    Intended Use

    The Reprocessed HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic, thoracic procedures, and sealing and transection of lymphatic vessels. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand control button.

    Device Description

    The reprocessed HARMONIC 700 with Advanced Hemostasis are designed for soft tissue incisions requiring bleeding control and minimal thermal injury. The instruments serve as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in various procedures, including general, pediatric, gynecologic, urologic, and thoracic surgeries, as well as in the sealing and transection of lymphatic vessels. They enable the coagulation of vessels up to 7 mm in diameter using the Advanced Hemostasis hand control button.

    The instruments are available in three (3) shaft lengths: 23cm, 36cm, and 45cm lengths (HAR723, HAR736, and HAR745 respectively). Each shaft length has a diameter of 5 mm. The following features are essential to the control and performance of the device:

    • An actuating trigger that closes and releases the clamp arm, securing tissue against the scalpel rod.
    • MIN/MAX control buttons that adjust energy levels between minimum and maximum modes on the generator, enabling vessel sealing up to 5 mm.
    • An Advanced Hemostasis button that allows the clinician to activate an additional energy mode, enabling vessel sealing up to 7 mm.
    • A rotation knob to rotate the shaft 360° unless energy is being delivered.
    • A torque wrench, a sterile, single-use component, used to apply the correct amount of torque when attaching the Hand Piece to the device.

    The instruments connect to a generator and hand piece, which are essential for the device's functionality but are outside the scope of this submission.
    The hand piece is a reusable component that attaches to the device and plugs into the generator, allowing the device to interface with the generator. This component contains the transducer, which converts electrical power to ultrasonic mechanical energy.

    The generator is a reusable component that generates the electrical signal. Colored light indicators on the front panel of the generator visually communicate device status information to the user.

    AI/ML Overview

    This document is a 510(k) premarket notification for reprocessed medical devices, not an AI medical device. Therefore, the questions related to AI device performance, such as MRMC studies, training and test sets, ground truth establishment, and expert adjudication, are not applicable to the content provided.

    However, I can extract the acceptance criteria and the study that proves the device meets those criteria from the provided text, focusing on the reprocessed medical device context.

    Here's the information based on the provided FDA 510(k) letter for the Reprocessed HARMONIC 700 Shears:

    1. A table of acceptance criteria and the reported device performance:

    The document describes functional performance tests conducted to demonstrate safety and effectiveness. While explicit numerical acceptance criteria values are not provided, the general categories of testing imply the criteria for performance equivalence to the predicate device. The reported performance is that the device meets these criteria and is at least as safe and effective as the predicate.

    Acceptance Criteria CategoryReported Device Performance
    Validation of ReprocessingDemonstrated to be effective.
    Electrical Safety and Electromagnetic CompatibilityTested in accordance with IEC 60601-1 and IEC 60601-1-2; results deemed acceptable.
    Functional Performance Tests:
    Jaw Clamp ForceTested and found acceptable for intended function.
    Tissue Retention ForceTested and found acceptable for intended function.
    Burst Pressure (for vessel sealing)Tested and found acceptable for sealing vessels up to 7mm.
    Maximum Jaw and Shaft TemperatureTested and found acceptable.
    Device ReliabilityDemonstrated to be reliable.
    ATT Functionality and Transection TimeTested and found acceptable for intended function.
    BiocompatibilityPreviously identified and cleared materials and reprocessing methods were maintained, demonstrating continued biocompatibility.
    Sterilization/Shelf-LifeValidated to a 10-6 sterility assurance level (SAL) through an EO sterilization process in accordance with ISO 11135. Shelf-life not explicitly detailed but implied as validated.
    Preclinical Laboratory Evaluations (Animal Model):
    Thermal SpreadTested and found acceptable.
    Ability to achieve Hemostasis of VesselsTested and found acceptable for sealing vessels up to 7mm.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the exact sample sizes (number of devices or tests) for each bench and laboratory test.
    • Data Provenance: The studies were conducted as part of the manufacturer's premarket notification submission to the FDA. They are internal validation studies performed by Stryker Sustainability Solutions. The document does not specify country of origin for the data or whether it was retrospective or prospective, but these studies for 510(k) submissions are typically prospective and specifically designed for the submission.
      • For the preclinical laboratory evaluations, an "animal model" was used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not applicable to traditional reprocessing medical device submissions. Ground truth in this context is established through engineering specifications, material science, and performance testing against established standards and predicate device performance, not expert consensus on medical images or patient outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods like 2+1 or 3+1 refer to human expert review processes for AI model output or image interpretation, which is not relevant for this type of device submission. Device performance is determined through standardized engineering and biological tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a reprocessed surgical instrument, not an AI-assisted diagnostic or therapeutic device. MRMC studies are specific to evaluating human reader performance with and without AI assistance in diagnostic imaging.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI algorithm. Its performance is inherent to its physical properties and functionality.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is established by:

    • Engineering specifications and design requirements: The device must meet predefined physical, mechanical, and electrical parameters.
    • Performance of the original (new) predicate device: The reprocessed device must demonstrate substantial equivalence in performance to the original, legally marketed predicate device.
    • Established industry standards: Compliance with standards like ISO 10993-1 (Biocompatibility) and ISO 11135 (EO Sterilization) serves as a ground truth for safety aspects.
    • Preclinical (animal) studies: In vivo performance (e.g., thermal spread, hemostasis) in animal models serves as a proxy for clinical performance.

    8. The sample size for the training set:

    Not applicable. There is no "training set" in the context of a reprocessed medical device. The device itself is manufactured and reprocessed, not "trained" like a machine learning model.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241606

    Validate with FDA (Live)

    Date Cleared
    2024-07-03

    (29 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed HARMONIC 1100 Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic, thoracic, sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis.

    Device Description

    The Reprocessed HARMONIC 1100 Shears instrument is a sterile, single patient used for dissection, grasping, coagulation, and cutting between the blade and clamp arm. It consists of an ergonomic handle with an integrated hand piece and two energy delivery buttons. 1) Energy button for power levels 1-5, and 2) Energy with Advanced Hemostasis button for large vessel sealing. The instrument is available in two shaft lengths (20 cm and 36 cm). An integrated audible and tactile mechanism in the grip housing indicates full trigger closure. The instrument has a clamp and coated curved blade that are designed to work through a 5 mm trocar, through a 5 mm reducer cap in a larger diameter trocar, or through an incision without the use of a trocar. The instrument shafts can be rotated continuously to facilitate visualization and access to targeted tissue. The instrument are intended to represent relative vessel size. The Energy button is in diameter. When the Energy button is used, cutting speed is the fastest. The energy button with Advanced Hemostasis is designed for larger vessels and is indicated for vessels up to 7 mm in diameter. When the Energy button with Advanced Hemostasis is reduced and hemostasis is maximized. The instrument utilizes Adaptive Tissue Technology. This provides the generator with the ability to identify and monitor the instrument during use, which enables the generator to modulate and adjust its power output as well as provide audible feedback to the user as appropriate. The HARMONC 1100 Shears instrument is designed for use exclusively with the Generator 11 (GEN11) software version 2018-1 or later.

    AI/ML Overview

    The provided text is a 510(k) summary for reprocessed HARMONIC 1100 Shears. It describes the device, its intended use, and a study comparing it to predicate devices. However, it does not contain detailed information about specific acceptance criteria or the study design and results in a format that allows for the extraction of all the requested information.

    Here's an analysis of what can be extracted and what is missing:

    The submission is for Stryker Sustainability Solutions' Reprocessed HARMONIC 1100 Shears (HAR1120 and HAR1136). The focus of the provided document is the expansion of the indications for use to include coagulation of veins up to and including 7 mm in diameter.

    1. A table of acceptance criteria and the reported device performance

    • The document states: "Supplemental chronic 21-day survival studies were completed to assess the long-term seal quality of veins up to and including 7 mm in diameter. The data supports the expansion of the include coagulation of vessels up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis."
    • Missing from the text: Specific quantitative acceptance criteria (e.g., minimum seal burst pressure, failure rates) and the precise reported device performance metrics against those criteria are not detailed. The phrase "long-term seal quality" is qualitative without acceptance thresholds.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The study is described as "Supplemental chronic 21-day survival studies." This implies a prospective animal study.
    • Missing from the text: The number of animals or vessels tested (sample size) for this study is not provided. The country of origin for the data is also not specified, though it's implied to be a pre-clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This question is not applicable to this type of device and study. The "ground truth" for vessel sealing devices in pre-clinical studies is typically based on physiological measurements (e.g., burst pressure, histological analysis) rather than expert interpretation of images. The study involved "animal study data," which suggests direct observation and measurement in live subjects.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. As noted above, this study is not interpreting images or clinical outcomes that would require an adjudication method by human readers. It's a technical performance study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This document describes a medical device (surgical shears), not an AI-based diagnostic or assistive technology for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This document describes a medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The ground truth in this context would be the directly measured "long-term seal quality of veins" in the animal model. This would likely involve physiological parameters like burst pressure, histological examination of the sealed vessels for integrity, and potentially observation of complications over the 21-day survival period.
    • Missing from the text: The specific metrics and methods used to define "long-term seal quality" are not detailed.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device that requires a training set.

    Summary of available information regarding the device's performance study:

    Device: Reprocessed HARMONIC 1100 Shears (HAR1120, HAR1136)

    Purpose of Study: To support the expansion of indications for use to include coagulation of veins up to and including 7 mm in diameter.

    Study Type: Supplemental chronic 21-day survival studies in an animal model.

    Evaluation Focus: Assessment of "long-term seal quality" of veins up to and including 7 mm in diameter.

    Conclusion stated: "The data supports the expansion of the include coagulation of vessels up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis."

    Missing Critical Information from the provided text:

    • Detailed quantitative acceptance criteria.
    • Specific numerical results of the "long-term seal quality" assessment.
    • The exact sample size (number of animals or vessels) used in the study.
    • Specific ground truth metrics (e.g., burst pressure values, histological findings).
    • Details of the study protocol.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233471

    Validate with FDA (Live)

    Date Cleared
    2024-02-20

    (118 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed HARMONIC 1100 Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, thoracic, sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of veins up to and including 5 mm in diameter and arteries up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis.

    Device Description

    The Reprocessed HARMONIC 1100 Shears sterile instruments are used for dissection, grasping, coaqulation, and cutting between the blade and clamp arm. Each device model consists of an ergonomic handle, an integrated hand piece, and two energy delivery buttons:

    • Energy Button - User can adjust power levels from 1-5
    • Energy button with Advanced Hemostasis – For larger vessel sealing; user cannot adjust.
      The instruments are available in two shaft lengths, 20 cm and 36 cm. An integrated audible and tactile mechanism in the handle indicates full trigger closure. The instrument has a clamp arm and coated curved blade that are deigned to work through a 5 mm trocar, through a 5mm reducer cap in a larger diameter trocar, or through an incision without the use of a trocar. The instrument shafts can be rotated continuously to facilitate visualization and access to targeted tissue. The two dashes on the instrument are intended to represent relative vessel size. The energy button is indicated for vessels up to 5 mm in diameter. When the energy button is used, cutting speed is the fastest. The energy button with Advanced Hemostasis is designed for larger vessels and is indicated for vessels up to 7 mm in diameter. When the Energy button with Advanced Hemostasis in used, cutting speed in reduced and hemostasis is maximized. The instrument utilizes Adaptive Tissue Technology. This provides the generator with the ability to identify and monitor the instrument during use, which enables the generator to modulate and adjust its power output as well as provide audible feedback to the user as appropriate.
    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA regarding the Reprocessed HARMONIC 1100 Shears. It states that the device is substantially equivalent to a legally marketed predicate device. This type of FDA clearance does not involve an AI/ML component, and therefore, the information requested in your prompt regarding acceptance criteria and studies for AI/ML devices is not applicable to this document. The document describes physical and performance tests for a reprocessed surgical instrument, not an AI-powered diagnostic or therapeutic device.

    Therefore, I cannot provide the information requested in your prompt based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202554

    Validate with FDA (Live)

    Date Cleared
    2021-03-16

    (194 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed HARMONIC ACE®+7, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand control button.

    Device Description

    Reprocessed HARMONIC ACE®+7, 5mm Diameter Shears with Advanced Hemostasis are used for coagulation and mechanical transection of soft tissue during laparoscopic and open procedures. The devices consist of an ergonomic handle and 3 hand-controlled activation buttons. The handle includes a mechanism that provides both audible and tactile feedback indicating full closure. The instruments utilize Adaptive Tissue Technology which provides the generator with the ability to identify and monitor the instrument during use and enables the generator to modulate and adjust its power output as well as provide audible feedback to the user as appropriate. Each instrument is shipped with one sterile, single-use, disposable torque wrench.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a reprocessed medical device, specifically "Reprocessed HARMONIC ACE®+7, 5 mm Diameter Shears with Advanced Hemostasis". This document details the comparison to a predicate device and various tests conducted to demonstrate substantial equivalence, rather than a clinical study evaluating a diagnostic or AI-driven device's performance against specific acceptance criteria like accuracy, sensitivity, or specificity.

    Therefore, the requested information regarding acceptance criteria, reported device performance in those terms, sample sizes for test/training sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone algorithm performance, and types of ground truth for a diagnostic AI device is not applicable to this document.

    The document focuses on demonstrating that the reprocessed surgical shears are as safe and effective as the original, legally marketed predicate device through a series of engineering and performance tests.

    Here's what can be extracted regarding the device's performance demonstration:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document outlines functional performance tests and other validations, rather than specific diagnostic accuracy metrics. The "acceptance criteria" are implied to be achieving performance comparable to the predicate device and meeting safety standards.

    Acceptance Criteria CategorySpecific Test/ValidationReported Device Performance
    Safety & BiocompatibilityBiocompatibilityConducted
    Validation of ReprocessingConducted
    Sterilization ValidationConducted (Sterile by Ethylene Oxide)
    Pyrogen FreeNo (Same as Predicate)
    Electromagnetic Compatibility & Electrical Safety TestingConducted (in accordance with IEC 60601-1-2)
    Functional PerformanceBlade to Clamp Arm AngleConducted
    Actuating Trigger ForceConducted
    MIN/MAX Button Activation ForceConducted
    Advanced Hemostasis (AH) Button Activation ForceConducted
    Rotation Knob Force – Actuating Trigger DisengagedConducted
    Rotation Knob Force Actuating Trigger EngagedConducted
    Jaw Clamp ForceConducted
    Tissue Retention ForceConducted
    Shaft StraightnessConducted
    ATT Functionality and Transection TimeConducted. Evaluation of device's performance and ability to seal and divide vessels up to 7mm, including transection time.
    Burst PressureConducted. Evaluation of device's performance and ability to seal and divide vessels up to 7mm, including burst pressure.
    Maximum Jaw and Shaft TemperatureConducted. Evaluation of device's performance and ability to seal and divide vessels up to 7mm, including thermal spread.
    Reliability TestingConducted
    Packaging & LabelingPackaging ValidationConducted
    Clinical EquivalencePre-Clinical Testing (Acute Animal Study, Chronic Animal Study)Conducted. "The performance testing demonstrates that reprocessed devices are as safe and effective as the predicate devices and operate as originally intended."

    2. Sample size used for the test set and the data provenance:

    • This document does not specify exact sample sizes for each test. It generally states that "Bench and laboratory testing was conducted" and includes "verification/comparative testing (to the predicate device)."
    • Data provenance: Not explicitly stated, but assumed to be internal laboratory testing by Stryker Sustainability Solutions. It is not clinical data from specific countries or indicated as retrospective/prospective in a diagnostic sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is not a diagnostic device relying on expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a diagnostic device requiring adjudication of interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a reprocessed surgical instrument, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a reprocessed surgical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the performance tests, the "ground truth" would be established by engineering specifications, validated test methods, and comparison against the performance of the original, new predicate device. For animal studies, the "ground truth" would be physiological outcomes observed by veterinarians or pathologists.

    8. The sample size for the training set:

    • Not applicable. This is a reprocessed physical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190610

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-08-15

    (157 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, otorhinolaryngologic (ENT), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space) and other open procedures.

    Device Description

    The Reprocessed Harmonic FOCUS Shears + Adaptive Tissue Technology is a sterile, single-patient use surgical instrument consisting of a soft grip scissor handle housing assembly with two hand controls (MIN for minimum power level and MAX for maximum power level). The instrument's working length is 9 cm in length with a 16 mm active blade length. The instrument allows for the cutting and coagulation of vessels up to and including 5 mm in diameter.

    AI/ML Overview

    The provided text describes the regulatory clearance for a reprocessed medical device, specifically the "Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology." It details the device's indications for use, its characteristics, and the non-clinical tests performed to demonstrate its substantial equivalence to an original predicate device.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" with quantitative targets and reported performance in the format typically seen for AI device studies (e.g., sensitivity, specificity thresholds). Instead, for this reprocessed surgical instrument, the acceptance criteria are implicitly defined by demonstrating that the reprocessed device performs as "originally intended" and is "substantially equivalent" to the original predicate device through a series of functional, safety, and sterilization tests.

    From the "Summary of Non-Clinical Tests Conducted" section, the performance tests conducted include:

    Test PerformedReported Device Performance
    Grasping ForcePerformed and found to be equivalent
    Artery Seal Burst and Tissue AdhesionPerformed and found to be equivalent
    Vein Seal Burst and Tissue AdhesionPerformed and found to be equivalent
    Electrical and Thermal Safety IEC 60601Performed and found to be equivalent
    Tissue (Jaw) Pad LifePerformed and found to be equivalent
    Tissue (Jaw) Pad Removal ForcePerformed and found to be equivalent
    Thermal Spread and Transection TimePerformed and found to be equivalent
    Cleaning ValidationValidated
    Sterilization VerificationVerified
    Ethylene Oxide Residual Testing (ISO 10993-7)Performed and found to be compliant
    Packaging Validation (ASTM D4169, ASTM F88, ASTM F2096)Validated
    Shelf-Life Validation (ASTM 1980-07)Validated
    Biocompatibility Testing (ISO 10993-1)Compliant (Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogenicity)

    The overall conclusion states: "Performance testing shows the Harmonic FOCUS Shears + Adaptive Tissue Technology performs as originally intended." and "Results demonstrated substantial equivalence to the predicate devices with respect to safety and effectiveness."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample sizes used for each performance test. It mentions "Reprocessed devices were tested" and "Each device and accessory are marked and tracked through the reprocessing cycle." The testing appears to be conducted on samples of the reprocessed devices. The provenance is not explicitly stated in terms of country of origin of data or whether it was retrospective or prospective. It is implied to be prospective testing of reprocessed devices manufactured by SterilMed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is a physical medical device, not an AI or imaging device that requires expert interpretation for ground truth. Therefore, the concept of "experts" establishing ground truth in this context is not applicable. The performance is assessed through objective engineering and biological tests, not human interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is objective performance testing of a physical device, not an interpretative task requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI or imaging device, and there is no human reader component. The study compares the performance of the reprocessed device against its original predicate device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is a physical surgical instrument, not an algorithm. The "standalone performance" in this context refers to the device's functional integrity and safety.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's reprocessed performance is established by:

    • Predicate Device Performance: The original HARMONIC FOCUS® Shears + Adaptive Tissue Technology (Ethicon K133314) serves as the benchmark for expected performance.
    • Accepted Standards: Compliance with various international and national standards (e.g., ISO 10993-7, ISO 10993-1, IEC 60601, ASTM D4169, ASTM F88, ASTM F2096, ASTM 1980-07) for sterilization, biocompatibility, packaging, and electrical safety.
    • Engineering Specifications: The device's original design specifications for grasping force, seal burst, tissue adhesion, jaw pad life, removal force, thermal spread, and transection time.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182272

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-04-15

    (236 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterilmed Reprocessed HARMONIC ACE®+ 7 Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter using the Advanced Hemostasis hand control button.

    Device Description

    The Sterilmed Reprocessed HARMONIC ACE+7 Shears with Advanced Hemostasis are used for coagulation and mechanical transection of soft tissue during laparoscopic and open procedures. The devices allow the surgeon to grasp, coagulate, and transect soft tissue with a single instrument. The devices are hand-actuated with a shaft and tissue effector that can be rotated. The energy delivery can be activated with hand activation or with an optional generator foot switch. The two models are identical to each other in materials, form, fit and function except for the length of the shafts which are 23cm and 36 cm.

    AI/ML Overview

    This document describes the validation of reprocessed HARMONIC ACE®+ 7 Shears with Advanced Hemostasis, demonstrating their substantial equivalence to the original equipment manufacturer (OEM) predicate device. Given that the provided text is a 510(k) summary for a reprocessed medical device (surgical shears), the acceptance criteria and study design are focused on demonstrating that the reprocessed device performs as safely and effectively as the original device. This is a non-clinical submission, thus the information regarding AI/human reader studies, expert consensus on images, etc., is not applicable here.

    Here's a breakdown of the requested information based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for a reprocessed device are fundamentally centered on demonstrating that it performs equivalently to the new, original device in terms of safety and functional effectiveness. The performance of the reprocessed device is compared against the OEM predicate device. Specific quantitative acceptance criteria or threshold values are not explicitly listed in this summary, but the general principle is that the reprocessed device "performs as originally intended" and is "substantially equivalent" to the predicate.

    Test Parameter / Acceptance CriteriaReported Device Performance (Reprocessed Device)
    Functional PerformanceSimulated use, visual inspection, fatigue testing. Must perform as originally intended."Performance testing shows the Reprocessed HARMONIC ACE+ 7 Shears with Advanced Hemostasis performs as originally intended."
    Cleaning ValidationValidation of cleaning efficacy."Process validation testing was performed to validate cleaning..."
    Sterilization VerificationValidation of sterilization efficacy."...and sterilization..."
    Ethylene Oxide Residual Testing(ISO 10993-7 compliant for residuals)"...ethylene oxide residual testing (ISO 10993-7)..."
    Packaging Validation(ASTM D4169, ASTM F88, ASTM F2096 compliant)"...packaging validation (ASTM D4169, ASTM F88, ASTM F2096)..."
    Shelf-Life Validation(ASTM 1980-07 compliant)"...and shelf-life validation (ASTM 1980-07)."
    Electrical SafetyMust meet safety standards."Testing performed: Electrical Safety"
    Vein Seal Burst (1-7mm) & Tissue AdhesionEquivalent sealing performance to predicate."Vein Seal Burst (1-7mm) and Tissue Adhesion" (Implied equivalent to predicate)
    Artery Seal Burst (1-7mm) & Tissue AdhesionEquivalent sealing performance to predicate."Artery Seal Burst (1-7mm) and Tissue Adhesion" (Implied equivalent to predicate)
    Lymphatics Seal PerformanceEquivalent sealing performance to predicate."Lymphatics Seal Performance" (Implied equivalent to predicate)
    Vessel Seal Thermal SpreadMinimize thermal injury, equivalent to predicate."Vessel Seal Thermal Spread" (Implied equivalent to predicate)
    Drop Fluid IngressMaintain integrity against fluid ingress."Drop Fluid Ingress" (Implied satisfactory)
    Pad Retention, Pad Life, Grasping ForceMaintain mechanical integrity and function."Pad Retention, Pad Life, Grasping Force" (Implied satisfactory)
    Biocompatibility(ISO 10993-1 compliant: Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogenicity)"Biocompatibility testing included: Cytotoxicity Sensitization Irritation/Intracutaneous Reactivity Acute Systemic Toxicity Pyrogenicity" (Implied compliant)
    Overall Comparison to Predicate DeviceSubstantially equivalent with respect to safety and effectiveness."Results demonstrated substantial equivalence to the predicate devices with respect to safety and effectiveness."

    Study Details

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document states that "representative samples of reprocessed devices were tested," and for functional testing, "all products produced" undergo visual and validated functional testing. However, specific numerical sample sizes for each test (e.g., how many devices were tested for vein seal burst, or fatigue testing) are not provided in this summary.
    • Data Provenance: The testing was conducted by Sterilmed, Inc., a reprocessor based in Plymouth, MN, and Maple Grove, MN, USA. The data is from prospective bench testing and validation studies of their reprocessed devices. It is not human clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This information is not applicable as this is a submission for reprocessed surgical shears based on non-clinical, bench testing, and engineering validation, not an AI or imaging-based device. No "experts" in the sense of medical professionals interpreting images were used to establish ground truth for the device's functional performance. Ground truth was established through validated engineering and laboratory test methods, often comparing performance against the OEM predicate.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving multiple human readers interpreting data, typically in diagnostic imaging. For a reprocessed surgical device, the "adjudication" is inherent in the robust, validated test methods and comparison to established predicate performance. If a test failed, it would need to be investigated and resolved, but there isn't a "reader-based" adjudication process.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. MRMC studies are specific to diagnostic imaging devices, often involving AI, where human readers evaluate cases. This submission is for reprocessed surgical shears, which do not involve human readers for diagnostic interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable. This device is a mechanical/ultrasonic surgical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is established by:

    • Engineering specifications and performance characteristics of the OEM predicate device (Ethicon HARMONIC ACE®+7 Shears with Advanced Hemostasis, K132612).
    • Validated industry standards and test methods (e.g., ISO 10993, ASTM standards for packaging, electrical safety).
    • Successful functional performance benchmarks (e.g., vein/artery seal burst pressure, thermal spread, grasping force), which show the device "performs as originally intended."
      In essence, the ground truth is that the reprocessed device must meet or exceed the safety and functional performance of the new, original device.

    8. The sample size for the training set:

    This is not applicable. This is not an AI/machine learning device; there is no "training set" in the computational sense. The "training" for the manufacturing process involves process validation and quality control measures to ensure consistent reprocessing.

    9. How the ground truth for the training set was established:

    This is not applicable for the same reason as point 8. The "ground truth" for the overall manufacturing and reprocessing process is established through rigorous validation of cleaning, sterilization, packaging, and functional testing protocols to ensure the device is consistently reprocessed to meet its intended performance specifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170456

    Validate with FDA (Live)

    Date Cleared
    2017-04-05

    (49 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, otorhinolaryngologic (ENT), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space) and other open procedures.

    Device Description

    Reprocessed HARMONIC FOCUS® Shears + Adaptive Tissue Technology are a sterile, single patient use instrument consisting of a soft grip scissor handle housing assembly with two hand controls (MIN for minimum power level and MAX for maximum power level). The instrument has a curved blade and clamp arm with Teflon pad. The instrument is 9 cm in length with a 16 mm active blade length. The Reprocessed HARMONIC FOCUS®+ Shears instrument allows for the cutting and coagulation of vessels up to and including 5 mm in diameter. Each Reprocessed HARMONIC FOCUS®+ Shears instrument is packaged with one sterile, single patient use, disposable Torque Wrench.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a reprocessed medical device, the "Reprocessed HARMONIC FOCUS Shears + Adaptive Tissue Technology." This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the de novo effectiveness of a novel device. Therefore, the concept of "acceptance criteria" and "study that proves the device meets the acceptance criteria" as typically applied to AI/ML or diagnostic devices with quantifiable performance metrics (like sensitivity, specificity, AUC) is not directly applicable in the same way.

    However, we can interpret acceptance criteria in the context of this 510(k) as demonstrating that the reprocessed device performs as well as the original predicate device, and that the reprocessing itself does not introduce new safety or effectiveness concerns.

    Here's an attempt to structure the information based on your request, acknowledging the limitations of the provided document for each specific point:

    Acceptance Criteria and Study for Reprocessed HARMONIC FOCUS Shears + Adaptive Tissue Technology

    In the context of this 510(k) submission for a reprocessed device, the "acceptance criteria" are implicitly tied to demonstrating that the reprocessed device is substantially equivalent to the original predicate device and that the reprocessing process does not compromise safety or effectiveness. The "study" refers to the performance data submitted to the FDA to support this claim.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a reprocessed device, the acceptance criteria are not explicitly defined as numerical thresholds for specific performance metrics like those for a diagnostic AI. Instead, they are comparative, aiming to show that the reprocessed device performs equivalently to the new, original device.

    Acceptance Criterion (Implicit)Reported Device Performance
    Safety and Effectiveness Equivalence to Predicate Device"The performance testing demonstrates that reprocessed devices are as safe and effective as the predicate and operate as originally intended." This generally means that the reprocessed device should not introduce new safety risks and should achieve the same clinical outcomes as the original device. Specific tests contributing to this include: - Electrical Safety: Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. - Electromagnetic Compatibility: Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. - Functional Performance (Comparative): Evaluation of ability to seal and divide vessels up to 5mm, including: - Thermal spread - Transection time - Burst pressure - Device functionality - Device reliability - Preclinical Laboratory Evaluations (Animal Model): Acute and chronic survival studies to evaluate thermal spread and ability to achieve hemostasis.
    BiocompatibilityTesting was conducted. (Assumed acceptance: demonstrates no new biocompatibility risks after reprocessing).
    Validation of ReprocessingTesting was conducted. (Assumed acceptance: demonstrates the reprocessing method effectively cleans, restores, and prepares the device for reuse without degradation or contamination). The description mentions "removal of adherent visible soil and decontamination."
    Sterilization ValidationTesting was conducted. (Assumed acceptance: demonstrates the reprocessed device is sterile).
    Packaging ValidationTesting was conducted. (Assumed acceptance: demonstrates packaging maintains sterility and device integrity).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample sizes for the "bench and laboratory testing" or the "preclinical laboratory evaluations." It mentions "animal model" but not the number of animals or trials. The data provenance is internal to the manufacturer's testing, presumably conducted at their facilities or certified labs. No indication of country of origin of data beyond where Stryker Sustainability Solutions operates (Tempe, Arizona). The studies are inherently prospective in nature, as they involve testing the reprocessed devices under controlled conditions.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable and not provided in the document. For a reprocessed surgical instrument, "ground truth" is not established by expert consensus on image interpretation, but rather by objective physical, chemical, and biological testing as outlined above (e.g., burst pressure measurements, bacterial counts, visual inspection criteria, etc.). There would be engineers, microbiologists, and other technical experts involved in conducting and interpreting these tests, but they are not "experts establishing ground truth" in the diagnostic sense.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for human expert interpretation of data (e.g., medical images) to establish a consensus ground truth. For instrument performance testing, the results are typically quantitative measurements or pass/fail criteria based on defined specifications, not subjective interpretation requiring adjudication among experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. MRMC studies are conducted for diagnostic devices (often AI-assisted) to assess how human readers' performance (e.g., diagnostic accuracy) changes with or without AI assistance. This document is for a physical surgical instrument, not a diagnostic tool where human interpretation is involved.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    Not applicable. There is no AI algorithm being evaluated for its standalone performance in a diagnostic capacity. The "Adaptive Tissue Technology" refers to algorithms within the generator that communicate with the device to optimize energy delivery, but this is an integrated system functionality, not a standalone diagnostic algorithm. The functional performance tests implicitly evaluate the device's performance, which includes the integrated technology, without specific focus on "algorithm-only" performance as a separate entity.

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance is established through:

    • Physical and Mechanical Measurements: Such as thermal spread, transection time, burst pressure, device functionality, and reliability, measured against predetermined specifications or compared to the predicate device's performance.
    • Biological/Chemical Testing: Biocompatibility, sterilization validation (e.g., sterility assurance level), and validation of cleaning (e.g., residual protein levels, visual cleanliness).
    • Preclinical (Animal Model) Outcomes: Evaluation of hemostasis and thermal spread in vivo.
      This is distinct from "expert consensus," "pathology," or "outcomes data" in the context of diagnostic performance.

    8. Sample Size for the Training Set

    Not applicable. This is a reprocessed physical device, not an AI/ML algorithm that learns from a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML algorithm in this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161693

    Validate with FDA (Live)

    Date Cleared
    2017-02-23

    (248 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed HARMONIC ACE®+7, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand control button.

    Device Description

    Reprocessed HARMONIC ACE®+7, 5mm Diameter Shears with Advanced Hemostasis are used for coaqulation and mechanical transection of soft tissue during laparoscopic and open procedures. The devices consist of an erqonomic handle and 3 hand-controlled activation buttons. The handle includes a mechanism that provides both audible and tactile feedback indicating full closure. The instruments utilize Adaptive Tissue Technology which provides the generator with the ability to identify and monitor the instrument during use and enables the generator to modulate and adjust its power output as well as provide audible feedback to the user as appropriate. The only difference between the three (3) model numbers subject of this submission is the shaft length detailed in the table below.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a reprocessed medical device, the Reprocessed HARMONIC ACE®+7, 5mm Diameter Shears with Advanced Hemostasis. This is crucial as the acceptance criteria and supporting studies are focused on demonstrating that the reprocessed device is substantially equivalent to the original, legally marketed predicate device, and operates as originally intended.

    Therefore, the studies and acceptance criteria are not about establishing the primary clinical efficacy of a new device, but rather validating the safety and functionality of a reprocessed version against an established benchmark.

    Here's a breakdown of the requested information based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table with specific quantitative acceptance criteria or detailed reported performance values for each test. Instead, it broadly states that the performance testing "demonstrates that reprocessed devices are as safe and effective as the predicates and operate as originally intended."

    The types of tests conducted suggest the implicit acceptance criteria would be that the reprocessed device's performance is comparable or equivalent to the predicate device within acceptable limits for:

    Acceptance Criteria (Implied)Reported Device Performance (Summary)
    Biocompatibility: Device materials are not cytotoxic, irritating, sensitizing, or otherwise harmful.Not explicitly stated, but validated by "Biocompatibility" testing.
    Validation of Reprocessing: The reprocessing procedure effectively cleans, sterilizes, and restores the device, without compromising material integrity.Not explicitly stated, but validated by "Validation of Reprocessing" testing.
    Sterilization Validation: The sterilization method achieves the required Sterility Assurance Level (SAL).Not explicitly stated, but validated by "Sterilization Validation" testing.
    Functional Performance (General): Device functions as specified, with no degradation in operation."Device functionality" and "device reliability" were evaluated. Performance "as originally intended" was concluded.
    Electrical Safety: Meets relevant electrical safety standards.Validated in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2.
    Electromagnetic Compatibility: Does not interfere with, or is not affected by, electromagnetic fields.Validated in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2.
    Software Validation: Any embedded software functions correctly and reliably.Validated by "Software Validation" testing.
    Packaging Validation: Packaging maintains sterility and device integrity until use.Validated by "Packaging Validation" testing.
    Thermal Spread: Comparable thermal spread to the predicate device.Evaluated in bench and acute/chronic pre-clinical testing. Concluded as "safe and effective as the predicates."
    Transection Time: Comparable transection time to the predicate device.Evaluated in bench testing. Concluded as "safe and effective as the predicates."
    Burst Pressure: Comparable burst pressure to the predicate device for sealed vessels.Evaluated in bench testing. Concluded as "safe and effective as the predicates."
    Ability to Seal and Divide Vessels (up to 7mm): Maintained efficacy in sealing and dividing vessels of specified size.Evaluated in bench and acute/chronic pre-clinical testing. Concluded as "safe and effective as the predicates."
    Hemostasis: Ability to achieve hemostasis is maintained.Evaluated in acute/chronic pre-clinical testing ("ability to achieve hemostasis of vessels and tissues").

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify exact sample sizes for each test. It generally refers to "bench and laboratory testing" and "acute and chronic pre-clinical testing."
    • Data Provenance: The studies were conducted as part of a 510(k) submission by Stryker Sustainability Solutions. The data would be internally generated through laboratory and pre-clinical animal studies. No information is provided regarding the country of origin of the data beyond the applicant company's location (Tempe, Arizona, USA). These studies are inherently prospective as they are specifically designed and executed to support the regulatory submission for the reprocessed device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and their Qualifications

    This information is not applicable in the context of this 510(k) submission. For reprocessed devices, the "ground truth" for performance is typically established by comparing the reprocessed device's objective physical and functional parameters against those of the original, new predicate device, and ensuring it meets its performance claims. This involves engineering and scientific testing rather than expert-derived ground truth based on cases (e.g., in medical image analysis).

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like "2+1" or "3+1" are relevant in studies where human readers are interpreting data (e.g., medical images) and their decisions need to be reconciled to form a consensus ground truth. The studies described here are primarily objective, bench, and pre-clinical tests comparing device performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. MRMC studies are typically used for diagnostic or screening devices (e.g., AI in radiology) to evaluate the impact of a system on human reader performance. This submission is for a reprocessed surgical instrument.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable in the way it's typically understood for AI/software devices. The device is a surgical instrument. However, the "Functional Performance Tests" and "Electrical Safety Testing," "Electromagnetic Compatibility Testing," and "Software Validation" can be considered "standalone" evaluations of the device's inherent physical and functional properties, independent of a human operator, to ensure it meets its specifications.

    7. The type of ground truth used

    The "ground truth" for the performance evaluations performed for this reprocessed device is established through:

    • Engineering and Scientific Measurements: Objective laboratory measurements of device characteristics (e.g., thermal spread, transection time, burst pressure, electrical parameters).
    • Pre-clinical Animal Model Observations: Direct observation of the device's effects in acute and chronic animal studies (e.g., hemostasis, tissue effects).
    • Comparison to Predicate Device Performance: The underlying ground truth is that the reprocessed device must perform equivalently to the original predicate device (HARMONIC ACE+ Shears with Advanced Hemostasis, K132612) and meet its stated indications.

    8. The Sample Size for the Training Set

    This information is not applicable. This device is a reprocessed physical medical instrument, not an AI/ML algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable per the reason stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161086

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-09-20

    (155 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed Harmonic Shears with Adaptive Tissue Technology are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures.

    Device Description

    The Reprocessed Harmonic Shears with Adaptive Tissue Technology are used for coagulation and mechanical transection of soft tissue during laparoscopic and open procedures. The devices allow the surgeon to grasp, coagulate and transect soft tissue with a single instrument. The devices are hand-actuated with shaft and tissue effector that can be rotated. The energy delivery can be activated with hand activation or with an optional generator foot switch. The device includes a torque wrench as an accessory piece (the torque wrench is designed to ensure that the hand piece is properly secured to the device.

    AI/ML Overview

    The provided text describes the 510(k) summary for Sterilmed's Reprocessed Harmonic Shears with Adaptive Tissue Technology. It details various tests conducted to demonstrate substantial equivalence to the predicate device.

    Here's a breakdown of the requested information based on the text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with numerical targets. Instead, it states that "Performance testing shows the reprocessed Harmonic Shears with Adaptive Tissue Technology to perform as originally intended" and "Results demonstrated substantial equivalence to the predicate devices with respect to safety and effectiveness."

    The types of tests performed, which imply the areas where performance was assessed, are listed under "Summary of Non-Clinical Tests Conducted" and "Functional and Safety Testing." These tests aimed to demonstrate that the reprocessed devices perform as well as the original predicate devices.

    Acceptance Criteria (Implied from Tests)Reported Device Performance
    Cleaning validation successValidated to meet cleaning requirements.
    Sterilization verification successValidated to meet sterilization requirements.
    Ethylene oxide residual levels (meeting ISO 10993-7)Met ISO 10993-7 requirements.
    Packaging integrity (meeting ASTM D4169, ASTM F88, ASTM F2096)Validated to maintain packaging integrity.
    Shelf-life validation success (meeting ASTM 1980-07)Validated to maintain shelf-life.
    Functional performance (bench testing) equivalent to predicate devicePerformed "as originally intended" and demonstrated "substantial equivalence" to predicate devices for: - Electrical Safety - Device Functionality - Vessel Seal Burst (Static and Burst Pressure) - Vessel Seal Thermal Spread - Drop - Fluid Ingress - Pad Retention - Pad Failure - Clamp Force
    Biocompatibility (meeting ISO 10993-1 for short duration contact with blood)Passed tests for: - Cytotoxicity - Sensitization - Irritation/Intracutaneous Reactivity - Acute Systemic Toxicity - Hemocompatability - Chemical Pyrogens - Hemolysis - Immune Response

    2. Sample Size Used for the Test Set and the Data Provenance

    The document mentions "Representative samples of reprocessed devices were tested" for functional characteristics. However, specific sample sizes for each test are not provided in this summary.

    The data provenance is from non-clinical tests conducted by Sterilmed, Inc., a reprocessing company. This is an in-house prospective testing of their reprocessed devices. The country of origin is implicitly the USA where Sterilmed is located and where the FDA regulatory submission was made.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information is not applicable to this document. The "ground truth" for this device (a reprocessed surgical instrument) is established by objective engineering and safety standards (e.g., sterilization parameters, mechanical performance, biocompatibility) rather than expert interpretation of data like images or patient outcomes. The performance is compared against the original (predicate) device.

    4. Adjudication Method for the Test Set

    This is not applicable as the evaluation is based on objective measurements against engineering standards and comparison to a predicate device, not on subjective assessments requiring adjudication among experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This device is a surgical instrument, not an AI-powered diagnostic or assistive technology that would involve human readers or AI.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a reprocessed surgical instrument and does not involve an algorithm or AI.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is based on defined engineering specifications, performance standards, regulatory requirements (e.g., ISO for biocompatibility and residuals), and the performance characteristics of the legally marketed predicate device. The goal is to demonstrate that the reprocessed device performs identically or equivalently to the new/original device.

    8. The Sample Size for the Training Set

    This question is not applicable. This is a reprocessed medical device, not an AI or machine learning algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132566

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-06-14

    (303 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The reprocessed harmonic shears are intended to be used for cutting soft tissue and when control of bleeding and minimal thermal injury is desired. The instruments can be used as un udjunct to or a substitute for clectrosurgery, lasers, and steel scalpels in general, plastic, pediatic, gyncoologic, thoracic, exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures.

    Device Description

    The harmonic shears are used in combination with a handpiece, generator and torque wrench and are intended to be used for cutting and coagulation of soft tissue during laparoscopic and open procedures. The instruments allow the surgeon to grasp, coagulate and cut soft tissue with one instrument. The instruments allow for the coagulation of vessels up to and including 5 mm in diameter. The instruments are hand-actuated with a shaft and distal tip that can be rotated 360°. The ultrasonic energy is delivered by hand activation or with an optional generator foot switch.

    AI/ML Overview

    The provided document pertains to the reprocessed Harmonic Shears and does not describe a study involving an AI device or its acceptance criteria. The document is a 510(k) summary for a medical device (reprocessed harmonic shears) and focuses on demonstrating substantial equivalence to predicate devices, rather than the performance of an AI algorithm.

    Therefore, I cannot provide the requested information regarding AI device acceptance criteria and study details based on the provided text.

    Ask a Question