510(k) Data Aggregation

    K Number
    K140742

    Validate with FDA (Live)

    Date Cleared
    2014-04-25

    (31 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reliance Anterior Cervical Plate System is indicated for stabilization of the anterior cervical spine from C2 to C7 employing unicortical screw fixation at the anterior face of the vertebral bodies. Specific clinical indications for anterior plating include:

    • instability caused by trauma or fracture;
    • instability associated with correction of cervical lordosis and kyphosis deformity;
    • instability associated with pseudoarthosis as a result of previously failed cervical spine surgery;
    • instability associated with major reconstructive surgery for primary tumors or metastatic malignant tumors of the cervical spine;
    • instability associated with single or multiple level corpectomy in advanced degenerative disk disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), spinal canal stenosis and cervical myelopathy.
      This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.
    Device Description

    The Reliance Anterior Cervical Plate System device is intended to be used as an anterior cervical plate device. The Reliance Anterior Cervical Plate System is comprised of implant and instrument components. The Reliance Anterior Cervical Plate implant device is manufactured from Titanium alloy as specified by ASTM F-136. The Reliance Anterior Cervical Plate is a combination of the plate, cover plate, set screw, and bone screw components. The cover plate is attached to the plate by means of the set screw.

    AI/ML Overview

    The provided text describes the 510(k) submission for the Reliance Anterior Cervical Plate System. However, it does not contain any information regarding a study involving artificial intelligence (AI) or a specific AI device.

    Instead, the document details a mechanical device (an anterior cervical plate system) and its substantial equivalence to a predicate device based on:

    • Mechanical Testing: Performed according to ASTM F-1717 (static compression bending, static torsion, and dynamic compression bending).
    • Sterilization and Biocompatibility.
    • Risk Analysis.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves an AI device meets its acceptance criteria using the provided input, as the input clearly pertains to a medical implant (hardware) and not an AI-driven software or device.

    The questions regarding AI performance criteria, sample sizes for AI test sets, expert ground truth establishment, MRMC studies, standalone algorithm performance, AI training sets, and their ground truth methods are not applicable to this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122216

    Validate with FDA (Live)

    Date Cleared
    2012-09-05

    (42 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reliance Anterior Cervical Plate System is indicated for stabilization of the anterior cervical spine from C2 to C7 employing unicortical screw fixation at the anterior face of the vertebral bodies. Specific clinical indications for anterior plating include:

    • instability caused by trauma or fracture; .
    • instability associated with correction of cervical lordosis and kyphosis deformity; ●
    • . instability associated with pseudoarthosis as a result of previously failed cervical spine surgery;
    • instability associated with major reconstructive surgery for primary tumors or metastatic . malignant tumors of the cervical spine;
    • instability associated with single or multiple level corpectomy in advanced degenerative . · disk disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), spinal canal stenosis and cervical myelopathy.

    This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

    Device Description

    The Reliance Anterior Cervical Plate System device is intended to be used as an anterior cervical plate device. The Reliance Anterior Cervical Plate System is comprised of implant and instrument components. The Reliance Anterior Cervical Plate implant device is manufactured from Titanium alloy as specified by ASTM F-136. The Reliance Anterior Cervical Plate is a combination of the plate, cover plate, set screw, and bone screw components. The cover plate is attached to the plate by means of the set screw.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Study Type)Reported Device Performance
    Static Compression Bending (ASTM F-1717)Substantially equivalent to predicate devices
    Static Torsion (ASTM F-1717)Substantially equivalent to predicate devices
    Dynamic Compression Bending (ASTM F-1717)Substantially equivalent to predicate devices
    SterilizationSubstantially equivalent to predicate devices
    BiocompatibilitySubstantially equivalent to predicate devices

    2. Sample size used for the test set and the data provenance:

    The provided document describes mechanical and other non-clinical performance testing. It does not mention a "test set" in the context of clinical data, human subjects, or data provenance from specific regions. The testing was conducted on samples of the Reliance Anterior Cervical Plate System itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    Not applicable. This was a non-clinical performance study involving mechanical testing and assessments of material properties, not a study requiring expert clinical review to establish ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods are relevant for clinical studies where expert consensus is needed to establish ground truth. This document describes non-clinical performance testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a spinal implant, not an AI-assisted diagnostic or imaging device, so an MRMC study is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical medical implant, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this performance study was established through standardized mechanical testing protocols (ASTM F-1717) and assessments of material properties (sterilization, biocompatibility). The comparison was made against the performance of legally marketed predicate devices.

    8. The sample size for the training set:

    Not applicable. This document describes non-clinical performance testing of a physical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as no training set for a machine learning model was used.