table of numerical thresholds for AI performance, but rather as meeting the design requirements and demonstrating similar performance characteristics to the predicate device. The primary goal is to show the Reverso device is "as safe and effective as the predicate device, and that no new safety or effectiveness issues are raised."

    The document focuses on comparing the Reverso with the predicate device (Venus Viva MD) across various technical specifications and performance aspects.

    Here's a table summarizing the "acceptance criteria" (implied equivalence to predicate) and the reported performance:

    Criterion/CharacteristicImplied Acceptance Criteria (Equivalence to Predicate)Reported Device Performance (Reverso)
    Operational PrincipleSame underlying technology of fractional RFFractional RF: Use of RF energy delivered through a matrix of multiple pin electrodes allocated on the applicator tip. Treatment is based on fractional sub-necrotic heating of papillary and reticular dermis triggering slow collagen remodeling.
    Intended UseNon-invasive dermatological procedures requiring ablation and resurfacing of the skin"The Reverso is a non-invasive device intended for use in dermatological procedures requiring ablation and resurfacing of the skin." (Identical to predicate's relevant indication)
    RF Performance SpecsIdentical/Equivalent RF output specifications (RF Frequency, Energy per pin)RF Frequency: 0.46 MHz. Energy per pin: 62 mJ/pin (160 pin tip), 124 mJ/pin (80 pin tip). "The RF performance specifications... were shown to be identical to those of the predicate device as demonstrated in the bench performance testing."
    Tissue Effect (Ex-vivo)Similar ablation and coagulation characteristics / thermal effect on tissues"The ex-vivo study results demonstrated that the Reverso device is safe for use and effective in achieving the specified indications of dermatological skin ablation and skin resurfacing. Furthermore, the histology results of the subject device tissue samples were compared with the histology results of the predicate device tissue samples. The comparison demonstrated an equivalency in ablation and coagulation characteristics." "similar thermal effect over ex-vivo human tissues following a single treatment performed under similar RF performance specifications and similar operation conditions."
    Safety Features & Standards ComplianceSimilar to predicate, compliance with relevant standardsComplies with ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-1-6. "The safety features and compliance with safety standards in the subject device are similar to the safety features and compliance with safety standards found in the predicate device."
    BiocompatibilityAll materials are biocompatible"All materials are biocompatible" (identical to predicate)
    SterilityProvided non-sterile; single-use tip head to be sterilized by user"Provided non-sterile. The detachable, single use tip head is to be sterilized by the user close to treatment onset, using autoclaving sterilization technique." (identical to predicate)
    Hazards/RisksLow risk, no new safety/effectiveness issues"All of the identified device related hazards are of low risk and are mostly related with transient symptoms of erythema and edema which are resolved 24-72 hours post treatment." "no new safety or effectiveness issues are raised in regard to the subject device functionality."

    Study Details (Applicable to this type of device)

    The studies conducted are primarily for engineering validation and comparison to a predicate device, not for typical AI/ML performance evaluation.

    1. Sample Size and Data Provenance for Test Set:

      • Bench Tests: No specific "sample size" for a data set in the AI sense is mentioned. These are engineering measurements.
      • Ex-Vivo Tissue Testing: "The study was conducted on human abdomen tissues following an abdominoplasty surgery". A "single treatment" was performed. While it doesn't state the number of tissue samples or individual treatments, it implies a small, illustrative set rather than a large statistical sample.
      • Data Provenance: Not specified, but standard for ex-vivo human tissue. Retrospective/Prospective is not applicable here as it's not a clinical study on living subjects. The tissues were obtained after abdominoplasty surgery.
    2. Number of Experts and Qualifications for Ground Truth:

      • Experts: Not explicitly stated as "experts" for ground truth adjudication in the AI sense.
      • Qualifications: The histology section mentions "biopsy sampling of slices trimmed along the pin's penetration path and collection immediately stained by H&E staining to visualize the tissue ablation and coagulation patterns." This implies histopathological analysis, which would typically be performed by trained personnel (e.g., pathologists or histotechnicians). There is no expert consensus or panel adjudication as would be common for subjective AI image analysis.
      • Ground Truth Type: For the ex-vivo study, the ground truth was histology (H&E staining) to observe ablation and coagulation patterns. This is an objective measurement of tissue response.
    3. Adjudication Method for Test Set:

      • Method: Not applicable. There was no expert panel or subjective assessment requiring adjudication. The comparison was based on objective histological findings.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • Performed? No. This device is not an AI/ML diagnostic aid, so an MRMC study comparing human reader performance with and without AI assistance is not relevant or required.
    5. Standalone (Algorithm Only) Performance:

      • Performed? Not applicable. This is a physical electrosurgical device, not a software algorithm providing diagnostic output.
    6. Type of Ground Truth Used:

      • For the ex-vivo study, the ground truth was histology (H&E stained samples) to visually assess and compare tissue ablation and coagulation patterns between the subject and predicate devices.
    7. Sample Size for Training Set & How Ground Truth Was Established (for AI/ML):

      • Not applicable. This document describes a traditional medical device, not an AI/ML algorithm that requires training data and ground truth labeling. The "training" here refers to the device's design, manufacturing, and internal testing, not machine learning model training.
    Ask a Question

    Ask a specific question about this device

    K Number
    K994306

    Validate with FDA (Live)

    Device Name
    REVERSO
    Manufacturer
    Date Cleared
    2000-03-20

    (90 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medisystems Blood Tubing Set Accessories are indicated for use as a part of an extracorporeal blood circuit for hemodialysis. The specific indications for use each accessory are as follows:
    a. The Reverso™ is indicated for use as an access flow reversing valve to reverse the blood flow to and from the arterial and venous access devices during hemodialysis procedures without the need to clamp and disconnect bloodlines.
    b. The Recirculation Connector is indicated for use as a male to male Luer adaptor.
    c. The Parallel Dialyzer Connector is indicated for use to interconnect two hemodialyzers in parallel as prescribed by the physician as part of the extracorporeal blood system.
    d. The Series Dialyzer Connector is indicated for use to interconnect two hemodialyzers in series as prescribed by the physician as part of the extracorporeal blood system.

    Device Description

    A series of blood tubing set accessories to be used as part of the extracorporeal circuit.
    A. Medisystems Blood Tubing Set Accessory, Access Flow Reversing Valve, (Reverso): The technical characteristics of the device consist of a manually operated valve with a gasket to prevent leakage. The device is also equipped with tubing and Luer connectors to allow it to interconnect between the blood tubing set and the blood access devices. The device allows reversal to the arterial and venous blood access without the need to disconnect the blood tubing set from the access devices. The reversal of the arterial and venous accesses is necessary to facilitate certain common diagnostic procedures and can be used to reverse flow in central venous or femoral catheters that have patency problems.
    B. Medisystems Blood Tubing Set Accessory, Recirculation Connector: The device is a singly packed recirculation connector (male-to male Luer adaptor) used to interconnect the arterial and venous bloodlines.
    C. Medisystems Blood Tubing Set Accessory, Parallel Dialyzer Interconnector: The device is used with a standard blood tubing set to provide a means to connect two dialyzers in parallel into the extracorporeal circuit.
    D. Medisystems Blood Tubing Set Accessory, Series Dialyzer Interconnector: The device is used with a standard blood tubing set to provide a means to connect two dialyzers in series into the extracorporeal circuit.

    AI/ML Overview

    The provided text describes a 510(k) submission for "Blood Tubing Set Accessories" by Medisystems Corporation. The document outlines the device's intended use and technological characteristics but does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of device performance metrics like sensitivity, specificity, accuracy, or similar measures typically found in AI/ML device submissions.

    Instead, the submission focuses on establishing substantial equivalence to a legally marketed predicate device (Medisystems' Blood Tubing Sets described in K953823) and details the safety and effectiveness measures in terms of manufacturing processes and quality control.

    Here's a breakdown of the requested information based on the provided text, highlighting what is missing:


    Acceptance Criteria and Device Performance Study for K994306 (Medisystems Blood Tubing Set Accessories)

    Given the nature of the device (blood tubing set accessories for hemodialysis) and the regulatory context (510(k) submission from 2000), the testing described is primarily for manufacturing quality assurance and functional integrity rather than AI/ML performance metrics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (from text)Reported Device Performance (from text)
    Manufacturing/Quality ControlSterility"all finished products are tested and must meet all required release specifications before distribution"
    Pyrogenicity"all finished products are tested and must meet all required release specifications before distribution"
    Physical Testing"all finished products are tested and must meet all required release specifications before distribution"
    Visual Examination (in-process and finished product)"all finished products are tested and must meet all required release specifications before distribution"
    Functional IntegrityGasket to prevent leakage (for Reverso™ valve)The device description states it has a "gasket to prevent leakage," implying this is a design feature meant to be met. The general statement about meeting release specifications would cover this.
    Ability to interconnect between blood tubing set and access devices (for Reverso™)The device description states it is "equipped with tubing and Luer connectors to allow it to interconnect..." Implied to be met through design and quality checks.
    Ability to reverse arterial and venous access without disconnection (for Reverso™)This is the intended function. No specific performance metric is given beyond the description of functionality.
    Proper interconnection as male-to-male Luer adaptor (Recirculation Connector)Implied through design and quality checks. Functionality as described is the performance.
    Ability to connect two dialyzers in parallel (Parallel Dialyzer Interconnector)Implied through design and quality checks. Functionality as described is the performance.
    Ability to connect two dialyzers in series (Series Dialyzer Interconnector)Implied through design and quality checks. Functionality as described is the performance.

    Missing Information: The document does not specify quantitative acceptance criteria for device performance (e.g., "leakage rate must be < X%," or "reversal mechanism must operate smoothly for Y cycles"). The "acceptance criteria" are implied by the necessity to conform to "product design specifications" and "required release specifications" through general quality control tests.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified. The document states "all finished products are tested," which implies 100% inspection or batch testing based on internal procedures, but no specific sample size for a "test set" in a clinical/performance study context is mentioned.
    • Data Provenance: Not applicable in the context of clinical/performance data. The testing described is part of the manufacturing process and quality control, not from patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable. This device is a mechanical accessory, not an AI/ML diagnostic or prognostic device that requires expert-established ground truth from images or clinical data. The "ground truth" for this device's safety and effectiveness relies on meeting engineering specifications and quality control parameters.

    4. Adjudication Method for the Test Set

    • Not Applicable. There is no "adjudication method" described as it relates to expert review of output or clinical outcomes. Testing is physical and functional.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI assistance

    • Not Applicable. This device is a mechanical medical accessory, not an AI-assisted diagnostic tool. No MRMC study was performed or is relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not Applicable. This is not an algorithm-based device.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's safety and effectiveness is based on engineering specifications, manufacturing standards, and quality control procedures (e.g., demonstration of sterility, lack of pyrogens, physical integrity, and functional performance as designed). There is no "expert consensus," "pathology," or "outcomes data" ground truth described for its basic functional performance.

    8. The Sample Size for the Training Set

    • Not Applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. This is not an AI/ML device that requires a training set and associated ground truth.

    Summary Conclusion:

    The K994306 submission describes a mechanical medical device designed to facilitate hemodialysis. The "safety and effectiveness" section focuses entirely on standard manufacturing quality control and adherence to design specifications (sterility, pyrogenicity, physical testing, visual examination). It establishes substantial equivalence based on its intended use and technological characteristics being comparable to previously cleared predicate devices, rather than through performance metrics derived from clinical studies or AI/ML evaluation. Therefore, most of the requested information (related to AI/ML device evaluation) is not present in this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1