Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K072012

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-08-22

    (30 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reflexion Spiral catheter can be used for recording intracardiac signals and for cardiac stimulation during electrophysiology studies. The Reflexion Spiral catheter is to be used to map the atrial regions of the heart.

    Device Description

    The St. Jude Medical Reflexion Spiral™ Bi-directional Variable Radius Mapping Catheter (Reflexion Spiral™ Decapolar EP Catheter) is a flexible, bi-directional, variable radius loop electrophysiology catheter constructed of a polymer shaft that incorporates platinum electrodes.

    The Reflexion Spiral™ Decapolar EP Catheter has a loop and a proximal handle (the ComfortGrip™ handle) that contains: 1) A shaft actuator mechanism for varying the distal portion of the shaft; 2) A loop actuator mechanism for varying the loop diameter from approximately 20mm to approximately 15mm; and 3) An electrical connector.

    AI/ML Overview

    This 510(k) summary describes a medical device, the Reflexion Spiral™ Variable Radius Catheter, which is an electrode recording catheter used for cardiac mapping. The submission primarily focuses on demonstrating substantial equivalence to a predicate device (Reflexion Spiral™ Variable Radius Catheter K062251) through non-clinical bench testing.

    Based on the provided text, there is no acceptance criteria or detailed study information presented in the format requested. The document primarily focuses on establishing substantial equivalence to a predicate device through bench testing rather than demonstrating performance against specific numerical acceptance criteria.

    Therefore, many of the requested fields cannot be filled.

    Here's an attempt to extract what is available and note what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (e.g., Sensitivity, Specificity, Accuracy, Mechanical Properties)Reported Device Performance
    Not explicitly stated as numerical criteria. The general criteria are implied to be "meets product specifications and intended uses" and "does not adversely affect the safety and effectiveness" compared to the predicate."Results of the testing demonstrates that the Reflexion Spiral™ Variable Radius Catheter design meets product specifications and intended uses."
    Functional equivalence to predicate device in terms of recording intracardiac signals and cardiac stimulation.Claimed substantial equivalence; "same intended use and fundamental scientific technology."
    BiocompatibilityClaimed substantial equivalence.
    SterilizationClaimed substantial equivalence.
    PackagingClaimed substantial equivalence.
    LabelingClaimed substantial equivalence.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: Not specified. The study is described as "Bench testing," which implies testing of physical devices/components, but the number of units tested is not provided.
    • Data Provenance: Not applicable in this context; it's bench testing, not human subject data.

    3. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This was bench testing, not an expert-driven performance evaluation.

    4. Adjudication method for the test set

    • Not applicable. This was bench testing, not a study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is an electrode recording catheter, not an AI software. An MRMC study is not relevant to its stated purpose or evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This device is a physical catheter, not an algorithm.

    7. The type of ground truth used

    • For bench testing, the "ground truth" would be established by engineering specifications, validated test methods, and potentially comparisons against known performance characteristics of the predicate device. Details are not provided.

    8. The sample size for the training set

    • Not applicable. This device is not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This device is not an AI model that requires a training set.

    Summary of Study:

    The primary study mentioned is "Summary of Non-Clinical Testing: Bench testing."

    • Objective: To support substantial equivalence to the predicate device (Reflexion Spiral™ Variable Radius Catheter K062251).
    • Methodology: Bench testing was performed to evaluate the Reflexion Spiral™ Variable Radius Catheter.
    • Findings: The testing demonstrated that "the Reflexion Spiral™ Variable Radius Catheter design meets product specifications and intended uses." It also showed that "differences [between proposed and predicate device] do not adversely affect the safety and effectiveness."
    • Conclusion: The device is substantially equivalent to the predicate.

    This document is a typical 510(k) submission for a Class II medical device, which often relies on demonstrating substantial equivalence to a predicate device through non-clinical (bench) testing, rather than extensive clinical trials with detailed performance metrics and human-in-the-loop studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062251

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-10-20

    (78 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reflexion Spiral™ catheter can be used for recording intracardiac signals and for cardiac stimulation during electrophysiology studies. The Reflexion Spiral™ catheter is to be used to map the atrial regions of the heart.

    Device Description

    The St. Jude Medical (SJM) Reflexion Spiral™ Variable Radius Catheter (Reflexion Spiral catheter) is a flexible, asymmetric, bi-directional, radiopaque variable radius loop electrophysiology catheter constructed of a polymer shaft that incorporates platinum electrodes.

    The Reflexion Spiral catheter has a proximal handle (ComfortGrip™) that contains: 1) A shaft actuator mechanism for varying the asymmetrical sweep (90° sweep) and curl (180° curl) of the distal portion of the shaft. 2) A loop actuator mechanism for varying the loop diameter from approximately 25mm to approximately 15mm. 3) An electrical connector fitted into the proximal end of the handle.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document does not explicitly state specific acceptance criteria in terms of numerical performance targets. Instead, it relies on demonstrating equivalence to a predicate device. The performance is described in a comparative manner rather than against pre-defined numerical thresholds.

    Criterion TypeAcceptance Criteria (as implied)Reported Device Performance
    Intended UseThe device must be suitable for recording intracardiac signals and cardiac stimulation during electrophysiology studies, specifically for mapping atrial regions."The Reflexion Spiral catheter can be used for recording intracardiac signals and for cardiac stimulation during electrophysiology studies. The Reflexion Spiral catheter is to be used to map the atrial regions of the heart." (Matches intended use)
    Equivalence to Predicate DeviceEquivalent performance in loop radius control and sweep/curl positioning control compared to the predicate device."The different proximal handles show equivalent performance of the loop radius control and sweep/curl positioning control and does not affect the intended use or the scientific technology of the device."
    Safety and EffectivenessDoes not raise new issues of safety, effectiveness, or performance."The Reflexion Spiral catheter does not raise any new issues of safety, effectiveness or performance of the device."
    Conformance to StandardsConformance to FDA Guidance "Electrode Recording Catheter Preliminary Guidance, Draft Version," March 1995 and ISO 10555-1."The test results indicate conformance to the standards and reliable performance when used in conformance with the device Instructions for Use."
    Technological CharacteristicsThe technological characteristics should be comparable to the predicate device, not affecting intended use or scientific technology."The Reflexion Spiral catheter and the predicate IBI Inquiry™ Optima™ catheter are both intended for electrogram recording and stimulation during electrophysiological studies. The different proximal handles show equivalent performance...and does not affect the intended use or the scientific technology of the device." (Confirms comparability and functional equivalence despite design differences)

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: The document does not specify a patient or clinical sample size for the test set. The testing described is primarily non-clinical (bench and animal data).
    • Data Provenance: The data comes from bench testing and animal studies conducted by St. Jude Medical. No information on country of origin of data or whether it was retrospective or prospective is provided for the non-clinical tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Number of Experts: This information is not provided in the document.
    • Qualifications of Experts: This information is not provided in the document.

    Given that the studies were primarily non-clinical (bench and animal), "ground truth" would likely refer to engineering specifications or physiological measurements in the animal model, rather than expert interpretation of clinical data.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set:

    • Adjudication Method: Not applicable/Not mentioned. Since the studies were non-clinical (bench and animal) and no expert review of clinical cases is described, there would be no need for a human adjudication method for establishing ground truth as would be relevant in a clinical imaging or diagnostic study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. The device in question is a physical medical catheter, not an AI-assisted diagnostic tool for human readers. Therefore, the concept of human readers improving with AI assistance is not relevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Standalone Performance: Not applicable/Not done in the context of an algorithm. The device is a physical catheter, not an algorithm. The "standalone performance" was demonstrated through bench testing and animal data to show its functional capabilities and safety, independent of human interpretive assistance (as it's a tool for direct physiological measurement/stimulation).

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    • Ground Truth Type: For the non-clinical tests, the "ground truth" would be established by:
      • Engineering Specifications: For bench testing of loop radius control, sweep/curl positioning, and electrical characteristics.
      • Physiological Measurements/Observation: In animal studies, to ensure the catheter can record intracardiac signals and provide cardiac stimulation as intended, and to assess biocompatibility and safety.
      • Compliance with Standards: Meeting the requirements outlined in FDA Guidance and ISO 10555-1.

    8. The Sample Size for the Training Set:

    • Sample Size for Training Set: Not applicable. As this is a physical medical device and not an AI/machine learning model, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    • Ground Truth for Training Set Establishment: Not applicable. There is no training set for this device.

    Summary of Study and Device Performance:

    The study described for the Reflexion Spiral™ Variable Radius Catheter is a non-clinical study primarily consisting of:

    • Bench Testing: To evaluate functional performance aspects like loop radius control and sweep/curl positioning, comparing them to the predicate device.
    • Animal Data: Likely to assess in-vivo performance, safety, and biocompatibility.

    The purpose of these studies was to demonstrate substantial equivalence to a currently marketed predicate device (Irvine Biomedical, Inc., (IBI) Inquiry™ Optima™ Steerable Electrophysiology Catheter K042775) and conformance to relevant industry standards (FDA Guidance and ISO 10555-1). The document asserts that the device achieves functional equivalence to the predicate and does not introduce new safety or effectiveness concerns, thereby meeting the requirements for 510(k) clearance by the FDA.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1