Found 2 results

510(k) Data Aggregation

    K Number
    K031702

    Validate with FDA (Live)

    Date Cleared
    2003-08-08

    (67 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reflex™ ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (including fractures), tumors, deformities or curvatures (including kyphosis, lordosis or scoliosis), pseudoarthrosis, failed previous fusion, decompression of the spinal cord following total or partial cervical vertebrectomy, spondylolisthesis and spinal stenosis.

    WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

    Device Description

    The Reflex " ACP System consists of bone plates and screws. Both components are available in a variety of lengths in order to accommodate patient physiology. The components are fabricated from titanium alloy. The implants will be provided non-sterile.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Reflex™ Anterior Cervical Plate System, focusing on adding indications to the device. However, it does not contain the detailed information necessary to answer all the questions regarding acceptance criteria and a study proving the device meets those criteria, especially in the context of AI/ML or a clinical study with human readers and ground truth.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Synthes' Cervical Spine Locking Plate (CSLP) System) through mechanical properties.

    Here's an attempt to answer the questions based only on the provided text, highlighting where information is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical PropertiesReflex ACP System demonstrated comparable mechanical properties to the predicate device (Synthes' CSLP System).
    Intended Use EquivalenceSubstantially equivalent for the stated indications to the predicate device.
    Material EquivalenceBoth components (bone plates and screws) fabricated from titanium alloy, similar to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the document. The study described is a mechanical test comparing the Reflex ACP System to a predicate device. There is no mention of a "test set" in the context of patient data, nor any data provenance (country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided. Ground truth in the context of clinical outcomes or expert consensus is not relevant to the type of study (mechanical testing) described.

    4. Adjudication Method for the Test Set

    This information is not provided. Adjudication methods are typically used in clinical studies involving interpretation or outcome assessment, which is not the case here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission focuses solely on demonstrating substantial equivalence through mechanical testing, not on clinical performance or the impact of AI assistance on human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone algorithm performance study was not done. This device is a physical medical implant (anterior cervical plate system), not an AI/ML software device.

    7. The Type of Ground Truth Used

    The "ground truth" used for this device's evaluation was primarily the mechanical performance and material characteristics of the legally marketed predicate device (Synthes' Cervical Spine Locking Plate (CSLP) System). The stated criterion is "comparable mechanical properties" to the predicate.

    8. The Sample Size for the Training Set

    This information is not provided. The concept of a "training set" is not applicable to the mechanical testing described for this physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided. The concept of "ground truth for a training set" is not applicable. The basis for comparison was the established mechanical properties of the predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K010115

    Validate with FDA (Live)

    Date Cleared
    2001-04-16

    (90 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reflex Anterior Cervical Plate (ACP) System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (including fractures), tumors, deformities or curvatures (including kyphosis, lordosis or scoliosis), pseudoarthrosis, failed previous fusion and decompression of the spinal cord following total or partial cervical vertebrectomy.

    WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

    Device Description

    The Reflex ACP System consists of a bone plates (an anti-migration mechanism is preassembled to plates) and screws. Both components are available in a variety of lengths in order to accommodate patient physiology. The components are fabricated from titanium alloy as described in ASTM F-136, ISO 5832-3 and ASTM F-1813. The implants will be provided nonsterile.

    AI/ML Overview

    The provided text (K010115) is a 510(k) Premarket Notification for the Reflex Anterior Cervical Plate System. This document is a regulatory submission for a medical device and describes its intended use, design, and claim of substantial equivalence to already marketed devices.

    However, the 510(k) summary does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of diagnostic performance or clinical effectiveness studies as typically found for AI/imaging devices. Instead, it focuses on demonstrating substantial equivalence through similarities in materials, design, and intended use, and indicates "Testing has been conducted demonstrating substantial equivalence to a currently marked device." Such testing for an implantable device like an anterior cervical plate system would typically involve non-clinical performance testing (e.g., mechanical strength, fatigue, biocompatibility) rather than a clinical study with human readers or AI algorithms.

    Therefore, most of the requested information regarding clinical studies, ground truth, expert opinions, sample sizes for test/training sets, or MRMC studies is not applicable to this 510(k) document.

    Here's a breakdown of the information that can be extracted or inferred, and what is not present in the document based on your questions:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria TypeAcceptance Criteria (Not Explicitly Stated for Clinical Performance)Reported Device Performance (Inferred/Implied)
    Substantial EquivalenceTo a legally marketed predicate device (materials, design, intended use)."Equivalency of this device is based on similarities in intended use, materials, and design to other currently marketed anterior cervical plate systems."
    Mechanical/Biocompatibility Performance(Not explicitly detailed, but implied through standard medical device testing for implants)."Testing has been conducted demonstrating substantial equivalence to a currently marked device." (This would typically include bench testing for mechanical properties and biocompatibility).
    Clinical Performance (e.g., accuracy, sensitivity, specificity, patient outcomes)(Not applicable for this type of 510(k), as it's not a diagnostic or AI device and no clinical efficacy study is described).Not reported in this document.

    2. Sample size used for the test set and the data provenance

    • Not applicable for a clinical test set. The "testing" mentioned refers to non-clinical performance (bench testing) to demonstrate substantial equivalence, not a clinical study with a patient test set.
    • The document does not provide details on the specifics of this non-clinical testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This device is an implant, not a diagnostic tool requiring expert interpretation or ground truth establishment in a clinical sense.

    4. Adjudication method for the test set

    • Not applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI or diagnostic imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is not an AI or algorithm device.

    7. The type of ground truth used

    • Not applicable in the context of clinical "ground truth." For substantial equivalence of an implant, the "ground truth" equates to the established performance characteristics of the predicate device, which would be based on its intended use, materials, design, and previously demonstrated safety and effectiveness (likely through non-clinical testing and clinical history).

    8. The sample size for the training set

    • Not applicable. This device does not involve an AI algorithm or a training set.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.

    Summary of what the 510(k) document states about "proving" the device meets criteria:

    The document states:

    • "Equivalency of this device is based on similarities in intended use, materials, and design to other currently marketed anterior cervical plate systems."
    • "Testing has been conducted demonstrating substantial equivalence to a currently marked device."

    This means the device's "acceptance criteria" revolve around demonstrating that it is substantially equivalent to a device already on the market (the predicate device) in terms of its intended use, design, materials, and performance (likely mechanical and biocompatibility). The study proving this involves comparing these aspects of the Reflex Anterior Cervical Plate System to a predicate device and conducting relevant non-clinical tests to show similar performance characteristics. The specific details of these non-clinical tests (e.g., fatigue testing, tensile strength, biocompatibility studies) are not disclosed in this 510(k) summary but would be part of the full 510(k) submission.