    K Number
    K100909

    Validate with FDA (Live)

    Date Cleared
    2010-07-23

    (113 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When coupled with an appropriate inflation system, compression devices are intended for use in preventing deep vein thrombosis (DVT), diminishing post-operative pain and swelling, enhancing blood circulation, and reducing wound healing time.

    Device Description

    Compression sleeves are part of an external compression system, in which intermittent or sequential compression is provided using a pump/controller and limb garment. The system consists of the following three main components: a control unit, inflatable limb sleeves and conduit tubing with connectors for pump attachment. Only the compression sleeves are reprocessed.

    AI/ML Overview

    Ascent Healthcare Solutions' Reprocessed Compression Sleeves were tested against several acceptance criteria through bench and laboratory testing. The study proved that the reprocessed sleeves perform as originally intended, demonstrating safety and effectiveness compared to predicate devices.

    1. Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    BiocompatibilityPerformance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended.
    Validation of ReprocessingPerformance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended.
    Function Test(s)Performance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended.

    2. Sample Size and Data Provenance:

    The document does not specify the exact sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data). It broadly states that "Bench and laboratory testing was conducted," implying a controlled experimental environment rather than patient data.

    3. Number of Experts and Qualifications for Ground Truth:

    This information is not provided in the document. Given that the testing involved "bench and laboratory testing" rather than clinical assessment or diagnostic evaluations, the concept of "ground truth established by experts" as typically understood in AI/medical imaging studies may not directly apply. The evaluation would likely have been based on established engineering and materials science standards.

    4. Adjudication Method:

    The document does not specify an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was mentioned. The device is a reprocessed physical medical device, not a diagnostic algorithm or AI system intended to assist human readers.

    6. Standalone Performance (Algorithm Only):

    Not applicable. The device is a physical medical device (compression sleeve), not an algorithm or AI system.

    7. Type of Ground Truth Used:

    The "ground truth" for the performance testing was based on established engineering and safety standards for medical device function and material properties, as evidenced by the mention of "Biocompatibility," "Validation of reprocessing," and "Function test(s)." It was not based on expert consensus, pathology, or outcomes data in the typical sense of a diagnostic or predictive device. The goal was to show that the reprocessed device met the performance characteristics of its original, non-reprocessed counterparts.

    8. Sample Size for the Training Set:

    Not applicable. The device is a reprocessed physical medical device, not an AI model requiring a training set. The "reprocessing" itself involves a set of procedures that are validated, rather than an algorithm that is "trained."

    9. How Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device. The reprocessing process itself would be validated against established standards to ensure the device performs as intended after reprocessing, which serves a similar role to "ground truth" for the manufacturing/reprocessing process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060049

    Validate with FDA (Live)

    Date Cleared
    2006-02-24

    (49 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reprocessed Compression Sleeves are intended to help prevent deep vein thrombosis and pulmonary embolism by supplying a measured, intermittent pressure into the compression sleeves worn on the lower extremities of a recumbent patient, resulting in a gradient, sequential, repetitive squeezing and relaxing action, simulating normal muscle contractions.

    When coupled with an appropriate inflation system, compression devices are intended for use in preventing deep vein thrombosis (DVT), diminishing post-operative pain and swelling, enhancing blood circulation, and reducing wound healing time.

    Device Description

    Compression sleeves are part of an external compression system, in which intermittent or sequential compression is provided using a pump/controller and limb garment. The system consists of the following three main components: a control unit, inflatable limb sleeves and conduit tubing with detachable connections. Only the compression sleeves are reprocessed.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the Reprocessed Compression Sleeves, focusing on acceptance criteria and the supporting study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific quantitative acceptance criteria or detailed performance metrics in a tabular format. Instead, it makes a general statement about the testing performed and the conclusion reached.

    Acceptance Criterion (Implicit)Reported Device Performance
    Safety: BiocompatibilityTesting conducted to demonstrate safety.
    Effectiveness: Function as originally intendedPerformance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended.
    Effectiveness: Equivalence to predicate device functionIdentical design, materials, sizes, mechanism of action, and performance as predicate devices.
    Processing Validation: Removal of soil & decontaminationReprocessing includes removal of adherent visible soil and decontamination.
    Processing Validation: Appropriate function of componentsEach individual compression sleeve is tested for appropriate function of its components prior to packaging, labeling, and E.O. exposure.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document states that "Bench and laboratory testing was conducted," but no sample sizes for these tests are provided.
    • Data Provenance: Not specified. The document does not mention the country of origin of the data or whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. This type of device (reprocessed compression sleeves) typically relies on objective physical, chemical, and functional testing rather than expert interpretation of images or clinical data for ground truth establishment in this context.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. As noted above, the assessment of this device type does not typically involve human expert adjudication in the way medical imaging software might. The performance testing would likely involve standardized measurements and observations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study Done? No. This type of study is not relevant for this device. MRMC studies are typically used to assess the impact of AI on human reader performance, especially in diagnostic imaging.

    6. Standalone (Algorithm Only) Performance Study

    • Standalone Study Done? Not applicable. This device is not an algorithm or AI system. Its performance is assessed through physical and functional testing of the reprocessed sleeves.

    7. Type of Ground Truth Used

    The ground truth for this device's performance would be established through a combination of:

    • Objective Measurements: Physical dimensions, material integrity tests, pressure characteristics, flow rates, etc.
    • Chemical Analysis: To confirm cleanliness and absence of contaminants after reprocessing.
    • Microbiological Testing: To confirm effective decontamination/sterilization.
    • Functional Testing: To ensure the sleeve inflates and deflates correctly, holds pressure, and delivers compression as designed, mimicking original device performance.
    • Comparison to Predicate Device Specifications: The reprocessed device is intended to perform identically to the original, so the "ground truth" implicitly references the established performance characteristics of a new, predicate device.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This product does not involve machine learning or AI, and therefore no "training set" in that sense. The manufacturing process itself could be seen as "trained" through process validation, but this term is not used in the context of device performance data for this product.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set Established: Not applicable, as there is no "training set" for an AI algorithm. The methods used to validate the reprocessing procedure would involve standard engineering and quality control practices to ensure consistent and safe reprocessing outcomes.
    Ask a Question

    Ask a specific question about this device

    K Number
    K053316

    Validate with FDA (Live)

    Date Cleared
    2006-02-13

    (76 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reprocessed Compression Sleeves are intended to help prevent deep vein thrombosis and pulmonary embolism by supplying a measured, intermittent pressure into the compression sleeves worn on the lower extremities of a recumbent patient, resulting in a gradient, sequential, repetitive squeezing and relaxing action, simulating normal muscle contractions.

    When coupled with an appropriate inflation system, compression devices are intended for use in preventing deep vein thrombosis (DVT), diminishing post-operative pain and swelling, enhancing blood circulation, and reducing wound healing time.

    Device Description

    Compression sleeves are part of an external compression system, in which intermittent or sequential compression is provided using a pump/controller and limb garment. The system consists of the following three main components: a control unit, inflatable limb sleeves and conduit tubing with detachable connections. Only the compression sleeves are reprocessed.

    AI/ML Overview

    The provided text is a 510(k) summary for Reprocessed Compression Sleeves. It describes the device, its intended use, and the testing conducted to demonstrate its substantial equivalence to a predicate device. However, it does not contain the detailed information necessary to fully answer all aspects of your request, particularly regarding specific numerical acceptance criteria, detailed study designs, or expert qualifications for ground truth establishment.

    Here's a breakdown of the available information and where details are missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance (Summary)Numerical Acceptance Criteria (if available)
    BiocompatibilityDemonstrated performanceNot specified
    Function Test(s)Demonstrated performanceNot specified
    Validation of ProcessingDemonstrated performanceNot specified

    Explanation of Device Performance: The document states, "Performance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended." It also mentions, "Each individual compression sleeve is tested for appropriate function of its components prior to packaging, labeling, and EO exposure operations." However, specific numerical targets or detailed metrics for "appropriate function" are not provided.


    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for the test set in the performance studies (biocompatibility, function tests, validation of processing). It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective nature). The studies appear to be bench and laboratory tests, implying they were conducted in a controlled environment rather than involving patient data.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable and not provided. The studies described are bench and laboratory tests, not clinical studies requiring expert interpretation of medical images or patient outcomes to establish ground truth. The "ground truth" here would likely be defined by engineering specifications and objective measurements of device function and material properties.


    4. Adjudication Method for the Test Set

    This information is not applicable and not provided as the tests are not observational studies requiring adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating the performance of the reprocessed sleeves itself and its substantial equivalence to the predicate, not on human-in-the-loop performance or the effect of AI assistance.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the studies described are standalone performance tests of the device itself (biocompatibility, function, and processing validation). There is no "algorithm" in the context of this device; it is a physical medical device. The "standalone" performance refers to the device's inherent functional and material characteristics without human intervention during the test aside from operating the test equipment.


    7. The Type of Ground Truth Used

    For the performance studies (biocompatibility, function tests, validation of processing), the ground truth would be established based on:

    • Engineering Specifications/Standards: Performance metrics likely compared against established design specifications for new compression sleeves and relevant industry standards.
    • Material Science Properties: For biocompatibility, adherence to material safety standards.
    • Functional Parameters: Objective measurements of pressure, inflation/deflation cycles, integrity of components, etc., compared against the predicate device's known performance or design criteria.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. The device is a physical medical device, not an AI/machine learning algorithm, so there is no "training set" in the computational sense. The studies focused on verifying the reprocessing methods and the resulting device's performance.


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K024087

    Validate with FDA (Live)

    Date Cleared
    2003-03-11

    (90 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reprocessed Compression Sleeves are intended to help prevent deep vein thrombosis and pulmonary embolism by supplying a measured, intermittent pressure into the compression sleeves worn on the lower extremities of a recumbent patient, resulting in a gradient, sequential, repetitive squeezing and relaxing action, simulating normal muscle contractions.

    Reprocessed Compression Sleeves are indicated for use in providing external compression therapy to reduce the incidence of deep vein thrombosis and resulting pulmonary embolism in patients at risk for thrombosis formation.

    Device Description

    Compression sleeves are part of an external compression system, in which intermittent or sequential compression is provided using a pump/controller and limb garment. The system consists of the following three main components: a control unit, inflatable limb sleeves and conduit tubing with detachable connections. Only the compression sleeves are reprocessed.

    Alliance Medical Corporation's reprocessing of compression sleeves includes removal of adherent visible soil and decontamination. Each individual compression sleeve is tested for appropriate function of its components prior to packaging, labeling, and sterilization operations.

    AI/ML Overview

    The provided text is for a 510(k) premarket notification for reprocessed compression sleeves. This type of regulatory submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving novel clinical effectiveness through extensive clinical trials. Therefore, much of the typical information requested for AI/ML device studies (like sample sizes for test sets with ground truth, expert qualifications, adjudication methods, MRMC studies, and training set details) is not applicable in this context.

    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryReported Device Performance
    CleaningAdherent visible soil removed.
    DecontaminationDecontamination performed.
    FunctionalityEach individual compression sleeve is tested for appropriate function of its components. Performance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended. The design, materials, and intended use are identical to predicate devices. Mechanism of action is identical (standard mechanical design, materials, and sizes). No changes to claims, intended use, clinical applications, patient population, performance specifications, or method of operation.
    BiocompatibilityBench and laboratory testing was conducted to demonstrate biocompatibility.
    Sterilization ValidationBench and laboratory testing was conducted to demonstrate sterilization validation.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The filing is for a reprocessed medical device, where the focus is on maintaining the original device's performance through reprocessing steps. "Bench and laboratory testing" is mentioned, but specific sample sizes for these tests are not detailed. It's highly likely this was internal, retrospective testing, but the document does not specify.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable/not provided. For device reprocessing, the "ground truth" is typically the performance specifications of the original device. The testing conducted (biocompatibility, sterilization, function) are objective engineering and laboratory tests, not subjective expert assessments.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. Adjudication methods are relevant for subjective assessments (e.g., image interpretation by multiple readers) to establish a consensus ground truth. The tests for reprocessed sleeves are objective performance measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This is not an AI/ML device, and no MRMC study was conducted or required for this type of submission.

    6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done

    This information is not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the reprocessed compression sleeves is based on:

    • Original Device Specifications: The reprocessed devices are compared against the performance of the legally marketed predicate devices, ensuring "identical" design, materials, intended use, and mechanism of action. The implicitly accepted "ground truth" is that the original, new device performed safely and effectively.
    • Objective Performance Metrics: For the specific tests mentioned (biocompatibility, sterilization validation, function tests), the ground truth would be established by validated test methods and predefined quantitative specifications inherent to those tests (e.g., sterility assurance levels, material integrity, pressure output ranges).

    8. The sample size for the training set

    This information is not applicable/not provided. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    This information is not applicable/not provided. As there is no training set for an AI/ML model, this question is irrelevant to this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K024074

    Validate with FDA (Live)

    Date Cleared
    2003-03-07

    (87 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reprocessed Compression Sleeves are intended to help prevent deep vein thrombosis and pulmonary embolism by supplying a measured, intermittent pressure into the compression sleeves worn on the lower extremities of a recumbent patient, resulting in a gradient, sequential, repetitive squeezing and relaxing action, simulating normal muscle contractions.

    Reprocessed Compression Sleeves are indicated for use in providing external compression therapy to reduce the incidence of deep vein thrombosis and resulting pulmonary embolism in patients at risk for thrombosis formation.

    Device Description

    Compression sleeves are part of an external compression system, in which intermittent or sequential compression is provided using a pump/controller and limb garment. The system consists of the following three main components: a control unit, inflatable limb sleeves and conduit tubing with detachable connections. Only the compression sleeves are reprocessed.

    AI/ML Overview

    The provided text describes the 510(k) summary for "Reprocessed Compression Sleeves" by Alliance Medical Corporation. This submission focuses on demonstrating substantial equivalence to a predicate device rather than a new AI-powered diagnostic device. Therefore, many of the requested categories for AI-specific studies (like sample size, ground truth, expert adjudication, MRMC studies, standalone performance, and training set details) are not applicable or not detailed in this document.

    However, I can extract the relevant information regarding acceptance criteria and the studies performed to meet them for this reprocessed medical device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Performance Tests)Reported Device Performance (Conclusion)
    BiocompatibilityPerformance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended. Alliance Medical Corporation concludes that the modified devices (Reprocessed Compression Sleeves) are safe, effective and substantially equivalent to the predicate devices.
    Sterilization ValidationPerformance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended. Alliance Medical Corporation concludes that the modified devices (Reprocessed Compression Sleeves) are safe, effective and substantially equivalent to the predicate devices.
    Function Test(s)Each individual compression sleeve is tested for appropriate function of its components prior to packaging, labeling, and sterilization operations. Performance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended. Alliance Medical Corporation concludes that the modified devices (Reprocessed Compression Sleeves) are safe, effective and substantially equivalent to the predicate devices.
    Packaging ValidationsPerformance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended. Alliance Medical Corporation concludes that the modified devices (Reprocessed Compression Sleeves) are safe, effective and substantially equivalent to the predicate devices.
    Overall Safety and Effectiveness"Alliance Medical Corporation concludes that the modified devices (Reprocessed Compression Sleeves) are safe, effective and substantially equivalent to the predicate devices as described herein."

    2. Sample size used for the test set and the data provenance:

    • The document states that "Bench and laboratory testing was conducted." It does not specify the sample sizes used for these tests.
    • Data provenance is not explicitly mentioned (e.g., country of origin, retrospective/prospective), as the testing appears to be internal laboratory and bench testing of the reprocessed sleeves themselves.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable as this is not a diagnostic device relying on expert-established ground truth from patient data. The ground truth for performance tests (like function, sterility) would be established by validated test methods and specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring expert adjudication of patient cases.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI-powered device or a study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm or AI device. The "standalone" performance refers to the reprocessed sleeve's ability to meet its functional and safety specifications. The document states that "Each individual compression sleeve is tested for appropriate function of its components prior to packaging, labeling, and sterilization operations."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for this device's performance is based on established specifications and validated test methods for biocompatibility, sterilization, functionality, and packaging integrity. For example, for "Function Test(s)", the ground truth would be that the sleeve performs according to its original design specifications (e.g., pressure output, inflation/deflation cycles).

    8. The sample size for the training set:

    • Not applicable. This is not an AI device trained on data.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K021654

    Validate with FDA (Live)

    Date Cleared
    2003-02-05

    (261 days)

    Product Code
    Regulation Number
    870.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reprocessed Compression Sleeves are intended to help prevent deep vein thrombosis and pulmonary embolism by supplying a measured, intermittent pressure into the compression sleeves worn on the lower extremities of a recumbent patient, resulting in a gradient, sequential, repetitive squeezing and relaxing action, simulating normal muscle contractions.

    Reprocessed Compression Sleeves are indicated for use in providing external compression therapy to reduce the incidence of deep vein thrombosis and resulting pulmonary embolism in patients at risk for thrombosis formation.

    Device Description

    Compression sleeves are part of an external compression system, in which intermittent or sequential compression is provided using a pump/controller and limb garment. The system consists of the following three main components: a control unit, inflatable limb sleeves and conduit tubing with detachable connections.

    Alliance Medical Corporation's reprocessing of compression sleeves includes removal of adherent visible soil and decontamination. Each individual compression sleeve is tested for appropriate function of its components prior to packaging and labeling operations.

    AI/ML Overview

    The provided document is a 510(k) summary for a reprocessed medical device, specifically compression sleeves, and does not contain the detailed information typically found in a study report for novel device performance or AI/software validation. Therefore, many of the requested elements for acceptance criteria and study details cannot be fully extracted.

    Based on the available text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Criteria (Inferred)Reported Device Performance (as stated)
    Functional EquivalencePerform "as originally intended" (of predicate devices)"Performance testing demonstrates that Reprocessed Compression Sleeves perform as originally intended."
    SafetyNo adverse impact on patient safety from reprocessingImplied by "demonstrate performance (safety and effectiveness)" and substantial equivalence conclusion
    EffectivenessMaintain original therapeutic effect from reprocessingImplied by "demonstrate performance (safety and effectiveness)" and substantial equivalence conclusion
    Material IntegrityMaintain design, materials, and mechanical characteristics"The design, materials, and intended use of Reprocessed Compression Sleeves are identical to the predicate devices."
    Component FunctionIndividual compression sleeve components function appropriately"Each individual compression sleeve is tested for appropriate function of its components prior to packaging and labeling operations."
    Cleanliness/DecontaminationRemoval of adherent visible soil and decontamination"Alliance Medical Corporation's reprocessing of compression sleeves includes removal of adherent visible soil and decontamination."

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified. The document only mentions "Bench and laboratory testing was conducted." It does not provide the number of reprocessed sleeves tested.
    • Data Provenance: Not specified. It's bench and laboratory testing, so it's not patient data from a specific country, nor is it explicitly stated as retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This was bench testing, not a study requiring expert clinical assessment for ground truth.

    4. Adjudication method for the test set:

    • Not applicable. This was bench testing, not a study requiring adjudication of expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a 510(k) for a reprocessed physical medical device (compression sleeves), not an AI-powered diagnostic or assistive technology.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. This device is not an algorithm.

    7. The type of ground truth used:

    • The ground truth for the performance claim appears to be the original intended performance of the predicate device. The testing confirms that the reprocessed device performs to these established specifications.

    8. The sample size for the training set:

    • Not applicable. This is a reprocessed physical device, not an AI/ML algorithm requiring a training set. The "reprocessing" procedure could be considered analogous to a "training" process for maintaining device specifications, but it's not a data-driven training set in the AI sense.

    9. How the ground truth for the training set was established:

    • Not applicable. (See #8)