510(k) Data Aggregation

    K Number
    K152223

    Validate with FDA (Live)

    Date Cleared
    2016-04-29

    (266 days)

    Product Code
    Regulation Number
    868.5470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The following indications which are listed on the Undersea and Hyperbaric Medical Society (UHMS) website: www.uhms.org are approved uses of hyperbaric oxygen therapy as defined by the Hyperbaric Oxygen Therapy Committee.

    1. Air or Gas Embolism
    2. Carbon Monoxide Poisoning
      a. Carbon Monoxide Poisoning Complicated by Cyanide Poisoning
    3. Clostridial Myositis and Myonecrosis (Gas Gangrene)
    4. Crush Injury, Compartment Syndrome and Other Acute Ischemias
    5. Decompression Sickness
    6. Arterial Insufficiencies
      a. Central Retinal Artery Occlusion
      b. Enhancement of Healing in Selected Problem Wounds
    7. Severe Anemia
    8. Intracranial Abscess
    9. Necrotizing Soft Tissue Infections
    10. Osteomyelitis (Refractory)
    11. Delayed Radiation Injury (Soft Tissue and Bony Necrosis)
    12. Compromised Grafts and Flaps
    13. Acute Thermal Burn Injury
    Device Description

    The OxyHeal® 5000, Rectangular Multiplace Hyperbaric Chamber System is comprised of a multiplace hyperbaric chamber and a number of major subsystems that support the overall system operation, control, and monitoring. An OxyHeal® 5000 multiplace hyperbaric chamber is a pressure vessel for human occupancy that is designed in a rectangular geometry. Configurations vary based on the needs of the end user and may consist of two (2), three (3), or more compartments. Capacities may range anywhere from six (6) to twenty-four (24) patients or more, dependent on chamber of compartments, or the direction of the customer to meet their needs. Lastly, maximum operating pressures range from 3ATA (~30psi) to 6ATA (~73.5psi), with each of the compartments designed to operate independently. Each OxyHeal® 5000 multiplace hyperbaric chamber is designed, inspected, tested, marked and stamped to meet the standards defined in the American Society of Mechanical Engineers / Pressure Vessel for Human Occupancy (ASME/PVHO-1). These chambers comply with the National Fire Protection Agency (NFPA) 99, Health Care Facilities. The OxyHeal 5000, Rectangular Multiplace Hyperbaric Chamber System consists of the hyperbaric chamber itself and the major subsystems briefly described below: Compressed Air System, Fire Suppression System, Bulk Oxygen System, Built-in Breathing System, HP Gas System, Environmental Control System (ECS), Control Console.

    AI/ML Overview

    The provided text details the 510(k) summary for the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family, demonstrating its substantial equivalence to predicate devices rather than proving it meets specific acceptance criteria through a "study" in the typical sense of a clinical trial or performance study comparing it against a predefined metric.

    Instead, the "study" mentioned here refers to a comparison of the new device's technological characteristics, design, and operational parameters against those of legally marketed predicate devices. The acceptance criteria are implicit in meeting or exceeding the standards and performance of these predicate devices, as well as adherence to relevant industry standards like ASME and NFPA.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence submission, the "acceptance criteria" are the characteristics of the predicate devices and relevant standards (ASME, NFPA). The "reported device performance" is the OxyHeal® 5000's characteristics in comparison to these.

    CharacteristicAcceptance Criteria (Predicate Devices & Standards)Reported Device Performance (OxyHeal® 5000)
    Product CodeCBF (Fink K031649, OxyHeal® 2000 K011866)CBF
    Regulation Number21 CFR 868.5470 (Fink K031649, OxyHeal® 2000 K011866)21 CFR 868.5470
    Regulation NameHyperbaric Chamber (Fink K031649, OxyHeal® 2000 K011866)Hyperbaric Chamber
    Indications for UseAs defined in the Hyperbaric Oxygen Therapy Committee Report, dated 1999 (for predicates)As defined in the Hyperbaric Oxygen Therapy Committee Report, dated 2008 (Substantially equivalent, updated report)
    Hyperbaric Chamber Code DesignASME Section VIII, Div. 1; ASME PVHO-1 (for predicates)ASME: Boiler and Pressure Code; ASME PVHO-1: Safety Standard for Pressure Vessels for Human Occupancy (Substantially equivalent)
    Hyperbaric Chamber System DesignNFPA 99, Chapter 19 – Hyperbaric Facilities (for predicates)NFPA 99, Chapter 14 – Hyperbaric Facilities (Substantially equivalent – different chapter but same core standard)
    Operating Pressure3.0 ATA (OxyHeal® 2000) to 6.0 ATA (Fink DL8/TL20)3.0ATA – 6.0ATA (Substantially equivalent to Fink DL8/TL20; OxyHeal 2000 is lower pressure range)
    Operating Temperature62°F - 100°F (Fink); Capable of operating within design temperature ranges based on patient comfort (OxyHeal® 2000)50°F-125°F (Substantially equivalent, OxyHeal 2000 designed for lower pressure range)
    Design Temperature62°F - 100°F (Fink); 50°F - 120°F (OxyHeal® 2000)50°F - 125°F (Substantially equivalent)
    Design Pressure30psig (OxyHeal® 2000) to 80.0 psi (Fink DL8/TL20)30psig - 75psig (Substantially equivalent to Fink DL8/TL20; OxyHeal 2000 for lower pressure)
    Design Life≥ 30 years (Fink, OxyHeal® 2000)90,000 cycles or 60 years, whichever happens first (≥ 30 years: Substantially equivalent)
    Hydrostatic Pressure45psi (OxyHeal® 2000) to 104.0 psi (Fink)39psi - 97.5psi (Substantially equivalent to Fink DL8/TL20; OxyHeal 2000 for lower pressure)
    Inspection AuthorityASME "U" Stamp (Fink); Independent 3rd Party ASME Authorized Inspector (AI) (OxyHeal® 2000)Independent 3rd Party ASME Authorized Inspector (AI). Affix ASME Stamp on chamber data plate (Substantially equivalent)
    Weight (lbs.)~ 7,000 (OxyHeal® 2000) to ~110,231 (Fink)15,000lbs to 120,000lbs (Substantially equivalent to Fink DL8/TL20; OxyHeal 2000 is smaller)
    Dimensions (Main Compartment)8.1'W x 7'H x 11'L to 10.3'W x 7'H x 19.3'L (Fink); Dia. = 6'; H = 7' (OxyHeal® 2000)Min: 8' W x 7' H x 10'L; Max: 11'W x 8'H x 20'L (Substantially equivalent to Fink DL8/TL20; OxyHeal 2000 intentionally smaller)
    Total Volume197.8ft³ (OxyHeal® 2000) to 2,546.2ft³ (Fink)600ft³ to 2,600ft³ (Substantially equivalent to Fink DL8/TL20; OxyHeal 2000 intentionally smaller)
    Medical Lock10" ID x 10"L (OxyHeal® 2000) to 13.8" x 13.8" x 19.7" (Fink)Cylindrical Min: 10 inch diameter; Max: 16 inch diameter (Substantially equivalent)
    Main Doorway Size32" x 83" (OxyHeal® 2000) to 39.4" x 75.6" (Fink)Minimum door frame size: 44" x 80"; Maximum 52" x 80" (Substantially equivalent in terms of safety and effectiveness of operation)
    LightingOne (1) internally mounted light (OxyHeal® 2000) to Six (6) to fourteen (14) external dimmable lights (Fink)LED lights Min: 4; Max: 15 (Comply with NFPA-99 and are substantially equivalent in terms of safety and effectiveness of operation)
    Viewports (PVHO-1)16" ID - Qty. 2 each (OxyHeal® 2000) to various (Fink)Minimum: One (1) per hyperbaric chamber. Maximum: Six (6) per compartment. Minimum: 8" Diameter, Maximum: 30" Diameter (Substantially equivalent)
    Capacity Main CompartmentFrom eight (8) seated patients / Four (4) wheelchairs (Fink) to Six (6) upright seated patients (OxyHeal® 2000)4 Patients Up to 24 Patients (Substantially equivalent to Fink DL8/TL20; OxyHeal 2000 is smaller)
    Fire SuppressionIAW NFPA 99 (Fink, OxyHeal® 2000)IAW NFPA 99 (Substantially equivalent)
    Life Support ControlsManual electropneumatic (pressurization/depressurization) (Fink); Automatic pressurization & depressurization with manual back-up from both inside and outside the chamber (OxyHeal® 2000)Automatic pressurization & depressurization with manual back-up from both inside and outside each chamber compartment (Substantially equivalent to OxyHeal 2000)
    Environmental ControlHeating (Fink); Heating and cooling (OxyHeal® 2000)Heating & cooling (Substantially equivalent to OxyHeal 2000)
    VentilationConstant air flow (Fink); Automatic chamber ventilation with manual back-up (OxyHeal® 2000)Automatic chamber ventilation with manual back-up. Min 6 cfm, Max 48cfm with a +/- 1 fsw stability (Substantially equivalent)
    BIBS with Overboard DumpTwo (2) - Four (4) (Fink); Four (4) (OxyHeal® 2000)Four (4) to Twenty-Four (24), on demand gas delivery (Substantially equivalent in terms of safety and effectiveness)
    Hoods with Overboard DumpSix (6) to Twenty two (22) (Fink); Four (4) (OxyHeal® 2000)Four (4) to Twenty-Four (24) 1-100LPM delivery flow meters. Minimum flow for hoods: 40-lpm (Substantially equivalent in terms of safety and effectiveness)
    Depth MeasurementDigital with analog backup (Fink, OxyHeal® 2000)Digital with analog backup (Substantially equivalent)
    Gas AnalysisOxygen (O2) and carbon dioxide (CO2) (Fink, OxyHeal® 2000)Oxygen (O2) 1-100% and carbon dioxide (CO2) 0-5000ppm (Substantially equivalent)
    CommunicationsInternal/external PA system, Sound powered backup (Fink); Primary: Wireless telephone, Secondary: Intercom (OxyHeal® 2000)Primary: Wireless telephone, Secondary: Intercom, Tertiary: Sound powered backup (Substantially equivalent)
    EntertainmentIndividual 4-channel selection for eight (8) persons (Fink); Individual 4-channel selection for up to six (6) persons (OxyHeal® 2000)Individual up to 4-channels. Varies based on customer specification for number of compartments and number of patients to be treated (Substantially equivalent)
    TV SystemExternal color with remote control AM/FM tuner/CD & DVD player (Fink); One (1) internally mounted TV System. AM/FM tuner/CD & DVD player (OxyHeal® 2000)24VDC LED TV Monitor. AM/FM Tuner/CD, DVD, and Cable TV Tuner (Comply with NFPA-99 and are substantially equivalent in terms of safety and effectiveness of operation)
    Compartment ReliefASME certified pressure relief valve (Fink, OxyHeal® 2000)One (1) ASME certified pressure relief valve per compartment. 30 psig to 75 psig (Substantially equivalent)
    Compartment DrainOne (1) manual drain in each compartment (Fink, OxyHeal® 2000)Minimum One (1) manual drain in each compartment (Substantially equivalent)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This document describes engineering tests and a comparison to predicate devices, not a clinical study on a patient test set in the conventional sense. The "test set" for the non-clinical testing comprises the physical components and integrated system of the OxyHeal® 5000 chamber itself.

    • Sample Size: The sample size is essentially one (the OxyHeal® 5000 Rectangular Multiplace Hyperbaric Chamber System Product Family, representing its design). The testing applies to the specific device models being submitted. The document evaluates the design and performance characteristics of new models within this product family against established standards and predicate devices.
    • Data Provenance: The testing details suggest that tests were conducted by the manufacturer (OxyHeal Medical Systems, Inc.) or by independent bodies (like ASME authorized inspectors) as part of the manufacturing and verification process. This is prospective testing for the submitted device design. The origin would be the manufacturing site or testing facilities. No country of origin for specific testing data is explicitly mentioned, but the manufacturer is based in National City, CA, USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as this is a device engineering and performance validation against standards and predicate devices, not an AI or diagnostic device requiring expert interpretation for "ground truth" establishment in a clinical imaging context. The "ground truth" for compliance to standards comes from the ASME and NFPA codes themselves, and expert judgment is inherent in the design, engineering, and inspection processes by qualified engineers and ASME authorized inspectors.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for this type of device submission. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies, especially those involving multiple readers interpreting data where a consensus or tie-breaking mechanism is needed to establish ground truth.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-based device. The device itself is the hyperbaric chamber system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context refers to established engineering standards, safety requirements, and the validated performance of predicate devices. Specifically:

    • Engineering Standards: ASME/PVHO-1 for pressure vessels, NFPA 99 for health care facilities (including fire suppression, electrical design, etc.).
    • Performance Specifications: User Design Specifications for the OxyHeal® 5000, which are verified through various tests (structural, fire suppression, operational, software validation, factory acceptance).
    • Predicate Device Characteristics: The established and legally marketed performance and design features of the Fink Engineering PTY LTD SL8, DL8, and TL20 Hyperbaric Oxygen Facility (K031649) and the OxyHeal Health Group®, Inc. OxyHeal® 2000 Hyperbaric Chamber Series (K011866).

    8. The sample size for the training set

    Not applicable. This is not an AI or machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082455

    Validate with FDA (Live)

    Date Cleared
    2009-05-22

    (269 days)

    Product Code
    Regulation Number
    868.5470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is the expressed, intended use of the Gulf Coast Hyperbaric rectangular hyperbaric chamber to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class II prescriptive device, it is further intended for physician involvement in their procurement and routine use.

    The UHMS is the professional medical organization chartered with setting the standards of care defining the appropriate use of hyperbaric oxygen. More specifically, the UHMS publishes a listing of medical conditions that have been clearly established as appropriate primary or adjunctive use of hyperbaric oxygen (HBO). The disorders on the list have been scientifically validated and verified through extensive data collection. It should be noted that the list is dynamic. Based on the strength of the scientific data, disorders are both added and removed from the list, depending on the outcomes of scientific pursuit.

    The conditions listed as appropriate for the use of HBO in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) is as follows:

    1. Air or gas embolism
    2. Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning
    3. Clostridial myositis and myonecrosis
    4. Crush injury, compartment syndrome, and other acute traumatic ischemias
    5. Decompression sickness
    6. Enhanced of selected problem wounds
    7. Exceptional blood loss anemia
    8. Necrotizing soft tissue infections
    9. Osteomyelitis (refractory)
    10. Delayed radiation injury (soft tissue and bony necrosis)
    11. Skin grafts and flaps (compromised)
    12. Thermal burns
    13. Intracranial abscess
    Device Description

    The Gulf Coast Hyperbarics rectangular multiplace hyperbaric chambers are designed and manufactured in accordance with the requirements of the ANSI/ASME Boiler and Pressure Vessel code, Section VIII, Division 1, Pressure Vessels ANSVASME-PVHO-1 (American Society of Mechanical Engineers- Pressure Vessels for Human Occupancy. and NFPA 99, Health Care Facilities; Chapter 20, Hyperbaric Facilities, 2005 Edition. The new Gulf Coast Hyperbaric multiplace rectangular chambers (K082455) are designed and manufactured using the same major components as in the cleared Gulf Coast Hyperbarics multiplace cylindrical chamber (K950957). The rectangular multiplace chamber includes as a key component, a new and user friendly rectangular design which is simple and simulates as closely as possible the clinical conditions found elsewhere in their working environment.

    The rectangular chamber(s) place major emphasis on patient and operator safety, and user-friendly operator controls and includes several unique features including:

    • A large, comfortable rectangular hyperbaric chamber that has been outfitted to appear like any other clinical room in a hospital to reduce patient anxiety
    • Large walk-through rectangular sliding doors that fit flush with the floor so that patients can be wheeled into the chamber without bumping over the doorjamb.
    • A large and easy to use control console designed to display all functions of the chamber
    • An oxygen hood supply and exhaust system designed for easy use and eliminates over or under pressurization of the hood and eliminates the possibility of "shrink wrapping" the patient hood.
    • Air conditioning system providing controllable cooling and heating of the chamber interior. The temperature and humidity is monitored at the control console.
    AI/ML Overview

    The provided text is a 510(k) summary for a hyperbaric chamber (K082455). It describes the device, its intended use, and its substantial equivalence to existing devices. However, it does not contain information about any studies conducted to prove the device meets acceptance criteria.

    The document lists the specifications of the device, which could be considered design acceptance criteria, but there is no mention of testing or studies performed to demonstrate adherence to these specifications beyond the statement that the device is "designed and manufactured in accordance with the requirements".

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, nor can I provide information on:

    • A table of acceptance criteria and reported device performance.
    • Sample size used for the test set or data provenance.
    • Number of experts or their qualifications for ground truth establishment.
    • Adjudication method for the test set.
    • MRMC comparative effectiveness study or its effect size.
    • Standalone performance study.
    • Type of ground truth used.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    The document is purely a premarket notification for substantial equivalence, which primarily relies on comparing the new device to existing legally marketed predicate devices, rather than conducting new performance studies for the new device as would be required for a novel device or a PMA submission.